CentriMag Pre-connected Pack

K222297 · Abbott · DTZ · Dec 1, 2022 · Cardiovascular

Device Facts

Record IDK222297
Device NameCentriMag Pre-connected Pack
ApplicantAbbott
Product CodeDTZ · Cardiovascular
Decision DateDec 1, 2022
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.4350
Device ClassClass 2
AttributesTherapeutic

Intended Use

The CentriMag™ Pre-connected Pack is indicated for use with the CentriMag™ Acute Circulatory Support System to provide physiologic gas exchange of the blood and to pump a patient's blood through an extracorporeal circuit for periods lasting less than six (6) hours for the purpose of providing either: (i) Full or partial cardiopulmonary bypass during open surgical procedures on the heart or great vessels; or (ii) Temporary circulatory bypass for diversion of flow around a planned disruption of the circulatory necessary for open surgical procedures on the aorta or vena cava.

Device Story

CentriMag Pre-connected Pack provides extracorporeal circulation for cardiopulmonary bypass; used with CentriMag Acute Circulatory Support System (console, motor, flow probe, monitor). System components: sterile disposable blood pump, oxygenator, priming bag, and tubing. Operates by pumping blood through extracorporeal circuit for gas exchange; used in OR by surgical team. Output: physiologic gas exchange and blood flow; assists clinical decision-making during cardiac/vascular surgery. Benefits: temporary circulatory support during planned surgical disruptions.

Clinical Evidence

Bench testing only. No clinical or animal studies required. Performance testing included gas exchange (O2/CO2 transfer), blood pathway pressure drop, in vitro hemolysis, pathway integrity, heat exchange efficiency, usable life/reliability, and packaging/sterilization validation. Biocompatibility testing performed per ISO 10993-1 for limited (≤ 24 hour) circulating blood contact, including genotoxicity, hemocompatibility, cytotoxicity, implantation, sensitization, irritation, systemic toxicity, particulate matter, and VOCs.

Technological Characteristics

Sterile disposable pack containing blood pump, oxygenator, priming bag, and tubing. Centrifugal pump principle. External communicating device with limited (≤ 24 hour) blood contact. EtO sterilization. Complies with ISO 10993-1 for biocompatibility. Designed for use with CentriMag console/motor/flow probe/monitor.

Indications for Use

Indicated for patients undergoing open surgical procedures on the heart, great vessels, aorta, or vena cava requiring full or partial cardiopulmonary bypass or temporary circulatory bypass for < 6 hours.

Regulatory Classification

Identification

A cardiopulmonary bypass oxygenator is a device used to exchange gases between blood and a gaseous environment to satisfy the gas exchange needs of a patient during open-heart surgery.

Special Controls

*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Guidance for Cardiopulmonary Bypass Oxygenators 510(k) Submissions.”

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo in blue, with the words "U.S. FOOD & DRUG" stacked on top of the word "ADMINISTRATION". December 1, 2022 Abbott Kim Bondarenko Manager, Regulatory Affairs 6035 Stoneridge Drive Pleasanton, California 94588 Re: K222297 Trade/Device Name: CentriMag Pre-connected Pack Regulation Number: 21 CFR 870.4350 Regulation Name: Cardiopulmonary bypass oxygenator Regulatory Class: Class II Product Code: DTZ, KFM, DTQ, DTR, DWA, DWF, KRI Dated: November 17, 2022 Received: November 18, 2022 Dear Kim Bondarenko: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for {1}------------------------------------------------ devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. # Nicole M. Gillette -S Nicole Gillette Assistant Director DHT2B: Division of Circulatory Support, Structural and Vascular Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K22297 Device Name CentriMag™ Pre-connected Pack ### Indications for Use (Describe) The CentriMag™ Pre-connected Pack is indicated for use with the CentriMag™ Acute Circulatory Support System to provide physiologic gas exchange of the blood and to pump a patient's blood through an extracorporeal circuit for periods lasting less than six (6) hours for the purpose of providing either: (i) Full or partial cardiopulmonary bypass during open surgical procedures on the heart or great vessels; or (ii) Temporary circulatory bypass for diversion of flow around a planned disruption of the circulatory necessary for open surgical procedures on the aorta or vena cava. | Type of Use (Select one or both, as applicable) | | |--------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------| | <div><span><span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D)</span></div> | <div><span>☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span></div> | # CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the Abbott logo, which is a blue stylized letter 'A'. Below the logo, the word "Abbott" is written in bold, followed by the text "CentriMag Pre-connected Pack" in a smaller font size. The text indicates that the image is related to Abbott's CentriMag Pre-connected Pack product. # 510(k) Summary This summary of 510(k) safety and effectiveness information is being submitted in accordance with the Safe Medical Devices Act (SMDA) of 1990 and Title 21 of the Code of Federal Regulations, Part 807 (particularly §807.92). #### l. SUBMITTER II. | Date Prepared: | July 29, 2022 | |--------------------------------|----------------------------------------------------------------------------------| | Submitter's Name & Address: | Abbott<br>6035 Stoneridge Drive<br>Pleasanton, CA 94588 | | Establishment Registration No. | 2916596 | | Contact Person: | Kim Bondarenko<br>Phone: (925) 460-7361 | | DEVICE | | | Trade or Proprietary Name: | CentriMag™ Pre-connected Pack | | Common or Usual Name: | Pre-connected blood pump and oxygenator<br>components for cardiopulmonary bypass | | Classification Name: | Cardiopulmonary bypass oxygenator | | Classification Regulation: | 21 CFR 870.4350 Cardiopulmonary bypass<br>oxygenator | | Product Code: | DTZ, KFM, DWA, DTQ, DTR, DWF, and KRI | | Regulatory Class: | Class II | {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the logo for Abbott, a healthcare company. The logo features a stylized letter "A" in blue, with a thick, rounded design. Below the symbol is the company name, "Abbott," in a bold, black font. #### III. LEGALLY MARKETED PREDICATE DEVICE # Predicate Device: - . OXY-1 System (K200109) # Reference Devices: - Rotaflow Centrifugal Pump with Softline Coating (K090515) - . Quadrox-iD Adult Oxygenator (K150267) - HLM Tubing Set with Softline Coating (K090533) - CentriMag Acute Circulatory Support System (K200306) - Eurosets PMP Sterile Oxygenator (K141492) #### DEVICE DESCRIPTION IV. CentriMag Pre-connected Pack is to be used with the CentriMag™ Acute Circulatory Support System components (console, motor, flow probe and monitor) to provide extracorporeal circulation for full or partial cardiopulmonary bypass support for less than 6 hours. The CentriMag™ Pre-connected Pack contains sterile disposable components: a blood pump, oxygenator, and priming bag pre-connected by tubing. The pack also includes disposable accessories commonly used in the setup of pump and oxygenator connections to the cardiopulmonary bypass extracorporeal circuit. #### V. INDICATIONS FOR USE The CentriMag™ Pre-connected Pack is indicated for use with the CentriMag™ Acute Circulatory Support System to provide physiologic gas exchange of the blood and to pump a patient's blood through an extracorporeal circuit for periods lasting less than 6 hours for the purpose of providing either: (i) Full or partial cardiopulmonary bypass during open surgical procedures on the heart or great vessels; or (ii) Temporary circulatory bypass for diversion of flow around a planned disruption of the circulatory pathway necessary for open surgical procedures on the aorta or vena cava. #### VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE Substantial equivalence of the CentriMag Pre-connected Pack to be used with the CentriMag Acute Circulatory Support System was demonstrated through technological characteristics with the identified predicate device; performance testing, and comparative testing using reference devices. CentriMag Pre-connected Pack and Abiomed's OXY-1 System (K200109) have the same intended use, clinical setting, target user, tubing connections and principle of operation. The minor technological differences do {5}------------------------------------------------ Image /page/5/Picture/0 description: The image shows the logo for Abbott Laboratories. The logo is a stylized letter 'A' in blue. The 'A' is formed by thick, rounded lines, giving it a modern and clean appearance. Below the logo, the company name 'Abbott' is written in a bold, sans-serif font. # CentriMag Pre-connected Pack not alter the intended therapeutic use of the device, nor do they affect the safety and effectiveness of the device relative to the predicates. - VII. PERFORMANCE DATA Performance tests were conducted on the CentriMag Pre-connected Pack to support the determination of substantial equivalence and demonstrate the integrity, durability of the devices over the intended shelf life. # Performance Testing - The following testing was conducted as recommended in "Guidance for Cardiopulmonary Bypass Oxygenators 510(k) Submissions Final Guidance for Industry and FDA Staff" dated November 13, 2000: - Gas Exchange (oxygen and carbon dioxide transfer) O - O Blood Pathway Pressure drop - In vitro Hemolysis O - Blood, Gas, and Water Pathway Integrity O - Heat Exchange Efficiency and Water Pathway Pressure Drop O - Usable Life (Reliability) Testing - . Functionality and Integrity Testing of Accessories - Shelf Life Testing - Validation of EtO Sterilization - Packaging Testing No animal studies or clinical studies are required. # Biocompatibility Testing Biocompatibility testing of the CentriMag Pre-connected Pack was conducted in accordance with FDA biocompatibility guidance Use of International Standard ISO 10993-1 Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process per ISO 10993-1. The results meet the requirements for external communicating medical device with limited (≤ 24 hour) circulating blood contact. - Genotoxicity - . Hemocompatibility - Cytotoxicity {6}------------------------------------------------ Image /page/6/Picture/0 description: The image shows the logo for Abbott Laboratories. The logo consists of a stylized lowercase letter 'a' in a light blue color. Below the symbol, there is text, but it is not clear enough to read. # Abbo # CentriMag Pre-connected Pack - Implantation - Sensitization, Irritation - Systemic Toxicity - Particulate Matter - . Volatile Organic Compounds #### CONCLUSION VIII. CentriMag Pre-connected Pack to be used with CentriMag Acute Circulatory Support System has the same intended use as the identified predicate device OXY-1 System. The comparison of technological characteristics between subject and predicate device, and performance testing conducted on the subject device including comparator testing with identified reference devices do not raise different questions of safety and effectiveness. The subject device and the predicate device are substantially equivalent.
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%