TEMBO Embolic System

K240873 · Instylla, Inc. · KRD · Dec 16, 2024 · Cardiovascular

Device Facts

Record IDK240873
Device NameTEMBO Embolic System
ApplicantInstylla, Inc.
Product CodeKRD · Cardiovascular
Decision DateDec 16, 2024
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.3300
Device ClassClass 2
AttributesTherapeutic

Intended Use

The TEMBO Embolic System is a gelatin agent intended for use in the embolization of hypervascular tumors and blood vessels to occlude blood flow in the peripheral vasculature.

Device Story

TEMBO Embolic System consists of biocompatible, dry resorbable porcine gelatin particles provided in a 10 mL syringe. Used by interventional radiologists or physicians in clinical settings to occlude blood flow in peripheral vasculature or hypervascular tumors. Particles are hydrated with contrast media or saline and delivered via a standard microcatheter. The material acts as a mechanical barrier to blood flow. The device is intended for single use. Benefits include targeted vascular occlusion. Clinical decision-making is supported by confirmation angiography to verify embolic placement and persistence.

Clinical Evidence

No human clinical data. Evidence includes bench testing (visual verification, particle size, chemical compatibility, delivery ease, embolic persistence, pepsin digestion, pH, pot life, simulated use) and animal studies. Two swine studies (n=11 and n=8) evaluated acute embolization effects in renal arteries, systemic impact, target tissue response, and non-target migration over 28 days, with comparative analysis against EmboCube and Gel-Bead controls.

Technological Characteristics

Biocompatible, resorbable porcine gelatin particles. Dry particle size 85-255μm. Delivered via microcatheter. Sterilized via gamma radiation (ISO 11137-1/2) to SAL 10^-6. Gelatin raw material compliant with ISO 22442-1. Biocompatibility per ISO 10993-1 for long-term (>30 days) blood contact. Endotoxin level <20 EU/device.

Indications for Use

Indicated for embolization of hypervascular tumors and blood vessels to occlude blood flow in the peripheral vasculature.

Regulatory Classification

Identification

A vascular embolization device is an intravascular implant intended to control hemorrhaging due to aneurysms, certain types of tumors (e.g., nephroma, hepatoma, uterine fibroids), and arteriovenous malformations. This does not include cyanoacrylates and other embolic agents, which act by polymerization or precipitation. Embolization devices used in neurovascular applications are also not included in this classification, see § 882.5950 of this chapter.

Special Controls

*Classification.* Class II (special controls.) The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance Document: Vascular and Neurovascular Embolization Devices.” For availability of this guidance document, see § 870.1(e).

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the seal of the Department of Health & Human Services. To the right of the seal is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and the word "ADMINISTRATION" in a smaller font size below. December 16, 2024 Instylla, Inc. Jennifer Greer Senior Regulatory Affairs Manager 201 Burlington Rd North Building Bedford, Massachusetts 01730 Re: K240873 Trade/Device Name: TEMBO Embolic System Regulation Number: 21 CFR 870.3300 Regulation Name: Vascular Embolization Device Regulatory Class: Class II Product Code: KRD Dated: November 13, 2024 Received: November 13, 2024 Dear Jennifer Greer: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" {1}------------------------------------------------ (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rue"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advice-comprehensiveregulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reporting-mdr-howreport-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). {2}------------------------------------------------ Sincerely, Digitally signed by Finn E. Finn E. Donaldson -S Donaldson -S Date: 2024.12.16 10:50:30 -05'00' Finn Donaldson Acting Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ # Indications for Use 510(k) Number (if known) K240873 Device Name TEMBO Embolic System Indications for Use (Describe) The TEMBO Embolic System is a gelatin agent intended for use in the embolization of hypervascular tumors and blood vessels to occlude blood flow in the peripheral vasculature. | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------|--| | | | X Prescription Use (Part 21 CFR 801 Subpart D) | | Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ Image /page/4/Picture/1 description: The image shows the word "instylla" in a lowercase sans-serif font. A curved line in a copper color starts above the "i" and extends over the "stylla" part of the word, creating a design element. The dot above the "i" is also in the same copper color, providing a visual accent. ## 510(k) Summary ### Submitter Information: Instylla, Inc. 201 Burlington Rd, North Building Bedford, MA 01730 ### Contact Person: Jennifer Greer Sr. Regulatory Affairs Manager Phone: 781-622-9293 Email: jennyg@instylla.com ### Date Prepared: December 13, 2024 ## Subject Device: | Proprietary Name: | TEMBO Embolic System | |------------------------|----------------------------------------------| | Common Name: | Embolization Device | | Classification Name: | Device, Vascular, For Promoting Embolization | | | (21 CFR 870.3300, Product Code KRD) | | Device Classification: | Class II | | Classification Panel: | Cardiovascular | ### Predicate Device: | Proprietary Name: | EmboCube™ Embolization Gelatin | |-------------------|--------------------------------| | Manufacturer: | Biosphere Medical, S.A. | | 510(k) Number: | K183120 | The Gel-Bead Embolization Spheres, cleared under 510(k) Premarket Notification K133237, was used as a reference device in this submission. ### Device Description: The TEMBO Embolic System consists of biocompatible, dry resorbable particles of porcine gelatin in a 10 mL syringe with a non-vented luer lock cap. {5}------------------------------------------------ The TEMBO Embolic System is a gelatin agent and a sterile, single use, medical device intended for the embolization of hypervascular tumors and blood vessels to occlude blood flow in the peripheral vasculature. The TEMBO Embolic System particles are hydrated using commercially available contrast media or contrast mixed with saline. Once hydrated, the material is injected into the target blood vessel via a commercially available microcatheter for occlusion of target vasculature. The dry TEMBO Embolic System gelatin particles are between 85 µm. ## Indications for Use: The TEMBO Embolic System is a gelatin agent intended for use in the embolization of hypervascular tumors and blood vessels to occlude blood flow in the peripheral vasculature. ## Comparison to Predicate and Reference Device: The TEMBO Embolic System is similar in design to the predicate device, EmboCube Embolization Gelatin (K183120), as both provide a mechanical barrier to blood flow and are delivered using microcatheters. The TEMBO Embolic System's indications for use are similar to EmboCube's, as it is within the scope of the predicate device's indications for use. They are both intended to embolize hypervascular tumors and blood flow to occlude blood vessels in peripheral vasculature. The subject device contains gelatin and sodium chloride, while the predicate only contains gelatin. The TEMBO Embolic System and EmboCube Embolization Gelatin differ in particle size and shape: the subject device has a dry particle size between 85-255μm and has non-uniform particle shape, while the predicate device has a particle size of 2.5mm and 5mm and has a cube particle shape. The reference device, Gel-Bead Embolization Spheres (K133237), contains similar materials to the subject device and the hydrated TEMBO Embolic System particle size falls within the reference device's particle size ranges. ## Performance Data Performance testing of the final, sterilized TEMBO Embolic System included bench testing and functional testing to verify specifications fundamental to the device. Testing included the following: - -Visual Verification - -Particle Size - Dry - -Iodinated Contrast Media Chemical Compatibility - -Delivery Ease - Embolic Persistence - {6}------------------------------------------------ - Pepsin Digestion - - pH of Hydrated Gelatin - Pot Life - - -Simulated Use Testing which included testing the following: - Ability to Create an Embolization - Depth of Penetration - Minimal Non-target Embolization - 트 Directed Delivery of Embolic Material - Embolic Material Withstands Confirmation Angiography - I Embolic Delivery Through Microcatheter System - 트 Ability to Start and Stop Embolic Delivery - Compatibility with Female Luer Connections - All performance testing met predetermined specifications. Comparative testing was also conducted between the subject and predicate device. No new safety or performance issues were raised during testing. ## Biocompatibility Testing A biocompatibility evaluation was conducted in accordance with the FDA Guidance Document Use of International Standard ISO 10993-1. "Biological Evaluation of Medical Devices Part 1: Evaluation and Testing within a Risk Management Process," consistent with an implant device in contact with blood for a long term (>30 days) duration. The following biocompatibility tests were successfully completed on the final, sterilized TEMBO Embolic System: - -Cytotoxicity - -Sensitization - Irritation - - Acute Systemic Toxicity - - -Material Mediated Pyrogenicity - -Subacute Systemic Toxicity - Implantation - - -Genotoxicity - Hemocompatibility - - -Chemical Characterization The gelatin raw material used in the TEMBO Embolic System is compliant with ISO 22442-1 Medical devices utilizing animal tissues and their derivatives Part 1: Application of risk management. {7}------------------------------------------------ # Sterility TEMBO is sterilized via gamma radiation to a Sterility Assurance Level of 10 °, in accordance with ISO 11137-1 Sterilization of health care products Radiation Part 1: Requirements for development, validation and routine control of a sterilization process for medical devices and ISO 11137-2 Sterilization of health care products Radiation Part 2: Establishing the sterilization dose. A bacterial endotoxins test (BET), also known as the Limulus amebocyte lysate (LAL) test, was validated to establish that the TEMBO endotoxin level will be <20 endotoxin units (EU)/device. ## Shelf Life TEMBO has a shelf-life of 6-months. Shelf-life studies have been conducted to demonstrate that the device maintains its performance and the packaging will maintain its sterile barrier over the entirety of the intended shelf life. ## Animal Testing Instylla has conducted an animal study, on eleven swine, to evaluate the acute embolization effect of the TEMBO Embolic System in the swine kidney via the renal arteries. Additionally, the systemic impact of embolic was monitored during the survival period. The impact of embolization to the target tissue and non-target embolization/device migration was assessed over a period of 28 days. An additional animal study was conducted on 8 swine to evaluate the embolic effect of TEMBO at 7 days, compared to EmboCube and Gel-Bead as controls. ## Conclusion Instylla has demonstrated that the TEMBO Embolic System is substantially equivalent in fundamental design, function, device materials, sterilization, operating principle, intended use/indication for use and fundamental technology to the legally marketed predicate device, EmboCube™ Embolization Gelatin (K183120) and reference device Gel-Bead Embolization Spheres (K133237).
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...