Cerebral Adaptive Index (CAI) Algorithm is intended to be used by qualified personnel or trained clinicians in a critical care environment in a hospital setting. The algorithm is intended to show the level of coherence between MAP and cerebral StO2.
Device Story
CAI algorithm is a software-based derived parameter quantifying the dynamic relationship between Mean Arterial Pressure (MAP) and cerebral tissue oxygenation (StO2). Inputs include continuous MAP data from HemoSphere Pressure Cable and StO2 data from ForeSight Oximeter Cable. The algorithm calculates coherence between these inputs, outputting an index value (0-100) and trended graph displayed at 20-second intervals. Used in hospital critical care settings (OR/ICU) by clinicians. High CAI (≥45) indicates strong coherence; low CAI (<45) indicates weak/moderate coherence. Output assists clinicians in identifying hemodynamic causes of cerebral desaturation events (e.g., hypotension vs. inadequate oxygen content) by providing continuous, objective assessment of MAP-StO2 relationship, reducing reliance on subjective visual/mental estimation.
Clinical Evidence
Retrospective clinical study of 145 adult patients (18+) across 4 US sites (cardiac/general surgery, ICU). Performance evaluated against Pearson's Correlation Coefficient (Corr) of MAP and StO2 time-series data. Threshold of 45 used to discriminate strong (Corr≥0.7) vs. weak/moderate (0<Corr<0.7) association. Results: ROC AUC 0.88 [0.85, 0.90], sensitivity 0.84 [0.78, 0.88], specificity 0.80 [0.76, 0.85]. Performance consistent across sites.
Technological Characteristics
Software-based algorithm processing hemodynamic signals (MAP, StO2). Operates on HemoSphere Advanced Monitoring Platform. Connectivity via HemoSphere Pressure Cable and ForeSight Oximeter Cable. Continuous display at 20-second intervals. No alarm ranges. Index range 0-100.
Indications for Use
Indicated for patients over 18 years of age receiving advanced hemodynamic monitoring to assess the level of coherence between Mean Arterial Pressure (MAP) and cerebral tissue oxygenation (StO2). Not indicated for treatment of any disease or condition; no therapeutic decisions should be made based solely on the CAI.
Regulatory Classification
Identification
An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.
Predicate Devices
HemoSphere ForeSight Oximeter Cable on HemoSphere Advanced Monitoring Platform (K213682)
{0}------------------------------------------------
May 26, 2023
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Edwards Lifesciences, LLC Sara Pesian Manager, Regulatory Affairs One Edwards Way Irvine, CA 92614
Re: K223651
Trade/Device Name: Cerebral Adaptive Index (CAI) Algorithm Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: Class II Product Code: MUD, QEM Dated: April 28, 2023 Received: April 28, 2023
Dear Sara Pesian:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Patrick Antkowiak -S
Patrick Antkowiak Acting Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K223651
Device Name Cerebral Adaptive Index (CAI) Algorithm
#### Indications for Use (Describe)
Cerebral Adaptive Index (CAI) Algorithm is an informational index to help assess the level of coherence or lack thereof between Mean Arterial Pressure (MAP) and the Absolute Levels of Blood Oxygenation (StO2) in patient's cerebral tissue. MAP is acquired by the HemoSphere Pressure Cable and StO2 is acquired by the ForeSight Oximeter Cable.
CAI is intended for use in patients over 18 years of age receiving advanced hemodynamic monitoring.
CAI is not indicated to be used for treatment of any disease or condition and no therapeutic decisions should be made based solely on the Cerebral Adaptive Index (CAI) Algorithm.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# K223651
| Sponsor: | Edwards Lifesciences LLC<br>One Edwards Way<br>Irvine, CA 92614 |
|------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Establishment<br>Registration<br>Number: | 2015691 |
| Contact<br>Person: | Sara Pesian<br>Manager, Regulatory Affairs<br>One Edwards Way<br>Irvine, CA 92614<br>sara_pesian@edwards.com<br>Telephone: (949) 250-2232 |
| Date: | May 26, 2023 |
| Trade Name: | Cerebral Adaptive Index (CAI) Algorithm |
| Common<br>Name: | Oximetry |
| Classification<br>Name: | Oximeter 21 CFR 870.2700 |
| Product<br>Code: | MUD, Class II<br>QEM, Class II |
| Primary<br>Predicate<br>Device: | HemoSphere ForeSight Oximeter Cable on HemoSphere Advanced Monitoring<br>Platform, manufactured by Edwards Lifesciences, K213682 cleared June 22, 2022.<br>is being utilized for substantial equivalence since it contains the same StO2<br>algorithm as the subject device, and similar indications for use/intended use and<br>principle of operation and technology as the subject CAI Algorithm. |
| Reference<br>Predicate<br>Device (s) | FORE-SIGHT ELITE Absolute Tissue Oximeter, manufactured by Edwards<br>Lifesciences, K190205 cleared August 29, 2019 and HemoSphere Pressure<br>Cable, manufactured by Edwards Lifesciences, K180881 cleared November 16,<br>2018 are being utilized as reference devices |
# 510(k) Summary – Cerebral Adaptive Index (CAI) Algorithm
{4}------------------------------------------------
| Device<br>Description: | Cerebral Adaptive Index (CAI) Algorithm is a derived parameter that quantifies<br>the dynamic relationship between two existing hemodynamic parameters, Mean<br>Arterial Pressure (MAP) and the Absolute Levels of Blood Oxygenation<br>Saturation (StO2) in the cerebral tissue. CAI is intended to show the level of<br>coherence between MAP and cerebral StO2. The output will be represented as an<br>index value and a trended graph.<br><br>MAP is acquired from the HemoSphere Pressure Cable (initially cleared in<br>K180881 on November 16, 2018). StO2 used for computing CAI is acquired from<br>the ForeSight Oximeter Cable (cleared in K201446 on October 1, 2020).<br><br>The CAI parameter can enhance clinician's understanding of the underlying<br>hemodynamic changes behind cerebral desaturation events. It helps the clinician<br>recognize/ identify possible causes of, for example, decrease in StO2 and clinical<br>events related to StO2 decrease (e.g., hypotension as opposed to inadequate<br>oxygen content).<br><br>CAI will be continuously displayed at 20-second rate. The parameter will not<br>have any alarm ranges and will only be represented as a number with a range<br>between 0 to 100. A high CAI value (CAI ≥45) means that MAP and StO2 have a<br>greater coherence and informs the clinician that alterations in MAP may result in<br>concomitant changes in cerebral oxygen saturation Whereas a low CAI value<br>(CAI < 45) means there is lesser coherence between the two parameters, and<br>therefore alterations in MAP may not result in concomitant changes in cerebral<br>oxygen saturation. |
|---------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications<br>for Use: | Cerebral Adaptive Index (CAI) Algorithm is an informational index to help<br>assess the level of coherence or lack thereof between Mean Arterial Pressure<br>(MAP) and the Absolute Levels of Blood Oxygenation Saturation (StO2) in<br>patient's cerebral tissue. MAP is acquired by the HemoSphere pressure cable and<br>StO2 is acquired by the ForeSight oximeter cable. CAI is intended for use in<br>patients over 18 years of age receiving advanced hemodynamic monitoring. CAI<br>is not indicated to be used for treatment of any disease or condition and no<br>therapeutic decisions should be made based solely on the Cerebral Adaptive<br>Index (CAI) Algorithm. |
| Intended Use: | Cerebral Adaptive Index (CAI) Algorithm is intended to be used by qualified<br>personnel or trained clinicians in a critical care environment in a hospital setting.<br>The algorithm is intended to show the level of coherence between MAP and<br>cerebral StO2. |
| Comparison<br>to Predicate<br>Device: | The HemoSphere ForeSight Oximeter Cable (K213682 most recently cleared June<br>22, 2022) is chosen as the predicate since it contains the same StO2 algorithm, and<br>similar indications for use/intended use and principle of operation and technology<br>as the subject CAI algorithm when used in conjunction with MAP data. The subject |
{5}------------------------------------------------
CAI and the existing StO2 algorithms both use the measurement of absolute regional hemoglobin oxygen saturation of blood under the sensor in their calculations. When sensors are applied to patient's cerebral region, CAI uses the StO2 signals in conjunction with MAP signals and calculates the coherence between the measured cerebral StO2 and MAP. Hence the HemoSphere ForeSight Oximeter Cable is being used to establish substantial equivalence for the subject CAI.
The following verification activities were performed in support of a substantial equivalence determination for CAI Algorithm and HemoSphere Foresight Oximeter Cable and to ensure the safety and effectiveness of the Cerebral Adaptive Index (CAI) Algorithm.
#### Performance Software Verification
Data (Bench and/or Clinical):
Software verification was performed per FDA's Guidance for Industry and FDA Staff, Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices (issued May 11, 2005). The Cerebral Adaptive Index was tested at the algorithm level to ensure the safety of the device. The validation of the CAI Algorithm was performed as follows:
- 1. Validation using simulated time-series data
- 2. Validation using data retrospectively collected from animals
The algorithm passed all testing.
#### Clinical Performance Data
The following clinical performance data was submitted to support a determination of substantial equivalence:
#### Demographics:
The demographics of CAI Algorithm validation dataset are captured in Table 1-1 below.
#### Patient Baseline Characteristics:
The validation dataset had no inclusion/exclusion criteria related to patient baseline characteristics. In general, adult surgical patients (cardiac surgery, general surgery and surgical ICU) whose StO2 and MAP were being monitored respectively via Foresight Sensors and Flotrac Sensors were randomly selected for the retrospective analysis.
#### Enrollment:
The validation dataset was retrospectively obtained from 4 different clinical sites within US (Northwestern University, Chicago; UC Davis, Sacramento; University of Minnesota, Minneapolis; Stanford University, Stanford). A total of 145 subjects aged 18 or older were randomly selected for inclusion in the CAI Algorithm
{6}------------------------------------------------
validation study. Table 1-1 provides the number of patients enrolled from each site as well as the patient demographics and the surgery types from all four sites.
| Site | Nº of<br>patients | Age<br>(years) | Gender | Height<br>(cm) | Weight<br>(Kg) | Clinical<br>Environment |
|-----------------------------------------------------|-------------------|-----------------------|-------------------------|------------------------------|----------------------------|-------------------------|
| Northwester<br>n University,<br>Chicago<br>(NWU) | 29 | 60.1±12.3<br>[31, 76] | 7 Females<br>22 Males | 174.1±9.7<br>[155.0, 191.0] | 84.8±15.0<br>[46, 111.6] | Cardiac<br>surgery |
| UC Davis,<br>Sacramento<br>(UCD) | 43 | 62.9±15.9<br>[26, 87] | 20 Females<br>23 Males | 170.3±10.3<br>[152.4, 195.6] | 84.8±19.0<br>[41.5, 119.2] | General<br>surgery |
| University of<br>Minnesota,<br>Minneapolis<br>(UoM) | 67 | 63.1±10.1<br>[39, 84] | 14 Females<br>53 Males | 175.4±7.9<br>[154.0, 188.0] | 93.8±20.5<br>[48.0, 139.1] | Cardiac<br>surgery |
| Stanford<br>University,<br>Stanford<br>(SU) | 6 | 54.0±17.4<br>[22, 66] | 3 Females<br>3 Males | 178.8±9.7<br>[167.0, 190.0] | 94.6±38.2<br>[56.0, 157.0] | Surgical ICU |
| All | 145 | 61.9±12.8<br>[22, 87] | 44 Females<br>101 Males | 173.6±9.3<br>[152.4, 195.6] | 89.0±20.2<br>[41.5, 157.0] | OR/ICU |
Table 1-1 Summary of patient demographic information for the 4 sites used in the clinical data performance analysis. Data for Age, Height and Weight are Mean ± 1std and full range [max. minl
## General Summary:
The following is a summary of the validation procedure performed:
MAP and StO2 time-series data obtained from adult (18 years and older) patients were used for the validation testing. To validate and assess the performance of CAI, a Receiver Operating Characteristic (ROC) analysis was performed based on weak/moderate and strong MAP-StO2 association. Labeling of weak/moderate versus strong MAP-StO2 association states is conducted based on the computed values of the Pearson's Correlation Coefficient (Corr) between MAP and StO2 in the utilized time-series clinical data. Specifically, 0<Corr<0.7 were labeled as weak/moderate MAP-StO2 association and Corr≥0.7 were labeled as strong MAP-StO2 association.
{7}------------------------------------------------
Performance Metrics:
Using this data, the following performance metrics were calculated:
• Sensitivity: The true positive rate; ratio of true positives to total number of positive events. (True positives (TP) are defined as positive events that have a CAI value greater or equal than a given threshold. True negatives (TN) are defined as negative events that have a CAI value smaller than a given threshold.)
· Specificity: The false positive rate; ratio of true negatives to total number of negative events (False positives (FP) are defined as negative events that have a CAI value greater or equal than a given threshold. False negatives (FN) are defined as positive events that have a CAI value smaller than a given threshold.)
· ROC AUC: the area under the ROC curve (AUC) summarizes the performance as a single number (0.5 to 1) with a higher AUC associated with a better performing algorithm.
#### Performance Goals:
The performance goals for CAI algorithm are defined as follows: Sensitivity and Specificity ≥ 80 % at the CAI threshold of 45.
Results of Clinical Performance Testing:
The performances of CAI for the chosen threshold of 45 on the 145 subjects enrolled, are reported in Table 1-2 below.
| ROC Analysis for Clinical Data | | | |
|--------------------------------|------------------|------------------|---------------|
| ROC AUC | Sensitivity | Specificity | CAI Threshold |
| 0.88 [0.85,0.90] | 0.84 [0.78,0.88] | 0.80 [0.76,0.85] | 45 |
Table 1-3 below provides the confusion matrix that was used to calculate sensitivity/specificity for CAI threshold of 45.
Table 1-3 Confusion matrix for CAI at the chosen threshold of 45
| | | Corr | |
|-----|------------------------|-----------|------------|
| | Positive<br>(CAI≥45) | 3176 (TP) | 18235 (FP) |
| CAI | Negative<br>(CAI < 45) | 625 (FN) | 74110 (TN) |
{8}------------------------------------------------
As shown in Table 1-2, at the chosen threshold of 45, CAI can accurately discriminate conditions where MAP-StO2 association is strong from conditions where MAP-StO2 association is weak/moderate.
In order to evaluate potential site effects on CAI performance, the ROC analysis was also repeated for each site individually. The results are summarized in Table 1-4.1-5.1-6 and 1-7 below. The results demonstrate that CAI performance is consistent across different sites, as shown by the almost identical AUCs. The confidence intervals of AUCs and the sensitivity and specificity vary across sites due to different number of patients enrolled in different sites, and different number of positive events and/or negative events in these patients.
| ROC Analysis - UCD Clinical Data | | | |
|----------------------------------|------------------|------------------|---------------|
| AUC | Sens | Spec | Threshold (%) |
| 0.87 [0.82,0.92] | 0.86 [0.76,0.93] | 0.73 [0.65,0.80] | 45 |
Table 1-4 ROC analysis results for UC Davis Clinical Data
#### Table 1-5 ROC analysis results for Northwestern University Clinical Data
| ROC Analysis - NWU Clinical Data | | | |
|----------------------------------|------------------|------------------|---------------|
| AUC | Sens | Spec | Threshold (%) |
| 0.87 [0.84,0.91] | 0.90 [0.83,0.94] | 0.71 [0.63,0.78] | 45 |
## Table 1-6 ROC analysis results for University of Minnesota Clinical Data
| AUC | Sens | Spec | Threshold (%) |
|--------------------------|-------------------|-------------------|---------------|
| <b>0.89</b> [0.85, 0.92] | 0.65 [0.53, 0.76] | 0.93 [0.92, 0.95] | 45 |
Table 1-7 ROC analysis results for Stanford University Clinical Data
| ROC Analysis - SU Clinical Data | | | |
|---------------------------------|-------------------|-------------------|---------------|
| AUC | Sens | Spec | Threshold (%) |
| <b>0.87</b> [0.70, 0.98] | 0.96 [0.90, 0.99] | 0.75 [0.52, 0.92] | 45 |
Benefit-Risk There was no exclusion of data across the spectrum of weak, moderate, and Analysis strong MAP-StO2 association states in the post-hoc statistical analysis. All patient conditions were included in the analysis.
> CAI quantifies the dynamic relationship between two existing hemodynamic parameters, MAP and StO2 in the cerebral tissue, to show the level of coherence between the two parameters. The ROC analysis against Corr demonstrates that
{9}------------------------------------------------
CAI can accurately differentiate weak/moderately MAP-StO2 association from strong MAP-StO2 association states at the chosen threshold of 45. Although the agreement between CAI and Corr is strong but not perfect, the benefits outweigh the risks, as CAI can enhance clinician's understanding of the underlying hemodynamic changes behind cerebral desaturation events that may not be easily identifiable if just visually looking at MAP and StO2 values individually (which is the current clinical practice). Additionally, CAI is continuously calculated and displayed, enabling clinicians to continuously assess the underlying hemodynamic changes instead of visually/mentally calculating the relationship which is more susceptible to error and subjective judgement.
While CAI will be shown on the monitor as an index value and a trended graph, both StO2 and MAP trends will still be individually displayed on the monitor for review by the clinician and to aid them make treatment decisions based on these individual values.
#### Conclusions Overall Conclusion:
The Cerebral Adaptive Index (CAI) Algorithm has successfully passed functional and performance testing, including software verification and validation, bench, and clinical analysis. Completion of all performance verification and validation activities demonstrated that the subject device meets its predetermined design and performance specifications. Verification activities performed confirmed that the differences in the features and design did not adversely affect the safety and effectiveness of the subject device. Note that the lower bound CI performance goal was not pre-specified, but the observed results demonstrate safety and effectiveness. The testing performed demonstrates that the Cerebral Adaptive Index (CAI) Algorithm is substantially equivalent to its legally marketed predicate.
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.