The BMEYE CC Nexfin is intended to, non-invasively and continuously, measure blood pressure, hemodynamic parameters, functional saturation of arterial hemoglobin (SpO2), and total hemoglobin concentration (SpHb) in adult patients. The pulse-oximetry component of CC Nexfin is indicated for use during both no motion and motion conditions and for patients who are well or poorly perfused. The CC Nexfin monitor does not feature (physiological) alarms, therefore the continuous availability of pulse-oximetry data should be treated as a series of spot-checks rather than continuous monitoring. The device is intended for use by physicians or other properly trained medical personnel in a hospital or other appropriate clinical setting.
Device Story
CC Nexfin is a non-invasive hemodynamic monitor for adult patients. It utilizes the Peňáz-Wesseling method to measure continuous blood pressure (systolic, diastolic, mean) and heart rate; cardiac output is derived from the blood pressure waveform. It integrates Masimo Rainbow SET technology for pulse-CO oximetry to measure SpO2 and SpHb. Used in hospitals or clinical settings by physicians or trained personnel. The device lacks physiological alarms; pulse-oximetry data functions as spot-checks. Output assists clinicians in hemodynamic assessment and patient management. Benefits include non-invasive monitoring of hemodynamic status and hemoglobin levels without arterial catheterization.
Clinical Evidence
Bench testing only. Verification and validation activities included risk analysis, design reviews, code inspections, and hardware/software implementation verification. SpO2 and SpHb functions were verified by Masimo.
Technological Characteristics
Non-invasive hemodynamic monitor. Employs Peňáz-Wesseling method for blood pressure/cardiac output and Masimo Rainbow SET technology for SpO2/SpHb. System includes pulse-CO oximetry sensor. Software-based analysis for hemodynamic parameter derivation.
Indications for Use
Indicated for adult patients requiring non-invasive, continuous measurement of blood pressure, hemodynamic parameters, SpO2, and SpHb. Suitable for patients with varying perfusion levels and during motion or no-motion conditions. Intended for use by trained medical personnel in clinical settings.
Regulatory Classification
Identification
A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.
{0}------------------------------------------------
K122381
## Section 5 - 510(k) Summary
APR 2 2 2013
# 510(k) Summary (per 21 CFR 807.92)
| Name of Submitter: | BMEYE BV<br>Hoogoorddreef 60<br>1101 BE Amsterdam<br>the Netherlands |
|---------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person | Mr. J van Goudoever,<br>BMEYE BV<br>Tel + 31 20 7512410<br>Fax + 31 20 7512419 |
| Date Prepared | August 1, 2012 |
| Trade Names | Nexfin Model 2, Trade name CC Nexfin |
| Classification | Class II |
| Classification Name | CFR 870.1130, System, measurement, blood pressure, non-invasive, DXN<br>CFR 870.2770, Plethysmograph, impedance, DSB<br>CFR 870.2700, Oximeter, DQA |
| Predicate | Nexfin Model 2, K101123 |
| Device Description | The BMEYE CC Nexfin hemodynamic monitor is a non-invasive monitor<br>that enables the continuous assessment of a patient's hemodynamic<br>function based on the scientific method of Peňáz - Wesseling. The device<br>measures continuous non-invasive blood pressure (Systolic, Diastolic and<br>Mean) and heart rate as well as a Cardiac Output (CO), which is derived,<br>non-invasively, from the blood pressure waveform. The CC Nexfin also<br>enables the simultaneous measurement of SpO2 and SpHb using a pulse-<br>CO oximetry sensor. The operation of the blood pressure, cardiac output,<br>SpO2 and SpHb measurement are identical to the operation in the<br>predicate (K101123). |
| Intended Use: | The BMEYE CC Nexfin is intended to, non-invasively and<br>continuously, measure blood pressure, hemodynamic parameters,<br>functional saturation of arterial hemoglobin (SpO2), and total hemoglobin<br>concentration (SpHb) in adult patients. The pulse-oximetry component of<br>CC Nexfin is indicated for use during both no motion and motion<br>conditions and for patients who are well or poorly perfused. The CC<br>Nexfin monitor does not feature (physiological) alarms, therefore the<br>continuous availability of pulse-oximetry data should be treated as a<br>series of spot-checks rather than continuous monitoring. The device is<br>intended for use by physicians or other properly trained medical<br>personnel in a hospital or other appropriate clinical setting. |
| Technology | The device employs identical technology compared to the predicate<br>(K101123) for blood pressure and cardiac output measurement and the<br>measurement of SpHb and SpO2 using the Masimo Rainbow SET<br>technology. |
NEXFIN MODEL 2, Traditional 510(k) Filing BMEYE BV - August 2012 .
·
Commission of the count
.
..
.
.
5-2] Page | : | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
·
{1}------------------------------------------------
#### K122381
| Functional / Safety<br>Testing | The CC Nexfin has successfully undergone safety testing as well as<br>functional testing to demonstrate equivalence to its predicate devices. The<br>following quality assurance measures were applied to the device:<br>Risk Analysis Requirements Review Design reviews Code Inspections Verification and Validation H/W and S/W Implementation Verification Testing of the SpO2 and<br>SpHb functions by Masimo Biocompatibility Testing Safety Testing |
|--------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Conclusion | The results of this testing demonstrates that the device is safe and effective<br>and substantially equivalent to its predicate device. |
:
Comments of the country
.
NEXFIN MODEL 2, Traditional 510(k) Filing BMEYE BV - August 2012 Page | : | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
Page 19 of 664
{2}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name arranged in a circular fashion around a symbol. The symbol consists of a stylized caduceus, which is a traditional symbol of medicine, with a single staff entwined by a serpent.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 22, 2013
Edward Lifesciences BMEYE c/o Mr. Jeroen van Goudoever Hoogoorddreeef 60 1101 BE Amsterdam The Netherlands
Re: K122381
> Trade/Device Name: Nexfin Model 2 Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive blood pressure measurement system Regulatory Class: Class II (two) Product Code: DXN, DSB, DQA Dated: April 3, 2013 Received: April 08, 2013
Dear Mr. van Goudoever:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
#### Page 2 – Mr. Jeroen van Goudoever
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
OwenParis -S
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
### Indications for Use
510(k) Number (if known):
Device Name: Nexfin Model 2 (trade name CC Nexfin)
Indications For Use:
The BMEYE CC Nexfin is intended to, non-invasively and continuously, measure blood pressure, hemodynamic parameters, functional saturation of arterial hemoglobin (SpO2), and total hemoglobin concentration (SpHb) in adult patients. The pulse-oximetry component of CC Nexfin is indicated for use during both no motion and motion conditions and for patients who are well or poorly perfused. The CC Nexfin monitor does not feature (physiological) alarms, therefore the continuous availability of pulse-oximetry data should be treated as a series of spot-checks rather than continuous monitoring. The device is intended for use by physicians or other properly trained medical personnel in a hospital or other appropriate clinical setting.
Prescription Use _ V AND/OR Over-The-Counter Use _
(Part 21 CFR 801 Subpart D)
(21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Owen P. Faris -S
2013.04.22
16:04:32 -04'00'
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.