K141313 · Reverse Medical Corporation · KRD · Jun 18, 2014 · Cardiovascular
Device Facts
Record ID
K141313
Device Name
REVERSE MEDICAL MICRO VASCULAR PLUG SYSTEM
Applicant
Reverse Medical Corporation
Product Code
KRD · Cardiovascular
Decision Date
Jun 18, 2014
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.3300
Device Class
Class 2
Attributes
Therapeutic
Intended Use
The Reverse Medical® Corporation MVP® Micro Vascular Plug System is indicated to obstruct or reduce the rate of blood flow in the peripheral vasculature.
Device Story
The MVP Micro Vascular Plug System is a vascular embolization device used to obstruct or reduce blood flow in peripheral vessels. The device consists of a self-expandable, ovoid-shaped nitinol frame with a PTFE cover, secured by platinum marker bands. It is delivered to the treatment site via a microcatheter using a composite delivery wire. The device is operated by a physician. The system features a mechanical detachment mechanism; the plug is detached from the delivery wire by rotating the wire counter-clockwise. This mechanical method replaces a previous electrolytic detachment system, eliminating the need for external power boxes or cables. The device is intended for use in clinical settings where peripheral vascular embolization is required to manage blood flow.
Clinical Evidence
Bench testing only. No clinical data was provided. Verification testing included dimensional and visual inspection, microcatheter compatibility (flexibility, kinking, deployment/retraction forces), detachment evaluations (torque strength, number of turns), galvanic corrosion (ASTM G71), MRI compatibility (ASTM F-2503), and package integrity/shelf-life testing. All tests met acceptance criteria.
Indicated for patients requiring obstruction or reduction of blood flow in the peripheral vasculature.
Regulatory Classification
Identification
A vascular embolization device is an intravascular implant intended to control hemorrhaging due to aneurysms, certain types of tumors (e.g., nephroma, hepatoma, uterine fibroids), and arteriovenous malformations. This does not include cyanoacrylates and other embolic agents, which act by polymerization or precipitation. Embolization devices used in neurovascular applications are also not included in this classification, see § 882.5950 of this chapter.
Special Controls
*Classification.* Class II (special controls.) The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance Document: Vascular and Neurovascular Embolization Devices.” For availability of this guidance document, see § 870.1(e).
K123803 — REVERSE MEDICAL MICRO VASCULAR PLUG (MVP) SYSTEM · Reverse Medical Corporation · Jul 2, 2013
K150108 — Reverse Medical Micro Vascular Plug System · Reverse Medical Corporation · Apr 22, 2015
K133282 — REVERSE MEDICAL MICRO VASCULAR PLUG (MVP) SYSTEM · Reverse Medical Corporation · Nov 27, 2013
K031810 — AMPLATZER VASCULAR PLUG · Aga Medical Corp. · Sep 9, 2003
K191680 — AZUR Vascular Plug · MicroVention, Inc. · Mar 20, 2020
Submission Summary (Full Text)
{0}------------------------------------------------
| Sponsor/Applicant: | Reverse Medical Corporation<br>13700 Alton Parkway, Ste. 167<br>Irvine, CA 92618 |
|-------------------------|------------------------------------------------------------------------------------------------------------|
| Date Prepared: | May 17, 2014 |
| Contact Person: | Jane Metcalf<br>Vice President, Quality, Regulatory and Clinical Affairs<br>Email: jmetcalf@reversemed.com |
| Trade Name: | Reverse Medical® MVP® Micro Vascular Plug System |
| Common Name: | Vascular Embolization Device |
| Classification Name: | Vascular Embolization Device |
| Device Classification : | Class 2 |
| Regulation Number: | 870.3300 (product code: KRD) |
| Predicate Devices: | Reverse Medical MVP-3 (K123803) and Reverse Medical<br>MVP-5 (K133282) |
#### Purpose of Submission
The purpose of this special 510(k) submission is to obtain market clearance for two (2) modifications to the MVP System. The first is to replace the electrolytic detachment method with a mechanical detachment method. The second is to reduce the length of the delivery wire from 180 cm to 160 cm.
#### Indication for Use and Intended Use
The Reverse Medical® Corporation MVP® Micro Vascular Plug System is indicated to obstruct or reduce the rate of blood flow in the peripheral vasculature.
#### Device Description
Reverse Medical® MVP® Micro Vascular Plug System consists of a micro vascular occlusion plug that is attached to a composite delivery wire and is intended to be delivered to the treatment site through a catheter. The MVP occlusion plug is a selfexpandable, ovoid-shaped frame made from nitinol and incorporates a PTFE cover over the proximal portion of the ovoid. The plug device is secured at both ends with platinum marker bands. The proximal marker band is attached to a delivery wire that is used to push the plug device through a commercially available catheter to the intended treatment site. After satisfactory deployment of the plug device at the treatment site, the implant is detached from the delivery wire by rotating the wire counter clockwise.
#### Technical Characteristics
The modified devices have the same technological characteristics as the predicate devices except for the detachment method and length of delivery wire. The mechanical
{1}------------------------------------------------
detachment method eliminates the need for a software controlled detachment box and cables. Table 5-1 compares characteristics of the modified devices to the predicates.
| | K123803 | K133282 | Subject of this Submission | |
|---------------------------------|----------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------|
| Feature | MVP-3<br>Predicate | MVP-5<br>Predicate | MVP-3 Modified | MVP-5 Modified |
| Indications for Use | To obstruct or reduce<br>the rate of blood flow<br>in the peripheral<br>vasculature. | To obstruct or reduce<br>the rate of blood flow<br>in the peripheral<br>vasculature. | To obstruct or reduce<br>the rate of blood flow<br>in the peripheral<br>vasculature. | To obstruct or reduce<br>the rate of blood flow<br>in the peripheral<br>vasculature. |
| Materials of<br>Construction | Nitinol, PTFE,<br>Platinum, SS 301,<br>Solder, Polypropylene<br>sheath, Urethane,<br>Cyanoacrylate | Nitinol, PTFE,<br>Platinum, SS 301,<br>Solder, Polypropylene<br>sheath, Urethane,<br>Cyanoacrylate | Nitinol, PTFE,<br>Platinum, Solder,<br>Polypropylene sheath,<br>Urethane<br>Cyanoacrylate | Nitinol, PTFE,<br>Platinum, Solder,<br>Polypropylene sheath,<br>Urethane<br>Cyanoacrylate |
| Plug (Implant)<br>descripton | Self-expandabable,<br>ovoid shaped frame<br>with a PTFE cover<br>over the proximal<br>portion | Self-expandabable,<br>ovoid shaped frame<br>with a PTFE cover<br>over the proximal<br>portion | Self-expandabable,<br>ovoid shaped frame<br>with a PTFE cover<br>over the proximal<br>portion | Self-expandabable,<br>ovoid shaped frame<br>with a PTFE cover<br>over the proximal<br>portion |
| Plug Diameter,<br>Unconstrained | 5.3 mm | 6.5 mm | 5.3 mm | 6.5 mm |
| Plug Length,<br>Unconstrained | 12 mm | 12 mm | 12 mm | 12 mm |
| Target Vessel<br>Diameter | 1.5-3.0 mm | 3.0-5.0 mm | 1.5-3.0 mm | 3.0-5.0 mm |
| Method of<br>Placement | Delivery wire through<br>a 0.021" ID<br>Microcatheter | Delivery wire through<br>a 0.027" ID<br>Microcatheter | Delivery wire<br>through a 0.021" to<br>0.027" ID<br>Microcatheter | Delivery wire<br>through a 0.027" ID<br>Microcatheter |
| Radiopaque<br>Markers | Platinum marker bands<br>at each end of the plug | Platinum marker<br>bands at each end of<br>the plug | Platinum marker<br>bands at each end of<br>the plug | Platinum marker<br>bands at each end of<br>the plug |
| Proximal End of<br>Plug Config. | Proximal marker band<br>attached to delivery<br>wire | Proximal marker band<br>attached to delivery<br>wire | Proximal marker<br>band attached to<br>delivery wire | Proximal marker<br>band attached to<br>delivery wire |
| Delivery Wire<br>Length | 180 cm | 180 cm | 160 cm | 160 cm |
| Detachment System | Electrolytic | Electrolytic | Mechanical | Mechanical |
| Sterilization<br>Process | EO | EO | EO | EO |
| Accessories | Electrolytic Box and<br>Cables | Electrolytic Box and<br>Cables | Torquer | Torquer |
# Table 5-1 Comparison to the Predicates
{2}------------------------------------------------
## Performance Tests - Non-clinical
Due to the change in detachment method and delivery wire length the following design verification tests were conducted in accordance with Reverse Medical Design Control procedures. All testing was performed on units that were sterilized and met all inspection criteria. Tests on the Reverse Medical MVP System included:
- . Dimensional Inspection
- Visual Inspection
- Microcatheter Compatibility within "Simulated Use Vascular Model" .
- Flexibility within microcatheter o
- Delivery wire kinking assessment o
- Multiple deployments and withdrawals through the microcatheter O
- Force required to deploy and retract device within the microcatheter O
- Detachment Evaluations
- o Number of turns required to detach
- Torque strength of detachment junction o
- Galvanic Corrosion per ASTM G71 .
- MRI Compatibility per ASTM F-2503
- Package Integrity and Shelf-life .
All tests successfully passed acceptance criteria. This demonstrates that the modified devices meet the product specification.
## Basis for Determination of Substantial Equivalence
Upon reviewing the performance data and comparing intended use, design, materials, principle of operation and overall technological characteristics, the modified Reverse Medical MVP System is determined to be substantially equivalent to the currently marketed Reverse Medical MVP System. Differences between the systems do not raise any issues of safety or effectiveness.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized graphic of three abstract shapes that resemble a human figure.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 18, 2014
Reverse Medical Corporation % Jane Metcalf VP, Quality Assurance, Regulatory and Clinical Affairs 13700 Alton Parkway, Suite 167 Irvine, California 92618
Re: K141313
> Trade/Device Name: MVP Micro Vascular Plug System Regulation Number: 21 CFR 870.3300 Regulation Name: Device Embolization, Vascular Regulatory Class: 11 Product Code: KRD Dated: May 16, 2014 Received: May 20, 2014
Dear Ms. Metcalf,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or 10 devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, howeyer, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{4}------------------------------------------------
Page 2 - Jane Metcalf
the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours
for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
### Indications for Use
510(k) Number (if known) K141313
#### Device Name
Reverse Medical(R) Corporation MVP(R) Micro Vascular Plug System
#### Indications for Use (Describe)
The Reverse Medical(R) Corporation MVP(R) Micro Vascular Plug System is indicated to obstruct or reduce the rate of blood flow in the peripheral vasculature.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY "
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.