The VitalFlow Set with Balance Biosurface is indicated for respiratory/cardiopulmonary support up to 48 hours that provides assisted extracorporeal circulation and physiologic gas exchange (oxygenation) of the patient's blood in adults with acute respiratory failure or acute cardiopulmonary failure, where other available treatment options have failed, and continued clinical deterioration is expected or the risk of death is imminent.
Device Story
VitalFlow Set with Balance Biosurface is a preassembled extracorporeal circuit kit for ECMO procedures. It includes tubing assemblies, a venous reservoir, a one-way valve, the Nautilus VF Oxygenator, and the VitalFlow Centrifugal Pump. The device is used in clinical settings to provide assisted circulation and gas exchange for patients in acute respiratory or cardiopulmonary failure. The circuit is primed and connected to the patient by clinical staff. The system functions as a mechanical support device; it does not utilize software or algorithms for operation. It benefits patients by providing temporary life support when other treatments have failed. The device is single-use and sterilized via ethylene oxide.
Clinical Evidence
Clinical evidence consisted of a retrospective analysis of 195 adult patients from the ELSO Registry, representing over 39,000 hours of ECMO time. The average ECMO duration was 201.8 hours. The study compared complication rates against a large cohort of non-subject ECMO systems (n=61,176). Results showed an overall complication rate of 38.5% for the VitalFlow Set versus 35.1% for the comparator group. Specific mechanical complication rates (e.g., oxygenator failure, hemolysis, thrombosis) were generally comparable or lower in the VitalFlow group, demonstrating acceptable clinical performance.
Technological Characteristics
The device is an extracorporeal circuit kit comprising tubing, a centrifugal pump, and an oxygenator. Materials are biocompatible per ISO 10993-1. The system is designed for single-use, ethylene oxide sterilization, and has a 2-year shelf life. It operates via mechanical extracorporeal circulation and gas exchange. No software or electronic algorithms are involved in the device's function.
Indications for Use
Indicated for adults with acute respiratory or cardiopulmonary failure requiring extracorporeal circulation and gas exchange for up to 48 hours, where other treatments failed and death risk is imminent.
Regulatory Classification
Identification
An extracorporeal circuit and accessories for long-term respiratory/cardiopulmonary support (>6 hours) is a system of devices and accessories that provides assisted extracorporeal circulation and physiologic gas exchange of the patient's blood in patients with acute respiratory failure or acute cardiopulmonary failure, where other available treatment options have failed, and continued clinical deterioration is expected or the risk of death is imminent. The main devices and accessories of the system include, but are not limited to, the console (hardware), software, and disposables, including, but not limited to, an oxygenator, blood pump, heat exchanger, cannulae, tubing, filters, and other accessories (e.g., monitors, detectors, sensors, connectors).
Special Controls
*Classification* —Class II (special controls). The special controls for this device are:(1) The technological characteristics of the device must ensure that the geometry and design parameters are consistent with the intended use, and that the devices and accessories in the circuit are compatible;
(2) The devices and accessories in the circuit must be demonstrated to be biocompatible;
(3) Sterility and shelf-life testing must demonstrate the sterility of any patient-contacting devices and accessories in the circuit and the shelf life of these devices and accessories;
(4) Non-clinical performance evaluation of the devices and accessories in the circuit must demonstrate substantial equivalence of the performance characteristics on the bench, mechanical integrity, electromagnetic compatibility (where applicable), software, durability, and reliability;
(5) In vivo evaluation of the devices and accessories in the circuit must demonstrate their performance over the intended duration of use, including a detailed summary of the clinical evaluation pertinent to the use of the devices and accessories to demonstrate their effectiveness if a specific indication (patient population and/or condition) is identified; and
(6) Labeling must include a detailed summary of the non-clinical and in vivo evaluations pertinent to use of the devices and accessories in the circuit and adequate instructions with respect to anticoagulation, circuit setup, performance characteristics with respect to compatibility among different devices and accessories in the circuit, and maintenance during a procedure.
K023132 — JOSTRA MECC SYSTEM · Jostra AG · Dec 17, 2002
K191407 — Novalung System · Fresenius Medical Care Renal Therapies Group, LLC · Feb 21, 2020
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a symbol representing the Department of Health & Human Services - USA. To the right of this symbol, there is a blue square with the letters "FDA" in white. Next to the blue square, the words "U.S. FOOD & DRUG ADMINISTRATION" are written in blue.
September 12, 2024
Medtronic Inc. Anna Wetherille Senior Regulatory Affairs Specialist 8200 Coral Sea St. NE Mounds View, Minnesota 55112
Re: K240880
Trade/Device Name: VitalFlow™ Set with Balance™ Biosurface Regulation Number: 21 CFR 870.4100 Regulation Name: Extracorporeal Circuit And Accessories For Long-Term Respiratory/Cardiopulmonary Failure Regulatory Class: Class II Product Code: QJZ, QNR, BYS, QWF Dated: July 30, 2024 Received: July 30, 2024
Dear Anna Wetherille:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}------------------------------------------------
Sincerely,
# Nicole M. Gillette -S
Nicole Gillette Assistant Director DHT2B: Division of Circulatory Support, Structural, and Vascular Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K240880
Device Name VitalFlow™ Set with Balance™ Biosurface
#### Indications for Use (Describe)
The VitalFlow Set with Balance Biosurface is indicated for respiratory/cardiopulmonary support up to 48 hours that provides assisted extracorporeal circulation and physiologic gas exchange (oxygenation) of the patient's blood in adults with acute respiratory failure or acute cardiopulmonary failure, where other available treatment options have failed, and continued clinical deterioration is expected or the risk of death is imminent.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------|
| <span style="text-decoration: overline;">X</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="text-decoration: overline;"> </span> Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary of Safety and Effectiveness
| Date Prepared: | March 29, 2024 |
|----------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Applicant: | Medtronic, Inc.<br>Medtronic Perfusion Systems<br>7611 Northland Drive<br>Minneapolis, MN 55428<br>Establish Registration Number: 2184009 |
| Contact Person: | Anna Wetherille<br>Senior Regulatory Affairs Specialist<br>Phone: (763) 514-9842<br>Email: anna.wetherille@medtronic.com |
| | Ryan Calabrese (Alternate)<br>Vice President, Regulatory Affairs<br>Phone: (763) 234-3574<br>Email: ryan.s.calabrese@medtronic.com |
| Trade Name: | VitalFlow™ Set with Balance™ Biosurface |
| Common Name: | Extracorporeal Life Support Circuit and Accessories |
| Classification Name: | Extracorporeal circuit and accessories for long-term<br>respiratory/cardiopulmonary failure |
| Classification: | Class II (with special controls) |
| Regulation Number: | 21 CFR 870.4100 |
| Product Codes: | QJZ, QNR, BYS, QWF |
| Name of Predicate Device: | Novalung System (K191407) |
| Name of Reference Devices: | Tubing and Accessories Sets for Extracorporeal Membrane<br>Oxygenation (ECMO) with Balance™ Biosurface (K223168)<br>Nautilus VF ECMO Oxygenator (K232767)<br>VitalFlow Centrifugal Pump (K223898)<br>Tubing Pack (K171979) |
{5}------------------------------------------------
#### Device Description:
VitalFlow Sets with Balance™ Biosurface contain components used to prepare an extracorporeal circuit for ECMO procedures. The components include the VitalFlow Tubing Sets packaged together with the legally marketed Nautilus VF Oxygenator and VitalFlow Centrifugal Pump as a standard kit. The components are connected together to prepare an extracorporeal circuit for ECMO procedures.
The VitalFlow Tubing Set contains a preassembled drainage and return loop "ECMO Circuit," tubing assemblies, and other components used to prepare a basic extracorporeal circuit. The tubing assembly is provided pre-connected to the inlet of the VitalFlow Centrifugal Pump. The "Priming Circuit" contains components to supplement the basic extracorporeal circuit, as needed per hospital protocols for setting up the ECMO circuit. A venous reservoir and one-way valve are included in the priming circuit to facilitate ease of priming the ECMO circuit.
This product is nonpyrogenic, is intended for single use, and has been sterilized using ethylene oxide.
#### Indications for Use:
The VitalFlow Set with Balance Biosurface is indicated for respiratory/cardiopulmonary support up to 48 hours that provides assisted extracorporeal circulation and physiologic gas exchange of the patient's blood in adults with acute respiratory failure or acute cardiopulmonary failure, where other available treatment options have failed, and continued clinical deterioration is expected or the risk of death is imminent.
#### Substantial Equivalence:
Substantial equivalence analysis includes both comparison to the predicate device and clinically relevant reference devices. The VitalFlow Sets design, principles of operation, and fundamental scientific technology are substantially equivalent to the primary predicate device. Novalung System.
#### Comparison to Predicate:
A comparison of the VitalFlow™ Set with Balance™ Biosurface to the predicate device indicates the following similarities:
- Same intended use
- Similar technological characteristics ●
- Same operating principle
- Similar design features ●
- Similar materials ●
{6}------------------------------------------------
#### Comparison to Reference Devices:
The components of the VitalFlow™ Set with Balance™ Biosurface are substantially equivalent with respect to the following technological characteristics of the reference devices:
- Same intended use
- Same technological characteristics .
- Same operating principle ●
- Same design features ●
- Same materials ●
- Same shelf-life, the subject device has a 2-year shelf life ●
The VitalFlow™ Set with Balance™ Biosurface device meets all special controls identified in 21 CFR 870.4100, as follows:
- Technological Characteristics: The geometry and design parameters of the subject ● device are consistent with the device's intended use in extracorporeal support procedures, and the device is compatible with the other devices and accessories in the extracorporeal circuit.
- Biocompatibility: The subject device is demonstrated to be biocompatible in ● accordance with ISO 10993-1.
- o Sterility and Shelf-life: Sterilization adoption evaluation and shelf-life assessment demonstrate that the subject device maintains its sterility, integrity, durability, and reliability over the stated shelf life of the device.
- Non-clinical Performance: Substantial equivalence of the performance ● characteristics is demonstrated on bench, mechanical integrity, durability, and reliability testing.
- Clinical Evidence of Performance: A summary of real-world evidence of the ● clinical experience with this device was prepared by ELSO and presented which demonstrates acceptable performance of the device in ECMO patients.
- Labeling: The Instructions for Use include a detailed summary of the non-clinical ● evaluations pertinent to use of the device in an extracorporeal circuit and adequate instructions with respect to anticoagulation, circuit setup, performance characteristics with respect to compatibility among different devices and accessories in the circuit, and maintenance during a procedure.
## Summary of Performance Data
Pre-clinical bench testing was used to verify the performance characteristics of this device.
The following preclinical bench studies were conducted:
- Simulated use durability testing ●
- Tensile strength after life conditioning (long term use) ●
- Pressure test after life conditioning (long term use) ●
- Functional testing ●
{7}------------------------------------------------
- Kink testing ●
- Blood trauma testing ●
- Coating coverage ●
#### Clinical Summary:
A summary of real-world evidence (195 reports) of the clinical experience with the VitalFlow Set from the ELSO Registry was presented which demonstrated acceptable long-term performance of the subject device in ECMO patients. The performance of the device was shown to be excellent with the data representing over 39,000 hours of ECMO time with this device, with the ECMO duration averaging 201.8 hours in the adult group.
Complications on ECMO (adult): The overall complication rate in the adult ECMO population was 38.5% (75/195) for the VitalFlow Set group versus 35.1% (21501/61176) in the nonsubject device group. For specific mechanical complications, the proportion of runs with an oxygenator failure was 4.6% [9/195] in the VitalFlow Set group versus 8.0% [4878/61176] in the non-subject device group, and the pump failure was 2.6% [5/195] in the VitalFlow Set group versus 0.6 [368/61176] in the non-subject device group. The proportion of runs with hemolysis was 2.1% [4/195] in the VitalFlow Set group. lower than the corresponding percentage of 5.2% [3197/611762] in the non-subject device group. The proportion of runs with thrombosis or clots in the circuit was 1.0% [2/195] in the VitalFlow Set group. lower than the corresponding percentage of 2.1% [3197/611762] in the non-subject device group.
| Complication | VitalFlow Sets<br>(n=195 patients) | All other ECMO Systems<br>(n=61176) |
|----------------------------------------|------------------------------------|-------------------------------------|
| ≥ 1 of any | 38.5% (75) | 35.1% (21501) |
| Mechanical | | |
| Oxygenator Failure (Prevalence)a | 4.6% (9)a | 8.0% (4878)a |
| Oxygenator Failure (Rate per 1000 Hrs) | 0.30 | 0.35 |
| Pump Failure (Prevalence)b | 2.6% (5)b | 0.6% (368)b |
| Pump Failure (Rate per 1000 Hrs) | 0.16 | 0.02 |
| Heat Exchanger Malfunction | 0.0% (0) | 0.0% (27) |
| Air in Circuit | 1.0% (2) | 0.8% (507) |
| Cannula Problems | 2.6% (5) | 4.6% (2825) |
| Circuit Change | 6.2% (12) | 4.6% (2896) |
| Emboli (Clots or Air) | 0.5% (1) | 0.2% (150) |
| Thrombosis/Clots in Circuit Component | 1.0% (2) | 2.1% (1272) |
| Neurologic | | |
| CNS Infarction (US/CT/MRI) | 2.6% (5) | 2.8% (1688) |
| CNS Hemorrhage (US/CT/MRI) | 3.6% (7) | 3.8% (2296) |
| CNS Diffuse Ischemia (CT/MRI) | 2.1% (4) | 1.9% (1145) |
| Hemolysis | | |
| Moderate Hemolysis | 1.0% (2) | 3.4% (2052) |
| Severe Hemolysis | 1.5% (3) | 2.6% (1605) |
| Moderate or Severe Hemolysis | 2.1% (4) | 5.2% (3197) |
Row Notes:
a) Prevalence of failures corresponding to the initial equipment
b) Rate of failures per 1000 hours corresponding to the initial equipment
{8}------------------------------------------------
#### Conclusion:
The data included in this submission are sufficient to demonstrate that the VitalFlow™ Set with Balance™ Biosurface is substantially equivalent to the predicate device, Novalung System, and the legally marketed reference devices, Tubing and Accessories Sets for Extracorporeal Membrane Oxygenation (ECMO) with Balance™ Biosurface, Nautilus VF ECMO Oxygenator, and VitalFlow Centrifugal Pump, and does not raise different questions of safety or effectiveness.
{9}------------------------------------------------
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.