ADVANCED MEMBRANCE GAS EXCHANGE PMP STERILE (A.M.G PMP STERILE)
Applicant
Eurosets S.R.L
Product Code
DTZ · Cardiovascular
Decision Date
Feb 6, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.4350
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Advanced Membrane Gas Exchange PMP STERILE (A.M.G. PMP STERILE) is intended for use in adult Surgical procedures requiring extracorporeal gas exchange support and blood temperature control for periods of up to 6 hours. The Advanced Membrane Gas Exchange PMP STERILE (A.M.G. PMP STERILE) for extracorporeal circulation is a microporous hollow-fiber oxygenator with an integral heat exchanger used to perform cardiopulmonary bypass. It includes a detachable 4.5 liter blood reservoir.
Device Story
Microporous hollow-fiber oxygenator with integral heat exchanger and detachable 4.5L blood reservoir; performs cardiopulmonary bypass by substituting lung function (oxygenation/CO2 removal) and controlling blood temperature. Utilizes polymethyl pentene (PMP) fibers for improved hydrophobicity compared to polyethylene. Used in surgical settings by clinical staff. Device provides gas exchange and thermal regulation during extracorporeal circulation; supports patient physiological stability during procedures up to 6 hours.
Clinical Evidence
Bench testing only. Evaluated gas transfer, pressure drop, fiber hydrophilization, coating uniformity, plasma hemoglobin, heat exchanger efficiency, structural integrity, and biocompatibility (ISO 10993). Sterilization validated per EN ISO 10993-7. No clinical data presented.
Technological Characteristics
Microporous hollow-fiber oxygenator with integral heat exchanger; polymethyl pentene (PMP) fibers; detachable 4.5L reservoir. Sterilized via Ethylene Oxide (EtO). Complies with ISO 7199:2009 and EN 12022:1998 standards.
Indications for Use
Indicated for adult patients undergoing surgical procedures requiring extracorporeal gas exchange support and blood temperature control for up to 6 hours, including cardiopulmonary bypass.
Regulatory Classification
Identification
A cardiopulmonary bypass oxygenator is a device used to exchange gases between blood and a gaseous environment to satisfy the gas exchange needs of a patient during open-heart surgery.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Guidance for Cardiopulmonary Bypass Oxygenators 510(k) Submissions.”
Predicate Devices
A.M.G. advanced membrane gas exchange and accessories (K102109)
{0}------------------------------------------------
Public Health Service
Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized representation of three human profiles facing right, stacked on top of each other. The profiles are black and have a flowing, abstract design. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the left side of the logo.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 6, 2015
EUROSETS s.r.1. % Mario Gennari Official Correspondent, Gemar s.r.1. Via G. Puccini, 1 Medolla, Modena, IT I-43036
Re: K141492
> Trade/Device Name: Advanced Membrane Gas Exchange PMP Sterile (A.M.G. PMP Sterile) Regulation Number: 21 CFR 870.4350 Regulation Name: Cardiopulmonary Bypass Oxygenator Regulatory Class: Class II Product Code: DTZ Dated: December 4, 2014 Received: January 14, 2015
Dear Mario Gennari,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply
{1}------------------------------------------------
with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
M.A. Hillebrenner
for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K141492
Device Name
Advanced Membrane Gas Exchange PMP (A.M.G. PMP STERILE)
Indications for Use (Describe)
The Advanced Membrane Gas Exchange PMP STERILE (A.M.G. PMP STERILE) is intended for use in adult Surgical procedures requiring extracorporeal gas exchange support and blood temperature control for periods of up to 6 hours.
The Advanced Membrane Gas Exchange PMP STERILE (A.M.G. PMP STERILE) for extracorporeal circulation is a microporous hollow-fiber oxygenator with an integral heat exchanger used to perform cardiopulmonary bypass. It includes a detachable 4.5 liter blood reservoir.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------|---------------------------------------------------------------------------------|
| <div> <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(K) SUMMARY, AS REQUIRED BY 21 CFR 807.92
| Submitter's Name | Eurosets s.r.l. |
|--------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Address | Strada Statale 12, nº143<br>41036 Medolla (MO) Italy |
| Establishment<br>Registration Number | 3003752502 |
| Summary<br>Preparation Date | Feruary 05, 2015 |
| Contact Person<br>Telephone Number<br>Fax Number | Dr. Katia Vescovini, Regulatory Affairs Manager<br>0039 0535660334<br>0039 0535 51248 |
| Name of the Device | Advanced Membrane Gas Exchange PMP STERILE (A.M.G.<br>PMP STERILE) |
| Name of the<br>Oxygenating module | A.M.G. MODULE PMP STERILE<br>A.M.G. MODULE PMP NO T.P. STERILE |
| Name of the<br>Cardiotomy<br>reservoir | VCR 4500 PMP STERILE |
| Common name of<br>the device | Advanced Membrane Gas Exchange and Accessories A.M.G.<br>PMP STERILE |
| Common name of<br>the Oxygenating<br>module | A.M.G. MODULE PMP STERILE<br>A.M.G. MODULE PMP NO T.P. STERILE |
| Common name of<br>the Cardiotomy<br>reservoir | Venous Cardiotomy Reservoir 4500 PMP STERILE |
| Classification Name | Cardiopulmunary device<br>Classification Name: Advanced membrane gas exchange<br>Device Class: II<br>Product Code: DTZ<br>Regulation Number: 21 CFR §870.4350 |
| Performance<br>Standards | No performance standards applicable to Advanced Membrane<br>Gas Exchange PMP STERILE (A.M.G. PMP STERILE) have<br>been established by the FDA |
{4}------------------------------------------------
## 510(k) SUMMARY, continued
| DESCRIPTION: | The Advanced Membrane Gas Exchange and accessories PMP<br>STERILE (A.M.G. PMP STERILE) is a microporous hollow-fiber<br>oxygenator with an integral heat exchanger, used to perform<br>cardiopulmonary bypass or surgical procedures requiring<br>extracorporeal gas exchange support and blood temperature<br>control for periods of up to 6 hours. It includes a detachable 4.5<br>liter blood reservoir.<br>The A.M.G. PMP STERILE is equipped with polymethyl pentene<br>fibres, more hydrophobic than the polyethylene used for the<br>predicate device and so considered a more reliable barrier<br>between the blood pathway and gas pathway of the oxygenator.<br>The A.M.G. PMP STERILE is provided in different configurations:<br>- microporous hollow-fiber oxygenator with temperature probe<br>and pre-connected venous cardiotomy reservoir (“A.M.G. PMP<br>STERILE”),<br>- microporous hollow-fiber oxygenator with temperature probe<br>and without pre-connected venous cardiotomy reservoir<br>(“A.M.G. MODULE PMP STERILE”),<br>- microporous hollow-fiber oxygenator without temperature<br>probe and without pre-connected venous cardiotomy reservoir<br>(“A.M.G. MODULE PMP NO TP STERILE”),<br>- Venous Cardiotomy Reservoir 4500 PMP (“VCR 4500 PMP<br>STERILE”).<br>The device is used to temporarily substitute the functions of the |
|-----------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | lung as it supplies oxygen and removes carbon dioxide from the<br>blood. The device can be use with adult patients. |
| Indications for Use | The Advanced Membrane Gas Exchange PMP STERILE (A.M.G.<br>PMP STERILE) is intended for use in adult Surgical procedures<br>requiring extracorporeal gas exchange support and blood<br>temperature control for periods of up to 6 hours.<br>The Advanced Membrane Gas Exchange PMP STERILE (A.M.G.<br>PMP STERILE) for extracorporeal circulation is a microporous<br>hollow-fiber oxygenator with an integral heat exchanger used to<br>perform cardiopulmonary bypass.<br>It includes a detachable 4.5 liter blood reservoir. |
| Identification of<br>Predicate Device | Proprietary Name: AMG<br>Classification Name: 21 CFR 870.4350, Oxygenator,<br>Cardiopulmonary Bypass<br>Registered Establishment Name:<br>Eurosets srl<br>Strada Statale 12, n. 143<br>41036 Medolla (MO) Italy |
| Registered Establishment Number: | 3003752502<br>Owner/Operator: Eurosets srl<br>Establishment Operations: Manufacturer<br>510 (k): K102109<br>Device Name: A.M.G. advanced membrane gas exchange and<br>accessories<br>Applicant: Eurosets srl |
| | Please note that regarding the polymethyl pentene fibres used by<br>the A.M.G. PMP STERILE, Eurosets will refer to the following<br>predicate device: |
| | Proprietary Name: MEDOS HILITE 7000 & 7000 LT Hollow Fiber<br>Oxygenator |
| | Classification Name: 21 CFR 870.4350, Oxygenator,<br>Cardiopulmonary Bypass<br>Registered Establishment Name: GISH BIOMEDICAL, INC. |
| | Registered Establishment Number: 3002845651<br>Owner/Operator: GISH BIOMEDICAL, INC.<br>Owner/Operator Number: 3002845651<br>Establishment Operations: Manufacturer 510 (k): K082403 |
| | Device Name: 7000 LT Hollow Fiber Oxygenator (K082403) |
| | Applicant: GISH BIOMEDICAL, INC. |
| Common name of<br>the predicate device<br>for principles of<br>operatations | A.M.G. advanced membrane gas exchange and accessories |
| Classification Name<br>of the device | Oxygenator, Cardiopulmonary Bypass |
| Classification Name<br>of the Oxygenating<br>module | Device Class: II<br>Product Code: DTZ<br>Regulation Number: 21 CFR §870.4350<br>Oxygenator cardiopulmonary bypass Cardiovascular<br>Device Class: II<br>Product Code: DTZ |
| Classification Name<br>of the Cardiotomy<br>reservoir | Regulation Number: 21 CFR §870.4350<br>Hard shell Venous/cardiotomy reservoir Cardiovascular<br>Device Class: II<br>Product Code: DTZ<br>Regulation Number: 21 CFR §870.4350 |
| Applicant name and<br>address: | Eurosets s.r.l.<br>Strada Statale 12, n°143<br>41036 Medolla (MO) Italy |
| | MEDOS MEDIZINTECHNIK AG<br>Obere Steinfurt 8-10 |
| Comparison of<br>Technological<br>Characteristics | In vitro bench testing was performed to support a determination of<br>substantial equivalence (refer to performance testing below)<br>between the new Advanced Membrane Gas Exchange PMP<br>STERILE (A.M.G. PMP STERILE) and the predicate.<br>The results of these tests provide reasonable assurance that the<br>proposed device has been designed and tested to assure<br>conformance to the requirements for its intended use and<br>performs comparably to the existing predicate devices. |
| Performance<br>Testing (non-<br>clinical) | In vitro bench tests were carried out to demonstrate equivalence,<br>according to the requirements of FDAs document "Guidance for<br>Cardiopulmunary Bypass Oxygenators 510(k) Submission",<br>issued on November 13, 2000, the ISO 7199:2009 and of the EN<br>12022:1998 "Blood-gas exchangers".<br>The following areas have been tested and/or evaluated:<br>- Gas Transfer and Pressure Drop,<br>- Fiber Hydrophilization,<br>- Cracking test external body,<br>- Coating Uniformity,<br>- Plasma hemoglobin concentration,<br>- Heat exchanger Efficiency and Filling Volume,<br>- Seal Connection,<br>- Structural and Mechanical Integrity,<br>- Bioburden tests,<br>- Sterility tests,<br>- LAL-test,<br>- Validation of the EtO Sterilization process,<br>- Sterility tests,<br>- Packaging evaluation,<br>- Labelling evaluation,<br>- EtO Residual, according to EN ISO 10993-7,<br>- Biocompatibility, according to ISO 10993 series<br>requirements.<br>The results from these performance evaluations demonstrated<br>that the AMG PMP STERILE met the acceptarfce criteria defined<br>in the product specification and performed comparably to the<br>predicate device. |
| SUBSTANTIAL<br>EQUIVALENCE: | The A.M.G PMP STERILE are identical to the predicate device in<br>terms of intended use, indications for use and surgical technique.<br>Based on the safety and performance testing, technological<br>characteristics and the indications for use for the device, the<br>proposed A.M.G PMP STERILE has been demonstrated to be<br>appropriate for its intended use and is considered substantially<br>equivalent to the previously cleared A.M.G. advanced membrane<br>gas exchange and accessories (K102109). |
{5}------------------------------------------------
-
{6}------------------------------------------------
Stolberg, GERMANY 52222
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.