EECP THERAPY SYSTEM MODEL TS3 WITH PULSE OXIMETRY AND ENHANCED EXTERNAL COUNTERPULSATION EECP MC-2 EXTERNAL COUNTER
K020857 · Vasomedical, Inc. · DRN · Jun 14, 2002 · Cardiovascular
Device Facts
Record ID
K020857
Device Name
EECP THERAPY SYSTEM MODEL TS3 WITH PULSE OXIMETRY AND ENHANCED EXTERNAL COUNTERPULSATION EECP MC-2 EXTERNAL COUNTER
Applicant
Vasomedical, Inc.
Product Code
DRN · Cardiovascular
Decision Date
Jun 14, 2002
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.5225
Device Class
Class 2
Attributes
Therapeutic, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K020857 · Jun 14, 2002
EECP THERAPY SYSTEM MODEL TS3 WITH PULSE OXIMETRY AND ENHANCED EXTERNAL COUNTERPULSATION EECP MC-2 EXTERNAL COUNTER
Vasomedical, Inc.
Clinical registries
Registry-based clinical investigations were used to demonstrate clinical benefit (e.g., peak oxygen consumption, exercise duration, quality of life) in patients treated with the EECP therapy system.
Peak oxygen consumption, exercise duration, pre-load adjusted maximal left ventricular power, quality of life, functional ability
Indications for Use
Vasomedical, Inc.'s EECP® Therapy System Model TS3 with Pulse Oximetry is a non-invasive external counterpulsation device intended for use in the treatment of patients with congestive heart failure, stable or unstable angina pectoris, acute myocardial infarction, or cardiogenic shock.
Device Story
Non-invasive external counterpulsation system comprising control console, treatment table, and patient cuff set. Inputs include ECG, finger plethysmograph, and oxygen saturation signals. Microprocessor-controlled pneumatic system inflates/deflates cuffs on lower extremities in synchronization with cardiac cycle; inflation during diastole increases coronary perfusion and venous return; rapid deflation before systole reduces afterload. Used in professional clinical settings by trained operators. System monitors treatment pressure via internal sensors and closed-loop control. Output includes real-time display of ECG, inflation/deflation timing, plethysmography, heart rate, pressure ratios, and oxygen saturation. Benefits include improved coronary blood flow and reduced cardiac workload, supporting treatment of heart failure and angina.
Clinical Evidence
Clinical evaluation included multi-center, single-center, and registry-based investigations. Results demonstrated clinical benefit in patients with congestive heart failure. Improvements observed in objective measures (peak oxygen consumption, exercise duration, pre-load adjusted maximal left ventricular power) and subjective measures (quality of life, functional ability).
Technological Characteristics
System components: control console (air compressor, microprocessor, signal module), treatment table (motorized lift, pneumatic valves), and cuff set. Materials: cotton and Lycra (Spandex) treatment pants. Sensing: ECG, finger plethysmography, pulse oximetry. Connectivity: internal data storage, printer, transferable media. Software: proprietary custom software for system monitoring and closed-loop pressure control. Standards: EN 14971-1, IEC 601-1-4, ISO 10993-1/5/10/12.
Indications for Use
Indicated for adult patients with congestive heart failure, stable or unstable angina pectoris, acute myocardial infarction, or cardiogenic shock.
Regulatory Classification
Identification
An external counter-pulsating device is a noninvasive, prescription device used to assist the heart by applying positive or negative pressure to one or more of the body's limbs in synchrony with the heart cycle.
Special Controls
*Classification.* (1) Class II (special controls) when the device is intended for the treatment of chronic stable angina that is refractory to optimal anti-anginal medical therapy and without options for revascularization. The special controls for this device are:(i) Nonclinical performance evaluation of the device must demonstrate a reasonable assurance of safety and effectiveness for applied pressure, synchronization of therapy with the appropriate phase of the cardiac cycle, and functionality of alarms during a device malfunction or an abnormal patient condition;
(ii) Reliabilities of the mechanical and electrical systems must be established through bench testing under simulated use conditions and matched by appropriate maintenance schedules;
(iii) Software design and verification and validation must be appropriately documented;
(iv) The skin-contacting components of the device must be demonstrated to be biocompatible;
(v) Appropriate analysis and testing must be conducted to verify electrical safety and electromagnetic compatibility of the device; and
(vi) Labeling must include a detailed summary of the device-related and procedure-related complications pertinent to use of the device.
(2) Class III (premarket approval) for the following intended uses: Unstable angina pectoris; acute myocardial infarction; cardiogenic shock; congestive heart failure; postoperative treatment of patients who have undergone coronary artery bypass surgery; peripheral arterial disease associated with ischemic ulcers rest pain or claudication, threatened gangrene, insufficient blood supply at an amputation site, persisting ischemia after embolectomy or bypass surgery, and/or pre- and post-arterial reconstruction to improve runoff; diabetes complicated by peripheral arterial disease or other conditions possibly related to arterial insufficiency including nocturnal leg cramps and/or necrobiosis diabeticorum; venous diseases, including prophylaxis of deep vein thrombophlebitis, edema (e.g., chronic lymphedema) and/or induration (e.g., stasis dermatitis) associated with chronic venous stasis, venous stasis ulcers, and/or thrombophlebitis; athletic injuries, including Charley horses, pulled muscles and/or edematous muscles; necrotizing cellulitis.
(c)
*Date premarket approval application (PMA) or notice of completion of product development protocol (PDP) is required.* A PMA or notice of completion of a PDP is required to be filed with FDA on or before March 31, 2014, for any external counter-pulsating device, with an intended use described in paragraph (b)(2) of this section, that was in commercial distribution before May 28, 1976, or that has, on or before March 31, 2014, been found to be substantially equivalent to any external counter-pulsating device, with an intended use described in paragraph (b)(2) of this section, that was in commercial distribution before May 28, 1976. Any other external counter-pulsating device with an intended use described in paragraph (b)(2) of this section shall have an approved PMA or declared completed PDP in effect before being placed in commercial distribution.
Nonin® PalmSAT®, Model 2500, Pulse Oximeter (K002690)
Submission Summary (Full Text)
{1}------------------------------------------------
# 510(k) Summary
## JUN 1 4 2002
### Vasomedical, Inc. EECP® Therapy System Model TS3 with Pulse Oximetry
| 1. | Date Prepared: | 13 June 2002 |
|----|----------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 2. | Submitter's Name:<br>and Address | Vasomedical, Inc.<br>180 Linden Ave.<br>Westbury, NY 11590 |
| 3. | Contact Person: | Thomas R. Varricchione, MBA, RRT<br>Vice President, Clinical Research and Regulatory Affairs<br>Vasomedical, Inc.<br>Telephone: (516) 997-4600<br>Facsimile: (516) 997-2299<br>E-mail: TVarricchione@Vasomedical.com |
| 4. | Device Name: | Vasomedical, Inc.<br>EECP® Therapy System Model TS3 with Pulse Oximetry<br>EECP® Therapy System Model TS3<br>Enhanced External Counterpulsation EECP® MC-2 |
| | Proprietary Name: | EECP Therapy System Model TS3 with Pulse Oximetry<br>EECP® Therapy System Model TS3<br>Enhanced External Counterpulsation EECP® MC-2 |
| | Common Name: | Enhanced External Counterpulsation EECP® MC-2<br>Enhanced External Counterpulsation (EECP®) Therapy<br>System |
| | Classification Name: | Device, Counter-pulsating, External |
| 5. | Predicate Device: | The EECP® Therapy System Model TS3 with Pulse<br>Oximetry is substantially equivalent to the EECP® Therapy<br>System Model TS3 (K003469) and, by reference, to the<br>EECP® MC-2 (K940264). |
| 6. | Device Description: | The EECP® Therapy System Model TS3 with Pulse<br>Oximetry is comprised of three major components, a<br>Control Console, a Treatment Table, and a patient Cuff Set. |
{2}------------------------------------------------
The Control Console accommodates the air compressor and reservoir, a signal module panel, a power module, a microprocessor with touch screen/keyboard interface, data storage drives and printer, and components for acquiring and processing ECG, finger plethysmograph and oxygen saturation signals. The microprocessor is used to operate and monitor the system by means of proprietary custom software, with the operator using the touch screen/keyboard interface to control its operation. The screen displays information pertinent to operating the system, as well as treatment parameters and patient waveforms during use. Treatment pressure is monitored with an internal pressure sensor and the operator-selected set-point maintained by a closed-loop control system. The touch screen employs "hardware-less keys" which the operator touches to select a function or execute a command and the keyboard enables alphanumeric text entries. An internal hard disk drive is used to store data on the system, an internal floppy disk drive is used to record data onto transferable media, and a printer is used to produce hard copy of site and patient identifiers and physiologic data.
The Treatment Table accommodates a motorized lifting mechanism, mattress and the pneumatic circuit valve assembly. The motorized lifting mechanism is used to move the mattress up and down, providing a convenient height for patient and operator use. The valve assembly consists of three pairs of inflation/deflation valves that open and close on command to inflate or deflate the patient Cuff Set with air. The valve assembly is connected to the air compressor and reservoir components in the Control Console via connecting air hoses.
External pressure is applied via the patient Cuff Set to the lower extremities of the patient in synchronization with the heart, i.e. the cuffs compress vascular beds in the calves, lower thighs and upper thighs/buttocks on inflation. When the heart is in its relaxed state during the diastolic period, pressure is applied sequentially from the calves, to the lower thighs, to the upper thighs and buttocks, forcing blood back to the heart, increasing coronary perfusion pressure and coronary blood flow (diastolic augmentation), as well as venous return. Immediately before the heart begins to eject blood during the next systolic phase, the cuffs are rapidly deflated and all externally applied pressure is eliminated. The vasculature in the lower extremities
{3}------------------------------------------------
reconforms and is able to receive the output of the heart with lessened resistance, thereby reducing systolic pressure and the workload of the heart (decreased afterload). Stretchable treatment pants comprised of cotton and Lycra (Spandex) are worn by the patient under the Patient Cuff Set to allow for greater comfort during treatment.
The Model TS3 with Pulse Oximetry incorporates a noninvasive sensor and electronic module to acquire and process the patient's oxygen saturation. These same components functions can be used separately or simultaneously. The Nonin oxygen saturation metering device has been an added feature to the following devices:
- NONIN® Pulse Oximeter and Carbon Dioxide Detector (K982969)
- 9303 Neonatal/Adult Vital Signs Monitor (K982776)
- MTS Option for the ESCORT® II Monitor (K970763)
- Nonin® PalmSAT®, Model 2500, Pulse Oximeter with 2500C & 2500B (K002690)
Key differences between the current device and the modified device include:
- Congestive heart failure has been added to the indications for use of the device. .
- . Contraindications for left ventricular hypertrophy and cardiomyopathy have been removed from the device's labeling.
- . A pulse oximetry function has been added to measure and display oxygen saturation in adult patients in a professional setting.
- . The display has been modified to allow the operator to simultaneously display ECG, inflation/deflation time, finger plethysmography, heart rate, ratio of pressure amplitudes, ratio of area under pressure waveform, treatment time and functional oxygen saturation.
- . Modifications to the Pressure Control components to enable the operator to set the desired treatment pressure with less presses of up/down touch keys, automatic maintenance of the treatment pressure set point, and holding of the treatment pressure set point when the system is placed in "Standby" mode.
- The EECP® Therapy System Model TS3 with Pulse 7. Intended Use: Oximetry is a non-invasive external counterpulsation device intended for the use in the treatment of patients with congestive heart failure, stable or unstable angina pectoris, acute myocardial infarction, or cardiogenic shock.
- Technological and functional characteristics of the 8. Comparison of
{4}------------------------------------------------
9.
| Technological<br>Characteristics: | modified device are essentially the same as those of the<br>predicate device. Principles of operation are the same. |
|-----------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| . Non-clinical Tests: | Non-clinical testing on the EECP® Therapy System Model<br>TS3 with Pulse Oximetry included the following: |
| • | Software verification and validation, as follows: |
| | - functional requirements as defined in the EECP®<br>Therapy System Model TS3 with Pulse Oximetry<br>System Requirements Specification. |
| | - boundary values and stress testing as defined by<br>FDA's Guidance for the Content of Premarket<br>Submission for Medical Devices Containing<br>Software, CDRH, ODE, FDA, May, 1998. |
| | - safety requirements as identified in the safety risk<br>analysis performed in accordance with EN 14971-1,<br>Medical Device Risk Analysis, October, 1994, the<br>"Essential Requirements of the Medical Devices<br>Directives", 14 June, 1993, and IEC 601-1-4, Medical<br>Electrical Equipment Part 1: General requirements for<br>safety, 4 Collateral Standard: Programmable electrical<br>medical systems, 1996-05 |
- testing in support of validation in accordance with the FDA's General Principles of Software Validation, Final Guidance for Industry and FDA Staff, January 11, 2002, and IEC 601-1-4, Medical Electrical Equipment Part 1: General requirements for safety, 4 Collateral Standard: Programmable electrical medical systems. 1996-05.
- Verification of System operation, as follows: ●
- functional requirements as defined in the EECP® Therapy System Model TS3 with Pulse Oximetry System Requirements Specification to verify the performance of the modified device at the system level.
- safety requirements as identified in the safety risk analysis performed in accordance with EN 14971-1, Medical Device Risk Analysis.
- additional verification tests, as indicated, to verify performance at the component level.
- biocompatibility testing in conformity with recognized standards ISO 10993-1, ISO 10993-5, ISO 10993-10 and ISO 10993-12.
{5}------------------------------------------------
Clinical evaluation of EECP® in patients with congestive 10. Clinical Evaluation: heart failure has been performed in multi-center, single center and registry-based clinical investigations. Results of these investigations have demonstrated clinical benefit in patients treated with Vasomedical EECP® Therapy Systems. Objective measures such as peak oxygen consumption, exercise duration and pre-load adjusted maximal left ventricular power are improved following EECP® therapy, as well as subjective measures of patient response to therapy, such as quality of life and functional ability measures.
{6}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized design of an eagle or bird-like figure, composed of three curved lines that suggest the shape of a bird in flight.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
#### JUN 1 8 2002
Vasomedical, Inc. c/o Mr. Thomas R. Varricchione, MBA, RRT Vice President, Clinical Research and Regulatory Affairs 180 Linden Avenue Wesbury, NY 11590
Re: K020857
Trade Name: EECP® Therapy System Model TS3, Model TS3 with Pulse Oximetry , and Enhanced External Counterpulsation EECP® MC-2 External Counterpulsating Devices Regulation Number: 21 CFR 870.5225 Regulation Name: Device, Counter-pulsating, External Regulatory Class: Class III (three) Product Code: DRN Dated: March 15, 2002 Received: March 18, 2002
Dear Mr. Varricchione:
This letter corrects our substantially equivalent letter of June 14, 2002, regarding the trade names referenced in your submission.
We have reviewed your Section 510(k) premarket notification of intent to market the device w f nave a rowe and have determined the device is substantially equivalent for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, do noos mat have been . You may, therefore, market the device, subject to the general controls and Cosment + e. ( . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . profistion, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{7}------------------------------------------------
#### Page 2 - Mr. Thomas R. Varricchione, MBA, RRT
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to continue marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4646 Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at their toll free number (800) 638-2041 or at (301) 443-6597 or at its Internet address
http://www.fda.qov/cdrh/dsma/dsmamain.html.
Sincerely yours.
R. R. Till Rl. R.
Donna-Bea Tillman, Ph.D. Acting Director Division of Cardiovascular and Respiratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosures
{8}------------------------------------------------
# Indication for Use Statement
| 510(k) Number: | K020857 |
|----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name: | EECP® Therapy System Model TS3 with Pulse Oximetry<br>External Counter-pulsating Device |
| Indications for Use: | Vasomedical, Inc.'s EECP® Therapy System Model TS3 with Pulse<br>Oximetry is a non-invasive external counterpulsation device intended for<br>use in the treatment of patients with congestive heart failure, stable or<br>unstable angina pectoris, acute myocardial infarction, or cardiogenic<br>shock. |
#### PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED
#### Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use
(Per 21 CFR 801.109) ✓
OR
Over-The Counter Use
(Rel 21 CFR 801.109)
Division of Cardiovascular & Respiratory Devices
510(k) Number 12020857
{9}------------------------------------------------
# Indication for Use Statement
| 510(k) Number: | K020857 |
|----------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name: | EECP <sup>®</sup> Therapy System Model TS3<br>External Counter-pulsating Device |
| Indications for Use: | Vasomedical, Inc.'s EECP <sup>®</sup> Therapy System Model TS3 is a non-invasive<br>external counterpulsation device intended for use in the treatment of<br>patients with congestive heart failure, stable or unstable angina pectoris,<br>acute myocardial infarction, or cardiogenic shock. |
### PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use
(Per 21 CFR 801.109)
OR
Division of Cardiovascular & Respiratory Devices
510(k) Number//K020857
Over-The Counter Use
ﺘﺠﻬﺰ
{10}------------------------------------------------
# Indication for Use Statement
| 510(k) Number: | K020857 |
|----------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name: | Enhanced External Counterpulsation EECP® MC-2<br>External Counter-pulsating Device |
| Indications for Use: | Vasomedical, Inc.'s Enhanced External Counterpulsation EECP® MC-2 is<br>a non-invasive external counterpulsation device intended for use in the<br>treatment of patients with congestive heart failure, stable or unstable<br>angina pectoris, acute myocardial infarction, or cardiogenic shock. |
#### PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED
### Concurrence of CDRH, Office of Device Evaluation (ODE)
OR
Prescription Use
(Per 21 CFR 801.109)
Division of Cardiovascular & Respiratory Devices
510(k) Number_KO 2085 F
Over-The Counter Use
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.