K223898 · Michigan Critical Care Consultants, Inc. (D.B.A Mc3 Inc.) · QNR · Aug 25, 2023 · Cardiovascular
Device Facts
Record ID
K223898
Device Name
VitalFlowTM Centrifugal Pump
Applicant
Michigan Critical Care Consultants, Inc. (D.B.A Mc3 Inc.)
Product Code
QNR · Cardiovascular
Decision Date
Aug 25, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.4100
Device Class
Class 2
Attributes
Therapeutic, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K223898 · Aug 25, 2023
VitalFlowTM Centrifugal Pump
Michigan Critical Care Consultants, Inc. (D.B.A Mc3 Inc.)
ELSO (Extracorporeal Life Support Organization) registry
The sponsor provided a summary of real-world clinical experience from the ELSO registry to demonstrate the long-term performance and safety of the device in adult ECMO patients, comparing outcomes against other pumps.
Mechanical complications, hemolysis, neurologic complications, and pump failure rates
Indications for Use
The VitalFlow™ Centrifugal Pump is intended to pump blood through the extracorporeal circuit for circulatory support up to 48 hours, in adult patients with acute respiratory failure or acute cardiopulmonary failure, where other available treatment options have failed, and continued clinical deterioration is expected or the risk of death is imminent. The VitalFlow™ Centrifugal Pump is driven by the VitalFlow™ Console and Drive Motor, or the Emergency Handcrank.
Device Story
Sterile, single-use centrifugal blood pump; moves blood through extracorporeal circuit via centrifugal force. Input: blood via inlet port; cone with impeller blades accelerates blood toward outlet. Driven by external console/drive motor or emergency handcrank via magnetic coupling. Used in clinical settings for ECMO; operated by healthcare professionals. Output: blood flow; monitored/displayed by coupled console. Benefits: provides circulatory support for patients in acute failure. Affects clinical decision-making by enabling life-sustaining extracorporeal support.
Clinical Evidence
Bench testing (14-day durability, mechanical/pressure integrity, hydraulic performance, blood trauma) and 96-hour animal study (n=13 sheep) confirmed functionality and safety. Clinical evidence included real-world data (ELSO registry) of 1,048 reports (200,000+ hours of ECMO time). Complication rates (mechanical 1.2%, hemolysis 1.1%) were comparable or superior to non-AP40 pumps.
Technological Characteristics
Centrifugal blood pump; single-use; sterile. 3/8" inlet/outlet connectors. Magnetic coupling to external drive motor. Biocompatible per ISO 10993-1:2018. Max flow 7 L/min; max RPM 3800. Shaft seal design. Compatible with extracorporeal circuit accessories.
Indications for Use
Indicated for adult patients with acute respiratory failure or acute cardiopulmonary failure requiring circulatory support up to 48 hours when other treatments fail and clinical deterioration or death is imminent.
Regulatory Classification
Identification
An extracorporeal circuit and accessories for long-term respiratory/cardiopulmonary support (>6 hours) is a system of devices and accessories that provides assisted extracorporeal circulation and physiologic gas exchange of the patient's blood in patients with acute respiratory failure or acute cardiopulmonary failure, where other available treatment options have failed, and continued clinical deterioration is expected or the risk of death is imminent. The main devices and accessories of the system include, but are not limited to, the console (hardware), software, and disposables, including, but not limited to, an oxygenator, blood pump, heat exchanger, cannulae, tubing, filters, and other accessories (e.g., monitors, detectors, sensors, connectors).
Special Controls
*Classification* —Class II (special controls). The special controls for this device are:(1) The technological characteristics of the device must ensure that the geometry and design parameters are consistent with the intended use, and that the devices and accessories in the circuit are compatible;
(2) The devices and accessories in the circuit must be demonstrated to be biocompatible;
(3) Sterility and shelf-life testing must demonstrate the sterility of any patient-contacting devices and accessories in the circuit and the shelf life of these devices and accessories;
(4) Non-clinical performance evaluation of the devices and accessories in the circuit must demonstrate substantial equivalence of the performance characteristics on the bench, mechanical integrity, electromagnetic compatibility (where applicable), software, durability, and reliability;
(5) In vivo evaluation of the devices and accessories in the circuit must demonstrate their performance over the intended duration of use, including a detailed summary of the clinical evaluation pertinent to the use of the devices and accessories to demonstrate their effectiveness if a specific indication (patient population and/or condition) is identified; and
(6) Labeling must include a detailed summary of the non-clinical and in vivo evaluations pertinent to use of the devices and accessories in the circuit and adequate instructions with respect to anticoagulation, circuit setup, performance characteristics with respect to compatibility among different devices and accessories in the circuit, and maintenance during a procedure.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in a sans-serif font.
August 25, 2023
Michigan Critical Care Consultants, Inc. (d.b.a MC3 Inc.) Martha Rumford VP Regulatory Affairs 2555 Bishop Circle West Dexter, Michigan 48130
Re: K223898
Trade/Device Name: VitalFlowTM Centrifugal Pump Regulation Number: 21 CFR 870.4100 Regulation Name: Extracorporeal Circuit And Accessories For Long-Term Respiratory/Cardiopulmonary Failure Regulatory Class: Class II Product Code: ONR Dated: August 11, 2023 Received: August 11, 2023
Dear Martha Rumford:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{2}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Nicole M. Gillette -S
Nicole Gillette Assistant Director DHT2B: Division of Circulatory Support, Structural and Vascular Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known)
Device Name VitalFlow™ Centrifugal Pump
#### Indications for Use (Describe)
The VitalFlow™ Centrifugal Pump is intended to pump blood through the extracorporeal circuit for circulatory support up to 48 hours, in adult patients with acute respiratory failure or acute cardiopulmonary failure, where other available treatment options have failed, and continued clinical deterioration is expected or the risk of death is imminent.
The VitalFlow™ Centrifugal Pump is driven by the VitalFlow™ Console and Drive Motor, or the Emergency Handcrank.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|------------------------------------------------------------------------------------------------------|
| | <span style="display:inline-block; width:20px;"></span> Prescription Use (Part 21 CFR 801 Subpart D) |
| | <span style="display:inline-block; width:20px;"></span> Over-The-Counter Use (21 CFR 801 Subpart C) |
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/3/Picture/0 description: The image contains the logo for MC3 Cardiopulmonary. The logo consists of a red abstract symbol on the left, followed by the text "MC3" in a dark blue, bold font. Below "MC3" is the word "Cardiopulmonary" in a smaller, lighter font. The overall design is clean and professional.
Date Prepared:
# 510(k) Summary VitalFlow™ Centrifugal Pump
August 11, 2023
| Sponsor Information: | |
|------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Owner/Applicant/Submitter: | Michigan Critical Care Consultants, Inc<br>(dba MC3, Inc.)<br>2555 Bishop Circle West Dexter, MI 48130<br>1-734-995-9089<br>Registration number: 3011468686 |
| Contact Person: | Martha Rumford<br>Vice President of Regulatory Affairs<br>2555 Bishop Circle West<br>Dexter, MI 48130<br>734.995.9089 |
| Device Names/Classification: | |
| Device Trade Name: | VitalFlow™ Centrifugal Pump |
| Device Common Name: | Blood Pump For ECMO, Long-Term (> 6 Hours) Use |
| Regulation Name: | Extracorporeal circuit and accessories for long-term<br>respiratory/cardiopulmonary failure |
| Regulation Number: | 21 CFR 870.4100 |
| Product Code: | QNR |
| Predicate: | TandemHeart Pump (K202751) |
| Reference Device 1: | Affinity CP Centrifugal Blood Pump BBAP40<br>(K111657) |
| Reference Device 2: | Maquet Rotaflow Centrifugal Pump (K991864) |
# Indications for Use:
The VitalFlow™ Centrifugal Pump is intended to pump blood through the extracorporeal circuit for circulatory support up to 48 hours, in adult patients with acute respiratory failure or acute cardiopulmonary failure, where other available treatment options have failed, and continued clinical deterioration is expected or the risk of death is imminent.
The VitalFlow™ Centrifugal Pump is driven by the VitalFlow™ Console and Drive Motor, or the Emergency Handcrank.
{5}------------------------------------------------
Image /page/4/Picture/0 description: The image contains the logo for MC3 Cardiopulmonary. The logo consists of a red abstract symbol on the left, followed by the text "MC3" in large, bold, dark blue letters. Below "MC3" is the word "Cardiopulmonary" in a smaller, lighter blue font. The logo is clean and professional, suggesting a medical or healthcare-related organization.
### Device Description:
The VitalFlow Centrifugal Pump is a sterile, single-use centrifugal blood pump. It is a non-invasive, nonpyrogenic device designed to move blood through the extracorporeal circuit by centrifugal force. The pump is the disposable portion of the pumping system and it is electromagnetically coupled to an instrument that monitors and displays the flow and pressure of the blood. Blood enters the inlet port of the pump, where a cone with impeller blades within the pump housing rotates and the blood is gently accelerated toward the outlet of the pump.
The VitalFlow Centrifugal Pump can be driven through magnetic coupling by an External Drive Motor or the Emergency Handcrank.
# Bench and Animal Performance Evaluations:
Design verification studies were performed by the original manufacturer of the subject device and supplemented by MC3 bench performance testing, including long-term characterization. This data is coupled with animal studies to support a finding of substantial equivalence to the Predicate with respect to the Special Controls described by 21CFR 870.4100. The following testing demonstrated the pump's duration of use:
- Bearing wear
- Coating coverage ●
- Mechanical integrity
- Pressure integrity
- . Hydraulic performance
- High flow blood trauma testing
- In-vivo performance testing ●
- Magnetic coupling
- Port and Pull strength
- Prime volume
# Substantial Equivalence:
Substantial equivalence analysis includes both comparison to the predicate device and clinically relevant Reference Devices. The VitalFlow Centrifugal Pump is the same as the Affinity TM CP Centrifugal Blood Pump with Balance™ Biosurface in design, principles of operation, materials of construction, performance, and fundamental scientific technology. The design, principles of operation, materials of construction, performance, and fundamental scientific technology are substantially equivalent to the TandemHeart Pump.
# Comparison of Technological Characteristics with the Predicate Device
The subject and predicate devices are centrifugal pumps that are designed to move blood through the extracorporeal circuit. Blood enters the inlet port of the pump, where a cone with impeller blades within the pump housing rotates and the blood is gently accelerated toward the outlet of the pump is the disposable portion of the pumping system that is electromagnetically coupled to an instrument that monitors and displays the blood flow.
At a high level, the subject and predicate devices are based on the following same technological elements:
- . Principle of operation
{6}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for MC3 Cardiopulmonary. The logo consists of a red abstract symbol on the left, followed by the text "MC3" in blue. Below "MC3" is the word "Cardiopulmonary" in a smaller font size, also in blue. The red symbol appears to be composed of three curved shapes arranged in a circular pattern.
- . Design characteristics
- . Coupled to a motor for control via a pump console
- Connection to extracorporeal circuit via 3/8" inlet and outlet connectors
- . Sterilization method
The following technological differences exist between the subject and predicate devices:
- . The subject device has a shaft seal around the bearing, whereas the predicate device has a hydrodynamic bearing with an infusate line included in the device.
- . The capacity of the pump of the Predicate (prime volume 10 ml) is lower compared to the Subject pumping system (prime volume 40 ml) because TandemHeart pump head is smaller in diameter and volume. As a result, the predicate pump requires a higher RPM to achieve similar flows with output pressure of 450 mmHg at pump speed of 7500 RPM. The maximum RPM for Subject device is 3800 RPM with an output pressure of 680 mmHg.
- . Maximum flow rate of Subject device is 7 L/min, maximum flow rate of Predicate device is 5 L/min (percutaneous connection) or 8 L/min (surgical connection).
The VitalFlow Centrifugal Pump meets all special controls required by 21 CFR 870.4100. Special Controls met are:
- Technological Characteristics: Geometry and design parameters are consistent with the devices intended use in extracorporeal life support procedures. The subject device is designed to be compatible with other extracorporeal circuit devices and accessories.
- Biocompatibility: The subject device is demonstrated to be biocompatible in accordance with ISO 10993-1:2018 and in accordance with GLP (21 CFR 58).
- . Sterility and Shelf-life: Testing demonstrates the sterility of the subject device as provided and that it maintains its sterility, integrity, durability, and reliability over the stated Shelf-life of the device.
- . Non-clinical Performance: Substantial equivalence is demonstrated by performance characteristics on the bench, mechanical integrity, durability, and reliability.
- In vivo Evaluation: Animal testing demonstrates the subject device's performance over a longterm duration of use in a biologic test system.
- Clinical Evidence of Performance: A summary of real-world evidence of the clinical experience with this pump was prepared by ELSO.org and presented which demonstrated acceptable longterm performance of the device in ECMO patients.
- . Labeling: The instructions for use includes a detailed summary of the non-clinical evaluations pertinent to the device's use in an extracorporeal circuit. Adequate instructions are also included with respect to installation, circuit setup, maintenance during a procedure, adverse effects, and performance characteristics relevant to compatibility among different devices and accessories in the circuit.
# ECMO Performance Evidence
- . Bench studies demonstrating a blood pump use duration of 14 days were performed, including a bench durability study at 37° C in blood analog to assure mechanical stability with a substantial
{7}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for MC3 Cardiopulmonary. The logo consists of a red abstract symbol on the left, followed by the text "MC3" in large, bold, dark blue letters. Below "MC3" is the word "Cardiopulmonary" in a smaller, lighter blue font. The abstract symbol appears to be four stylized, curved shapes arranged in a circular pattern.
margin of safety. The device was mounted in the orientation as described by these instructions and met all critical flow performance attributes after this study, including pressure-flow performance and product integrity.
- . A 96-hour animal study was conducted to evaluate the safety and performance of the VitalFlow Centrifugal Pump used for extracorporeal membrane oxygenation (ECMO) when used for long term use with a target ACT of 180-220 sec. The study was conducted on a total of 13 sheep. High (5 L/min) and low (2 L/min) blood flow rates were sustained for 96 hours and there were no clots in any location in the device and blood pump functionality was confirmed.
- . Clinical Performance
A summary of real-world evidence (1048 reports) of the clinical experience with this pump was presented which demonstrated acceptable long-term performance of the device in ECMO patients. The performance of the pump was shown to be excellent with the data representing over 200,000 hours of ECMO time with this device, with the ECMO duration averaging 225.6 hours in the adult group.
Complications on ECMO (adult): For the complications that were specified to be of interest to pump performance, the overall complication rate in the adult ECMO population was 34.7% [364/1048] in the AP40 group versus 35.0% [17852/51032] in the non-AP40 group. For specific mechanical complications, the proportion of runs with a pump complication was 1.2% [13/1048] in the AP40 group versus 0.7 [355/51032] in the non-AP40 group, and the proportion of runs with at least one hemolysis complication was 1.1% [12/1048] in the AP40 group, lower than the corresponding percentage of 4.6% [2322/51032] in the non-AP40 group. The proportion of runs with the neurologic complication CNS infarction by US/ CT/MRI was 2.8% [29/1048] in the AP40 group versus 2.3% [1199/51032] in the non-AP40 group.
| Complication | VitalFlow Pump<br>(AP40)<br>(n=1048) | All Other Pumps<br>(Non-AP40)<br>(n=51032) |
|-----------------------------------|--------------------------------------|--------------------------------------------|
| ≥ 1 of any | 34.7% (364) | 35.0% (17852) |
| Mechanical | | |
| Oxygenator Failure | 5.1%(53) | 7.1%(3632) |
| Pump failure (Prevalence)ª | 1.2% (13) | 0.7% (355) |
| Pump failure (Rate per 1000 hrs)b | 0.06 | 0.03 |
| Heat Exchanger Malfunction | 0.1%(1) | 0.0% (22) |
| Air in circuit | 1.3% (14) | 0.9% (444) |
| Circuit change | 7.8% (82) | 5.6% (2851) |
| Emboli (clots or air) | 0.2% (2) | 0.2% (94) |
| Thrombosis/Clots in circuit | 2.9% (30) | 2.8% (1429) |
| Hemorrhagic | | |
| Cannulation site bleeding | 9.4%(98) | 8.1%(4157) |
| Surgical site bleeding | 10.8%(113) | 7.9%(4019) |
| Neurologic | | |
| CNS infarction (US/CT/MRI) | 2.8% (29) | 2.3% (1199) |
| CNS Hemorrhage (US/CT/MRI) | 3.6%(38) | 3.7%(1875) |
| CNS Diffuse Ischemia (CT/MRI) | 1.5%(16) | 1.6%(801) |
| Hemolysis | | |
| Moderate Hemolysis | 0.7%(7) | 2.4%(1242) |
| Severe Hemolysis | 0.4%(4) | 1.8%(925) |
| Moderate or Severe Hemolysis | 1.1% (12) | 4.6% (2322) |
b: Rate of failures per 1000 hours corresponding to the initial pur
{8}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for MC3 Cardiopulmonary. The logo consists of a red abstract symbol on the left, followed by the letters "MC3" in a dark blue, bold font. Below the letters, the word "Cardiopulmonary" is written in a smaller, lighter blue font.
# Conclusion:
The VitalFlow Centrifugal Pump device is substantially equivalent to the predicate device and the clinically used reference extracorporeal blood pump devices. The risks of this device are mitigated by meeting the Special Controls required by the regulation, 21 CFR 870.4100.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.