K020715 · Philips Medical Systems · MKJ · Nov 8, 2002 · Cardiovascular
Device Facts
Record ID
K020715
Device Name
PHILIPS M5066A AND M5068A
Applicant
Philips Medical Systems
Product Code
MKJ · Cardiovascular
Decision Date
Nov 8, 2002
Decision
ST
Submission Type
Abbreviated
Regulation
21 CFR 870.5310
Device Class
Class 3
Attributes
Therapeutic, Pediatric
Indications for Use
The M5066A is designed to be used on a person in sudden cardiac arrest, who is: - unresponsive when shaken, and . - not breathing normally. . If in doubt, apply the pads. If the victim is an infant or child younger than eight years old or weighs less than 55 lbs (25 kg), you should use the special infant/child pads. If the child appears older/larger, use the adult pads. Do not delay treatment to determine the child's exact age or weight. The M5066A is intended for use by people who have been specifically trained in its operation. A M5066A user should also have training in cardiopulmonary resuscitation (CPR) or another physician-authorized emergency medical response program in accordance with local and state requirements. Caution: Federal Law (USA) restricts this device to the sale by or on the order of a physician. The M5068A is designed to be used on a person who is in sudden cardiac arrest and who is: - unresponsive when shaken, and . - not breathing normally. . The defibrillator should not be used on a person who is: - responsive when shaken, or ● - breathing normally. . If you are not certain if the person is in sudden cardiac arrest, apply the defibrillator and follow its instructions. For children 8 years or older, or who weigh 55 pounds or more, use the M5068A with the adult pads that come with it. For younger children to those who weigh less than 55 pounds, the special infant/child pads should be used. When used with these pads, the M5068A delivers a lower energy appropriate for infants and small children. The M5068A is intended for use by people who have been specifically trained in its operation. The user should also have training in cardiopulmonary resuscitation (CPR) or another physician-authorized emergency response program in accordance with local and state requirements. Caution: Federal Law (USA) restricts this device to the sale by or on the order of a physician.
Device Story
Automated external defibrillator (AED) for sudden cardiac arrest (SCA) treatment. Inputs: patient ECG via electrode pads. Operation: Patient Analysis System (PAS) continuously monitors ECG; detects shockable rhythms; compensates for impedance; delivers biphasic truncated exponential waveform shock. Features: self-testing/calibration; infrared data port for event retrieval; training mode with simulated scripts. Used by trained lay responders or emergency personnel in out-of-hospital settings. Output: voice/visual prompts to responder; shock delivery. Clinical impact: facilitates rapid defibrillation; improves survival outcomes for SCA victims.
Clinical Evidence
Bench testing only. Performance verified against predefined specifications.
Indicated for treatment of sudden cardiac arrest (SCA) in unresponsive, non-breathing patients. Suitable for adults and children; infant/child pads required for patients <8 years or <55 lbs. Contraindicated for responsive or normally breathing individuals.
Regulatory Classification
Identification
An automated external defibrillator (AED) system consists of an AED and those accessories necessary for the AED to detect and interpret an electrocardiogram and deliver an electrical shock (e.g., battery, pad electrode, adapter, and hardware key for pediatric use). An AED system analyzes the patient's electrocardiogram, interprets the cardiac rhythm, and automatically delivers an electrical shock (fully automated AED), or advises the user to deliver the shock (semi-automated or shock advisory AED) to treat ventricular fibrillation or pulseless ventricular tachycardia.
Predicate Devices
Heartstream FR2 AED
Cardiac Science/SurVivaLink FirstSave AED
Katecho Adult/Defibrillator/Pacing Electrodes
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows a handwritten sequence of numbers and letters. The sequence starts with the letters 'ic', followed by the numbers '020715'. The handwriting is somewhat rough, with the characters appearing slightly uneven and the ink distribution varying in thickness.
NOV 8 2002
## 510(k) Summary
## 510(k) Number (if known): K020715
## 510(k) Summary for the Philips M5066A/M5068A
#### 1. Date Summary Prepared
November 6, 2002
#### 2. Submitter's Name and Address
Philips Medical Systems Heartstream 2301 Fifth Avenue, Suite 200 Seattle, WA 98121
#### 3. Contact Person
Tamara Yount Philips Medical Systems Heartstream Telephone: (206) 664-5000 Facsimile: (206) 664-5001
#### 4. Device Name
Proprietary Name: M5066A/M5068A Automated external defibrillator Common Name: Classification Names: Low-Energy Defibrillator
#### 5. Predicate Devices
The legally marketed device to which Philips Medical Systems. Heartstream claims equivalence for the M5066A/M5068A is the Heartstream FR2 AED. The modified device is also substantially equivalent to the Cardiac Science/SurVivaLink FirstSave AED and the Katecho Adult/Defibrillator/Pacing Electrodes.
The design and intended use of the M5066A/M5068A is substantially equivalent in safety and performance to the devices named above.
#### 6. Device Description
The M5066A/M5068A is an automated external defibrillator indicated to treat victims of sudden cardiac arrest (SCA). Features of the M5066A/M5068A
{1}------------------------------------------------
includes self-testing and self-calibration, an impedance-compensating biphasic truncated exponential therapy waveform, a multi-parameter Patient Analysis System (PAS) for determining if a shock is required and integrated human factors designs to facilitate use by lay responders.
A non-rechargeable lithium manganese dioxide battery powers the M5066A/M5068A with a minimum capacity of 40 shocks and 30 minutes of operating time.
Except for specific programmed periods when a responder needs to deliver uninterrupted CPR, the M5066A/M5068A continuously and automatically analyzes the patient's ECG and alerts the responder when the ECG changes to a possible shockable rhythm. Analysis continues even after the M5066A/M5068A advises a shock and arms - if the ECG spontaneously converts to a non-shockable rhythm prior to a responder pressing the shock button, the M5066A/M5068A disarms.
If significant artifact is detected in the ECG, the PAS suspends further analysis until reliable data is available. When a shockable rhythm is detected, the M5066A/M5068A directs the responder to press the shock button to deliver a biphasic shock to the patient.
The M5066A/M5068A has an infrared communication port to facilitate data management. Event and incident data is recorded during use that can be retrieved later for viewing, printing, annotating and forwarding. Factory settings of the M5066A/M5068A can be customized. In addition, the M5066A/M5068A converts into a training device with selectable training "scripts" that simulate different SCA scenarios.
#### 7. Intended Use
The M5066A is designed to be used on a person in sudden cardiac arrest, who is:
- unresponsive when shaken, and .
- not breathing normally. .
If in doubt, apply the pads.
If the victim is an infant or child younger than eight years old or weighs less than 55 lbs (25 kg), you should use the special infant/child pads. If the child appears older/larger, use the adult pads. Do not delay treatment to determine the child's exact age or weight.
The M5066A is intended for use by people who have been specifically trained in its operation. A M5066A user should also have training in cardiopulmonary
{2}------------------------------------------------
resuscitation (CPR) or another physician-authorized emergency medical response program in accordance with local and state requirements.
The M5068A is designed to be used on a person who is in sudden cardiac arrest and who is:
- unresponsive when shaken, and .
- not breathing normally. .
The defibrillator should not be used on a person who is:
- responsive when shaken, or ●
- breathing normally. .
If you are not certain if the person is in sudden cardiac arrest, apply the defibrillator and follow its instructions.
For children 8 years or older, or who weigh 55 pounds or more, use the M5068A with the adult pads that come with it. For younger children to those who weigh less than 55 pounds, the special infant/child pads should be used. When used with these pads, the M5068A delivers a lower energy appropriate for infants and small children.
The M5068A is intended for use by people who have been specifically trained in its operation. The user should also have training in cardiopulmonary resuscitation (CPR) or another physician-authorized emergency response program in accordance with local and state requirements.
#### 8. Comparison of Technology Characteristics
The M5066A/M5068A employs the same fundamental scientific technologies as the commercially available FR2 AED. The device is also similar to the Cardiac Science/SurVivaLink FirstSave AED and Katecho Adult Electrodes.
#### 9. Data Used in Determination of Substantial Equivalence
The M5066A/M5068A employs the similar technologies to provide similar performance characteristics as the predicate devices.
Testing demonstrated that the performance of the M5066A/M5068A meets its specifications. Furthermore, additional bench testing will demonstrate that the M5066A/M5068A will meet its predefined criteria.
#### 10. Conclusion
The introduction of the M5066A/M5068A does not present new issues of safety or effectiveness.
{3}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image is a black and white seal for the Department of Health & Human Services - USA. The seal is circular with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the seal is a stylized image of three human profiles facing right, with flowing lines representing hair or clothing.
### Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Philips Medical Systems c/o Ms. Tamara Yount 2401 Fourth Avenue, Suite 500 Seattle. WA 98121-1436
8 2002 NOV
Re: K020715
Philips M5066A/M5068A Regulation Number: 870.1025 Regulation Name: Arrythmia Detector and Alarm Regulatory Class: III (three) Product Code: MKJ Dated: September 3, 2002 Received: September 5, 2002
Dear Ms. Yount:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
· If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
### Page 2 - Ms. Tamara Yount
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must compliv with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4646. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
signature
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## 12.Indications for Use
November 1, 2002 -
510(k) Number (if known): #K020715
Device Name: Philips M5066A/M5068A
Indications for Use: The M5066A is designed to be used on a person in sudden cardiac arrest, who is:
- unresponsive when shaken, and
- not breathing normally. .
If in doubt, apply the pads.
If the victim is an infant or child younger than eight years old or weighs less than 55 lbs (25 kg), you should use the special infant/child pads. If the child appears older/larger, use the adult pads. Do not delay treatment to determine the child's exact age or weight.
The M5066A is intended for use by people who have been specifically trained in its operation. A M5066A user should also have training in cardiopulmonary resuscitation (CPR) or another physician-authorized emergency medical response program in accordance with local and state requirements.
Caution: Federal Law (USA) restricts this device to the sale by or on the order of a physician.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Division of Cardiovascular & Respiratory Devices
510(k) Number K020715
Prescription Use ***_*****_ or Over-The-Counter Use**
{6}------------------------------------------------
# Indications for Use (continued)
November 1, 2002
510(k) Number (if known): #K020715
Device Name: Philips M5066A/M5068A
Indications for Use:. The M5068A is designed to be used on a person who is in sudden cardiac arrest and who is:
- unresponsive when shaken, and
- not breathing normally. .
The defibrillator should not be used on a person who is:
- responsive when shaken, or ●
- breathing normally.
If you are not certain if the person is in sudden cardiac arrest, apply the defibrillator and follow its instructions.
For children 8 years or older, or who weigh 55 pounds or more, use the M5068A with the adult pads that come with it. For younger children to those who weigh less than 55 pounds, the special infant/child pads should be used. When used with these pads, the M5068A delivers a lower energy appropriate for infants and small children.
The M5068A is intended for use by people who have been specifically trained in its operation. The user should also have training in cardiopulmonary resuscitation (CPR) or another physician-authorized emergency response program in accordance with local and state requirements.
Caution: Federal Law (USA) restricts this device to the sale by or on the order of a physician.
## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Division of Cardiovascular & Respiratory Devices
510(k) Number IC926 7/5
Prescription Use
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.