K031810 · Aga Medical Corp. · KRD · Sep 9, 2003 · Cardiovascular
Device Facts
Record ID
K031810
Device Name
AMPLATZER VASCULAR PLUG
Applicant
Aga Medical Corp.
Product Code
KRD · Cardiovascular
Decision Date
Sep 9, 2003
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.3300
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The AMPLATZER Vascular Plug is indicated for arterial and venous embolizations in the peripheral vasculature.
Device Story
AMPLATZER Vascular Plug is a self-expandable, cylindrical embolization device composed of Nitinol wire mesh with platinum marker bands at both ends. It is preloaded in a PTFE loader and delivered via standard 5 or 6 French guiding catheters using a stainless steel delivery cable. The device is deployed percutaneously by a physician in a clinical setting to occlude vessels in the peripheral vasculature. The plug is mechanically released from the delivery cable via an unthreading mechanism. By providing a stable, self-expanding structure that conforms to the vessel wall, the device facilitates rapid occlusion and endothelialization, offering a more precise positioning alternative to traditional embolization coils. The healthcare provider uses fluoroscopic guidance to position and deploy the plug, with the output being the physical occlusion of the target vessel to treat conditions requiring embolization.
Clinical Evidence
Animal study (canine model) compared AMPLATZER Vascular Plug to standard coils. Results: Plug showed significantly shorter occlusion time (5.0 ± 3.5 min vs 10.2 ± 9.8 min, P < 0.001) and superior positioning precision. Recanalization rates were lower for plugs at 1 week (15% vs 33%) and 1 month (78% vs 80%). At 2 months, 88% of plug-occluded vessels remained occluded vs 0% for coils. Plugs showed firm attachment and endothelialization. Bench testing confirmed mechanical performance, including pull strength, luer compatibility, and deployment force.
Indicated for arterial and venous embolizations in the peripheral vasculature for patients requiring vessel occlusion.
Regulatory Classification
Identification
A vascular embolization device is an intravascular implant intended to control hemorrhaging due to aneurysms, certain types of tumors (e.g., nephroma, hepatoma, uterine fibroids), and arteriovenous malformations. This does not include cyanoacrylates and other embolic agents, which act by polymerization or precipitation. Embolization devices used in neurovascular applications are also not included in this classification, see § 882.5950 of this chapter.
Special Controls
*Classification.* Class II (special controls.) The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance Document: Vascular and Neurovascular Embolization Devices.” For availability of this guidance document, see § 870.1(e).
{0}------------------------------------------------
# 6. 510(k) Summary
This 510(k) summary is being submitted in accordance with the requirements of Safe Medical Devices Act 1990 and 21 CFR 807.92.
| Contact Person: | Patrice Stromberg<br>Regulatory Affairs Associate<br>AGA Medical Corporation<br>682 Mendelssohn Avenue<br>Golden Valley, MN 55427 |
|-----------------|-----------------------------------------------------------------------------------------------------------------------------------|
| Date Prepared: | June 11, 2003 |
| Device Name: | AMPLATZER® Vascular Plug<br>Arterial Embolization Device (79KRD)<br>(21 CFR 870.3300) |
#### Device Description:
The AMPLATZER Vascular Plug is provided sterile and is intended for one-time use. The AMPLATZER Vascular Plug is a self-expandable, cylindrical device made from a Nitinol wire mesh. The device is secured on both ends with platinum marker bands. The delivery system is comprised of a 135 cm long, stainless steel delivery cable and a PTFE loader. The AMPLATZER Vascular Plug is preloaded in a loader and delivered through currently available standard 5 or 6 French guiding catheters. The device will be available in the following sizes.
| Order No. | Vascular | Device | Catheter Requirements* | | |
|------------|----------|--------|------------------------|----------------|---------|
| | Plug | Length | Minimum | Minimum | Maximum |
| | Diameter | | Size | Inner Diameter | Length |
| 9-PLUG-004 | 4 mm | 7 mm | 5 French | .056" | 100 cm |
| 9-PLUG-006 | 6 mm | 7 mm | 5 French | .056" | 100 cm |
| 9-PLUG-008 | 8 mm | 7 mm | 5 French | .056" | 100 cm |
| 9-PLUG-010 | 10 mm | 7 mm | 6 French | .067" | 100 cm |
| 9-PLUG-012 | 12 mm | 8 mm | 6 French | .067" | 100 cm |
| 9-PLUG-014 | 14 mm | 8 mm | 6 French | .067" | 100 cm |
| 9-PLUG-016 | 16 mm | 8 mm | 6 French | .067" | 100 cm |
* For use with standard guiding catheters meeting minimum internal diameter requirements.
#### Intended Use:
The AMPLATZER Vascular Plug is indicated for arterial and venous embolizations in the peripheral vasculature.
## Predicate Devices:
The AMPLATZER Vascular Plug is similar to many devices already in commercial distribution for arterial and venous embolization. These devices include an Embolization Coil Positioner Set (COOK, Incorporated), Flipper Detachable Coil (COOK, Incorporated), Fibered Platinum Coil (Boston Scientific Target), and Guglielmi Detachable Coil (Boston Scientific Target). The devices
{1}------------------------------------------------
are introduced percutaneously using a catheter or microcatheter introducer. The following is a comparison of technological characteristics. Also refer to Table 6-1 for summary of Technological Characteristics.
The Embolization Coil Positioner Set was reviewed as substantially equivalent under D.C. K940189 and is indicated for arterial and venous embolization. The device is constructed of stainless steel and synthetic fiber with a coil wire diameter of 0.018 to 0.038 inches. The coils are available in straight or curled shapes with an emboli size range of 2 to 20 mm. A push-button release mechanism is the method of deployment.
The Flipper Detachable Coil was reviewed as substantially equivalent under D.C. K993455 and is indicated for arterial and venous embolization for the peripheral vasculature. The device is constructed of stainless steel and synthetic fiber with a coil wire diameter of 0.035 inches. The coils are available in curled shapes with an emboli diameter range of 3 to 8 mm. Deployment is achieved when interlocking threads between the coils and the delivery wire are unscrewed.
The Fibered Platinum Coil was reviewed as substantially equivalent under D.C. K955293 and is indicated for arterial and venous embolization in the peripheral vasculature. The device is constructed of platinum and synthetic fiber with a coil diameter of 0.010 to 0.035 inches. The coils are available in the following shapes: straight, C-shaped, helical and complex helical. The emboli size range is 2 to 30 mm. Deployment is achieved by a wire guide which pushes the coil out of the catheter.
The Guglielmi Detachable Coil was reviewed as substantially equivalent under D.C. K951256, K960705 and K962503 and is indicated for embolization of intracranial aneurysms, arteriovenous malformations, arteriovenous fistulae and arterial venous embolizations in the peripheral vascular. The device is constructed of platinum with a coil wire diameter of 0.010 to 0.018 inches. The coils are available in a helical shape with an emboli size range of 2 to 20 mm. The coils are deployed by electrolytic detachment from the wire guide.
·
{2}------------------------------------------------
| Device | Design | Coil wire<br>diameter | Shape |
|--------------------------------------------------------|----------------------------------------|-----------------------|--------------------------------------------------------|
| COOK, Embolization Coil<br>Positioner Set | Stainless steel and synthetic<br>fiber | 0.018-0.038" | Straight or curled |
| COOK, Flipper Detachable<br>Embolization Coil | Stainless steel and synthetic<br>fiber | 0.035" | Curled |
| Boston Scientific Target,<br>Fibered Platinum Coil | Platinum and synthetic fiber | 0.010 - 0.035" | Straight, C-shaped,<br>helical; and complex<br>helical |
| Boston Scientific Target,<br>Guglielmi Detachable Coil | Platinum | 0.010-0.018" | Helical |
| AMPLATZER Vascular<br>Plug | Nitinol wire | .0015" - .003" | Cylindrical, "plug" |
Table 6-1 Summary of Technological Characteristics Compared to Predicate Device
## Safety and Performance:
Animal Testing was conducted to evaluate the efficacy and recanalization rates of the AMPLATZER Vascular Plug as compared to standard coils. All procedures were performed successfully without misplacement. The following results were obtained:
- . Occlusion time was significant shorter with the AMPLATZER Vascular Plugs as compared to Cook coils (5.0 ± 3.5 minutes vs. 10.2 ± 9.8 minutes, P < 0.001).
- . Technically, positioning of the coils was far less precise than positioning of the AMPLATZER Vascular Plugs. Recanalization occurred with both the plugs and coils, but was more prominently seen in the coil group (33% and 80% at 1 week and 1 month vs. 15% and 78% for plugs) although it is well known from clinical practice that recanalization of coils embolized vessels in human subjects is exceedingly rare. The higher incidence of recanalizations in canines is due to the more active fibernolitic system of this species.
- Occluder stability was assessed immediately following deployment. No . migrations occurred.
- At 2 months, all vessels occluded with coils (100%) had become . recanalized compared to 16 out of 18 (88%) with the AMPLATZER Vascular Plug.
- Gross examination of occluded vessels showed all AMPLATZER . Vascular Plugs to be well endothelialized and firmly attached to the vessel wall, whereas spring coils were easy to shell out.
{3}------------------------------------------------
Bench testing was done to evaluate the ability of the AMPLATZER Vascular Plug and Delivery System to perform in accordance with the requirements of the design plan.
# AMPLAZER Vascular Plug Test Results:
- Vascular Plug Elongation Study Linear measurements of the Vascular ● Plug prior to and under a compression restraint.
- Pull Testing Laser Welded Marker Bands on Vascular Plug ●
- Pull Testing Screw Attachment to Marker Band Laser Weld .
- Threaded Connection Pull Testing Pull testing to verify the strength . property of the connection in between the delivery cable screw and the device end screw attachment.
- Vascular Plug detachment (unthreading) from delivery cable. .
## AMPLATZER Delivery System Test Results:
- Pull Test Weld Joint Cable to Cable Screw Verify weld strength ●
- Luer Testing Loader hubs were subjected to ISO 594 luer testing. .
- Loader Hub to PTFE tubing tensile strength -- Pull tested until hub . separation from the loader tubing.
- Loader Hub Luer Fitting Hoop dispenser and loader will allow . flushing with a syringe.
- Passage and Deployment through 5F or 6F Guide Catheters Measure . advancement force in the catheters.
The following Biocompatibility Testing was performed.
- . Cytotoxicity
- Hemolysis .
- Sensitization ●
- Ames Salmonela ●
- Intracutaneous Injection .
- Systemic Toxicity ●
The results of these tests provide reasonable assurance that the device has been designed and tested to assure conformance to the requirements for its use for arterial and venous embolization in the peripheral vasculature.
#### Statement of Equivalence:
The AMPLATZER Vascular Plug is substantially equivalent in intended use to currently marketed coils indicated for arterial and venous embolizations in the peripheral vascular. The proposed device is substantially equivalent to the Embolization Coil Positioner Set (COOK, Incorporated), Flipper Detachable Embolization Coil (COOK, Incorporated), the Fibered Platinum Coil (Boston Scientific Target), and the Guglielmi Detachable Coil (Boston Scientific Target).
{4}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Image /page/4/Picture/2 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with three lines representing its wings.
SEP - 9 2003
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
AGA Medical Corporation c/o Mr. Patrice M. Stromberg 682 Mendelssohn Avenue Golden Valley, MN 55427
Re: K031810
AMPLATZER® Vascular Plug Regulation Number: 21 CFR 870.3300 Regulation Name: Arterial embolization device Regulatory Class: Class III (three) Product Code: 79 KRD Dated: June 11, 2003 Received: June 12, 2003
Dear Mr. Stromberg:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Page 2 - Mr. Patrice M. Stromberg
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4646. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
Page 1 __of_1
510(k) Number (if known): K031810
Device Name: AMPLATZER Vascular Plug
Indications For Use:
.
The AMPLATZER Vascular Plug is indicated for arterial and venous embolizations in the peripheral vasculature.
(Please Do NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use Only
(Optional Format 3-10-98)
N. DeCortin
(Division Sign-Off)
Division of Cardiovascular Devices
510(k) Number K03810
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.