CARDIASSIST COUNTER PULSATION SYSTEM SERIES 4000/5000
K050172 · Cardiomedics, Inc. · DRN · Mar 31, 2005 · Cardiovascular
Device Facts
Record ID
K050172
Device Name
CARDIASSIST COUNTER PULSATION SYSTEM SERIES 4000/5000
Applicant
Cardiomedics, Inc.
Product Code
DRN · Cardiovascular
Decision Date
Mar 31, 2005
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.5225
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Cardiomedics, Inc., CardiAssist Counter Pulsation System - Series 4000 is intended to provide external counterpulsation (ECP) for the treatment of ischemic heart disease by increasing perfusion during diastole in people with chronic angina pectoris, congestive heart failure, myocardial infarction and cardiogenic shock. Use of this device may reduce pain and impairment associated with angina pectoris, congestive heart failure or myocardial infarction and may enhance coronary function.
Device Story
Non-invasive circulatory assist device; provides external counterpulsation (ECP). Inputs: patient ECG (5-lead or 12-lead), finger plethysmograph, blood pressure. Operation: microprocessor-controlled system triggers inflation of three sets of leg cuffs (calves, thighs, buttocks) during diastole; sequential compression 40ms apart; pressures up to 310 mmHg. Output: retrograde pressure transmitted through vascular system to increase diastolic aortic pressure; displayed on console touch screen. Used in clinical settings; operated by trained personnel. Benefits: enhanced coronary perfusion, reduced cardiac workload, improved collateral circulation. System includes portable console, computer, pumps, printer, cuffs, and hoses.
Clinical Evidence
No clinical data provided; substantial equivalence based on technological characteristics and functional comparison to predicate devices.
Technological Characteristics
Microprocessor-based control system; ECG R-wave triggering; pneumatic cuff system (no vacuum required); touch screen interface; through-hole circuit board; safety interlock requiring ECG signal; max pressure 310 mmHg.
Indications for Use
Indicated for patients with ischemic heart disease, including chronic angina pectoris, congestive heart failure, myocardial infarction, and cardiogenic shock, to increase diastolic perfusion and enhance coronary function.
Regulatory Classification
Identification
An external counter-pulsating device is a noninvasive, prescription device used to assist the heart by applying positive or negative pressure to one or more of the body's limbs in synchrony with the heart cycle.
Special Controls
*Classification.* (1) Class II (special controls) when the device is intended for the treatment of chronic stable angina that is refractory to optimal anti-anginal medical therapy and without options for revascularization. The special controls for this device are:(i) Nonclinical performance evaluation of the device must demonstrate a reasonable assurance of safety and effectiveness for applied pressure, synchronization of therapy with the appropriate phase of the cardiac cycle, and functionality of alarms during a device malfunction or an abnormal patient condition;
(ii) Reliabilities of the mechanical and electrical systems must be established through bench testing under simulated use conditions and matched by appropriate maintenance schedules;
(iii) Software design and verification and validation must be appropriately documented;
(iv) The skin-contacting components of the device must be demonstrated to be biocompatible;
(v) Appropriate analysis and testing must be conducted to verify electrical safety and electromagnetic compatibility of the device; and
(vi) Labeling must include a detailed summary of the device-related and procedure-related complications pertinent to use of the device.
(2) Class III (premarket approval) for the following intended uses: Unstable angina pectoris; acute myocardial infarction; cardiogenic shock; congestive heart failure; postoperative treatment of patients who have undergone coronary artery bypass surgery; peripheral arterial disease associated with ischemic ulcers rest pain or claudication, threatened gangrene, insufficient blood supply at an amputation site, persisting ischemia after embolectomy or bypass surgery, and/or pre- and post-arterial reconstruction to improve runoff; diabetes complicated by peripheral arterial disease or other conditions possibly related to arterial insufficiency including nocturnal leg cramps and/or necrobiosis diabeticorum; venous diseases, including prophylaxis of deep vein thrombophlebitis, edema (e.g., chronic lymphedema) and/or induration (e.g., stasis dermatitis) associated with chronic venous stasis, venous stasis ulcers, and/or thrombophlebitis; athletic injuries, including Charley horses, pulled muscles and/or edematous muscles; necrotizing cellulitis.
(c)
*Date premarket approval application (PMA) or notice of completion of product development protocol (PDP) is required.* A PMA or notice of completion of a PDP is required to be filed with FDA on or before March 31, 2014, for any external counter-pulsating device, with an intended use described in paragraph (b)(2) of this section, that was in commercial distribution before May 28, 1976, or that has, on or before March 31, 2014, been found to be substantially equivalent to any external counter-pulsating device, with an intended use described in paragraph (b)(2) of this section, that was in commercial distribution before May 28, 1976. Any other external counter-pulsating device with an intended use described in paragraph (b)(2) of this section shall have an approved PMA or declared completed PDP in effect before being placed in commercial distribution.
{1}------------------------------------------------
MAR 3 1 2005
510(k) Notification for New Device Cardiomedics CardiAssist Counter Pulsation System – 4000 Series December 2004
## Attachment 7
## 510(k) Summary
## Cardiomedics, Inc. CardiAssist Counter Pulsation System - Series 4000
- Date Prepared: 1.
January 20, 2005
- Cardiomedics, Inc. 2. Submitter's Name: 18872 Bardeen Avenue and Address Irvine, CA 92612
- John Hutchins, Contact Person: 3. Vice President, Marketing and Clinical Affairs Cardiomedics, Inc. Ph: (949) 863-2500 x104 Fax: (949) 474-2446 Email: hutch@Cardiomedics.com
- CardiAssist Counter Pulsation System 4. Device Name:
- CardiAssist Counter Pulsation System Series 4000 Proprietary Name:
Device, Counter-pulsating, External Classification Name:
- The CardiAssist Series 4000 Counter Pulsation System is equivalent 5. Predicate Device: in function and intended use to the currently marketed CardiAssist ECP System, Mark 3000 cleared for market entry under 510(k) K023427 on January 7, 2003 as well as technologically and functionally equivalent to the Nicore NCP-1 Device, cleared for market entry under 510(k) K023016 on December 4, 2002.
{2}------------------------------------------------
510(k) Notification for New Device Cardiomedics Series 4000 - CardiAssist Counter Pulsation System December 2004
- The Cardiomedics CardiAssist External Counter Pulsation System -6. Device Description: Series 4000 is a non-invasive circulatory assist device, which provides increased circulation via external counterpulsation (ECP) for the treatment of ischemic heart disease including congestive heart failure, chronic angina pectoris, acute myocardial infarction and cardiogenic shock. External counterpulsation therapy improves cardiac function by enhancing the perfusion of the coronary vasculature, the development of coronary collateral circulation, and by reducing the workload of the heart. The CardiAssist External Counter Pulsation System - Series 4000 consists of the portable console containing the computer and pumps with a touch screen for user interface, an integral patient chart printer, and leg cuffs and hoses. Additional components provided with the System include a finger plethysmograph and 5-lead ECG cable and leads (12 lead optional) and blood pressure cuff. This CardiAssist ECP System sequentially compresses the legs from the calves, thighs and buttocks, 40 milliseconds apart, by inflating three sets of flexible fabric cuffs during diastole. This results in movement of blood from the legs to the heart and entire upper body. Pressure, up to 310 mmHg, is applied with the timing and duration of each pulse, synchronized with the patient's ECG. When properly triggered, the pressure pulses applied to the vascular bed of the legs and buttocks transmit retrograde pressure through the entire vascular system. At the aorta, the aortic valve prevents retrograde flow into the left ventricle. Thus, a peak pulse of diastolic pressure occurs at or above systolic levels, which increases the driving pressure in the coronary vasculature. The Cardiomedics, Inc., CardiAssist External Counter Pulsation 7. Intended Use: System - Series 4000, is intended to provide external counterpulsation (ECP) for the treatment of ischemic heart disease by increasing perfusion during diastole in people with chronic angina pectoris, congestive heart failure, myocardial infarction and cardiogenic shock. Use of this device may reduce pain and impairment associated with angina pectoris, congestive heart failure or myocardial infarction and may enhance coronary function. Technological characteristics of this device which are similar to those 8. Comparison of of the Cardiomedics 3000 Series predicate system include: the Technological triggering mechanism (patient's ECG R-wave); microprocessor based Characteristics control system; emergency system power down; limited maximum
pressure used for treatment; through-hole circuit board; safety interlock requiring external ECG signal before treatment can start; and the Cardiomedics cuff system which eliminates the need for vacuum cuff deflation.
Technological characteristics of this device which are similar to those
{3}------------------------------------------------
510(k) Notification for New Device Cardiomedics Scries 4000 - CardiAssist Counter Pulsation System December 2004
> of the Nicore predicate device include: the triggering mechanism (patient's ECG R-wave), microprocessor based control system, limited maximum pressure used for treatment, operator's console and data display with mouse based interface.
{4}------------------------------------------------
MAR 3 1 2005
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Cardiomedics, Inc.. c/o Mr. John Hutchins 18872 Bardeen Avenue Irvine, CA 92612
Re: K050172
Cardiomedics CardiAssist Counter Pulsation System – Series 4000 Regulation Number: 21 CFR 870.5225 Regulation Name: External Counter Pulsation Device Regulatory Class: Class III (three) Product Code: DRN Dated: January 20, 2005 Received: February 1, 2005
Dear Mr. Hutchins:
We have reviewed your Section 510(k) premarket notification of intent to market the device we nave reviewed your bected. 310(t) pe device is substantially equivalent (for the indications ferenced above and nave acterings ally marketed predicate devices marketed in interstate for use stated in the encrosure) to regary and eath of the Medical Device Amendments, or to connineree prior to May 20, 1978, the excordance with the provisions of the Federal Food, Drug, devices that have been reclassined in assee approval of a premarket approval application (PMA). and Cosment Act (Act) that do not requently of the general controls provisions of the Act. The You may, merenne, market the dorres, basjoct on the morements for annual registration, listing of general controls provisions of the rice labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device is classified (see above) into existing major regulations affecting your device can Illay be subject to suen additions, Title 21, Parts 800 to 898. In addition, FDA may be Numd in the Oode of I ourself in the Federal Register.
{5}------------------------------------------------
Page 2- Mr. John Hutchins
Please be advised that FDA's issuance of a substantial equivalence determination does not mean I toase to advisou that I Dr usean that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must or uny I edetails all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. product in the low you to begin marketing your device as described in your Section 510(k) I mis lotter with and in Joint FDA finding of substantial equivalence of your device to a legally promaticated predicated. " Thesults in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire of Compliance at (301) 594-4648. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small other golleral micronational and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address
(2017) 115 097 fda.gov/cdrh/dsma/dsmamain.html
Sincerely vours,
Mocel Mayr
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
510(k) Notification for New Device 910(k) Notycation Jor 1000 - CardiAssist Counter Pulsation System December 2004
## Attachment 8 - Statement of Intended Use
510(k) Number: _______________________________________________________________________________________________________________________________________________________________
Device Name: CardiAssist Counter Pulsation System - Series 4000
Indications for Use:
The Cardiomedics, Inc., CardiAssist Counter Pulsation System - Series 4000 is intended to The Cardiomedics, Inc., CalulAssist Counter Fuloaten of isschemic heat disease by provide external Counterpulsation CEOT / Tell the chronic angina perforis, congestive heat.
Increasing perfusion during diastole in people with chrise may reduce pain and increasing perfusion during diastole in poople with of lease this device may request and fallure, myocardial infarction and cardiogenie oneon. "So on the libre or myocardial infarction and may enhance coronary function.
Prescription Use _____________________________________________________________________________________________________________________________________________________________ (Per 21 CFR 801.109)
Over-The Counter Use_
(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)
OR
Concurrence of CDRH, Office of Device Evaluation (ODE)
Mokey Moyer
scular Devices Oit L 510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.