HEARTSTREAM/HEARTSTART FR2 AED ECG CABLE, MODELS M3860A, M3861A, M3840A, M3841A

K013425 · Philips Medical Systems · MKJ · Jan 14, 2002 · Cardiovascular

Device Facts

Record IDK013425
Device NameHEARTSTREAM/HEARTSTART FR2 AED ECG CABLE, MODELS M3860A, M3861A, M3840A, M3841A
ApplicantPhilips Medical Systems
Product CodeMKJ · Cardiovascular
Decision DateJan 14, 2002
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.5310
Device ClassClass 3
AttributesTherapeutic, Pediatric

Indications for Use

The Heartstream FR2 is intended for use by personnel who have been trained in its operation. The user should be qualified by training in basic life support, or advanced life support, or other physician-authorized emergency medical response. The Heartstream FR2 is indicated for use on victims of sudden cardiac arrest exhibiting the following signs: . Unresponsiveness . Absence of breathing At the discretion of emergency care personnel, the M3860A FR2 with ECG display enabled can also be used with the FR2 ECG Cable to display the rhythm of a responsive or breathing patient, regardless of age. The FR2 and ECG Cable system provides a nondiagnostic display for attended patient monitoring. While connected to the FR2 ECG Cable, the FR2 evaluates the patient's ECG and disables its shock capability.

Device Story

Heartstream FR2 is an automated external defibrillator (AED) for use by trained BLS/ALS personnel. Input: ECG signals via defibrillation pads or optional 3-lead ECG cable. Operation: Patient Analysis System (PAS) continuously analyzes ECG for shockable rhythms; delivers impedance-compensating biphasic truncated exponential waveform shocks. When ECG cable is attached, device provides non-diagnostic ECG display and disables shock capability. Features include self-testing, voice/text prompts, and optional data card for event recording. Used in emergency settings to treat SCA; provides monitoring for responsive patients. Benefits: rapid rhythm analysis, automated shock delivery, and patient monitoring capability.

Clinical Evidence

Bench testing only. Performance of the ECG cable was verified against specifications to ensure appropriate function for its intended application.

Technological Characteristics

Impedance-compensating biphasic truncated exponential waveform; 3-lead ECG cable (AAMI/IEC conventions); lithium manganese dioxide battery; crystal display screen; infrared communication port; PAS algorithm for rhythm analysis.

Indications for Use

Indicated for victims of sudden cardiac arrest (SCA) exhibiting unresponsiveness and absence of breathing. For patients under 8 years or <55 lbs (25 kg), use attenuated pediatric pads. ECG monitoring mode indicated for responsive or breathing patients of any age for non-diagnostic attended monitoring.

Regulatory Classification

Identification

An automated external defibrillator (AED) system consists of an AED and those accessories necessary for the AED to detect and interpret an electrocardiogram and deliver an electrical shock (e.g., battery, pad electrode, adapter, and hardware key for pediatric use). An AED system analyzes the patient's electrocardiogram, interprets the cardiac rhythm, and automatically delivers an electrical shock (fully automated AED), or advises the user to deliver the shock (semi-automated or shock advisory AED) to treat ventricular fibrillation or pulseless ventricular tachycardia.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K013425 # JAN 1 4 2002 ## 14.PreMarket Notification Summary ## 510(k) Summary for the Heartstream FR2 AED #### Date Summary Prepared 1. October 15, 2001 #### 2. Submitter's Name and Address Philips Medical System Heartstream 2401 Fourth Avenue, Suite 500 Seattle, WA 98121 #### 3. Contact Person Tamara Yount Philips Medical System Heartstream Telephone: (206) 664-5141 Facsimile: (206) 664-5001 #### 4. Device Name Proprietary Name: Heartstream FR2 AED with ECG Cable Common Name: Automated external defibrillator Classification Names: Low-Energy Defibrillator #### 5. Predicate Devices The legally marketed device to which Philips Medical Systems, Heartstream claims equivalence for the Heartstream FR2 AED with ECG Cable is the Heartstream FR2 AED, Laerdal Heartstart 3000, Heartstream XLT Defibrillator/Monitor and the Active Corporation ActiveECG device. The design and intended use of the modified Heartstream FR2 AED is substantially equivalent in safety and performance to the devices named above. #### 6. Device Description The Heartstream FR2 is an automated external defibrillator available in two models, including one with ECG display and manual shock capability. Features include selftesting, impedance-compensating biphasic truncated exponential waveform, multiparameter Patient Analysis System (PAS), and human factor designs to facilitate use by lay responders. {1}------------------------------------------------ A non-rechargeable lithium manganese dioxide battery powers the FR2 with a typical capacity of 300 shocks or 12 hours of operating time. Except for specific programmed periods when a responder needs to deliver uninterrupted CPR, the FR2 continuously and automatically analyzes the ECG and alerts the responder when the ECG changes to a possible shockable rhythm. Analysis continues even after the FR2 advises a shock and arms - if the ECG spontaneously converts to a nonshockable rhythm prior to a responder pressing the shock button, the FR2 disarms. If significant artifact is detected in the ECG, Heartstream's PAS suspends further analysis until reliable data is available. When a shockable rhythm is detected, the FR2 directs the responder to press the shock button to deliver a biphasic shock to the patient. Event and incident data can be recorded during FR2 use with an optional data card having a recording capacity of four hours of event and ECG data (or thirty minutes with voice recording). The FR2 has an optional Training and Administration Pack that is used for device training and for customizing FR2 set-up options. Use of the Training and Administration Pack converts the FR2 to a training device with ten training "scripts" that simulate different SCA scenarios. The FR2 also has an infrared communication port to facilitate communication of set-up parameters. An optional reusable ECG Cable can be inserted into the AED's connector port to permit the viewing of lead II ECG and heart rate on the FR2's main crystal display screen. Three lead wires, connected to standard disposable ECG electrodes, are labeled according to AAMI or IEC conventions. If a potentially shockable rhythm is detected (using the existing FR2 PAS algorithm) or if the heart rate drops below 30 beats per minute, voice and text prompts are generated. ### 7. Intended Use The Heartstream FR2 is intended for use by personnel who have been trained in its operation. The user should be qualified by training in basic life support, or advanced life support, or other physician-authorized emergency medical response. The Heartstream FR2 is indicated for use on victims of sudden cardiac arrest exhibiting the following signs: - . Unresponsiveness - . Absence of breathing At the discretion of emergency care personnel, the M3860A FR2 with ECG display enabled can also be used with the FR2 ECG Cable to display the rhythm of a responsive {2}------------------------------------------------ or breathing patient, regardless of age. The FR2 and ECG Cable system provides a nondiagnostic display for attended patient monitoring. While connected to the FR2 ECG Cable, the FR2 evaluates the patient's ECG and disables its shock capability. ### Comparison of Technology Characteristics 8. The modified Heartstream FR2 AED with the ECG Cable employs the same fundamental scientific technologies as the Laerdal Heartstart 3000, the Heartstream XLT Defibrillator/Monitor and the Active Corporation ActiveECG device. ### Data Used in Determination of Substantial Equivalence 9. The Heartstream FR2 employs the same technologies as the predicate devices used for comparison. The FR2 acquires and analyzes ECG signals, and utilizes the same shock advisory criteria like the predicates and the FR2 without the ECG Cable. Bench testing demonstrates that the performance of the ECG Cable meets specifications is appropriate for its intended application. #### 10. Conclusion The modifications proposed to include an optional ECG Cable to the FR2 accessory offerings do not present new issues of safety or effectiveness. {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three curved lines. Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 JAN 1 4 2002 Ms. Tamara Yount Philips Medical Systems 2401 Fourth Avenue, Suite 500 Seattle, WA 98121-1436 Re: K013425 Trade Name: Heartstream FR2 AED with an ECG Cable Regulation Number: 21 CFR 870.1025 Regulation Name: Arrythmia Detector and Alarm Regulatory Class: III (three) Product Code: MKJ and DPS Dated: October 15, 2001 Received: October 16, 2001 Dear Ms. Yount: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {4}------------------------------------------------ Page 2 - Ms. Tamara Yount Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4646. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely yours, Bram D. Zuckerman, M.D. Acting Director Division of Cardiovascular and Respiratory Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ ## 11.Indications for Use A013425 510(k) Number (if known): To be assigned Device Name: Philips Medical Systems, Heartstream FR2 Automated External Defibrillator (AED) with ECG Cable Indications For Use: The FR2 is intended to be used with disposable Heartstream defibrillation pads applied to a person who is experiencing the symptoms of sudden cardiac arrest (SCA): unresponsiveness and absence of breathing. When the patient is under 8 years or weighs less than 55 pounds (25 kg), the FR2 should be used with attenuated pediatric defibrillation pads. The FR2 is intended for use by personnel who have been trained in its operation. The user should be qualified by training in basic life support (BLS), advanced life support (ALS) or other physician-authorized emergency medical response. At the discretion of emergency care personnel, the M3860A FR2 with ECG display enabled can also be used with the FR2 ECG Cable to display the rhythm of a responsive or breathing patient, regardless of age. The FR2 and ECG Cable system provides a non-diagnostic display for attended patient monitoring. While connected to the FR2 ECG Cable, the FR2 evaluates the patient's ECG and disables it's shock capability. (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE) Concurrence of CDRH, Office of Device Evaluation (ODE) Or Division of Cardiovascular & Respiratory Devices 510(k) Number R013423 Prescription Use Over-The-Counter Use (Per 21 CFR 801.109)
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...