Multicenter Perioperative Outcomes Group (MPOG) registry
A retrospective historical control group from the MPOG registry was used to compare the effectiveness of the Acumen HPI feature in reducing the duration of intraoperative hypotension against a prospective study arm.
HPI Study (Historical Control Arm); Retrospective observational analysis of registry data; Follow-up/Duration: Not applicable; Study Period: January 1, 2017 to December 31, 2017
Patients undergoing non-cardiac surgery at hospitals contributing to the MPOG consortium; Sample Size: 22,109; Number of Sites: Multiple (United States)
Not applicable for this study
Duration of intraoperative hypotension (MAP < 65 mmHg)
AI Performance
Output
Algorithm
Acceptance
Observed
Dev DS
Dev Readers
Test DS
Test Readers
Intraoperative Hypotension Prediction
—
Reduction of mean duration of IOH by at least 25%
57.6% reduction in mean IOH duration (p<0.0001)
—
—
HPI Study (Prospective, Single-Arm, Open-Label, Multicenter Study): 460 subjects (406 in Full Analysis Set) across 11 study sites in the United States.
—
Indications for Use
The Edwards Lifesciences Acumen Lifesciences Acumen Hypotension Index feature provides the clinician with physiological insight into a patient's likelihood of future hypotensive events (defined as mean arterial pressure < 65 mmHg for at least one minute in duration) and the associated hemodynamics. The Acumen HPI feature is intended for use in surgical or nonsurgical patients receiving advanced hemodynamic monitoring. The Acumen HPI feature is considered to be additional quantitative information regarding the patient's physiological condition for reference only and no therapeutic decisions should be made based solely on the Hypotension Prediction Index (HPI) parameter.
Device Story
The Acumen Hypotension Prediction Index (HPI) is a software feature integrated into the HemoSphere Advanced Monitoring Platform. It processes arterial pressure waveform data from an Acumen IQ extravascular blood pressure transducer and peripheral arterial catheter. The device uses an algorithm to calculate an HPI parameter (0-100 scale), representing the likelihood of a hypotensive event (MAP < 65 mmHg) occurring within the next 15 minutes. Used in OR or clinical settings by physicians/nurses, the system displays the HPI and associated hemodynamic parameters on a graphical user interface. It serves as a decision support tool, providing quantitative physiological insight to assist clinicians in identifying hemodynamic instability and managing intraoperative hypotension. It does not replace clinical judgment; therapeutic decisions must not be based solely on the HPI output. The device aims to reduce the duration of intraoperative hypotension, potentially improving patient outcomes by enabling proactive hemodynamic management.
Clinical Evidence
Prospective, single-arm, open-label, multicenter study (n=460) compared to a retrospective historical control (MPOG, n=22,109). Primary endpoint: reduction in duration of intraoperative hypotension (IOH, MAP < 65 mmHg). Results showed a 57.6% reduction in mean IOH duration (11.97 mins vs 28.20 mins, p<0.0001). Safety endpoints included 30-day post-operative composite events (cardiac arrest, death, stroke, AKI, MINS), with a composite event rate of 4.75%. No device-related serious adverse events were reported.
Technological Characteristics
Software-based feature running on the HemoSphere Advanced Monitoring Platform. Inputs: arterial pressure waveform via Acumen IQ extravascular blood pressure transducer and peripheral arterial catheter. Output: HPI parameter (0-100 index). Connectivity: integrated into HemoSphere platform. Algorithm: predictive model for hemodynamic instability.
Indications for Use
Indicated for surgical or nonsurgical patients receiving advanced hemodynamic monitoring to provide insight into the likelihood of future hypotensive events (MAP < 65 mmHg for ≥ 1 minute).
Regulatory Classification
Identification
The adjunctive predictive cardiovascular indicator is a prescription device that uses software algorithms to analyze cardiovascular vital signs and predict future cardiovascular status or events. This device is intended for adjunctive use with other physical vital sign parameters and patient information and is not intended to independently direct therapy.
Special Controls
*Classification.* Class II (special controls). The special controls for this device are:(1) A software description and the results of verification and validation testing based on a comprehensive hazard analysis and risk assessment must be provided, including:
(i) A full characterization of the software technical parameters, including algorithms;
(ii) A description of the expected impact of all applicable sensor acquisition hardware characteristics and associated hardware specifications;
(iii) A description of sensor data quality control measures;
(iv) A description of all mitigations for user error or failure of any subsystem components (including signal detection, signal analysis, data display, and storage) on output accuracy;
(v) A description of the expected time to patient status or clinical event for all expected outputs, accounting for differences in patient condition and environment; and
(vi) The sensitivity, specificity, positive predictive value, and negative predictive value in both percentage and number form.
(2) A scientific justification for the validity of the predictive cardiovascular indicator algorithm(s) must be provided. This justification must include verification of the algorithm calculations and validation using an independent data set.
(3) A human factors and usability engineering assessment must be provided that evaluates the risk of misinterpretation of device output.
(4) A clinical data assessment must be provided. This assessment must fulfill the following:
(i) The assessment must include a summary of the clinical data used, including source, patient demographics, and any techniques used for annotating and separating the data.
(ii) The clinical data must be representative of the intended use population for the device. Any selection criteria or sample limitations must be fully described and justified.
(iii) The assessment must demonstrate output consistency using the expected range of data sources and data quality encountered in the intended use population and environment.
(iv) The assessment must evaluate how the device output correlates with the predicted event or status.
(5) Labeling must include:
(i) A description of what the device measures and outputs to the user;
(ii) Warnings identifying sensor acquisition factors that may impact measurement results;
(iii) Guidance for interpretation of the measurements, including a statement that the output is adjunctive to other physical vital sign parameters and patient information;
(iv) A specific time or a range of times before the predicted patient status or clinical event occurs, accounting for differences in patient condition and environment;
(v) Key assumptions made during calculation of the output;
(vi) The type(s) of sensor data used, including specification of compatible sensors for data acquisition;
(vii) The expected performance of the device for all intended use populations and environments; and
(viii) Relevant characteristics of the patients studied in the clinical validation (including age, gender, race or ethnicity, and patient condition) and a summary of validation results.
Predicate Devices
Acumen Hypotension Prediction Index (HemoSphere Advanced Monitoring Platform) (K201446)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square, the words "U.S. FOOD & DRUG ADMINISTRATION" are written in blue.
July 30, 2021
Edwards Lifesciences, LLC Lisa Gilman Distinguished Regulatory Affairs Program Manager One Edwards Way Irvine, California 92614
Re: K203224
Trade/Device Name: Acumen™ Hypotension Prediction Index Regulation Number: 21 CFR 870.2210 Regulation Name: Adjunctive predictive cardiovascular indicator Regulatory Class: Class II Product Code: QAQ Dated: October 30, 2020 Received: November 2, 2020
Dear Lisa Gilman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for
Stephen Browning Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K203224
Device Name Acumen Hypotension Prediction Index
#### Indications for Use (Describe)
The Edwards Lifesciences Acumen Hypotension Index feature provides the clinician with physiological insight into a patient's likelihood of future hypotensive events (defined as mean arterial pressure < 65 mmHg for at least one minute in duration) and the associated hemodynamics. The Acumen HPI feature is intended for use in surgical or nonsurgical patients receiving advanced hemodynamic monitoring. The Acumen HPI feature is considered to be additional quantitative information regarding the patient's physiological condition for reference only and no therapeutic decisions should be made based solely on the Hypotension Prediction Index (HPI) parameter.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------|
| <div style="display:flex; align-items:center;"><span style="margin-right:5px;">☑</span>Prescription Use (Part 21 CFR 801 Subpart D)</div> | <div style="display:flex; align-items:center;"><span style="margin-right:5px;">☐</span>Over-The-Counter Use (21 CFR 801 Subpart C)</div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# SECTION 5 – 510(k) SUMMARY
| | Acumen™ Hypotension Prediction Index |
|-------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Sponsor | Edwards Lifesciences LLC<br>One Edwards Way<br>Irvine, CA 92614 |
| Establishment<br>Registration<br>Number | 2015691 |
| Contact<br>Person | Lisa Gilman<br>Distinguished Regulatory Affairs Program Manager<br>One Edwards Way<br>Irvine, CA 92614<br>Telephone: (949) 250-1478<br>Fax: (949) 809-2996 |
| Date<br>Prepared | June 16, 2021 |
| Trade Name | Acumen™ Hypotension Prediction Index |
| Common<br>Name | Adjunctive Predictive Cardiovascular Indicator |
| Classification<br>Name | Adjunctive Predictive Cardiovascular Indicator |
| Regulation<br>Class /<br>Product Code | 21 CFR 870.2210/Class II/QAQ |
| Predicate<br>Device(s) | Acumen Hypotension Prediction Index (HemoSphere Advanced Monitoring<br>Platform, K201446 (Primary Predicate)<br>Acumen Hypotension Prediction Index, DEN160044 (Reference Predicate) |
| | The Acumen Hypotension Prediction Index feature consists of software<br>running on the Edwards Lifesciences HemoSphere Advanced Monitoring<br>Platform paired with the Acumen IQ extravascular blood pressure transducer<br>(K152980) and a peripheral arterial catheter. The monitoring system includes<br>the Acumen Hypotension Prediction Index (HPI), and graphical user<br>interface features displaying hemodynamic parameters relevant to assessing<br>the root cause of a potential hypotensive event.* |
| Device<br>Description | The Acumen Hypotension Prediction Index is an index related to the<br>likelihood of a patient experiencing hemodynamic instability defined as a<br>hypotensive event* within 15 minutes, where zero (0) indicates low<br>likelihood and one hundred (100) indicates a hypotensive event is occurring.<br>The Acumen Hypotension Prediction Index parameter (HPI), should not be<br>used exclusively to treat patients. A review of the patient's hemodynamics is<br>recommended prior to initiating treatment. |
| | * A hypotensive event is defined as mean arterial pressure (MAP) < 65<br>mmHg for one minute in duration |
| Indications<br>for<br>Use/Intended<br>Use | The Edwards Lifesciences Acumen Hypotension Prediction Index feature<br>provides the clinician with physiological insight into a patient's likelihood of<br>future hypotensive events (defined as mean arterial pressure < 65 mmHg for<br>at least one minute in duration) and the associated hemodynamics. The<br> |
| Comparison<br>to Predicate<br>Device | The Acumen Hypotension Prediction Index feature as implemented on the<br>HemoSphere Advanced Monitoring Platform (K201446, October 1, 2020) is<br>identical to itself as there are no changes to the Acumen™ Hypotension<br>Prediction Index feature indications for use, algorithm, nor the hardware and<br>software of the HemoSphere Advanced Monitoring Platform.<br>The Acumen Hypotension Prediction Index (HPI parameter) algorithm as<br>implemented on the HemoSphere Advanced Monitoring Platform (K201446,<br>October 1, 2020) is identical to Acumen Hypotension Prediction Index (HPI<br>parameter) algorithm granted in DEN160044.<br>Modifications<br>The modifications are to the labeling of the HemoSphere Advanced<br>Monitoring Platform to provide additional data demonstrating the<br>utility of the Acumen™ Hypotension Prediction Index Feature<br>Software in the detection of hemodynamic instability and the<br>treatment of intraoperative hypotension (IOH) during non-cardiac<br>surgery. |
| | Clinical Performance |
| | A Prospective, Single-Arm, Open-Label, Multicenter Study of the<br>Hypotension Prevention and Treatment in Patients Receiving Arterial<br>Pressure Monitoring with Acumen Hypotension Prediction Index Feature<br>(HPI Study) was undertaken to further understand the impact that the<br>Acumen Hypotension Prediction Index (HPI) feature may have in the<br>detection of hemodynamic instability and the treatment of intraoperative<br>hypotension in non-cardiac surgery. The comparison group was a<br>retrospective historical control group with patient-level data from a non-<br>profit academic consortium group, the Multicenter Perioperative Outcomes<br>Group (MPOG), that collects perioperative data from hospitals across United<br>States. |
| Performance<br>Data | The primary objective of the HPI Study was to determine whether the use of<br>the Acumen HPI Feature to guide intraoperative hemodynamic management<br>in non-cardiac surgery reduces the duration of intraoperative hypotension<br>(defined as MAP < 65 mmHg for at least 1 minute) as compared with a<br>historic retrospective control group. The HPI Study was a single-arm,<br>prospective, unblinded study conducted in 485 eligible Subjects (460 pivotal<br>with an additional 25 roll-in cases) at the same 11 study sites in the United<br>States that contributed to the historical control group. The HPI arm to<br>determine if using the Acumen HPI Feature to predict hypotension within 15<br>minutes of an actual event could reduce the mean duration of IOH by at least<br>25%.1 The incidence of IOH in the MPOG group was 88%<br>(n=19,445/22,109) and the dates of treatment were between January 1, 2017<br>and December 31, 2017. The secondary effectiveness endpoint was the<br>determination of total area under the curve of the time and MAP for all time<br>periods for which MAP < 65 mmHg in each Subject. This endpoint is<br>correlated with the duration and a descriptive analysis of this endpoint was<br>presented with the mean, standard deviation (SD), median, minimum and<br>maximum. |
{4}------------------------------------------------
{5}------------------------------------------------
<sup>1</sup> Shah NJ, Mentz G, Kheterpal S. The incidence of intraoperative hypotension in moderate to high risk patients undergoing non-cardiac surgery: A retrospective multicenter observational analysis. J Clin Anest. 2020: 66; 109961.
{6}------------------------------------------------
The primary safety endpoint was the percentage of serious adverse events to include perioperative events, postoperative complications, and device-related serious adverse events. The secondary objective for this study (secondary safety endpoint) was to determine if the guidance provided by the Acumen HPI Feature reduced a composite measure of complications as indicated below:
- Post-operative episodes of non-fatal cardiac arrest;
- In-hospital death; ●
- . Stroke:
- Acute Kidney Injury (AKI) within 30 days of the procedure; ●
- Myocardial Injury in non-cardiac surgery (MINS) within 30 days of . the procedure.
| | Description | HPI<br>(Intent-to-Treat) | HPI<br>(Full Analysis Set) | MPOG<br>(Full Analysis Set) |
|----------------------------|-----------------------------------|--------------------------------|----------------------------|-----------------------------|
| # of patients (N) | | 460 | 406* | 22,109 |
| Gender (%) | Male | 51.7 (n=238) | 52.96 (n=215) | 57.8 (n=12,770) |
| Gender (%) | Female | 48.3 (n=222) | 47.04 (n=191) | 42.2 (n=9330) |
| Age (year) | Mean ± SD | 63.0 ± 13.0 | 62.8 ± 13.0 | 63.3 ± 13.8 |
| Age (year) | Median | 65 (19 - 94) | 65 (19 - 89) | 65 (18 - 90) |
| BMI | Median (25th and 75th percentile) | 28.09 (24.37, 32.81) | 28.09 (24.41, 32.86) | 28.1 (24.2, 32.9) |
| ASA Score (%) | II** | 0.2 (n=1) | 0.25 (n=1) | 0.0 (n=0) |
| ASA Score (%) | III | 91.5 (n=421) | 92.12 (n=374) | 80.83 (n=17,870) |
| ASA Score (%) | IV | 8.0 (n=37) | 7.64 (n=31) | 19.17 (n=4,239) |
| ASA Score (%) | Not specified | 0.2 (n=1) | 0.0 (n=0) | 0.0 (n=0) |
| Surgery duration (minutes) | Mean ± SD | 338.1 ± 145.38<br>(n = 458) | 363.6 ± 134.04 | 355.2 ± 145.8 |
| Surgery duration (minutes) | Median (25th and 75th percentile) | 315.5 (235 - 416)<br>(n = 458) | 336 (262 - 430) | 317 (245, 427) |
#### Patient Demographics
*The Full Analysis Set (FAS) represent those subjects from the Intent-to-Treat (ITT) population that had a surgery duration of ≥3 hours.
** ASA II subject was identified as a protocol deviation, though not excluded from ITT and FAS populations as this subject met the defined criteria ( surgery > 3 hours and hemodynamic monitoring data). This subject was included in the efficacy and safety analyses, although by inclusion/exclusion criteria should not have been enrolled in the study.
{7}------------------------------------------------
| | Procedure Type (HPI) |
|-----------------------------------------------------------------------------|----------------------|
| Procedure Type | % (n/N) |
| Spine Surgery | 18.5 (85 / 460) |
| Hepatectomy | 13.7 (63 / 460) |
| Whipple | 10.0 (46 / 460) |
| Major, vascular | 8.5 (39 / 460) |
| Other | 8.5 (39 / 460) |
| Nephrectomy | 5.7 (26 / 460) |
| Other Genitourinary Surgery | 5.4 (25 / 460) |
| Cystectomy | 5.0 (23 / 460) |
| Pancreatectomy | 5.0 (23 / 460) |
| Renal Transplant | 4.3 (20 / 460) |
| Head & Neck Surgery | 3.9 (18 / 460) |
| Complex Combined Oncologic Surgery<br>(including 2 or more distinct organs) | 3.0 (14 / 460) |
| Exploratory Laparotomy | 3.0 (14 / 460) |
| Colectomy | 2.8 (13 / 460) |
| Adrenalectomy | 2.6 (12 / 460) |
| Gastrectomy | 2.0 (9 / 460) |
| Other Gastrointestinal Surgery | 2.0 (9 / 460) |
| Hip Revision | 1.7 (8 / 460) |
| Prostatectomy | 1.7 (8 / 460) |
| HIPEC | 1.3 (6 / 460) |
| Hysterectomy with Debulking | 1.3 (6 / 460) |
| Cholecystectomy | 0.9 (4 / 460) |
| Reoperative Orthopedic Surgery | 0.9 (4 / 460) |
| Splenectomy | 0.9 (4 / 460) |
| Bariatric Surgery | 0.4 (2 / 460) |
| Liver Transplant | 0.4 (2 / 460) |
| Sigmoidectomy | 0.4 (2 / 460) |
| Not Specified | 0.2 (1 / 460) |
K203224
{8}------------------------------------------------
### Study Results
#### Effectiveness
The HPI Study was designed to evaluate the ability of the Acumen HPI Feature, as a decision support tool, to reduce the duration of IOH by at least 25% in surgical patients that require advanced hemodynamic monitoring. An episode of intraoperative hypotension (IOH) was defined as a mean arterial pressure (MAP) below 65 for three (3) or more consecutive 20 second events for each subject, across all sites.
The primary effectiveness endpoint is a weighted average of site means and standard deviations combined in the same proportion of subjects that were included in the MPOG cohort. This weighted average and its properly computed standard deviation was compared to the estimates obtained from the subjects of the MPOG cohort.
The HPI Study met its primary effectiveness endpoint. The HPI Pivotal Subjects of the full analysis set experienced a mean IOH duration of 11.97 ± 13.92 minutes compared with the MPOG historical control mean IOH of 28.20 ± 42.60 minutes. The table below demonstrates that this result was a reduction of 57.6% compared to the MPOG historical control (p<0.0001). When considering instances where there were zero episodes of IOH experienced during surgery, there was a 65% reduction of IOH (p<0.0001).
| Statistics | HPI<br>(Subject=406) | MPOG<br>(Subject=22,109) | p value |
|-------------------|----------------------|--------------------------|----------|
| Sample size (n) | 293 | 19,446 | - |
| Total IOH Minutes | 3,508 | 548,465 | - |
| IOH Mean (mins) | 11.97 | 28.20 | <0.0001* |
| IOH STD | 13.92 | 42.60 | - |
### Mean IOH Duration - Primary Effectiveness Endpoint
Note: IOH Estimated with Stand Method; STD Estimated with Pooled Method (Pivotal Subject with IOH Episode in Test Arm). Standard Method - IOH episode is defined with at least three consecutive observations having MAP<65. FAS pivotal Subjects, with at least 3-hour surgery time.
*One-sided unequal variances t-test was used in analysis. Nominal alpha for the test is 0.025.
{9}------------------------------------------------
The results of the secondary effectiveness endpoint, determination of total area under the curve (AUC) of the time, and MAP for all time periods for which MAP < 65 mmHg in each Subject, are included in the table below.
| Study Category | Subject | AUC<br>Mean | AUC<br>SD | AUC<br>Median | AUC<br>Range | AUC<br>Q3-Q1 |
|-----------------------------------------------------------------------------------------|---------|-------------|-----------|---------------|--------------|--------------|
| All pivotal Subjects | 457 | 46.38 | 82.75 | 16.67 | 833.00 | 54.00 |
| All pivotal Subjects with at<br>least one episode | 328 | 64.63 | 91.46 | 32.33 | 832.00 | 68.00 |
| All pivotal Subjects with ≥3<br>hours surgery duration | 406 | 47.07 | 85.30 | 16.83 | 833.00 | 51.00 |
| All pivotal Subjects with ≥ 3<br>hours surgery duration and<br>at least one IOH episode | 293 | 65.23 | 94.36 | 32.00 | 832.00 | 62.67 |
| All pivotal Subjects with <3<br>hours surgery duration | 51 | 40.89 | 58.94 | 12.33 | 291.00 | 71.33 |
| All pivotal Subjects with <3<br>hours surgery duration and<br>at least one IOH episode | 35 | 59.58 | 62.94 | 37.00 | 290.00 | 73.33 |
Intraoperative Hypotension AUC - ITT, Pivotal Subjects (AUC displayed as Min*mmHo)
Note: Standard Method - IOH episode is defined with at least three consecutive observations having MAP<65. ITT pivotal subjects, with valid surgery time.
## Safety
The Acumen HPI Feature was shown to be safe when used in surgical patients that require advanced hemodynamic monitoring.
- There were no Subjects with events adjudicated to have any ● relationship to the Acumen HPI Feature.
- There were no ADEs or SADEs adjudicated as related to the Acumen ● HPI Feature.
- There were no unanticipated ADEs (0%) related to the HPI Feature. ●
- There were no deaths that occurred whether related/unrelated to HPI ● Feature.
The secondary safety endpoint was a composite of 30-day post-operative AEs in the completed cases (CC) population. The table below shows the components of the 30-Day post-operative composite endpoint for the Completed Cases (CC) population. The results demonstrate that the composite event rate was 4.75% (composite events =19 [95% CI: 2.88. 7.32], with one subject experiencing more than one of the individual composite elements).
{10}------------------------------------------------
| Analysis Population (Pivotal Subjects, n=400) | | | | | |
|-----------------------------------------------|-------------|------------|-----------------------|--------|-------|
| Analysis Endpoint | AE Event | | POD Post-surgery Days | | |
| | Events n(%) | 95% CI | Mean | Median | Range |
| Postoperative Non-Fatal<br>Cardiac Arrest | 1 (0.25) | 0.01, 1.38 | 2.00 | 2.00 | 2, 2 |
| In-Hospital Death | 0 (0.00) | 0.00, 0.92 | N/A | N/A | N/A |
| Stroke | 0 (0.00) | 0.00, 0.92 | N/A | N/A | N/A |
| Acute Kidney Injury - Overall | 16 (4.00) | 2.30, 6.41 | 5.94 | 1.00 | 0, 27 |
| Acute Kidney Injury - Stage 1 | 11 (2.75) | 1.38, 4.87 | 6.82 | 1.00 | 0, 27 |
| Acute Kidney Injury - Stage 2 | 3 (0.75) | 0.15, 2.18 | 6.33 | 7.00 | 2, 10 |
| Acute Kidney Injury - Stage 3 | 2 (0.50) | 0.06, 1.79 | 0.50 | 0.50 | 0, 1 |
| Myocardial Injury (MINS) | 3 (0.75) | 0.15, 2.18 | 1.67 | 1.00 | 0, 4 |
HPI Study - 30 Days Post-Operative Composite Endpoint Components -CC Analy
CC=Complete (Evaluable) Group, CI=confidence interval, Post-surgery Days (POD)=AESTDT-SGDT
Analysis of in the intent-to-treat population (n=460) vielded 3 (.066%) instances of myocardial injury (MINS) and 17 (3.7%) incidents of acute kidnev injurv (AKI).
### Study Summary
These results demonstrate a reduction in mean IOH, that was consistent across most sites; most sites had a > 25% reduction in its mean duration of IOH, with all sites but one exceeding 35%; ranging from a23% to 72% mean IOH reduction. The findings of the study showed a reduction of the duration of IOH to 11.97 mins (SD 13.92), representing a 57.6% reduction (p<0.0001). This reduction is clinically relevant, as IOH lasting at least 1minute has been associated with perioperative complications and morbidity such as AKI, MINS and stroke1.
The results demonstrate that Acumen HPI Feature was shown to be safe when used in surgical patients that require advanced hemodynamic monitoring, with no device-related adverse events. Additionally, the composite event rate 4.75% (composite events =19 [95% CI: 2.88, 7.32] is low when considering that the subjects were ASA Physical Status 3 and 4 undergoing non-cardiac surgery.
### Conclusion
The results of this study provide valid scientific evidence that the Acumen HPI feature is safe and provided a clinically significant reduction in mean IOH. Acumen HPI is effective in detecting hemodynamic instability and reducing the amount ofintraoperative hypotension when used in surgical patients who require intraoperative hemodynamic monitoring during noncardiac surgery.
{11}------------------------------------------------
| Conclusion | Acumen Hypotension Prediction Index Feature as implemented on the<br>HemoSphere Advanced Monitoring Platform (K201446, October 1, 2020) is<br>equivalent to itself as there are no changes to the Acumen Hypotension<br>Prediction Index feature indications for use, algorithm, nor the hardware and<br>software of the HemoSphere Advanced Monitoring Platform. Additionally,<br>the Acumen Hypotension Prediction Index (HPI parameter) algorithm as<br>implemented on the HemoSphere Advanced Monitoring Platform (K201446,<br>October 1, 2020) is identical to Acumen Hypotension Prediction Index (HPI<br>parameter) algorithm (DEN160044, March 16, 2020).<br><br>The clinical performance data further demonstrate the utility of the Acumen<br>Hypotension Prediction Index Feature Software in the detection of<br>hemodynamic instability and in the treatment of intraoperative hypotension<br>(IOH) during non-cardiac surgery.<br><br>The Acumen Hypotension Prediction Index is substantially equivalent to the<br>Acumen Hypotension Prediction Index (HemoSphere Advanced Monitoring<br>Platform), K201446. |
|------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.