Affinity Pixie Oxygenator with Balance Biosurface, Affinity Pixie Oxygenator with Cardiotomy/Venous Reservoir and Balance Biosurface, Affinity Pixie Oxygenator with Cortiva BioActive Surface, Affinity Fusion Oxygenator with Cardiotomy/Venous Reservoir and Cortiva BioActive Surface

K172984 · Medtronic, Inc. · DTZ · Nov 20, 2017 · Cardiovascular

Device Facts

Record IDK172984
Device NameAffinity Pixie Oxygenator with Balance Biosurface, Affinity Pixie Oxygenator with Cardiotomy/Venous Reservoir and Balance Biosurface, Affinity Pixie Oxygenator with Cortiva BioActive Surface, Affinity Fusion Oxygenator with Cardiotomy/Venous Reservoir and Cortiva BioActive Surface
ApplicantMedtronic, Inc.
Product CodeDTZ · Cardiovascular
Decision DateNov 20, 2017
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.4350
Device ClassClass 2
AttributesTherapeutic, Pediatric

Intended Use

The Affinity Pixie hollow fiber oxygenator with Balance biosurface is indicated for use for neonate, infant, and small pediatric patients undergoing cardiopulmonary bypass (CPB) procedures requiring a blood flow rate up to 2.0 L/min. The Affinity Pixie hollow fiber oxygenator is intended to be used in an extracorporeal perfusion circuit to oxygenate and remove carbon dioxide from the blood and to cool or warm the blood during routine CPB procedures up to 6 hours in duration. The Affinity Pixie cardiotomy/ venous reservoir is intended to be used in an extracorporeal perfusion circuit to collect venous and cardiotomy suctioned blood during routine cardiopulmonary procedures up to 6 hours in duration. The CVR is also intended for use during vacuum-assisted venous drainage (VAVD) procedures.

Device Story

Microporous hollow-fiber gas-exchange device; used in extracorporeal perfusion circuits for neonate, infant, and small pediatric patients. Oxygenates/removes CO2 from blood; cools/warms blood via integrated heat exchanger. Cardiotomy/Venous Reservoir (CVR) collects venous/suctioned blood; supports vacuum-assisted venous drainage (VAVD). Used in clinical settings (e.g., OR) by perfusionists during CPB procedures up to 6 hours. Device connects to heater/cooler, recirculation, and cardioplegia circuits. Fluid pressure maintained across blood, water, and gas compartments. Change notification allows use of 330 ppm hydrogen peroxide in heater/cooler water path for microbial control.

Clinical Evidence

Bench testing only. Testing included heat exchanger material compatibility with hydrogen peroxide, structural integrity (pressure, burst, port break tests), toxicological risk assessment, and heat exchanger permeability testing.

Technological Characteristics

Microporous hollow-fiber oxygenator; polyethylene terephthalate (PET) heat exchanger; Cortiva BioActive or Balance Biosurface coatings. Operates via gas exchange and heat transfer. Connects to external heater/cooler, recirculation, and cardioplegia circuits via barbed/luer ports. Designed for neonate/infant/pediatric flow rates up to 2.0 L/min.

Indications for Use

Indicated for neonate, infant, and small pediatric patients undergoing cardiopulmonary bypass (CPB) procedures requiring blood flow rates up to 2.0 L/min.

Regulatory Classification

Identification

A cardiopulmonary bypass oxygenator is a device used to exchange gases between blood and a gaseous environment to satisfy the gas exchange needs of a patient during open-heart surgery.

Special Controls

*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Guidance for Cardiopulmonary Bypass Oxygenators 510(k) Submissions.”

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. November 20, 2017 Medtronic, Inc. Lisa Stone Principal Regulatory Affairs Specialist 8200 Coral Sea Street NE Mounds View, Minnesota 55112 # Re: K172984 Trade/Device Name: Affinity Pixie Oxvgenator with Balance Biosurface (Model BBP211). Affinity Pixie Oxygenator and Cardiotomy/Venous Reservoir with Balance Biosurface (Model BBP241). Affinity Pixie Oxygenator with Cortiva BioActive Surface (Model CBP211), Affinity Pixie Oxygenator and Cardiotomy/Venous Reservoir with Cortiva BioActive Surface (Model CBP241) Regulation Number: 21 CFR 870.4350 Regulation Name: Cardiopulmonary Bypass (CPB) Oxygenator Regulatory Class: Class II Product Code: DTZ, DTR, DTN, JOD Dated: September 25, 2017 Received: September 27, 2017 Dear Lisa Stone: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's {1}------------------------------------------------ requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, # Nicole G. Ibrahim -S for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ | DEPARTMENT OF HEALTH AND HUMAN SERVICES | | |-------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Food and Drug Administration | | | <b>Indications for Use</b> | Form Approved: OMB No. 0910-0120<br>Expiration Date: January 31, 2017<br>See PRA Statement below. | | 510(k) Number (if known) | K172984 | | Device Name | Affinity Pixie Oxygenator with Balance Biosurface (Model BBP211) | | Indications for Use (Describe) | The Affinity Pixie hollow fiber oxygenator with Balance biosurface is indicated for use for neonate, infant, and small pediatric patients undergoing cardiopulmonary bypass (CPB) procedures requiring a blood flow rate up to 2.0 L\min.<br>The Affinity Pixie hollow fiber oxygenator is intended to be used in an extracorporeal perfusion circuit to oxygenate and remove carbon dioxide from the blood and to cool or warm the blood during routine CPB procedures up to 6 hours in duration. | | Type of Use (Select one or both, as applicable) | | | | <label><input checked="checked" type="checkbox"/>Prescription Use (Part 21 CFR 801 Subpart D)</label> <label><input type="checkbox"/>Over-The-Counter Use (21 CFR 801 Subpart C)</label> | CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*** The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." FORM FDA 3881 (8/14) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF : {3}------------------------------------------------ | DEPARTMENT OF HEALTH AND HUMAN SERVICES | | |----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Food and Drug Administration | | | Indications for Use | Form Approved: OMB No. 0910-0120<br>Expiration Date: January 31, 2017<br>See PRA Statement below. | | 510(k) Number (if known) | K172984 | | Device Name | Affinity Pixie Oxygenator and Cardiotomy/Venous Reservoir with Balance Biosurface (Model BBP241) | | Indications for Use (Describe) | | | The Affinity Pixie hollow fiber oxygenator and cardiotomy/venous reservoir with Balance biosurface are indicated for use for neonate, infant, and small pediatric patients undergoing cardiopulmonary bypass (CPB) procedures requiring a blood flow rate up to 2.0 Lmin. | | | The Affinity Pixie hollow fiber oxygenator is intended to be used in an extracorporeal perfusion circuit to oxygenate and remove carbon dioxide from the blood and to cool or warm the blood during routine CPB procedures up to 6 hours in duration. | | | The Affinity Pixie cardiotomy/ venous reservoir is intended to be used in an extracorporeal perfusion circuit to collect venous and cardiotomy suctioned blood during routine cardiopulmonary procedures up to 6 hours in duration. The CVR is also intended for use during vacuum-assisted venous drainage (VAVD) procedures. | | | Type of Use (Select one or both, as applicable) | <label><input checked="checked" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> | | CONTINUE ON A SEPARATE PAGE IF NEEDED. | | | This section applies only to requirements of the Paperwork Reduction Act of 1995. | | | *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | | | The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: | | | | Department of Health and Human Services<br>Food and Drug Administration<br>Office of Chief Information Officer<br>Paperwork Reduction Act (PRA) Staff<br>PRAStaff@fda.hhs.gov | | "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | | | FORM FDA 3881 (8/14) | Page 1 of 1 | | | PSC Publishing Services (301) 443-6740 EF | {4}------------------------------------------------ | DEPARTMENT OF HEALTH AND HUMAN SERVICES | |-----------------------------------------| | Food and Drug Administration | **Indications for Use** | Form Approved: OMB No. 0910-0120 | |-----------------------------------| | Expiration Date: January 31, 2017 | | See PRA Statement below. | | 510(k) Number (if known) | K172984 | |--------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Device Name | Affinity Pixie Oxygenator with Cortiva BioActive Surface (Model CBP211) | | Indications for Use (Describe) | The Affinity Pixie hollow fiber oxygenator with Cortiva bioactive surface is indicated for use for neonate, infant, and small pediatric patients undergoing cardiopulmonary bypass (CPB) procedures requiring a blood flow rate up to 2.0 Lmin.<br>The Affinity Pixie hollow fiber oxygenator is intended to be used in an extracorporeal perfusion circuit to oxygenate and remove carbon dioxide from the blood and to cool or warm the blood during routine CPB procedures up to 6 hours in duration. | | Type of Use (Select one or both, as applicable) | |----------------------------------------------------------------------------------------------------| | <label><input checked="true" type="checkbox"/>Prescription Use (Part 21 CFR 801 Subpart D)</label> | | <label><input type="checkbox"/>Over-The-Counter Use (21 CFR 801 Subpart C)</label> | CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*** The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | FORM FDA 3881 (8/14) | Page 1 of 1 | PSC Publishing Services (301) 443-6740 | EF | |----------------------|-------------|----------------------------------------|----| |----------------------|-------------|----------------------------------------|----| {5}------------------------------------------------ | Form Approved: OMB No. | 0910-0120 | |--------------------------|------------------| | Expiration Date: | January 31, 2017 | | See PRA Statement below. | | DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration **Indications for Use** | 510(k) Number (if known) | K172984 | |--------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Device Name | Affinity Pixie Oxygenator and Cardiotomy/Venous Reservoir with Cortiva BioActive Surface (Model CBP241) | | Indications for Use (Describe) | The Affinity Pixie hollow fiber oxygenator and cardiotomy/venous reservoir with Cortiva bioactive surface are indicated for use for neonate, infant, and small pediatric patients undergoing cardiopulmonary bypass (CPB) procedures requiring a blood flow rate up to 2.0 Lmin. The Affinity Pixie hollow fiber oxygenator is intended to be used in an extracorporeal perfusion circuit to oxygenate and remove carbon dioxide from the blood and to cool or warm the blood during routine CPB procedures up to 6 hours in duration. The Affinity Pixie cardiotomy/ venous reservoir is intended to be used in an extracorporeal perfusion circuit to collect venous and cardiotomy suctioned blood during routine cardiopulmonary procedures up to 6 hours in duration. The CVR is also intended for use during vacuum-assisted venous drainage (VAVD) procedures. | | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------| | <label><input checked="checked" type="checkbox"/>Prescription Use (Part 21 CFR 801 Subpart D)</label> | <label><input type="checkbox"/>Over-The-Counter Use (21 CFR 801 Subpart C)</label> | CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*** The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." FORM FDA 3881 (8/14) Page 1 of 1 ,第 {6}------------------------------------------------ ### 510(K) SUMMARY 5 | Date Prepared: | September 25, 2017 | |-------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Submitter's Name and Address: | Medtronic, Inc.<br>Medtronic Perfusion Systems<br>8200 Coral Sea Street NE<br>Mounds View, MN 55112<br>Establishment Registration Number: 2184009 | | Contact Person: | Lisa J. Stone, RAC<br>Principal Regulatory Affairs Specialist<br>Medtronic Perfusion Systems<br>Phone: (763) 514-9866<br>Email: lisa.j.stone@medtronic.com | | Alternate Contact: | Jake Roeller<br>Senior Regulatory Affairs Manager<br>Medtronic Perfusion Systems<br>Phone: (763) 526-0404<br>Email: jake.w.roeller@medtronic.com | | Proprietary Name: | Affinity Pixie™ Oxygenator with Balance™<br>Biosurface (Model BBP211)<br>Affinity Pixie™ Oxygenator and<br>Cardiotomy/Venous Reservoir with Balance™<br>Biosurface (Model BBP241)<br>Affinity Pixie™ Oxygenator with Cortiva™<br>BioActive Surface (Model CBP211)<br>Affinity Pixie™ Oxygenator and<br>Cardiotomy/Venous Reservoir with Cortiva™м<br>BioActive Surface (Model CBP241) | | Common Name: | Oxygenator | | Classification Name: | Cardiopulmonary Bypass Oxygenator | | Classification: | Classification: Class II<br>Panel: Cardiovascular<br>Regulation: 21 CFR 870.4350<br>Product Code: DTZ | | Predicate Device: | Affinity Pixie™ Oxygenator with Balance™ Biosurface (Model BBP211) – K100645 | | | Affinity Pixie™ Oxygenator and<br>Cardiotomy/Venous Reservoir with Balance™ Biosurface (Model BBP241) – K100645 and K141233 | | | Affinity Pixie™ Oxygenator with Cortiva™ BioActive Surface (Model CBP211) – K100645 | | | Affinity Pixie™ Oxygenator and<br>Cardiotomy/Venous Reservoir with Cortiva™ BioActive Surface (Model CBP241) – K100645 and K141233 | {7}------------------------------------------------ ## Device Description The Affinity Pixie Oxygenators are microporous, hollow-fiber, gas-exchange devices available with Cortiva BioActive Surface or Balance Biosurface bonded blood contacting surface. The oxygenator is used in an extracorporeal perfusion circuit to oxygenate and remove carbon dioxide from the blood and to cool and warm the blood during routine cardiopulmonary bypass procedures up to six (6) hours in duration. The oxygenator has a low prime volume which is appropriate for use in neonate, infant and small pediatric patients. The Affinity Pixie Cardiotomy/Venous Reservoir (CVR) with Cortiva BioActive Surface or Balance Biosurface is a single use device designed to collect venous and cardiotomy suctioned blood during routine cardiopulmonary procedures up to six (6) hours in duration. Additionally, the Affinity Pixie CVR may be used during vacuum assisted venous drainage (VAVD) procedures and collect autologous blood from the chest and to aseptically return the blood to the patient for blood volume replacement during open heart surgery. The CVR allows a lower operating volume which is appropriate for use in neonate, infant and small pediatric patients. The oxygenator can be connected to a heater/cooler device, recirculation circuit, cardioplegia circuit, and the main blood path. These connections are made with tubing connected to barbed or {8}------------------------------------------------ luer ports. The oxygenator is under constant fluid pressure. There is pressure exerted on the blood-side of the device from the blood pump and patient, the water-side of the device due to the flow of the heater-cooler for water, and the gas-side of the device due to the flow of gases through the device. The three compartments (blood-side, water-side and gas-side) must not leak into one another for the oxygenator to function properly. The water-side of the oxygenator is connected to a heater/cooler device to enable temperature control of the blood. To prevent microbial growth within the heater/cooler, some manufacturers specify the addition of disinfectants to the heater/cooler water. During operation, the water path of the polyethylene terephthalate (PET) heat exchanger is exposed to these disinfectants. The purpose of this 510(k) Notification is to notify the FDA of a change to allow for the use of hydrogen peroxide (330 ppm) in the water path of the oxygenator. There are no actual changes to the oxygenators. #### Indications for Use There is no change to the intended use of the devices within the scope of the proposed change in this 510(k) Notification. The current Indications for Use statement for each device model are noted below: #### Affinity Pixie Oxygenator with Balance Biosurface (Model BBP211) The Affinity Pixie hollow fiber oxygenator with Balance biosurface is indicated for use for neonate, infant, and small pediatric patients undergoing cardiopulmonary bypass (CPB) procedures requiring a blood flow rate up to 2.0 L/min. The Affinity Pixie hollow fiber oxygenator is intended to be used in an extracorporeal perfusion circuit to oxygenate and remove carbon dioxide from the blood and to cool or warm the blood during routine CPB procedures up to 6 hours in duration. ## Affinity Pixie Oxygenator and Cardiotomy/Venous Reservoir with Balance Biosurface (Model BBP241) The Affinity Pixie hollow fiber oxygenator and cardiotomy/venous reservoir with Balance biosurface are indicated for use for neonate, infant, and small pediatric patients undergoing cardiopulmonary bypass (CPB) procedures requiring a blood flow rate up to 2.0 L/min. The Affinity Pixie hollow fiber oxygenator is intended to be used in an extracorporeal perfusion circuit to oxygenate and remove carbon dioxide from the blood and to cool or warm the blood during routine CPB procedures up to 6 hours in duration. {9}------------------------------------------------ The Affinity Pixie cardiotomy/ venous reservoir is intended to be used in an extracorporeal perfusion circuit to collect venous and cardiotomy suctioned blood during routine cardiopulmonary procedures up to 6 hours in duration. The CVR is also intended for use during vacuum-assisted venous drainage (VAVD) procedures. ## Affinity Pixie Oxygenator with Cortiva BioActive Surface (Model CBP211) The Affinity Pixie hollow fiber oxygenator with Cortiva bioactive surface are indicated for use for neonate, infant, and small pediatric patients undergoing cardiopulmonary bypass (CPB) procedures requiring a blood flow rate up to 2.0 L/min. The Affinity Pixie hollow fiber oxygenator is intended to be used in an extracorporeal perfusion circuit to oxygenate and remove carbon dioxide from the blood and to cool or warm the blood during routine CPB procedures up to 6 hours in duration. ## Affinity Pixie Oxygenator and Cardiotomy/Venous Reservoir with Cortiva BioActive Surface (Model CBP241) The Affinity Pixie hollow fiber oxygenator and cardiotomy/venous reservoir with Cortiva bioactive surface are indicated for use for neonate, infant, and small pediatric patients undergoing cardiopulmonary bypass (CPB) procedures requiring a blood flow rate up to 2.0 L/min. The Affinity Pixie hollow fiber oxygenator is intended to be used in an extracorporeal perfusion circuit to oxygenate and remove carbon dioxide from the blood and to cool or warm the blood during routine CPB procedures up to 6 hours in duration. The Affinity Pixie cardiotomy/ venous reservoir is intended to be used in an extracorporeal perfusion circuit to collect venous and cardiotomy suctioned blood during routine cardiopulmonary procedures up to 6 hours in duration. The CVR is also intended for use during vacuum-assisted venous drainage (VAVD) procedures. #### Comparison to Predicate Devices When compared to the predicate devices, the Affinity Pixie Oxygenators are substantially equivalent. The oxygenators within scope of this 510(k) Notification have the following similarities: - Same intended use - Same operating principle ● - Same fundamental technological characteristics ● {10}------------------------------------------------ - Same design, dimensions and performance - Same device materials - Same packaging materials and design ● - Same sterilization requirements ### Summary of Performance Data While these devices are subject to special controls, most of the special control requirements are not applicable for this proposed change. During use, the water path of the heat exchanger is the only part of the oxygenator that is exposed to a disinfectant. Because of this, it is not necessary to repeat all recommended testing by the special control guidance documents. It is however necessary to conduct testing to verify the chemical disinfectant has no adverse effects on the oxygenator water path. The following testing and analysis was performed: - Heat Exchanger Product Material Compatibility with Hydrogen Peroxide . - Structural Integrity Testing (pressure integrity, burst and port break tests) ● - Toxicological Risk Assessment for Hydrogen Peroxide - Heat Exchanger Permeability Testing . #### Conclusion In conclusion, the information provided within this submission demonstrates the Affinity Pixie Oxygenators with/without Cardiotomy/Venous Reservoir and Balance Biosurface or Cortiva BioActive Surface are substantially equivalent to the marketed predicate devices.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...