K031187 · Philips Medical Systems · MKJ · Aug 27, 2003 · Cardiovascular
Device Facts
Record ID
K031187
Device Name
HEARTSTART MRX, MODEL M3535A
Applicant
Philips Medical Systems
Product Code
MKJ · Cardiovascular
Decision Date
Aug 27, 2003
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 870.5310
Device Class
Class 3
Attributes
Therapeutic, Pediatric
Indications for Use
The HeartStart MRx is for use for the termination of ventricular tachycardia and ventricular fibrillation. The device is for use by qualified medical personnel trained in the operation of the device and qualified by training in basic life support, advanced cardiac support, or defibrillation. It must be used by or on the order of a physician. The SMART Biphasic waveform utilized in the Heartstream XL Defibrillator/Monitor has previously undergone clinical testing in adults. These trials support the waveform's effectiveness for defibrillation of ventricular tachyarrhythmias at 150J. There are currently no clinical studies related to the use of SMART Biphasic waveform in pediatric applications. Manual Defibrillation: Asynchronous defibrillation is the initial treatment for ventricular fibrillation and ventricular tachycardia in patients that are pulseless and unresponsive. Synchronized defibrillation is indicated for termination of atrial fibrillation. AED Therapy: To be used in the presence of a suspected cardiac arrest on patients of at least 8 years of age that are unresponsive, not breathing and pulseless. Non-Invasive External Pacing: The pacing option is intended for treating patients with symptomatic bradycardia. It can also be helpful in patients with asystole, if performed early. 12-Lead ECG: The 12-Lead ECG option is to be used where the clinician decides to evaluate the electrocardiogram of adult and pediatric patients as part of decisions regarding possible diagnosis, potential treatment, effectiveness of treatment or to rule out causes for symptoms. Non-Invasive Blood Pressure: The NIBP option is intended for noninvasive measurement of arterial blood pressure for adult and pediatric patients. Endtidal CO2: The EtCO2 option is intended for noninvasive monitoring of exhaled carbon dioxide and to provide a respiration rate for adult, pediatric and neonatal patients. Pulse Oximetry: The SpO2 option is intended for use when it is beneficial to assess the oxygen saturation level for adult and pediatric patients. The HeartStart MRx is a fully featured, external defibrillator/monitor intended for use by qualified medical personnel, trained in either Advanced Cardiac Life Support or Basic Life Support, in a hospital or pre-hospital environment.
Device Story
Lightweight, portable external defibrillator/monitor; operates in manual (ACLS clinicians) or semi-automatic AED mode (BLS providers). Inputs: ECG signals (3/5/12-lead), impedance, NIBP, EtCO2, SpO2. Operation: Impedance-compensating biphasic truncated exponential waveform for defibrillation; rhythm analysis for shock advisory in AED mode. Outputs: Defibrillation shocks, voice/text prompts, ECG reports, vital sign measurements. Used in hospital/pre-hospital settings. Clinicians use display/printed reports for diagnosis, treatment, and monitoring. Benefits: Rapid cardiac rhythm management and patient monitoring.
Clinical Evidence
Bench testing only. Hardware and software testing performed to demonstrate device meets specifications. No clinical data provided for the subject device.
Technological Characteristics
Portable defibrillator/monitor; impedance-compensating biphasic truncated exponential waveform. Features: 3/5/12-lead ECG, NIBP, EtCO2, SpO2, non-invasive pacing. Power: Lithium Ion battery, AC/DC modules. Connectivity: PCMCIA data card for event capture/transfer. User interface: Display, voice prompts, strip chart printer.
Indications for Use
Indicated for termination of ventricular tachycardia/fibrillation in adults and pediatric patients (AED mode ≥8 years). Supports non-invasive pacing for symptomatic bradycardia/asystole; 12-lead ECG, NIBP, EtCO2, and SpO2 monitoring for adult, pediatric, and neonatal patients (EtCO2 only). Contraindications: No clinical studies for SMART Biphasic waveform in pediatric applications.
Regulatory Classification
Identification
An automated external defibrillator (AED) system consists of an AED and those accessories necessary for the AED to detect and interpret an electrocardiogram and deliver an electrical shock (e.g., battery, pad electrode, adapter, and hardware key for pediatric use). An AED system analyzes the patient's electrocardiogram, interprets the cardiac rhythm, and automatically delivers an electrical shock (fully automated AED), or advises the user to deliver the shock (semi-automated or shock advisory AED) to treat ventricular fibrillation or pulseless ventricular tachycardia.
Predicate Devices
Philips Medical Systems (formally Agilent Technologies) Heartstream XL Defibrillator/Monitor
ZOLL Medical Corp. M Series Defibrillator
Submission Summary (Full Text)
{0}------------------------------------------------
#### Ro31187 510(k) Summary
# Date Summary Prepared
April 4, 2003
# Submitter's Name and Address
Philips Medical Systems Cardiac and Monitoring Systems Cardiac Resuscitation 3000 Minuteman Road Andover, MA 01810-1099
# Contact Person
Peter Ohanian Philips Medical Systems Cardiac and Monitoring Systems Cardiac Resuscitation Telephone: (978) 659-3397 Facsimile: (978) 659-7360
# Device Name
| Proprietary Name: | HeartStart MRx |
|-----------------------|-------------------------------------------------------------|
| Common Name: | Defibrillator/Monitor |
| Classification Names: | Low-Energy Defibrillator, Arrhythmia Detection & Alarms |
| | (Automatic External Defibrillator), External Transcutaneous |
| | Pacemaker (noninvasive), ECG, Non-Invasive Blood |
| | Pressure, End Tidal Carbon Dioxide, Pulse Oximeter, and |
| | Cardiac Monitor. |
# Predicate Devices
The legally marketed devices to which Philips Medical Systems claims equivalence for the HeartStart MRx are as follows:
- Philips Medical Systems (formally Agilent Technologies) Heartstream XL . Defibrillator/Monitor, and
- . ZOLL Medical Corp. M Series Defibrillator
The design of the HeartStart MRx is substantially equivalent in safety and performance to the devices listed above.
# Device Description
The HeartStart MRx is a lightweight, portable external defibrillator, offering two modes of operation for defibrillation: manual mode and semi-automatic mode (AED).
In manual mode, the HeartStart MRx is a full-featured manual defibrillator, designed for use by clinicians trained in Advanced Cardiac Life Support (ACLS). Manual operation
{1}------------------------------------------------
allows users to select energy levels for external and internal defibrillation, perform synchronized cardioversion and provide non-invasive external pacing.
In AED mode, the HeartStart MRx, allows the provider who is trained in Basic Life Support (BLS) to provide defibrillation therapy. The device analyzes a patient's rhythm and advises the user to provide a shock. Voice prompts guide the user through the defibrillation process by providing instructions and patient information. The voice prompts are reinforced by messages that appear on the display.
In both modes of operation, the HeartStart MRx utilizes impedance compensating biphasic truncated exponential therapy waveform.
The HeartStart MRx can also be used for ECG monitoring of a patient using either 3 or 5 lead cables.
Additionally, the HeartStart MRx is offered with the following optional functionality:
#### Non-Invasive External Pacing:
The pacing option is intended for treating patients with symptomatic bradycardia. This parameter is used by ACLS trained clinicians typically performed in a hospital environment.
#### 12-Lead ECG:
The 12-Lead ECG option is intended to provide a conventional diagnostic 12-Lead ECG report, which may include measurements and interpretative statements. This parameter is used in both the hospital and pre-hospital environment by ACLS and BLS trained clinicians.
#### Non-Invasive Blood Pressure:
The NIBP option is intended for noninvasive measurement of a patient's arterial blood pressure. This parameter is used in both the hospital and pre-hospital environment by ACLS and BLS trained clinicians.
# Endtidal CO2:
The EtCO2 option is intended for noninvasive monitoring of a patient's exhaled carbon dioxide and also provides a respiration rate. This parameter is used by ACLS trained clinicians and performed in both the pre-hospital and hospital environments.
#### Pulse Oximetry:
The SpO2 option is intended for use when it is beneficial to assess a patient's arterial oxygen saturation level. This parameter is used by trained clinicians and performed in both the pre-hospital and hospital environments.
# Features
- ECG monitoring through pads or separate monitoring electrodes .
- Alarms on Heart Rate Limits and shockable rhythms ●
- Built-in strip chart printer ●
- Display for viewing waveforms and messages ●
{2}------------------------------------------------
- Automated self test with indicator .
- Internally stored event summary which may be printed .
- Voice prompts in AED mode .
- Adiustable ECG size .
- Adjustable volume control .
- Setup mode, automatic self tests and error handling .
- . Lithium Ion battery
- Internal Defibrillation .
- External Paddles with patient contact indicator .
- 3, 5. and 12 Lead ECG cables .
- Battery Charging Kit .
- PCMCIA Data card for data and event capture .
- Data recording, management, and transfer .
- Event Review .
- AC Power Module .
- DC Power Module .
# Intended Use
The HeartStart MRx is for use for the termination of ventricular tachycardia and ventricular fibrillation.
The device is for use by qualified medical personnel trained in the operation of the device and qualified by training in basic life support, advanced cardiac support, or defibrillation. It must be used by or on the order of a physician.
The SMART Biphasic waveform utilized in the Heartstream XL Defibrillator/Monitor has previously undergone clinical testing in adults. These trials support the waveform's effectiveness for defibrillation of ventricular tachyarrhythmias at 150J. There are currently no clinical studies related to the use of SMART Biphasic waveform in pediatric applications.
Manual Defibrillation: Asynchronous defibrillation is the initial treatment for ventricular fibrillation and ventricular tachycardia in patients that are pulseless and unresponsive. Synchronized defibrillation is indicated for termination of atrial fibrillation.
AED Therapy: To be used in the presence of a suspected cardiac arrest on patients of at least 8 years of age that are unresponsive, not breathing and pulseless.
# Non-Invasive External Pacing:
The pacing option is intended for treating patients with symptomatic bradycardia. It can also be helpful in patients with asystole, if performed early.
# 12-Lead ECG:
The 12-Lead ECG option is to be used where the clinician decides to evaluate the electrocardiogram of adult and pediatric patients as part of decisions regarding possible diagnosis, potential treatment, effectiveness of treatment or to rule out causes for symptoms.
# Non-Invasive Blood Pressure:
The NIBP option is intended for noninvasive measurement of arterial blood pressure for adult and pediatric patients.
{3}------------------------------------------------
# Endtidal CO2:
The EtCO2 option is intended for noninvasive monitoring of exhaled carbon dioxide and to provide a respiration rate for adult, pediatric and neonatal patients.
# Pulse Oximetry:
The SpO2 option is intended for use when it is beneficial to assess the oxygen saturation level for adult and pediatric patients.
The HeartStart MRx is a fully featured, external defibrillator/monitor intended for use by qualified medical personnel, trained in either Advanced Cardiac Life Support or Basic Life Support, in a hospital or pre-hospital environment.
# Comparison of Technology Characteristics
The HeartStart MRx employs the same fundamental scientific technologies as the commercially available predicate devices used for comparison. The HeartStart MRx acquires and analyzes ECG signals, utilizes the same shock advisory criteria, and advises the user to deliver a shock when required utilizing voice prompts as in the Heartstream XL. Heart rate alarms, noninvasive pacing and pulse oximetry functions are provided, as in the Heartstream XL. The HeartStart MRx's 12-Lead ECG, NIBP, and EtCO2 technologies are substantially equivalent to the ZOLL M Series defibrillator.
# Tests Used in Determination of Substantial Equivalence
The tests used in the determination of substantial equivalence included only bench testing. Bench testing includes hardware and software testing demonstrating that the performance of the device meets its specification
# Conclusion from Testing
Based on the results of the testing described above, it is concluded that the HeartStart MRx does not raise any different questions regarding the safety or effectiveness as compared with the predicate devices. It is considered to be substantially equivalent to the predicate devices.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States of America. The seal is circular and features the department's name around the perimeter. In the center of the seal is a stylized caduceus, a symbol often associated with healthcare, consisting of a staff with two snakes coiled around it and wings at the top.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
AUG 2 7 2003
Phillips Medical Systems c/o Mr. Peter Ohanian Director, Quality and Regulatory Affairs 3000 Minuteman Road, MS 0222 Andover. MA 01810
> K031187 Heartstart MRX, Model M3535A Regulation Number: 870.1025, 870.5300, 870.5550 Regulation Name: Arrhythmia Detector and Alarm, DC-defribrillator, External Transcutaneous cardiac pacemaker Regulatory Class: Class III Product Code: MKJ, LDD, DRO Dated: July 24, 2003 Received: July 29, 2003
Dear Mr. Ohanian:
Re:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Page 2 - Mr. Peter Ohanian
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4646. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
{6}------------------------------------------------
#### Indications for Use 03/187
510(k) Number (if known): K031187
Device Name: Philips Medical Systems, HeartStart MRx
Indications for Use: The HeartStart MRx is for use for the termination of ventricular tachycardia and ventricular fibrillation.
The device is for use by qualified medical personnel trained in the operation of the device and qualified by training in basic life support, advanced cardiac support, or defibrillation. It must be used by or on the order of a physician.
The SMART Biphasic waveform utilized in the Heartstream XL Defibrillator/Monitor has previously undergone clinical testing in adults. These trials support the waveform's effectiveness for defibrillation of ventricular tachyarmythmias at 150J. There are currently no clinical studies related to the use of SMART Biphasic waveform in pediatric applications.
Manual Defibrillation: Asynchronous defibrillation is the initial treatment for ventricular fibrillation and ventricular tachycardia in patients that are pulseless and unresponsive. Synchronized defibrillation is indicated for termination of atrial fibrillation.
AED Therapy: To be used in the presence of a suspected cardiac arrest on patients of at least 8 years of age that are unresponsive, not breathing and pulseless.
#### Non-Invasive External Pacing:
The pacing option is intended for treating patients with symptomatic bradycardia. It can also be helpful in patients with asystole, if performed early.
#### 12-Lead ECG:
The 12-Lead ECG option is to be used where the clinician decides to evaluate the electrocardiogram of adult and pediatric patients as part of decisions regarding possible diagnosis, potential treatment, effectiveness of treatment or to rule out causes for symptoms.
#### Non-Invasive Blood Pressure:
The NIBP option is intended for noninvasive measurement of arterial blood pressure for adult and pediatric patients.
#### Endtidal CO2:
The EtCO2 option is intended for noninvasive monitoring of exhaled carbon dioxide and to provide a respiration rate for adult, pediatric and neonatal patients.
#### Pulse Oximetry:
The SpO2 option is intended for use when it is beneficial to assess the oxygen saturation level for adult and pediatric patients.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| Prescription Use | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>A ND/OR -------------------------------------------------------------------------------------------------------------------------------------------------------------------- |
|------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| or | Over-The-Counter Use |
|---------------------|----------------------|
| (Division Sign-Off) | |
Philips Medical Systems
| 510(k) Number | KC3 |
|---------------|-----|
|---------------|-----|
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.