The HemoSphere Advanced Monitoring Platform is intended to be used by qualified personnel or trained clinicians in a critical care environment in a hospital setting. The HemoSphere Advanced Monitoring Platform is intended for use with the Edwards Swan-Ganz and Oximetry Catheters and FloTrac, FloTrac IQ and TruWave DPT sensors.
Device Story
HemoSphere Advanced Monitoring Platform is a modular system for hemodynamic monitoring in hospital critical care settings. Inputs include data from Swan-Ganz catheters, oximetry catheters, pressure sensors (FloTrac/TruWave), and the FORE-SIGHT ELITE tissue oximeter module; analog ECG/pressure signals can be slaved from external monitors. The system processes these inputs to calculate and display hemodynamic parameters (e.g., CO, CI, SV, SVR, SvO2, StO2) and provides a Fluid Responsiveness Test (passive leg raise/fluid bolus). The Acumen Hypotension Prediction Index (HPI) feature analyzes arterial pressure waveforms to provide insight into the likelihood of future hypotensive events. The device is operated by clinicians/nurses. Output is displayed on the monitor to assist in clinical decision-making regarding fluid status and cardiac function. Benefits include continuous, real-time assessment of patient physiological status to guide goal-directed therapy.
Clinical Evidence
No clinical data was required. Evidence consists of bench testing, system-level verification, and a usability study involving 32 users (clinicians and nurses) conducted per FDA human factors guidance. All verification activities passed.
Technological Characteristics
Modular hemodynamic monitor. Interfaces with catheters/sensors via dedicated modules. Connectivity: analog signal input/output, wireless capability. Standards: IEC 60601-1, 60601-1-2, 60601-1-6, 60601-1-8, 62304, 62366, 60601-2-34, 60601-2-49. Software: Major/Moderate level of concern. Features include tissue oximetry (StO2), fluid responsiveness testing, and Acumen HPI.
Indications for Use
Indicated for adult and pediatric critical care patients requiring hemodynamic monitoring (cardiac output, venous oxygen saturation, pressure, and derived parameters) in hospital environments (OR, ICU, ER). Tissue oximetry module indicated for patients at risk for ischemic states (adults/adolescents >40kg; pediatric ≥3kg; cerebral pediatric <8kg; non-cerebral pediatric <5kg). Acumen HPI feature indicated for OR patients receiving advanced hemodynamic monitoring.
Regulatory Classification
Identification
A programmable diagnostic computer is a device that can be programmed to compute various physiologic or blood flow parameters based on the output from one or more electrodes, transducers, or measuring devices; this device includes any associated commercially supplied programs.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the U.S. Department of Health and Human Services and the U.S. Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. Food & Drug Administration" in blue.
August 29, 2019
Edwards Lifesciences, LLC Chirag Shah Associate Manager, Regulatory Affairs One Edwards Way Irvine, California 92614
Re: K190205
Trade/Device Name: HemoSphere Advanced Monitor, HemoSphere Swan-Ganz Module, HemoSphere Oximetry Cable, HemoSphere Pressure Cable, Acumen Hypotension Prediction Index feature and HemoSphere Tissue Oximetry Module Regulation Number: 21 CFR 870.1425 Regulation Name: Programmable Diagnostic Computer Regulatory Class: Class II Product Code: DOK, DOE, OAO, MUD Dated: July 18, 2019 Received: July 22, 2019
Dear Chirag Shah:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Stephen Browning Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K190205
#### Device Name
HemoSphere Advanced Monitor, HemoSphere Swan-Ganz Module, HemoSphere Oximetry Cable, HemoSphere Pressure Cable, Acumen Hypotension Prediction Index feature and HemoSphere Tissue Oximetry Module
#### Indications for Use (Describe)
HemoSphere Advanced Monitor with HemoSphere Swan-Ganz Module
The HemoSphere Advanced Monitor when used with the HemoSphere Swan-Ganz Module and Edwards Swan-Ganz Catheters is indicated for use in adult and pediatric critical care patients requiring monitoring of cardiac output [continuous (CO) and intermittent (iCO)] and derived hemodynamic parameters. It may be used for monitoring hemodynamic parameters in conjunction with a perioperative goal directed therapy protocol in a hospital environment. Refer to the Edwards Swan-Ganz catheter indications for use statement for information on target patient population specific to the catheter being used.
Refer to the Intended Use statement below for a complete list of measured and derived parameters available for each patient population.
#### HemoSphere Advanced Monitor with HemoSphere Oximetry Cable
The HemoSphere Advanced Monitor when used with the HemoSphere Oximetry cable and Edwards is indicated for use in adult and pediatic critical care patients requiring of yenous oxygen saturation (SvO2 and Scv02) and derived hemodynamic parameters in a hospital environment. Refer to the Edwards oximetry catheter indications for use statement for information on target patient population specific to the catheter being used.
Refer to the Intended Use statement for a complete list of measured and derived parameters available for each patient population.
#### HemoSphere Advanced Monitor with HemoSphere Pressure Cable
The HemoSphere Advanced Monitor when used with the HemoSphere Pressure Cable is indicated for use in critical care patients in which the balance between cardiac function, fluid status, vascular resistance and pressure needs continuous assessment. It may be used for monitoring hemodynamic parameters in conjunction with a perioperative goal directed therapy protocol in a hospital environment. Refer to the Edwards FloTrac, Acumen IQ and TruWave DPT sensor indications for use statement for information on target patient population specific to the sensor being used.
The Edwards Lifesciences Acumen Hypotension Prediction Index feature provides the clinician with physiological insight into a patient's likelihood of future hypotensive events (defined as mean arterial pressure < 65 mmHg for at least one minute in duration) and the associated hemodynamics. The Acumen HPI feature is intended for use in opeating room (OR) only patients receiving advanced hemodynamic monitoring. The Acumen HPI feature is considered to be additional quantitative information regarding the patient's physiological condition for reference only and no therapeutic decisions should be made based solely on the Hypotension Prediction Index (HPI) parameter.
Refer to the Intended Use statement for a complete list of measured and derived parameters available for each patient population.
HemoSphere Advanced Monitor with HemoSphere Tissue Oximetry Module
The noninvasive FORE-SIGHT ELITE tissue oximeter module is intended for use as an adjunct monitor of absolute regional hemoglobin oxygen saturation of blood under the Sensors in individuals at risk for reduced-flow or no-flow ischemic states. The FORE-SIGHT ELITE tissue oximeter module is intended to allow for the display of StO2 on the HemoSphere advanced monitor.
· When used with large sensors, the FORE-SIGHT ELITE tissue oximeter module is indicated for use on adults and
{3}------------------------------------------------
transitional adolescents >40 kg.
· When used with Medium Sensors, the FORE-SIGHT ELITE tissue oximeter module is indicated for use on pediatric subjects ≥3 kg.
· When used with Small Sensors, the FORE-SIGHT ELITE tissue oximeter module is indicated for cerebral use on pediatric subjects < 8 kg and non-cerebral use on pediatric subjects <5kg.
Refer to the Intended Use statement for a complete list of measured and derived parameters available for each patient population.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
K190205 Page 1 of 10
# 510(k) Summary – HemoSphere Advanced Monitoring Platform
| Sponsor: | Edwards Lifesciences LLC<br>One Edwards Way<br>Irvine, CA 92614 |
|------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Establishment<br>Registration<br>Number: | 2015691 |
| Contact Person: | Chirag Shah<br>Program Manager, Regulatory Affairs<br>One Edwards Way<br>Irvine, CA 92614<br>Telephone: (949) 250-1580<br>Fax: (949) 809-2972 |
| Date: | August 26, 2019 |
| Platform Name | HemoSphere Advanced Monitoring Platform |
| Trade Name: | HemoSphere Advanced Monitor<br>HemoSphere Swan-Ganz Module<br>HemoSphere Oximetry Cable<br>HemoSphere Pressure Cable<br>HemoSphere Tissue Oximetry Module<br>Acumen Hypotension Prediction Index feature |
| Common Name: | Cardiac Output/Oximetry/Ejection Fraction Computer |
| Classification<br>Name: | Programmable Diagnostic Computer<br>21 CFR 870.1425<br>Fiberoptic Oximeter Catheter<br>21 CFR 870.1230<br>Adjunctive Predictive Cardiovascular Indicator<br>21 CFR 870.2210<br>Oximeter<br>21 CFR 870.2700 |
| Product Code: | DQK, Class II<br>DQE, Class II<br>QAQ, Class II<br>MUD, Class II |
| Primary<br>Predicate Device: | HemoSphere Advanced Monitoring Platform manufactured by Edwards<br>Lifesciences, K180881, cleared on November 16, 2018. |
| Additional<br>Predicate<br>Devices: | CAS Medical System Inc.'s FORE-SIGHT ELITE Absolute Tissue<br>Oximeter Monitor (K143675 cleared April 10, 2015) utilized for the tissue<br>oximetry parameter and graphical user interface.<br><br>Cheetah NICOM/Starling SV (K101487, cleared on July 2, 2010) utilized<br>for the fluid responsiveness test feature (passive leg raise and fluid bolus) |
| Device<br>Description: | The HemoSphere Advanced Monitoring platform was designed to<br>simplify the customer experience by providing one platform with modular<br>solutions for their hemodynamic monitoring needs. The user can choose<br>from the available optional sub-system modules or use multiple sub-<br>system modules at the same time. This modular approach provides the<br>customer with the choice of purchasing and/or using specific monitoring<br>applications based on their needs. Users are not required to have all of the<br>modules installed at the same time for the platform to function.<br><br>The HemoSphere Advanced Monitoring Platform consists of the<br>HemoSphere Advanced Monitor that provides a means to interact with<br>and visualize hemodynamic and volumetric data on a screen and four<br>optional external modules: the HemoSphere Swan-Ganz Module<br>(existing), the HemoSphere Oximetry Cable (existing), the HemoSphere<br>Pressure Cable (existing) and the HemoSphere Tissue Oximetry Module<br>(subject of this submission). The platform also includes the Acumen<br>Hypotension Prediction Index feature.<br><br>The existing optional HemoSphere Swan-Ganz Module and HemoSphere<br>Oximetry Cable provide an interface to connect with currently cleared and<br>commercially available Edwards Lifesciences Swan-Ganz catheters and<br>Oximetry catheters (K803058, K822350, K905458, K924650, K934742,<br>K940795, K053609 and K110167 and K160884).<br><br>The HemoSphere Pressure Cable provides an interface to connect with<br>currently cleared and commercially available Edwards Lifesciences<br>FloTrac (K152980), FloTrac IQ (K152980) and TruWave DPT sensors<br>(K142749).<br><br>The HemoSphere Pressure Cable also enables the Acumen Hypotension<br>Prediction Index (HPI) feature when connected to an Acumen IQ sensor.<br><br>The HemoSphere Tissue Oximetry Module is an interface module<br>intended to be used with the Fore-Sight Elite Tissue Oximeter Module<br>(K180003, cleared May 10, 2018) to display continuous monitoring of<br>blood oxygen saturation in the tissue (StO2). |
| | Additionally, the HemoSphere Advanced Monitoring Platform includes<br>the Fluid Responsiveness Test feature (fluid bolus and passive leg raise). |
| | The HemoSphere Advanced Monitor has an input that can be connected to<br>an external vital sign patient monitor for slaving in an analog ECG and<br>pressure signals. The HemoSphere Platform uses this analog ECG input<br>signal to calculate a heart rate that is used by the HemoSphere Swan-Ganz<br>Module to calculate certain derived parameters (e.g. HRavg, SV, RVEF<br>and EDV). |
| | The HemoSphere Pressure-Out cable enables output of analog pressure<br>signals (AP, CVP or PAP) for display on an external patient monitor. |
| Indications for<br>Use: | HemoSphere Advanced Monitor with HemoSphere Swan-Ganz<br>Module |
| | HemoSphere Advanced Monitor with HemoSphere Swan-Ganz Module<br>The HemoSphere Advanced Monitor when used with the HemoSphere<br>Swan-Ganz Module and Edwards Swan-Ganz Catheters is indicated for<br>use in adult and pediatric critical care patients requiring monitoring of<br>cardiac output [continuous (CO) and intermittent (iCO)] and derived<br>hemodynamic parameters. It may be used for monitoring hemodynamic<br>parameters in conjunction with a perioperative goal directed therapy<br>protocol in a hospital environment. Refer to the Edwards Swan-Ganz<br>catheter indications for use statement for information on target patient<br>population specific to the catheter being used. |
| | Refer to the Intended Use statement below for a complete list of measured<br>and derived parameters available for each patient population. |
| | HemoSphere Advanced Monitor with HemoSphere Oximetry Cable<br>The HemoSphere Advanced Monitor when used with the HemoSphere<br>Oximetry Cable and Edwards oximetry catheters is indicated for use in<br>adult and pediatric critical care patients requiring monitoring of venous<br>oxygen saturation (SvO2 and ScvO2) and derived hemodynamic<br>parameters in a hospital environment. Refer to the Edwards oximetry<br>catheter indications for use statement for information on target patient<br>population specific to the catheter being used. |
| | Refer to the Intended Use statement for a complete list of measured and<br>derived parameters available for each patient population. |
| | HemoSphere Advanced Monitor with HemoSphere Pressure Cable<br>The HemoSphere Advanced Monitor when used with the HemoSphere<br>Pressure Cable is indicated for use in critical care patients in which the<br>balance between cardiac function, fluid status, vascular resistance and |
{5}------------------------------------------------
{6}------------------------------------------------
{7}------------------------------------------------
pressure needs continuous assessment. It may be used for monitoring hemodynamic parameters in conjunction with a perioperative goal directed therapy protocol in a hospital environment. Refer to the Edwards FloTrac, Acumen IQ and TruWave DPT sensor indications for use statement for information on target patient population specific to the sensor being used.
The Edwards Lifesciences Acumen Hypotension Prediction Index feature provides the clinician with physiological insight into a patient's likelihood of future hypotensive events (defined as mean arterial pressure < 65 mmHg for at least one minute in duration) and the associated hemodynamics. The Acumen HPI feature is intended for use in operating room (OR) patients receiving advanced hemodynamic monitoring. The Acumen HPI feature is considered to be additional quantitative information regarding the patient's physiological condition for reference only and no therapeutic decisions should be made based solely on the Hypotension Prediction Index (HPI) parameter.
Refer to the Intended Use statement for a complete list of measured and derived parameters available for each patient population.
# HemoSphere Advanced Monitor with HemoSphere Tissue Oximetry Module
The noninvasive FORE-SIGHT ELITE tissue oximeter module is intended for use as an adjunct monitor of absolute regional hemoglobin oxygen saturation of blood under the Sensors in individuals at risk for reduced-flow or no-flow ischemic states. The FORE-SIGHT ELITE tissue oximeter module is intended to allow for the display of StO2 on the HemoSphere advanced monitor.
- When used with large sensorsis indicated for use on adults and transitional adolescents >40 kg.
• When used with Medium Sensors, the FORE-SIGHT ELITE tissue oximeter module is indicated for use on pediatric subjects >3 kg.
· When used with Small Sensors, the FORE-SIGHT ELITE tissue oximeter module is indicated for cerebral use on pediatric subjects < 8 kg and non-cerebral use on pediatric subjects <5kg.
Refer to the Intended Use statement for a complete list of measured and derived parameters available for each patient population.
{8}------------------------------------------------
The HemoSphere Advanced Monitoring Platform is intended to be used Intended Use: by qualified personnel or trained clinicians in a critical care environment in a hospital setting.
> The HemoSphere Advanced Monitoring Platform is intended for use with the Edwards Swan-Ganz and Oximetry Catheters and FloTrac, FloTrac IQ and TruWave DPT sensors.
> A comprehensive list of parameters available while monitoring with the HemoSphere advanced monitor and a connected HemoSphere Swan-Ganz Module are listed in the table below. Only iCO, iCI, iSVR, and iSVRI are available to the pediatric patient population.
| Parameter | Description | Sub-System<br>Module<br>Used | Patient<br>Population | Hospital<br>Environment |
|-----------|------------------------------------------------------|-----------------------------------|------------------------|--------------------------------------------------------------------|
| CO | continuous cardiac output | | | |
| sCO | STAT cardiac output | | | |
| CI | continuous cardiac index | | | |
| sCI | STAT cardiac index | | Adult only | Operating<br>Room,<br>Intensive<br>Care Unit,<br>Emergency<br>Room |
| EDV | right ventricular end diastolic<br>volume | HemoSphere<br>Swan-Ganz<br>Module | | |
| sEDV | STAT right ventricular end<br>diastolic volume | | | |
| EDVI | right ventricular end diastolic<br>volume index | | | |
| sEDVI | STAT right ventricular end<br>diastolic volume index | | | |
| HRavg | averaged heart rate | | | |
| LVSWI | left ventricular stroke work index | | | |
| PVR | pulmonary vascular resistance | | | |
| PVRI | pulmonary vascular resistance<br>index | | | |
| RVEF | right ventricular ejection fraction | | | |
| sRVEF | STAT right ventricular ejection<br>fraction | | | |
| RVSWI | right ventricular stroke work index | | | |
| SV | stroke volume | | | |
| SVI | stroke volume index | | | |
| SVR | systemic vascular resistance | | | |
| SVRI | systemic vascular resistance index | | | |
| iCO | intermittent cardiac output | | Adult and<br>Pediatric | |
| iCI | intermittent cardiac index | | | |
| iSVR | intermittent systemic vascular<br>resistance | | | |
| iSVRI | intermittent systemic vascular<br>resistance index | | | |
A comprehensive list of parameters available for adult and pediatric patient populations while monitoring with the HemoSphere Advanced Monitor and a connected HemoSphere oximetry cable are as listed below:
{9}------------------------------------------------
| Parameter | Description | Sub-System<br>Module<br>Used | Patient<br>Population | Hospital<br>Environment |
|-----------|-------------------------------------|------------------------------|-----------------------|---------------------------------|
| SvO2 | Mixed Venous Oxygen Saturation | HemoSphere | Adult and | Operating<br>Room,<br>Intensive |
| ScvO2 | Central Venous Oxygen<br>Saturation | Oximetry<br>Cable | Pediatric | Care Unit,<br>Emergency<br>Room |
A comprehensive list of additional parameters that are available for adult and pediatric patient populations on the HemoSphere Advanced Monitor and a connected HemoSphere Swan-Ganz Module and a connected HemoSphere Oximetry Cable are as listed below:
| Parameter | Description | Sub-System<br>Module<br>Used | Patient<br>Population | Hospital<br>Environment |
|-----------|------------------------------------------------------------------------|--------------------------------------------------------------------------|------------------------|--------------------------------------------------------------------|
| DO2 | Oxygen Delivery | | | |
| DO2I | Oxygen Delivery Indexed | HemoSphere<br>Swan-Ganz<br>Module and<br>HemoSphere<br>Oximetry<br>Cable | Adult and<br>Pediatric | Operating<br>Room,<br>Intensive<br>Care Unit,<br>Emergency<br>Room |
| VO2 | Oxygen Consumption | | | |
| VO2e | Estimated Oxygen Consumption<br>when ScvO2 is being monitored | | | |
| VO2I | Oxygen Consumption Index | | | |
| VO2Ie | Estimated Oxygen Consumption<br>Index when ScvO2 is being<br>monitored | | | |
A comprehensive list of parameters available for adult patient populations while monitoring with the HemoSphere Advanced Monitor and a connected HemoSphere pressure cable are as listed below:
| Parameter | Description | Sub-System<br>Module<br>Used | Patient<br>Population | Hospital<br>Environment |
|-----------|------------------------------------------------------------------------|---------------------------------|-----------------------|--------------------------------------------------------------------|
| CO/CI | Continuous Cardiac Output/<br>Continuous Cardiac Index | HemoSphere<br>Pressure<br>Cable | Adult only | |
| CVP | Central Venous Pressure | | | |
| DIA | Diastolic Blood Pressure | | | |
| dP/dt | Maximal slope of the arterial<br>pressure upstroke | | | |
| Eadyn | Dynamic Arterial Elastance | | | |
| MAP | Mean Arterial Pressure | | | |
| MPAP | Mean Pulmonary Arterial Pressure | | | |
| SV/SVI | Stroke Volume/Stroke Volume<br>Index | | | |
| SVR/SVRI | Systemic Vascular Resistance/<br>Systemic Vascular Resistance<br>Index | | | |
| SVV | Stroke Volume Variation | | | |
| SYS | Systolic Blood Pressure | | | Operating<br>Room,<br>Intensive<br>Care Unit,<br>Emergency<br>Room |
| HPI | Acumen Hypotension Prediction<br>Index | | | Operating<br>Room only |
A comprehensive list of parameters available for adult patient populations while monitoring with the HemoSphere Advanced Monitor and a
{10}------------------------------------------------
| Parameter | Description | Sub-System<br>Module<br>Used | Patient<br>Population | Hospital<br>Environment |
|-----------|---------------------------------------------------------------|-----------------------------------------------------------------|-----------------------|------------------------------------------------------------|
| DO2 | Oxygen Delivery | | | |
| DO2I | Oxygen Delivery Indexed | | | |
| VO2 | Oxygen Consumption | HemoSphere<br>Swan-Ganz<br>Module and<br>HemoSphere<br>Oximetry | Adult only | Operating<br>Room,<br>Intensive<br>Care Unit,<br>Emergency |
| VO2e | Estimated Oxygen Consumption<br>when ScvO2 is being monitored | | | |
| VO2I | Oxygen Consumption Index | | | |
Cable
Room
Tissue oxygen saturation. StO2, can be monitored with the HemoSphere Advanced Monitor, a connected HemoSphere Tissue Oximetry Module, and the FORE-SIGHT ELITE Tissue Oximeter Module. Refer to the FORE-SIGHT ELITE module instructions for use for specific information on the intended use environment and patient population.
| Comparison to<br>Predicate Device: | The existing HemoSphere Advanced Monitoring Platform (K180881<br>cleared November 16, 2018) consists of: |
|------------------------------------|----------------------------------------------------------------------------------------------------------|
| | • HemoSphere Advanced Monitor |
Estimated Oxygen Consumption
Index when ScvO2 is being
monitored
- HemoSphere Swan-Ganz Module ●
- HemoSphere Oximetry Cable
VO2Ie
- HemoSphere Pressure Cable ●
- . Acumen Hypotension Prediction Index
The purpose of this 510(k) submission is to add the following to the HemoSphere Advanced Monitoring Platform (K180881, cleared November 16, 2018):
- HemoSphere Tissue Oximetry Module a new sub-system ● technology module that interfaces with the Fore-Sight Elite Module Tissue Oximeter (cleared in K180003 on May 10, 2018) to display absolute levels of blood oxygenation saturation in the tissue (StO2)
- Additional new features added to the HemoSphere Advanced Monitor include:
- O
- Fluid Responsiveness Test (fluid bolus and passive leg O raise) feature
- Focused Screens o
- Modifications to existing elements of the HemoSphere Advanced ● Monitor include:
connected HemoSphere pressure cable and a connected HemoSphere oximetry cable are as listed below:
{11}------------------------------------------------
- Graphical User Interface updates to the HemoSphere o Advanced Monitor for a more modernized look to make more optimal use of screen space including addition of screens for the new HemoSphere Tissue Oximetry Module
- Acumen Hypotension Prediction Index feature Graphical User o Interface modifications (no algorithm changes)
- Modification to allow the use of dual pressure channels о simultaneously
- Modification of existing time and target screens to allow inputs O from invasive Swan-Ganz mode and associated indications for use update
- Cybersecurity enhancements о
The following predicates are used to establish substantial equivalence:
- Primary Predicate: HemoSphere Advanced Monitor (K180881, cleared November 16, 2018) utilized for substantial equivalence to the HemoSphere Advanced Monitor in terms of the graphical user interface (GUI) used. wireless module incorporated. device modularity and basic device functionality
- . Additional Predicate: CAS Medical System Inc.'s Fore-Sight Elite Absolute Tissue Oximeter Monitor (K143675 cleared April 10, 2015) is being used a predicate device since it displays the tissue oximetry parameter and provides the Graphical User Interface (GUI) for the subject device.
- Additional Predicate: Cheetah Nicom/Starling SV cleared in K101487, on July 2, 2010 was chosen to show equivalency to for the fluid responsiveness test (patient leg raise and fluid bolus) feature being added to the HemoSphere Advanced Monitoring Platform. The Cheetah Nicom/Starling SV medical device is being used as an additional predicate since it also has capabilities to perform a fluid responsiveness test. The intended use for both the subject and predicate devices is similar and hence can be used as a predicate.
The HemoSphere Advanced Monitor when used with the HemoSphere Swan-Ganz Module and the HemoSphere Oximetry Cable allows for monitoring of hemodynamic parameters, including continuous and intermittent cardiac output, right ventricular ejection fraction, end diastolic volume and mixed or central venous oxygen saturation along with additional calculated parameters.
The HemoSphere Advanced Monitor when used with the HemoSphere Pressure Cable allows for monitoring of pressure and arterial pressure based cardiac output along with Acumen Hypotension Prediction Index
{12}------------------------------------------------
(HPI). The HemoSphere Pressure Cable also calculates additional derived parameters based on the key monitored parameters. Verification and validation testing was performed to compare the performance and functionality of the HemoSphere Advanced Monitoring Platform to its predicate devices. Testing included a side-by-side comparison of the output parameters using a bench test. Performance The following verification activities were performed in support of a substantial equivalence determination for the modifications being made as Data (Bench and/or Clinical): part of this submission. System Verification Measured and derived parameters were tested using a bench simulation. Additionally, individual modules were tested at a system level to verify
the safety of these modules. They were also integrated as a system and verified for their safety and effectiveness. All tests passed.
# Electrical Safety and Electromagnetic Compatibility (EMC)
The HemoSphere Advanced Monitor and the HemoSphere Pressure Cable were tested to the following standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-6, IEC 60601-1-8, IEC 62304, IEC 62366, IEC 60601-2-34 and IEC 60601-2-49. All tests passed.
# Wireless Coexistence Testing
Bench and simulated environment testing was performed on the entire HemoSphere Advanced Monitoring Platform, including all sub-system modules and interfacing analog inputs and outputs. All tests passed.
# Software Verification
The HemoSphere Advanced Monitor and HemoSphere Swan-Ganz Module are considered as software of Major Level of Concern. The HemoSphere Oximetry Cable, HemoSphere Pressure Cable and HemoSphere Tissue Oximetry Module are considered as software of Moderate Level of Concern.
Software verification was performed per FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices". Software on each of the individual modules was tested at a sub-system level to ensure the safety of the device. All tests passed.
# Usabilitv Study
A usability study was performed for the HemoSphere Advanced Monitoring Platform in accordance with FDA's guidance, "Applying
{13}------------------------------------------------
Human Factors and Usability Engineering to Medical Devices". Testing involved 32 users with a mix of clinicians and nurses. Test Passed.
## Clinical Performance
Clinical data was not required for this device.
### Non-Clinical Performance Conclusions:
Completion of all verification and validation activities demonstrated that the subject devices meet their predetermined design and performance specifications. Verification activities performed confirmed that the differences in the design and materials used did not adversely affect the safe…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.