The Novalung® surgical Lung Assist (sLA) Membrane Lung™ is intended to be used for extracorporeal circulation during cardiopulmonary bypass in the field of open-heart surgery. Within the indicated flow rates, blood is oxygenated and carbon dioxide is removed. The utilization period of this device is restricted to six hours: The application and use of the oxygenator is the sole responsibility of the attending physician.
Device Story
Novalung sLA Membrane Lung is a sterile, single-use cardiopulmonary bypass oxygenator. Device incorporates solid surface hollow polymethylpentene (PMP) fiber membrane with 1.3 m² surface area. Used in extracorporeal circuits during open-heart surgery to oxygenate blood and remove carbon dioxide. Operated by clinical staff within a flow range of 0.5 L/min to 4.5 L/min. Attending physician manages device application. Benefits include gas exchange support during bypass procedures.
Clinical Evidence
Bench testing only. Device characterized in accordance with FDA guidance. Failure mode and effects analysis (FMEA) performed; potential failure modes evaluated in vitro.
Indicated for patients weighing >20kg undergoing cardiopulmonary bypass during open-heart surgery requiring extracorporeal circulation for oxygenation and carbon dioxide removal. Utilization restricted to six hours.
Regulatory Classification
Identification
A cardiopulmonary bypass oxygenator is a device used to exchange gases between blood and a gaseous environment to satisfy the gas exchange needs of a patient during open-heart surgery.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Guidance for Cardiopulmonary Bypass Oxygenators 510(k) Submissions.”
Predicate Devices
Jostra Quadrox D Safeline Diffusion Membrane Oxygenator (K061628)
Related Devices
K071774 — JOSTRA QUADROX D DIFFUSION MEMBRANE OXYGENATOR WITH BIOLINE COATING, MODEL BE-HMOD 2000, BEQ-HMOD 2030 · Maquet Cardiopulmonary, AG · Jan 23, 2008
K141492 — ADVANCED MEMBRANCE GAS EXCHANGE PMP STERILE (A.M.G PMP STERILE) · Eurosets S.R.L · Feb 6, 2015
K143073 — Affinity NT Oxygenator, Affinity NT Oxygenator with Trillium Biosurface, Affinity NT Oxygenator with Carmeda Biosurface · Medtronic, Inc. · Mar 25, 2015
K991452 — COBE OPTIMIN HOLLOW FIBER MEMBRANE OXYGENATOR · Cobe Cardiovascular, Inc. · Oct 7, 1999
{0}------------------------------------------------
K072362
007 2 3 2007
# 510(k) Summary for the Novalung® GmbH Novalung® surgical Lung Assist (sLA) Membrane Lung™
#### 1. SUBMITTER/510(K) HOLDER
Novalung® GmbH Lotzenäcker 3 D-72379 Hechingen, Germany
Contact Person: Heiko Frerichs Telephone: +49 7471 98488-30
Date Prepared: August 21, 2007
#### 2. DEVICE NAME
| Proprietary Name: | Novalung® surgical Lung Assist (sLA) Membrane Lung™ |
|----------------------|-----------------------------------------------------|
| Common/Usual Name: | Cardiopulmonary bypass oxygenator |
| Classification Name: | Oxygenator, cardiopulmonary bypass |
#### 3. PREDICATE DEVICES
- Jostra Quadrox D Safeline Diffusion Membrane Oxygenator . (Maquet Cardiovascular AG. K061628)
#### 4. DEVICE DESCRIPTION
The Novalung® surgical Lung Assist (sLA) Membrane Lung™ is a sterile device for single use only and it is not intended to be resterilized by the user. The Novalung® surgical Lung Assist (sLA) Membrane Lung™ is used to deliver oxygen into the blood and remove carbon dioxide from the blood. It is intended to be used as a component in an extracorporeal circuit for patients undergoing cardiopulmonary bypass. The Novalung® surgical Lung Assist (sLA) Membrane Lung™ incorporates a solid surface hollow polymethylpentene (PMP) fiber membrane and has a 1.3 m² surface area. The Novalung® surgical Lung Assist (sLA) Membrane Lung™ has a nominal flow range from 0.5 L/min. to 4.5 L/min and is suitable for patients weighing >20kg.
{1}------------------------------------------------
## 5. INTENDED USE
The Novalung® surgical Lung Assist (sLA) Membrane Lung™ is intended to be used for extracorporeal circulation during cardiopulmonary bypass in the field of open-heart surgery. Within the indicated flow rates, blood is oxygenated and carbon dioxide is removed. The utilization period of this device is restricted to six hours:
The application and use of the oxygenator is the sole responsibility of the attending physician.
#### 6. TECENOLOGICAL CHARACTERISTICS AND SUBSTANTIAL EQUIVALENCE
The Novalung® surgical Lung Assist (sLA) Membrane Lung™ has the same intended use, design, principals of operation, and performance as the Jostra Quadrox D Safeline Diffusion Membrane Oxygenator. The Jostra Quadrox D Safeline Diffusion Membrane Oxygenator uses a tight diffusive membrane, where as the Novalung® surgical Lung Assist (sLA) Membrane Lung™ uses a solid surface membrane. The Quadrox D offers 1.8 m² surface area and a blood flow range of 0.5 L/min to 7 L/min. The Novalung® surgical Lung Assist (sLA) Membrane Lung™ provides the user with a smaller surface area of 1.3m² and a reduced flow range of 0 L/min to 4.5 L/min. Comparative testing has demonstrated that these differences do not raise new issues of safety and effectiveness.
## 7. PERFORMANCE TESTING
The Novalung® surgical Lung Assist (sLA) Membrane Lung™ has been fully characterized by bench testing in accordance with FDA Guidance. A failure mode and effects analysis has been performed to identify potential failure modes of the device. Potential failure modes directly related to the design and function of the Novalung® surgical Lung Assist (sLA) Membrane Lung™ have been evaluated in vitro and results are provided in the premarket notification.
{2}------------------------------------------------
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
OCT 2 3 2007
Novalung® GmbH c/o Mr. Heiko Frerichs Lotzenäcker 3 D-72379 Hechingen, Germany
Re: K072362
Novalung® Surgical Lung Assist (sLA) Membrane Lung™ Regulation Number: 21 CFR 870.4350 Regulation Name: Cardiopulmonary Bypass Oxygenator Regulatory Class: Class II Product Code: DTZ Dated: August 21, 2007 Received: August 22, 2007
Dear Mr. Frerichs:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, EDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 - Mr. Heiko Frerichs
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance. please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
uma R. bohner
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
510(k) Number (if known): Jo 7 2 7 2 3 6 2
Device Name: Novalung® surgical Lung Assist (sLA) Membrane Lung™
Indications for Use:
The Novalung® surgical Lung Assist (sLA) Membrane Lung™ is intended to be used for extracorporeal circulation during cardiopulmonary bypass in the field of open-heart surgery. Within the indicated flow rates blood is oxygenated and carbon dioxide is removed. The utilization period of this device is restricted to six hours.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(Please DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
una R. hlmey
(Division Chief)
ardiovascular Devices
510(k) Number_
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.