Affinity NT Oxygenator, Affinity NT Oxygenator with Trillium Biosurface, Affinity NT Oxygenator with Carmeda Biosurface
K143073 · Medtronic, Inc. · DTZ · Mar 25, 2015 · Cardiovascular
Device Facts
Record ID
K143073
Device Name
Affinity NT Oxygenator, Affinity NT Oxygenator with Trillium Biosurface, Affinity NT Oxygenator with Carmeda Biosurface
Applicant
Medtronic, Inc.
Product Code
DTZ · Cardiovascular
Decision Date
Mar 25, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.4350
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Affinity NT Hollow Fiber Oxygenator with Plasma Resistant Fiber is intended to be used in an extracorporeal perfusion circuit to oxygenate and remove carbon dioxide from the blood and to cool or warm the blood during routine cardiopulmonary bypass procedures up to 6 hours in duration. The Affinity NT Hollow Fiber Oxygenator with Plasma Resistant Fiber with Carmeda BioActive Surface is intended to be used in an extracorporeal perfusion circuit to oxygenate and remove carbon dioxide from the blood and to cool or warm the blood during routine cardiopulmonary bypass procedures up to 6 hours in duration. The Affinity NT Hollow Fiber Oxygenator with Plasma Resistant Fiber with Trillium Biosurface is intended to be used in an extracorporeal perfusion circuit to oxygenate and remove carbon dioxide from the blood during routine cardiopulmonary bypass procedures up to 6 hours in duration.
Device Story
Single-use gas exchange device; integral heat exchanger; plasma-resistant hollow fibers. Blood flows outside fibers; oxygen/CO2 exchange occurs across fiber membrane. Used in extracorporeal heart-lung bypass circuits during cardiac surgery. Operated by perfusionists/clinicians. Provides oxygenation, CO2 removal, and thermal control. Benefits patient by maintaining physiological gas exchange and temperature during bypass. Device available uncoated or with Carmeda BioActive or Trillium Biosurface coatings to improve biocompatibility.
Clinical Evidence
Bench testing only. No clinical data required. Testing included biocompatibility (ISO 10993-1), positive pressure integrity, gas transfer, pressure drop, plasma breakthrough, blood trauma (hemolysis, platelets, WBCs), heat exchanger performance, and coating integrity. All tests passed.
Technological Characteristics
Hollow fiber oxygenator with integral heat exchanger. Materials: plasma-resistant fiber, optional Carmeda BioActive or Trillium Biosurface coatings. Single-use, sterile, non-pyrogenic. Connectivity: None (mechanical).
Indications for Use
Indicated for patients undergoing routine cardiopulmonary bypass procedures requiring extracorporeal blood oxygenation, carbon dioxide removal, and temperature regulation (cooling/warming) for up to 6 hours.
Regulatory Classification
Identification
A cardiopulmonary bypass oxygenator is a device used to exchange gases between blood and a gaseous environment to satisfy the gas exchange needs of a patient during open-heart surgery.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Guidance for Cardiopulmonary Bypass Oxygenators 510(k) Submissions.”
Medtronic Affinity Oxygenator with Trillium Biosurface (K973760)
Medtronic Affinity Oxygenator with Carmeda BioActive Surface (K000430)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/12 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular emblem with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines representing hair or fabric.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 25, 2015
Medtronic, Inc. Jessica Sixberry Principal Regulatory Affairs Specialist 7611 Northland Drive Minneapolis, Minnesota 55428
Re: K143073
Trade/Device Name: Affinity NT Oxygenator, Affinity NT Oxygenator with Trillium Biosurface, Affinity NT Oxygenator with Carmeda Biosurface Regulation Number: 21 CFR 870.4350 Regulation Name: Cardiopulmonary Bypass Oxygenator Regulatory Class: Class II Product Code: DTZ Dated: February 13, 2015 Received: February 18, 2015
Dear Ms. Sixberry,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply
{1}------------------------------------------------
with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
MA-Kille
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K143073
Device Name Affinity NT Hollow Fiber Oxygenator
#### Indications for Use (Describe)
#### Model 511:
The Affinity NT Hollow Fiber Oxygenator with Plasma Resistant Fiber is intended to be used in an extracorporeal perfusion circuit to oxygenate and remove carbon dioxide from the blood and to cool or warm the blood during routine cardiopulmonary bypass procedures up to 6 hours in duration.
#### Model CB511:
The Affinity NT Hollow Fiber Oxygenator with Plasma Resistant Fiber with Carmeda BioActive Surface is intended to be used in an extracorporeal perfusion circuit to oxygenate and remove carbon dioxide from the blood and to cool or warm the blood during routine cardiopulmonary bypass procedures up to 6 hours in duration.
Model 511T:
The Affinity NT Hollow Fiber Oxygenator with Plasma Resistant Fiber with Trillium Biosurface is intended to be used in an extracorporeal perfusion circuit to oxygenate and remove carbon dioxide from the blood during routine cardiopulmonary bypass procedures up to 6 hours in duration.
| Type of Use (Select one or both, as applicable) | |
|-----------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------|
| <span style="text-decoration: underline;">Prescription Use (Part 21 CFR 801 Subpart D)</span> | <span style="text-decoration: underline;">Over-The-Counter Use (21 CFR 801 Subpart C)</span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW!*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the Medtronic logo. On the left side of the logo is a gray circle with a human figure inside. The human figure is in three different positions, as if they are rotating. To the right of the circle is the word "Medtronic" in blue, bold letters.
# 510(k) Summary
| Date Prepared: | March 18, 2015 |
|-----------------|--------------------------------------------|
| Submitter: | Medtronic, Inc. |
| | Medtronic Perfusion Systems |
| | 7611 Northland Drive |
| | Minneapolis, MN 55428 |
| | Establishment Registration Number: 2184009 |
| Contact Person: | Jessica Sixberry |
| | Principal Regulatory Affairs Specialist |
| | Phone: (763) 514-9849 |
| | Fax: |
| | Email: jessica.m.sixberry@medtronic.com |
### Device Name and Classification:
| Trade Name: | Affinity NT Oxygenator Uncoated or with Carmeda® BioActive |
|-----------------------|------------------------------------------------------------|
| | Surface or with Trillium® Biosurface |
| Common Name: | Oxygenator |
| Classification Name: | Cardiopulmonary bypass Oxygenator |
| Classification Panel: | Cardiovascular |
| Regulation Number: | 21 CFR 870.4350 |
| Product Code: | DTZ |
| Classification: | Class II |
## Predicate Devices
Medtronic Affinity Oxygenator (K932252) Medtronic Affinity Oxygenator with Trillium Biosurface (K973760) Medtronic Affinity Oxygenator with Carmeda BioActive Surface (K000430)
## Device Description
The Medtronic Affinity NT Hollow Fiber Oxygenator is a single use gas exchange device with plasma resistant fiber. It is designed to have the blood flow outside the fiber and includes an integral heat exchanger. The oxygenator is available uncoated or bonded on its blood contacting surfaces with either Carmeda BioActive Surface or Trillium Biosurface. The device is single-use, nontoxic, nonpyrogenic, and supplied sterile for clinical use.
The Affinity NT Hollow Fiber Oxygenator with Plasma Resistant Fiber is intended to be used in an extracorporeal perfusion circuit to oxygenate and remove carbon dioxide from the blood and to cool or warm the blood during routine cardiopulmonary bypass procedures up to 6 hours in duration.
{4}------------------------------------------------
Affinity NT Oxygenator is designed to be an integral part of the cardiopulmonary heart lung bypass circuit for use during cardiac surgery. Blood that comes from the patient is delivered through a blood pump to the oxygenator and other auxiliary devices, and back to the patient.
The purpose of this 510(k) Notification was to notify the FDA of an alternate material formulation for the luer caps used on the access port and the sample port as well as report previous changes on the Affinity NT Oxygenator models identified in the table below.
| Models | Description |
|--------|-------------------------------------------------------|
| 511 | Affinity NT Oxygenator |
| CB511 | Affinity NT Oxygenator with Carmeda BioActive Surface |
| 511T | Affinity NT Oxygenator with Trillium Biosurface |
## Indications for Use
The Affinity NT Hollow Fiber Oxygenator with Plasma Resistant Fiber (with Carmeda BioActive surface or Trillium Biosurface) is intended to be used in an extracorporeal perfusion circuit to oxygenate and remove carbon dioxide from the blood and to cool or warm the blood during routine cardiopulmonary bypass procedures up to 6 hours in duration.
## Comparison to Predicate Devices
The Affinity NT Oxygenators have the same intended use, design and materials, and principles of operation and technology when compared to the predicate Affinity Oxygenators (including previously implemented changes).
- Intended Use: The intended use is the same as the predicate devices.
- Design: The design is the same as the predicate devices.
- Materials: The materials of the Affinity NT Oxygenator are the same or equivalent to . the materials used in the predicate devices.
- Principles of Operation and Technology: The principles of operation are the same as ● the predicate devices.
- Performance: The performance of the device is the same as the predicate device. ●
## Summary of Performance Data
Bench testing was used to verify the performance characteristics of this device. In addition, real-time aging testing for devices incorporating the previously implemented changes was completed and supports substantial equivalence. Clinical testing was not required to establish substantial equivalence.
{5}------------------------------------------------
The following performance tests were conducted:
| Testing | | Description | Result |
|--------------------------------------------------|-----------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------|
| Subject of 510(k) | Biocompatibility | Ensures alternate material for the luer cap is<br>biocompatible according to ISO 10993-1,<br>externally communicating, circulating blood<br>contact, limited duration | Pass |
| | Positive Pressure<br>Integrity Testing | Ensures luer caps made with the alternate<br>material maintains appropriate seal on the<br>device ensure the device does not leak | Pass |
| Previously Implemented Changes (real time aging) | Gas Transfer<br>(oxygen and carbon<br>dioxide) | Demonstrates the devices transfer gas to<br>specification | Pass |
| | Pressure Drop<br>(Blood and Gas) | Demonstrates the devices meet pressure drop<br>requirements | Pass |
| | Plasma<br>Breakthrough | Demonstrates the devices meet Plasma<br>Breakthrough requirements | Pass |
| | Blood Trauma<br>(hemolysis,<br>Platelets, White<br>Blood Cells) | Demonstrates the devices meet Blood<br>Trauma requirements | Pass |
| | Pressure Integrity | Demonstrates the devices meet Pressure<br>Integrity requirements | Pass |
| | Heat Exchanger<br>Performance | Demonstrates the devices meet the heat<br>exchanger efficiency requirements | Pass |
| | Coating Testing | Coating testing demonstrates that the coated<br>devices meet the coating requirements | Pass |
# Conclusion
Medtronic has demonstrated that the Affinity NT Oxygenators are substantially equivalent to the predicate devices based upon design, test results, and indications for use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.