K143675 · Cas Medical Systems, Inc. · MUD · Apr 10, 2015 · Cardiovascular
Device Facts
Record ID
K143675
Device Name
FORE-SIGHT Elite Absolute Tissue Oximeter
Applicant
Cas Medical Systems, Inc.
Product Code
MUD · Cardiovascular
Decision Date
Apr 10, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2700
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The noninvasive FORE-SIGHT ELITE Absolute Tissue Oximeter is intended for use as an adjunct monitor of absolute regional hemoglobin oxygen saturation of blood under the sensors in individuals at risk for reduced flow or no-flow ischemic states and is indicated for use as follows: When used with Large Sensors, the FORE-SIGHT ELITE Oximeter is indicated for use on adults and transitional adolescents ≥40 kg. When used with Medium Sensors, the FORE-SIGHT ELITE Oximeter is indicated for use on pediatric subjects ≥3 kg. When used with Small Sensors, the FORE-SIGHT ELITE Oximeter is indicated for cerebral use on pediatric subjects <8 kg and non-cerebral use on pediatric subjects <5 kg.
Device Story
Noninvasive monitor of absolute regional hemoglobin oxygen saturation (StO2) in tissue microvasculature; uses multi-distance diffuse reflectance spectroscopy. Input: light absorption signals from tissue via sensors (660-900 nm). Processing: monitor unit controls LED currents, amplifies signals, and applies algorithm to determine StO2. Output: continuous numeric and real-time graphical display of StO2 values for up to four sensors. Used in clinical settings by healthcare providers; provides adjunct data for monitoring ischemic risk. Features touchscreen interface, audible/visual alarms, and data export (USB/RS-232). Benefits: provides clinicians with absolute oxygenation levels to inform decision-making regarding patient perfusion status.
Clinical Evidence
Clinical validation study on 136 subjects (ages 1 day to 17 years, including premature neonates). Measurements calibrated against weighted co-oximetry reference (30% arterial/70% venous blood samples). Tissue saturations ranged 48-96%. Accuracy (bias and precision) reported between 0.05 ± 5.06% and 0.03 ± 5.69% depending on body location. No adverse events reported.
Technological Characteristics
Multi-distance diffuse reflectance spectroscopy; 660-900 nm LED light source. Components: monitor unit, preamplifier, and sensors (Small, Medium, Large). Materials: medical adhesive and orthopedic foam/fabric (ISO 10993-1 compliant). Power: 100-240 VAC, 50/60 Hz with lead-acid battery backup. Connectivity: USB, RS-232, VGA video output. Software-based signal processing.
Indications for Use
Indicated for individuals at risk for reduced flow or no-flow ischemic states. Large sensors: adults and transitional adolescents ≥40 kg. Medium sensors: pediatric subjects ≥3 kg. Small sensors: cerebral use on pediatric subjects <8 kg; non-cerebral use on pediatric subjects <5 kg.
Regulatory Classification
Identification
An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.
Nonin Model 7600 Regional Oximeter Equanox (K113215)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/12 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular emblem with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the top half of the circle. Inside the circle is a stylized image of an eagle with three human profiles incorporated into its design.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 10, 2015
CAS Medical Systems, Inc. Ron Jeffrey Director, Regulatory Affairs 44 East Industrial Rd. Branford, Connecticut 06405
Re: K143675
Trade/Device Name: Fore-Sight Elite Absolute Tissue Oximeter Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: Class II Product Code: MUD Dated: March 18, 2015 Received: March 19, 2015
Dear Ron Jeffrey:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR
{1}------------------------------------------------
Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
\http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Mitchell Stein
forBram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
{2}------------------------------------------------
# Indications for Use
| 510(k) Number: | K143675 | | |
|---------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|
| Device Name: | FORE-SIGHT ELITE® Absolute Tissue Oximeter. | | |
| Indications for Use: | The noninvasive FORE-SIGHT ELITE Absolute Tissue Oximeter is<br>intended for use as an adjunct monitor of absolute regional hemoglobin oxygen<br>saturation of blood under the sensors in individuals at risk for reduced flow or<br>no-flow ischemic states and is indicated for use as follows: | | |
| | When used with Large Sensors, the FORE-SIGHT ELITE Oximeter is<br>●<br>indicated for use on adults and transitional adolescents ≥40 kg.<br>When used with Medium Sensors, the FORE-SIGHT ELITE Oximeter is<br>●<br>indicated for use on pediatric subjects ≥3 kg.<br>When used with Small Sensors, the FORE-SIGHT ELITE Oximeter is<br>●<br>indicated for cerebral use on pediatric subjects <8 kg and non-cerebral use on<br>pediatric subjects <5 kg. | | |
| Prescription Use ✔<br>(Part 21 CFR 801 Subpart D) | Over-the-Counter Use<br>AND/OR<br>(21 CFR 807 Subpart C) | | |
(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of ____________________________________________________________________________________________________________________________________________________________________
{3}------------------------------------------------
## 6. 510(k) Summary
Image /page/3/Picture/1 description: The image shows the word "CASMED" in a bold, red font. The letters are all capitalized and connected to each other. There is a registered trademark symbol to the right of the letter D.
## 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS K143675
| Submitter: | CAS Medical Systems, Inc. |
|----------------------|-----------------------------------------------------------------------------------------------------------------------------|
| Address: | 44 East Industrial Rd. Branford CT. 06405 USA |
| Contact: | Ron Jeffrey - Director, Regulatory Affairs<br>Phone - (203) 488-6056<br>Fax - (203) 488-9438<br>Email - rjeffrey@casmed.com |
| Prepared: | December 22, 2014 |
| Trade Name: | FORE-SIGHT ELITE® Absolute Tissue Oximeter |
| Common Name: | FORE-SIGHT Oximeter |
| Classification Name: | Oximeter, Tissue Saturation (870.2700) (MUD) |
| Predicate Device(s): | |
- � FORE-SIGHT ELITE Absolute Tissue Oximeter Monitor (K133879)
- � FORE-SIGHT Absolute Tissue Oximeter Monitor (K112820)
- ❖ Nonin Model 7600 Regional Oximeter Equanox (K113215)
{4}------------------------------------------------
#### DESCRIPTION
The FORE-SIGHT ELITE Absolute Tissue Oximeter measures hemoglobin under the Sensor, allowing the clinician to continuously and accurately determine absolute levels of blood oxygenation saturation in the tissue (StO2).
The Oximeter consists of a monitor unit, preamplifier assembly, and Small, Medium and Large Sensors. The Sensors use multiple wavelengths in the range of 660 to 900 nm to precisely measure light absorption in tissue. Sensors are sized to provide targeted penetration depths appropriate for the tissue and patient populations of interest. The monitor unit controls the measurement sequence, generating the sensor LED currents and processing the detected light signals after amplification by the dual-channel preamplifier assembly. The FORE-SIGHT algorithm determines the StO2 values for the tissue under the sensor from the light absorption values and measured patient characteristics. The monitor unit provides simultaneous measurements on up to four Sensors with both numeric and real-time graphical display formats.
The monitor unit is a mains-powered device with a field-replaceable battery backup module. A touchscreen user interface allows configuration of the Oximeter including audible, on-screen, and dedicated visual alarm indicators. The monitor display can be replicated for simultaneous remote viewing through an auxiliary VGA video output. Measurement data can be exported through various interfaces such as USB and RS-232.
#### FORE-SIGHT Oximeter Monitor Intended Use
The noninvasive FORE-SIGHT ELITE Absolute Tissue Oximeter is intended for use as an adjunct monitor of absolute regional hemoglobin oxygen saturation of blood under the sensors in individuals at risk for reduced flow or no-flow ischemic states and is indicated as follows:
When used with large sensors, the FORE-SIGHT ELITE Oximeter is indicated for use on adults and transitional adolescents ≥40 kg. When used with Medium Sensors, the FORE-SIGHT ELITE Oximeter is indicated for use on pediatric subjects ≥3 kg. When used with Small Sensors, the FORE-SIGHT ELITE Oximeter is indicated for cerebral use on pediatric subjects <8 kg and non-cerebral use on pediatric subjects <5kg.
#### FORE-SIGHT Monitor Technology Compared to Predicate Devices
The FORE-SIGHT ELITE Absolute Tissue Oximeter compares substantially to the cited predicate devices in that they use fundamentally the same optical operating principle, called multi-distance diffuse reflectance spectroscopy. All cited monitors use light to examine a cross-section tissue microvasculature (a mixed bed of arterioles, capillaries and venules). The FORE-SIGHT ELITE Monitor and predicate devices analyze the light that is returned after having passed through tissues. The spectroscopic analysis determines concentrations of hemoglobin in its oxygenated and deoxygenated states. All cited monitors calculate oxygen saturation which reflects the percentage of oxygenated hemoglobin in the sampled tissue.
{5}------------------------------------------------
The following table compares design, materials, energy and others characteristics.
| Characteristic | K143675 (this<br>Submission) | CASMED<br>ELITE | CASMED<br>K112820 | NONIN<br>K113215 | How They<br>Compare |
|-----------------|------------------------------|--------------------------|-------------------------------|-----------------------|---------------------|
| | | K133879 | | | |
| FDA | Class II, | Class II, | Class II, | Class II, | Same |
| Classification | 870.2700, | 870.2700, | 870.2700, | 870.2700. | |
| | MUD | MUD | MUD | MUD | |
| Purpose | Detect | Detect | Detect | Detect | Same |
| | changing levels | changing levels | changing levels | changing levels | |
| | in cerebral or | in cerebral or | in cerebral or | in cerebral or | |
| | somatic | somatic | somatic | somatic | |
| | oxygenated | oxygenated | oxygenated | oxygenated | |
| | hemoglobin in | hemoglobin in | hemoglobin in | hemoglobin in | |
| | sampled tissue | sampled tissue | sampled tissue | sampled tissue | |
| Indicated For | Adults and | Adult and | Adults and | Adults and | Same as |
| | Pediatric | Transitional | Pediatric | Pediatric | CASMED |
| | subjects<br>depending on | adolescents | subjects | subjects | predicate |
| | sensor size | | depending on<br>sensor size | | |
| Parameters | Single | Single | Single | Single | Same |
| Monitored | | | | | |
| | Near-infrared | Near-infrared | Near-infrared | Near-infrared | Same |
| Technology | Spectroscopy | Spectroscopy | Spectroscopy | Spectroscopy | |
| | (NIRS) | (NIRS) | (NIRS) | (NIRS) | |
| Monitoring | 4 - Channels | 4 - Channels | 2 - Channels | 4 - Channels | Same as |
| Channels | | | | | CASMED |
| | | | | | predicate |
| System | Monitor, | Monitor, | Monitor, | Monitor, | Same |
| Components | Preamp, | Preamp, | Preamp, | Preamp, | |
| | Sensor | Sensor | Sensor | Sensor | |
| Power | 100 to 240 | 100 to 240 | 110 to 240 | 100 to 240 | Same as |
| Requirements | VAC 50/60 Hz | VAC 50/60 Hz | VAC 50/60 Hz<br>and Lead Acid | VAC 50/60 Hz | CASMED<br>predicate |
| | and Lead Acid<br>Battery | and Lead Acid<br>Battery | Battery | and Li-ion<br>Battery | |
| Sensor Size(s) | (4) Large, | (1) Large | (4) Large, | (2) sizes, one | Same as |
| | Medium, Small | | Medium, Small | with adhesive, | CASMED |
| | and Small | | and Small | one without | predicate |
| | Non-adhesive | | Non-adhesive | | |
| Sensor Light | LED | LED | Laser Photo | LED | Same as most |
| Source | | | Diode | | predicates |
| Operating | Continuous | Continuous | Continuous | Continuous | Same |
| Modes | | | | | |
| Patient Contact | Medical | Medical | Medical | Unknown | Similar to |
| Material | Adhesive and | Adhesive | Adhesive and | | CASMED |
| | Orthopedic | | biocompatible | | predicate |
| | foam/fabric | | fabric (small | | |
| | composition | | non-adhesive) | | |
| | (small non- | | | | |
| | adhesive) | | | | |
{6}------------------------------------------------
## Non-Clinical Performance Testing to Demonstrate Substantial Equivalence
The FORE-SIGHT ELITE Absolute tissue Oximeter has successfully undergone extensive performance, (safety, electromagnetic, and system testing) to ensure it has been found to be substantially equivalent to the predicate devices. Evaluation for the safety of materials is assured by biological testing under the standard ISO10993-1. All testing is a product of the risk management process.
Test standards met:
| Category | Testing Summary |
|----------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Electrical Safety and<br>Usability | ANSI/AAMI ES60601-1: (2005) IEC 60601-1-8: (2006) IEC 60601-1-6: (2010) IEC 62366: (2007) |
| Electromagnetic Testing | IEC 60601-1-2: (2007) |
| System Functional and<br>Performance Testing | The FORE-SIGHT ELITE Oximeter was designed and developed in<br>accordance with CASMED development processes and was verified and<br>validated. Test results demonstrated that the oximeter complies with its<br>predetermined specification. |
## Clinical Testing to Show Substantial Equivalence
The FORE-SIGHT ELITE Absolute Tissue Oximeter was clinically validated for cerebral and noncerebral body locations on a demographically diverse population of 136 subjects at multiple institutions with ages ranging from 1 day to 17 years old including very low birth weight and premature neonates. Measurements of the tissue oxygenation were calibrated against a weighted co-oximetry reference of arterial and venous blood samples drawn simultaneously from vascular locations appropriate to the tissue of interest (jugular bulb, central venous, umbilical venous) with a 30% arterial and 70% venous contribution. Tissue saturations were obtained in the range of 48-96%. Accuracy was determined as a bias and precision to the weighted co-oximetry reference with values from 0.05 ± 5.06% to 0.03 ± 5.69% depending upon body location. No safety issues or adverse events related to the FORE-SIGHT ELITE Absolute Tissue Oximeter were encountered.
## Conclusions Drawn from Clinical and Non-Clinical Testing
Clinical evaluation, safety testing and software validation demonstrate the FORE-SIGHT ELITE Absolute Tissue Oximeter is substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.