The HemoSphere Advanced Monitoring Platform is intended to be used by qualified personnel or trained clinicians in a critical care environment in a hospital setting. The HemoSphere advanced monitoring platform is intended for use with compatible Edwards Swan-Ganz and oximetry catheters, FloTrac sensors, Acumen IQ sensors, TruWave DPTs, ForeSight Elite sensors, and ClearSight finger cuffs.
Device Story
HemoSphere Advanced Monitoring Platform is a modular system for hemodynamic and volumetric data visualization in hospital critical care/OR settings. It processes inputs from various modules: Swan-Ganz catheters (CO, volumetric parameters), oximetry catheters (SvO2, ScvO2), pressure cables (arterial/venous pressures), tissue oximetry sensors (StO2), and ClearSight finger cuffs (noninvasive BP/hemodynamics). The platform transforms these signals into real-time hemodynamic parameters and derived metrics. The Acumen Hypotension Prediction Index (HPI) feature provides quantitative insight into the likelihood of future hypotensive events for reference only. Clinicians use the display to monitor patient physiological status, guide goal-directed therapy, and support clinical decision-making. Benefits include continuous, noninvasive or invasive hemodynamic assessment to optimize patient management in critical care.
Clinical Evidence
Bench testing only. Verification and validation activities included system-level performance testing, usability studies per FDA human factors guidance, electrical safety (IEC 60601 series), EMC, and wireless coexistence testing. All tests passed, demonstrating that the platform meets design specifications and is substantially equivalent to predicates.
Technological Characteristics
Modular platform with HemoSphere monitor and external modules (Swan-Ganz, Oximetry, Pressure, Tissue Oximetry, ClearSight). Complies with IEC 60601-1, 60601-1-2, 60601-1-6, 60601-1-8, 62304, 62366-1, 60601-2-34, 60601-2-57, 60601-2-49, 80601-2-49. Connectivity via proprietary modules and cables. Software-based processing for hemodynamic parameter derivation.
Indications for Use
Indicated for adult and pediatric critical care patients in hospital environments requiring hemodynamic monitoring (cardiac output, venous oxygen saturation, pressure, fluid status, vascular resistance). ClearSight module indicated for patients >18 years for continuous noninvasive hemodynamic assessment. Acumen HPI indicated for OR patients receiving advanced hemodynamic monitoring to provide insight into likelihood of future hypotensive events (MAP < 65 mmHg). FORE-SIGHT ELITE indicated for individuals at risk for reduced-flow or no-flow ischemic states.
Regulatory Classification
Identification
A programmable diagnostic computer is a device that can be programmed to compute various physiologic or blood flow parameters based on the output from one or more electrodes, transducers, or measuring devices; this device includes any associated commercially supplied programs.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 16, 2021
Edwards Lifesciences, LLC Varad Raghuwanshi Sr. Specialist, Regulatory Affairs One Edwards Way Irvine, California 92614
Re: K201446
Trade/Device Name: HemoSphere Advanced Monitoring Platform, HemoSphere ClearSight Module Regulation Number: 21 CFR 870.1425 Regulation Name: Programmable diagnostic computer Regulatory Class: Class II Product Code: DQK, DQE, QAQ, MUD, DXN, DSB
Dear Varad Raghuwanshi:
The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated October 1, 2020. Specifically, FDA is updating this SE Letter as an administrative correction.
Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact LT Stephen Browning, OHT2: Office of Cardiovascular Devices, 240-402-5241, Stephen.Browning(@fda.hhs.gov.
Sincerely,
# Jennifer W. Shih -S
for LT Stephen Browning Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
October 1, 2020
Edwards Lifesiences, LLC Varad Raghuwanshi Sr. Specialist, Regulatory Affairs One Edwards Wav Irvine, California 92614
Re: K201446
Trade/Device Name: HemoSphere Advanced Monitoring Platform Regulation Number: 21 CFR 870.1425 Regulation Name: Programmable Diagnostic Computer Regulatory Class: Class II Product Code: DOK, DOE, OAO, MUD, DXN, DSB Dated: May 29, 2020 Received: June 1, 2020
Dear Varad Raghuwanshi:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{2}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Stephen C. Browning -S5
LT Stephen Browning Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K201446
Device Name
HemoSphere Advanced Monitoring Platform
Indications for Use (Describe)
HemoSphere Advanced Monitor with HemoSphere Swan-Ganz Module
The HemoSphere Advanced Monitor when used with the HemoSphere Swan-Ganz Module and Edwards Swan-Ganz Catheters is indicated for use in adult and pediatric critical care patients requiring of cardiac output [continuous (CO) and intermittent (iCO)] and derived hemodynamic parameters. It may also be used for monitoring hemodynamic parameters in conjunction with a perioperative goal directed therapy protocol in a hospital environment. Refer to the Edwards Swan-Ganz catheter indications for use statement for information on target patient population specific to the catheter being used.
Refer to the Intended Use statement below for a complete list of measured and derived parameters available for each patient population.
HemoSphere Advanced Monitor with HemoSphere Oximetry Cable
The HemoSphere Advanced Monitor when used with the HemoSphere Oximetry cathetry catheters is indicated for use in adult and pediativ critical care patients requiring of venous oxygen saturation (Sv02 and ScvO2) and derived hemodynamic parameters in a hospital environment. Refer to the Edwards oximeter indications for use statement for information on target patient population specific to the catheter being used.
Refer to the Intended Use statement for a complete list of measured and derived parameters available for each patient population.
HemoSphere Advanced Monitor with HemoSphere Pressure Cable
The HemoSphere Advanced Monitor when used with the HemoSphere Pressure Cable is indicated for use in critical care patients in which the balance between cardiac function, fluid status, vascular resistance and pressure needs continuous assessment. It may be used for monitoring hemodynamic parameters in conjunction with a perioperative goal directed therapy protocol in a hospital environment. Refer to the Edwards FloTrac, Acumen IQ and TruWave DPT sensor indications for use statement for information on target patient population specific to the sensor being used.
The Edwards Lifesciences Acumen Hypotension Prediction Index feature provides the clinician with physiological insight into a patient's likelihood of future hypotensive events (defined as mean arterial pressure < 65 mmHg for at least one minute in duration) and the associated hemodynamics. The Acumen HPI feature is intended for use in operating room (OR) patients receiving advanced hemodynamic monitoring. The Acumen is considered to be additional quantitative information regarding the patient's physiological condition for reference only and no therapeutic decisions should be made based solely on the Hypotension Prediction Index (HPI) parameter.
Refer to the Intended Use statement for a complete list of measured and derived parameters available for each patient population.
HemoSphere Advanced Monitor with HemoSphere Tissue Oximetry Module and FORE-SIGHT ELITE Tissue Oximeter Module
{4}------------------------------------------------
## Traditional 510(k) – HemoSphere Advanced Monitoring Platform
The noninvasive FORE-SIGHT ELITE tissue oximeter module is intended for use as an adjunct monitor of absolute regional hemoglobin oxygen saturation of blood under the Sensors in individuals at risk for reduced-flow or no-flow ischemic states. The FORE-SIGHT ELITE tissue oximeter module is intended to allow for the display of StO2 on the HemoSphere advanced monitor.
· When used with large sensorsis indicated for use on adults and transitional adolescents ≥40 kg.
· When used with Medium Sensors, the FORE-SIGHT ELITE tissue oximeter module is indicated for use on pediatric subjects >3 kg.
· When used with Small Sensors, the FORE-SIGHT ELITE tissue oximeter module is indicated for cerebral use on pediatric subjects < 8 kg and non-cerebral use on pediatric subjects <5kg.
Refer to the Intended Use statement for a complete list of measured and derived parameters available for each patient population.
HemoSphere Advanced Monitor with HemoSphere ClearSight Module
The HemoSphere advanced monitor when used with the HemoSphere ClearSight module, pressure controller and a compatible Edwards finger cuff are indicated for patients over 18 years of age in which the balance between cardiac function, fluid status and vascular resistance needs continuous assessment. It may be used for monitoring hemodynamic parameters in conjunction with a perioperative goal directed therapy protocol in a hospital environment. In addition, the noninvasive system is indicated for use in patients with co-morbidities for which hemodynamic optimization is desired and invasive measurements are difficult. The HemoSphere advanced monitor and compatible Edwards finger cuffs noninvasively measures blood pressure and associated hemodynamic parameters.
Refer to the ClearSight finger cuff indications for information on target patient population specific to the finger cuff being used.
Refer to the Intended Use statement for a complete list of measured and derived parameters available for each patient population.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------|
| <div> <span> <span style="text-decoration: underline;"></span>Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span>Over-The-Counter Use (21 CFR 801 Subpart C)</span> </div> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor and a person is not required to respond to a collection of
'An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
# 510(k) Summary – HemoSphere Advanced Monitoring Platform
### I. Submitter:
| Sponsor: | Edwards Lifesciences LLC<br>One Edwards Way<br>Irvine, CA 92614 | Product Code<br>and Regulatory<br>Class: | DQK, Class II |
|------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------|---------------|
| Establishment<br>Registration<br>Number: | 2015691 | | DQE, Class II |
| Contact Person: | Varad Raghuwanshi<br>Sr. Specialist, Regulatory Affairs<br>One Edwards Way<br>Irvine, CA 92614<br>Telephone: (949) 756-4502<br>Fax: (949) 809-2972 | | QAQ, Class II |
| Date Prepared: | May 29, 2020 | | MUD, Class II |
| II.<br>Device Information: | | | DXN, Class II |
| Platform Name<br>(Name of the<br>Device) | HemoSphere Advanced Monitoring Platform | | DSB. Class II |
| Trade Name: | HemoSphere Advanced Monitor<br>HemoSphere Swan-Ganz Module<br>HemoSphere Oximetry Cable<br>HemoSphere Pressure Cable<br>HemoSphere Tissue Oximetry Module and FORE-SIGHT ELITE Tissue<br>Oximeter Module<br>HemoSphere ClearSight Module (subject) | | |
| Common Name: | Cardiac Output/Oximetry/Ejection Fraction Computer | | |
| Classification<br>Name: | Programmable Diagnostic Computer<br>21 CFR 870.1425<br>Fiberoptic Oximeter Catheter<br>21 CFR 870.1230<br>Adjunctive Predictive Cardiovascular Indicator<br>21 CFR 870.2210<br>Oximeter<br>21 CFR 870.2700 | | |
{6}------------------------------------------------
#### III. Predicate Device
Primary HemoSphere Advanced Monitoring Platform manufactured by Edwards Predicate Device: Lifesciences, K190205 (cleared August 29, 2019).
Additional EV1000 Clinical Platform Non-Invasive NI and ClearSight™ Finger Cuffs Predicate or ClearSight™ System K160552 (cleared June 1, 2016) Devices: utilized for the ClearSight Module (Non-invasive) parameters.
> ClearSight finger cuff K190130 (cleared June 21, 2019) utilized for the ClearSight finger cuff used with ClearSight Module
EV1000 Clinical Platform Non-Invasive NI and ClearSight™ Finger Cuffs or ClearSight™ System K182245 (cleared November 30, 2018) utilized for the Heart Reference Sensor
FORE-SIGHT ELITE Module (smart cable) (K180003, cleared May 10, 2018) for tissue oximetry.
Nexfin Model 2 (cleared K122381 on April 22, 2013) utilized as a reference predicate for the ClearSight workflow option.
#### IV. Device Description
Device HemoSphere Advanced Monitoring Platform consists of the HemoSphere Advanced Monitor that provides a means to interact with and visualize Description: hemodynamic and volumetric data on a screen and five (5) optional external modules: the HemoSphere Swan-Ganz Module (K163381 Cleared, April 14, 2017), the HemoSphere Oximetry Cable (K163381 Cleared, April 14, 2017), HemoSphere Pressure Cable (K180881 Cleared, November 16, 2018), HemoSphere Tissue Oximetry Module (K190205 August 29, 2019), FORE-SIGHT ELITE Tissue Oximeter Module (K180003, May 10, 2018), and the subject HemoSphere ClearSight Module.
> The subject HemoSphere ClearSight module is a non-invasive subsystem module intended to continuously and noninvasively measure blood
{7}------------------------------------------------
pressure and associated hemodynamic parameters. HemoSphere ClearSight module is intended to provide ClearSight functionality (Non-Invasive) previously cleared in EV1000 Clinical Platform NI (K160552 June 1, 2017) to the HemoSphere Advanced Monitoring Platform. It provides power, communications, air and an interface to connect with currently cleared and commercially available Edwards Lifesciences ClearSight System (K160552 June 1, 2016) which includes ClearSight finger cuffs (K160552 June 1, 2016 and K190130 June 21, 2019), Pressure Controller/Wrist Unit (K160552 June 1, 2016), Heart Reference Sensor (K182245 Nov 30, 2018), and the HemoSphere Advanced Monitor (K190205 August 29, 2019).
#### V. Indications for Use:
Note: The indication for Use statements are identical for all the previously cleared technologies (K190205 cleared August 29, 2019). with the exception of the addition of indication for subject HemoSphere ClearSight Module.
## HemoSphere Advanced Monitor with HemoSphere Swan-Ganz Module
The HemoSphere Advanced Monitor when used with the HemoSphere Swan-Ganz Module and Edwards Swan-Ganz Catheters is indicated for use in adult and pediatric critical care patients requiring monitoring of cardiac output [continuous (CO) and intermittent (iCO)] and derived hemodynamic parameters. It may also be used for monitoring hemodynamic parameters in conjunction with a perioperative goal directed therapy protocol in a hospital environment. Refer to the Edwards Swan-Ganz catheter indications for use statement for information on target patient population specific to the catheter being used. Refer to the Intended Use statement below for a complete list of measured and derived parameters available for each patient population.
## HemoSphere Advanced Monitor with HemoSphere Oximetry Cable
The HemoSphere Advanced Monitor when used with the HemoSphere Oximetry Cable and Edwards oximetry catheters is indicated for use in adult and pediatric critical care patients requiring monitoring of venous oxygen saturation (SvO2 and ScvO2) and derived hemodynamic parameters in a hospital environment. Refer to the Edwards oximetry catheter indications for use statement for information on target patient population specific to the catheter being used.
{8}------------------------------------------------
Refer to the Intended Use statement for a complete list of measured and derived parameters available for each patient population.
## HemoSphere Advanced Monitor with HemoSphere Pressure Cable
The HemoSphere Advanced Monitor when used with the HemoSphere Pressure Cable is indicated for use in critical care patients in which the balance between cardiac function, fluid status, vascular resistance and pressure needs continuous assessment. It may be used for monitoring hemodynamic parameters in conjunction with a perioperative goal directed therapy protocol in a hospital environment. Refer to the Edwards FloTrac, Acumen IQ and TruWave DPT sensor indications for use statement for information on target patient population specific to the sensor being used.
The Edwards Lifesciences Acumen Hypotension Prediction Index feature provides the clinician with physiological insight into a patient's likelihood of future hypotensive events (defined as mean arterial pressure < 65 mmHg for at least one minute in duration) and the associated hemodynamics. The Acumen HPI feature is intended for use in operating room (OR) patients receiving advanced hemodynamic monitoring. The Acumen HPI feature is considered to be additional quantitative information regarding the patient's physiological condition for reference only and no therapeutic decisions should be made based solely on the Hypotension Prediction Index (HPI) parameter.
Refer to the Intended Use statement for a complete list of measured and derived parameters available for each patient population.
# HemoSphere Advanced Monitor with HemoSphere Tissue Oximetry Module
The noninvasive FORE-SIGHT ELITE tissue oximeter module is intended for use as an adjunct monitor of absolute regional hemoglobin oxygen saturation of blood under the Sensors in individuals at risk for reducedflow or no-flow ischemic states. The FORE-SIGHT ELITE tissue oximeter module is intended to allow for the display of StO2 on the HemoSphere advanced monitor.
- . When used with large sensorsis indicated for use on adults and transitional adolescents ≥40 kg.
- . When used with Medium Sensors, the FORE-SIGHT ELITE tissue oximeter module is indicated for use on pediatric subjects ≥3 kg.
- . When used with Small Sensors, the FORE-SIGHT ELITE tissue oximeter module is indicated for cerebral use on pediatric subjects <8 kg and non-cerebral use on pediatric subjects <5kg.
{9}------------------------------------------------
Refer to the Intended Use statement for a complete list of measured and derived parameters available for each patient population.
## HemoSphere Advanced Monitor with HemoSphere ClearSight Module
Note: The ClearSight indication is identical to the Predicate K160552 (Cleared June 1, 2016) with the exception of the addition of the HemoSphere Advanced Monitoring Platform.
The HemoSphere Advanced Monitor when used with the HemoSphere ClearSight module, pressure controller and a compatible Edwards finger cuff are indicated for patients over 18 years of age in which the balance between cardiac function, fluid status and vascular resistance needs continuous assessment. It may be used for monitoring hemodynamic parameters in conjunction with a perioperative goal directed therapy protocol in a hospital environment. In addition, the noninvasive system is indicated for use in patients with co-morbidities for which hemodynamic optimization is desired and invasive measurements are difficult. The HemoSphere advanced monitor and compatible Edwards finger cuffs noninvasively measures blood pressure and associated hemodynamic parameters.
Refer to the ClearSight finger cuff indications for use statements for information on target patient population specific to the finger cuff being used.
Intended Use: The HemoSphere Advanced Monitoring Platform is intended to be used by qualified personnel or trained clinicians in a critical care environment in a hospital setting. The HemoSphere advanced monitoring platform is intended for use with compatible Edwards Swan-Ganz and oximetry catheters, FloTrac sensors, Acumen IQ sensors, TruWave DPTs, ForeSight Elite sensors, and ClearSight finger cuffs.
> A comprehensive list of parameters available while monitoring with the HemoSphere advanced monitor and a connected HemoSphere Swan-Ganz module are listed below. Only iCO, iCI, iSVR, and iSVRI are available to the pediatric patient population.
| Parame<br>ter | Description | Sub-<br>System<br>Module<br>Used | Patient<br>Populat<br>ion | Hospital<br>Environm<br>ent |
|---------------|---------------------------------------------------------|----------------------------------|---------------------------|-----------------------------|
| CO | continuous cardiac output | | | |
| sCO | STAT cardiac output | HemoSp | Adult | Operating |
| CI | continuous cardiac index | here | only | Room, |
| sCI | STAT cardiac index | Swan- | | Intensive |
| EDV | right ventricular end<br>diastolic volume | Ganz | | Care Unit, |
| sEDV | STAT right ventricular<br>end diastolic volume | Module | | Emergency Room |
| EDVI | right ventricular end<br>diastolic volume index | | | |
| sEDVI | STAT right ventricular<br>end diastolic volume<br>index | | | |
| HRavg | averaged heart rate | | | |
| LVSW | left ventricular stroke<br>work index | | | |
| PVR | pulmonary vascular<br>resistance | | | |
| PVRI | pulmonary vascular<br>resistance index | | | |
| RVEF | right ventricular ejection<br>fraction | | | |
| sRVEF | STAT right ventricular<br>ejection fraction | | | |
| RVSW | right ventricular stroke<br>work index | | | |
| SV | stroke volume | | | |
| SVI | stroke volume index | | | |
| SVR | systemic vascular<br>resistance | | | |
| SVRI | systemic vascular<br>resistance index | | | |
| iCO | intermittent cardiac output | | Adult<br>and<br>Pediatric | |
| iCI | intermittent cardiac index | | | |
| iSVR | intermittent systemic<br>vascular resistance | | | |
| iSVRI | intermittent systemic<br>vascular resistance index | | | |
{10}------------------------------------------------
A comprehensive list of parameters available for adult and pediatric patient populations while monitoring with the HemoSphere Advanced Monitor and a connected HemoSphere oximetry cable are as listed below:
| Parameter | Description | Sub-<br>System<br>Module<br>Used | Patient<br>Population | Hospital<br>Environment |
|-----------|-------------|----------------------------------|-----------------------|-------------------------|
|-----------|-------------|----------------------------------|-----------------------|-------------------------|
{11}------------------------------------------------
| SvO2 | Mixed Venous Oxygen<br>Saturation | HemoSp<br>here<br>Oximetry<br>Cable | Adult<br>and<br>Pediatric | Operating<br>Room,<br>Intensive<br>Care Unit,<br>Emergency Room |
|-------|-------------------------------------|-------------------------------------|---------------------------|-----------------------------------------------------------------|
| ScvO2 | Central Venous Oxygen<br>Saturation | | | |
A comprehensive list of parameters available for adult and pediatric patient populations while monitoring with the HemoSphere advanced monitor and both a connected HemoSphere Swan-Ganz module and oximetry cable are listed below:
| Parame<br>ter | Description | Sub-<br>System<br>Module<br>Used | Patient<br>Populat<br>ion | Hospital<br>Environm<br>ent |
|---------------|------------------------------------------------------------------------|----------------------------------------------|---------------------------|-----------------------------------------------------------------|
| DO2 | Oxygen Delivery | | | |
| DO2I | Oxygen Delivery Indexed | HemoSp<br>here | | |
| VO2 | Oxygen Consumption | | | Operating<br>Room,<br>Intensive<br>Care Unit,<br>Emergency Room |
| VO2e | Estimated Oxygen<br>Consumption when<br>ScvO2 is being monitored | Swan-Ganz<br>Module<br>and<br>HemoSp<br>here | Adult<br>and<br>Pediatric | |
| VO2I | Oxygen Consumption<br>Index | | | |
| VO2Ie | Estimated Oxygen<br>Consumption Index when<br>ScvO2 is being monitored | Oximetry<br>Cable | | |
A comprehensive list of parameters available for adult patient populations while monitoring with the HemoSphere Advanced Monitor and a connected HemoSphere pressure cable are as listed below:
| Parame<br>ter | Description | Sub-<br>System<br>Module<br>Used | Patient<br>Populat<br>ion | Hospital<br>Environm<br>ent |
|---------------|-------------------------------------------------------------|-------------------------------------|---------------------------|-----------------------------------------------|
| CO/<br>CI | Continuous Cardiac<br>Output1/ Continuous<br>Cardiac Index1 | | | Operating<br>Room, |
| CVP | Central Venous Pressure | HemoSp<br>here<br>Pressure<br>Cable | Adult<br>only | Intensive<br>Care Unit,<br>Emergenc<br>y Room |
| DIA | Systemic arterial diastolic<br>blood<br>pressure | | | |
| DIAPAP | pulmonary artery diastolic<br>blood<br>pressure | | | |
{12}------------------------------------------------
| dP/dt | Systolic slope2 | | |
|-------|--------------------------------------------|--|--|
| Eadyn | Dynamic Arterial<br>Elastance2 | | |
| MAP | Mean Arterial Pressure | | |
| MPAP | Mean Pulmonary Arterial<br>Pressure | | |
| PPV | pulse pressure variation1 | | |
| PRART | Pulse rate | | |
| SV/ | Stroke Volume1/ | | |
| SVI | Stroke Volume Index1 | | |
| SVR/ | Systemic Vascular<br>Resistance1/ Systemic | | |
| SVRI | Vascular Resistance1<br>Index | | |
| SVV | Stroke Volume Variation1 | | |
| SYS | Systolic Blood Pressure | | |
| HPI | Acumen Hypotension<br>Prediction Index | | |
A comprehensive list of parameters available for adult patient populations while monitoring with the HemoSphere Advanced Monitor and a
connected HemoSphere pressure cable and a connected HemoSphere oximetry cable are as listed below:
| Parame<br>ter | Description | Sub-<br>System<br>Module<br>Used | Patient<br>Populat<br>ion | Hospital<br>Environm<br>ent |
|---------------|------------------------------------------------------------------------|----------------------------------|---------------------------|---------------------------------------------------------------------|
| DO2 | Oxygen Delivery | | | |
| DO2I | Oxygen Delivery Indexed | HemoSp<br>here | | |
| VO2 | Oxygen Consumption | | | Operating<br>Room,<br>Intensive<br>Care Unit,<br>Emergenc<br>y Room |
| VO2e | Estimated Oxygen<br>Consumption when ScvO2<br>is being monitored | Swan-<br>Ganz<br>Module | Adult | |
| VO2I | Oxygen Consumption<br>Index | and<br>HemoSp<br>here | only | |
| VO2Ie | Estimated Oxygen<br>Consumption Index when<br>ScvO2 is being monitored | Oximetry<br>Cable | | |
A comprehensive list of parameters available while monitoring with the HemoSphere advanced monitor and both a connected HemoSphere Swan-
{13}------------------------------------------------
| Parameter | Description | Sub-System<br>Module Used | Patient<br>Population | Hospital<br>Environment |
|-----------|-------------------------------------|---------------------------------------------------------|-----------------------|-------------------------|
| CO20s | 20-second<br>cardiac output | HemoSphere | Adult<br>only | Operating<br>room, |
| CI20s | 20-second<br>cardiac index | Swan-Ganz<br>module and<br>HemoSphere<br>pressure cable | | intensive<br>care unit, |
| SV20s | 20-second<br>stroke volume | | | emergency<br>room |
| SVI20s | 20-second<br>stroke volume<br>index | | | |
Ganz Module and a connected HemoSphere pressure cable are as listed below.
Tissue oxygen saturation, StO2, can be monitored with the HemoSphere Advanced Monitor, a connected HemoSphere Tissue Oximetry Module, and the FORE-SIGHT ELITE Tissue Oximeter Module. Refer to the FORE-SIGHT ELITE HemoSphere Advanced Monitoring Platform Operators Manual for specific information on the intended use environment and patient population.
| Parame<br>ter | Description | Sub-System<br>Module Used | Patient<br>Popula<br>tion | Hospital<br>Environm<br>ent |
|---------------|-----------------------------|------------------------------------------------------------------------------|-------------------------------|--------------------------------------------------------------------|
| StO2 | Tissue oxygen<br>saturation | HemoSphere<br>Tissue<br>Oximetry<br>Module and<br>Fore-Sight<br>Elite Module | Adult<br>and<br>Pediatri<br>c | Operating<br>Room,<br>Intensive<br>Care Unit,<br>Emergency<br>Room |
A comprehensive list of parameters available while monitoring with the HemoSphere advanced monitor and a connected HemoSphere ClearSight module are listed below
| Parame<br>ter | Description | Sub-<br>System<br>Module<br>Used | Patient<br>Popula<br>tion | Hospital<br>Environm<br>ent |
|---------------|-----------------------------------------------------------|--------------------------------------------|---------------------------|----------------------------------|
| CO/CI | Continuous Cardiac<br>Output/ Continuous<br>Cardiac Index | HemoSp<br>here<br>ClearSig<br>ht<br>Module | Adult<br>only | Operating<br>Room,<br>Intensive |
| DIA | noninvasive arterial<br>diastolic blood<br>pressure | | | Care Unit,<br>Emergenc<br>y Room |
{14}------------------------------------------------
| MAP | Noninvasive Mean Arterial<br>Pressure |
|------|--------------------------------------------------|
| PPV | pulse pressure variation |
| PR | Noninvasive Pulse rate |
| SV/ | Stroke Volume/ |
| SVI | Stroke Volume Index |
| SVR/ | Systemic Vascular |
| SVRI | Resistance Systemic<br>Vascular Resistance Index |
| SVV | Stroke Volume Variation |
| SYS | Systolic Blood Pressure |
A comprehensive list of parameters available for adult patient populations while monitoring with the HemoSphere advanced monitor and both a connected HemoSphere ClearSight module and oximetry cable are listed below
| Parame<br>ter | Description | Sub-<br>System<br>Module<br>Used | Patient<br>Populat<br>ion | Hospital<br>Environm<br>ent |
|---------------|------------------------------------------------------------------------|------------------------------------------------------------------------------------------|---------------------------|----------------------------------------------|
| DO2 | Oxygen Delivery | | | |
| DO2I | Oxygen Delivery Indexed | HemoSp<br>here<br>ClearSig<br>ht<br>Module<br>and<br>HemoSp<br>here<br>Oximetry<br>Cable | Adult<br>only | Operating<br>Room,<br>Intensive<br>Care Unit |
| VO2 | Oxygen Consumption | | | |
| VO2e | Estimated Oxygen<br>Consumption when ScvO2<br>is being monitored | | | |
| VO2I | Oxygen Consumption<br>Index | | | |
| VO2Ie | Estimated Oxygen<br>Consumption Index when<br>ScvO2 is being monitored | | | |
{15}------------------------------------------------
#### VI. Comparison of Technological Characteristics with the Predicate Devices:
The existing HemoSphere Advanced Monitoring Platform which is the primary predicate for this submission consists of:
- I HemoSphere Advanced Monitor
- 트 HemoSphere Swan-Ganz Module
- 트 HemoSphere Oximetry Cable
- 트 HemoSphere Pressure Cable
- I HemoSphere Tissue Oximetry Module and FORE-SIGHT ELITE Tissue Oximeter Module
- I Acumen Hypotension Prediction Index (use with FloTrac only)
The subject and predicate devices are based on the following same technological elements:
- Platform: The subject device uses the same platform as the Predicate (K190205 cleared August 29, 2019).
- I Non-Invasive technology: The subject and Predicate (K160552 June 1, 2016) are Noninvasive blood pressure measurement systems and use the same method of measurement.
- . Accessories/Components: The subject and the Predicate (K160552 June 1, 2016) device both use previously cleared Pressure controller, Heart Reference Sensor and ClearSight Cuffs for measurement of non-invasive parameters.
The following technological differences exist between the subject and predicate devices:
- Algorithm update: An updated optional continuous cardiac output algorithm has been incorporated into the HemoSphere Advanced Monitoring Platform. This algorithm provides the option for a more responsive Continuous Cardiac Output (averaged 20 seconds) parameter as compared to that from the existing algorithm (average of 3 to 7 minutes). The new Algorithm uses the same SwanGanz Module and Pressure Cable to provide already available cardiac output parameters as the predicate K190205 August 29, 2019.
{16}------------------------------------------------
- 트 Algorithm update: For the key blood pressure (Noninvasive) parameters (MAP, SYS, DIA and PR) the algorithm has been changed to the existing algorithm used by HemoSphere Advanced Monitoring Platform (K190205 cleared August 29, 2019) to compute hemodynamic parameters (CO, CVP. MAP, DIA, MPAP, SV, SYS) when used with HemoSphere Pressure Cable (minimally invasive).
- ' Algorithm update: The FORE-SIGHT ELITE Tissue Oximeter Module algorithm for determining StO2 was updated to make more robust to better handle sensor disruption scenarios caused by user or environmental factors and made more dynamic for pediatric applications by increasing StO2 responsiveness. There are no new features with this algorithm update, and the device has comparable clinical accuracy when compared to the predicate device.
The purpose of this 510(k) submission is to add the following additional new features and modifications to the HemoSphere Advanced Monitoring Platform (K190205, cleared August 29, 2019):
# Addition of the ClearSight™ (Non-Invasive) functionality to the HemoSphere Advanced Monitoring Platform (K190205, previously cleared August 29, 2019):
- HemoSphere ClearSight Module (HEMCSM10): A new sub-system module: ClearSight Module (subject of this 510(k) submission) is being added to the HemoSphere Advanced Monitoring Platform (K190205, previously cleared August 29, 2019.
- ClearSight Calibration feature (Optional feature): ClearSight Calibration provides an additional option to the user to manually calibrate blood pressure (BP) to an external reference.
- ClearSight workflow options: Allows the user flexibility to choose the mode of monitoring based on the patient's position throughout the monitoring session.
- * Modifications to existing elements of the HemoSphere Advanced Monitoring Platform (previously cleared in K190205, August 29, 2019):
{17}------------------------------------------------
- I Graphical User Interface Modifications: The graphical user interface (GUI) is being updated to include additional screens specifically for the non-invasive (or ClearSight™) blood pressure and hemodynamic parameters.
- I Fluid Responsiveness Test for ClearSight: This feature is currently cleared for the minimally invasive (FloTrac) modes (K190205, cleared August 29, 2019). This same feature is now being made available for non-invasive ClearSight mode.
- I Cybersecurity Updates: Cybersecurity enhancements have been made to the HemoSphere Advanced Monitoring Platform toaddress Cybersecurity risks.
- I New Continuous Cardiac Output Algorithm: An updated cardiac algorithm (FastCCO) has been implemented for continuous cardiac output. This algorithmprovides the option for a more responsive Continuous Cardiac Output parameter as compared to the existing algorithm cleared in K163381 on April 14, 2017.
- I FORESIGHT ELITE Tissue Oximeter Module Algorithm update: The FORE-SIGHT ELITE Tissue Oximeter Module algorithm for determining StO2 was updated to make more robust to better handle sensor disruption scenarios caused by user or environmental factors and made more dynamic for pediatric applications by increasing StO2 responsiveness. There are no new features with this algorithm update, and the device has comparable clinical accuracy when compared to the predicate device.
# & Modifications to the Labeling of HemoSphere Advanced Monitoring Platform (previously cleared in K190205, August 29, 2019):
- I The HemoSphere Advanced Monitoring Platform Operator's Manual (Cleared in K190205 August 29, 2019) is being updated to include and address the additional and modified features that are the subject of this subject 510(k) premarket notification.
- I The ClearSight™ Finger Cuffs (Cleared in K160552 June 1, 2016 and K190130 June 21, 2019) indications for use are being updated to add the HemoSphere Advanced
{18}------------------------------------------------
| Performance<br>Data: | monitoring Platform to the indications for use as part of this subject 510(k). No other changes are occurring to the ClearSight Finger Cuffs. The Pressure Controller/Wrist Unit ( <i>Cleared in K160552 June 1, 2016</i> ) indications for use are being updated to add the HemoSphere Advanced monitoring Platform to the indications for use as part of this subject 510(k). No other changes are occurring to the Pressure Controller/Wrist Unit. The Heart Reference Sensor ( <i>K182245 cleared November 30, 2018</i> ) indications for use are being updated to add the HemoSphere Advanced monitoring Platform to the indications for use as part of this subject 510(k). No other changes are occurring to the Heart Reference Sensor. The following verification activities were performed in support of a substantial equivalence determination for the modifications being made as part of this submission.<br><br><b>Usability Study</b><br><br>Usability study was conducted per FDA's guidance document " <i>Applying Human Factors and Usability Engineering to Medical Devices</i> " to investigates primary operating functions and critical tasks of the system for any usability issues regarding HemoSphere Advanced Monitoring Platform that may lead to patient or user harm.<br><br>The usability study demonstrated that the intended users can perform primary operating functions and critical tasks of the system without any usability issues that may lead to patient or user harm.<br><br><b>System Verification (Non-Clinical Performance):</b><br><br>Completion of all verification and validation activities demonstrated that the subject devices meet their predetermined design and performance specifications. Verification activities performed confirmed that the differences in the design and materials used did not adversely affect the safety and effectiveness of the subject device.<br><br>Measured and derived parameters were tested using a bench simulation. Additionally, individual modules were tested at a system level to verify the safety of these modules. They were also integrated as a system and verified for their safety and effectiveness. All tests passed. |
|----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
{19}------------------------------------------------
## Electrical Safety and Electromagnetic Compatibility (EMC)
Electrical safety and EMC testing were conducted on the subject HemoSphere Advanced Monitoring Platform, consisting of the HemoSphere Monitor, the ClearSight Module, Pressure Controller, Heart Reference Sensor and finger cuff. The system complies with the IEC 60601-1, IEC 60601-1-2, IEC 60601-1-6, IEC 60601-1-8, IEC 62304, IEC 62366-1, IEC 60601-2-34, IEC 60601-2-57, IEC 60601-2-49 and IEC 80601-2-49. All tests passed.
# Wireless Coexistence Testing
Bench and simulated environment testing were performed on the entire HemoSphere Advanced Monitoring Platform, including all sub-system modules and interfacing analog inputs and outputs. All tests passed.
# Software Verification
Software verification was performed per FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices". Software on each of the individual modules was tested at a sub-system level to ensure the safety of the device. All tests passed.
The technological characteristics of the predicates and subject device are Conclusions identical. The HemoSphere Advanced Monitoring Platform has successfully passed functional and performance testing, including software verification and validation, and bench studies. The test performed to demonstrate that the HemoSphere Advanced Monitoring Platform including the subject new features for the HemoSphere ClearSight Module and modified features for the platform is substantially equivalent to the legally marketed predicates.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.