The purpose and function of the N-3000E Patient Monitor is to: - noninvasively and continuously monitor functional arterial oxygen saturation and pulse rate (using an accessory SpO2 sensor); - noninvasively and continuously monitor ECG and heart rate (using accessory ECG leads). When connected to the N-3200 Display/Printer, the purpose and function of the combined device is expanded to: - display and print out associated ECG and Plethysmographic waveforms, SpO2, pulse rate and heart rate. When further connected to the N-3100 Blood Pressure Monitor, the intended use of the interconnected N-3000E/N-3100/N-3200 is also to: - noninvasively and automatically measure systolic, diastolic and mean arterial blood pressure and to derive pulse rate from these measurements (using accessory blood pressure cuffs); and display and print out systolic, diastolic and mean arterial blood pressure (on the N-3200 display/printer) When the N-3100 is connected to the N-3200 Display/Printer, the purpose and function of the combined device (compared to K945947) is expanded to: - display and print out associated pulse rate and systolic, diastolic and mean arterial blood pressure. The N-3000E Patient Monitor is intended for use in hospital, hospital-type and hospital transport environments as a "standalone" product and when connected to the N-3200 display/printer and/or the N-3100 blood pressure monitor. It is also intended for use to monitor neonatal, pediatric, or adult patients. The N-3000E, N-3200 and N-3100 are for prescription use only.
Device Story
N-3000E Patient Monitor system; monitors SpO2, pulse rate, ECG, heart rate, and NIBP (when connected to N-3100). Inputs: surface ECG electrodes, SpO2 sensor, NIBP cuff. Processing: ECG signal amplification, noise filtering, and QRS detection algorithm; SpO2 processing via established Nellcor algorithm. Outputs: numerical heart rate, SpO2, blood pressure; graphical waveforms via N-3200 Display/Printer. Used in hospital/transport settings by clinicians. Benefits: continuous physiological monitoring for patient status assessment and clinical decision-making.
Clinical Evidence
Bench testing only. Compliance with AAMI/ANSI EC-13 standard (except clause 3.2.6.1) for ECG module. Verification testing confirmed that adding ECG and display/printer functionality did not adversely affect the safety or effectiveness of the predicate pulse oximetry and NIBP functions.
Technological Characteristics
ECG module: 3-lead surface electrodes, amplification, noise filtering. SpO2/NIBP: identical to predicate N-3000/N-3100. Connectivity: modular interconnection between N-3000E, N-3100, and N-3200. Standards: AAMI/ANSI EC-13 (except 3.2.6.1).
Indications for Use
Indicated for neonatal, pediatric, and adult patients in hospital, hospital-type, and hospital transport environments for continuous noninvasive monitoring of functional arterial oxygen saturation, pulse rate, ECG, heart rate, and automatic noninvasive blood pressure measurement.
Regulatory Classification
Identification
An electrocardiograph is a device used to process the electrical signal transmitted through two or more electrocardiograph electrodes and to produce a visual display of the electrical signal produced by the heart.
Predicate Devices
Datascope PASSPORT
Protocol Systems PROPAQ 104
NELLCOR SYMPHONY N-3000 Pulse Oximeter
NELLCOR SYMPHONY N-3100 Blood Pressure Monitor
Related Devices
K963400 — NELLCOR SYMPHONY N-3000 PATIENT MONITOR WITH SPO2 AND ECG AND N-3200 DISPLAY/PRINTER · Nellcor Puritan Bennett, Inc. · Nov 26, 1996
K151063 — Monitor B40 · Ge Medical Systems China Co., Ltd. · Dec 15, 2015
K980411 — NPB3900 SERIES PATIENT MONITORS · Analogic Corp. · Aug 20, 1998
K970445 — VITALMAX 4000 CONFIGURED/VITALMAX 4000 MODULAR/VITALMAX 4100/MINIPACK 3100/3000/VITALMAX 800+/MINIPACK 300 · Pace Tech Medical Monitors, Inc. · Jan 30, 1998
Submission Summary (Full Text)
{0}
K955642
# ATTACHMENT 7
JUL 2 1998
# 510(k) SUMMARY, REVISED
## (f) Intended Use:
The purpose and function of the N-3000E Patient Monitor is to:
- noninvasively and continuously monitor functional arterial oxygen saturation and pulse rate (using an accessory $\mathrm{SpO}_2$ sensor);
- noninvasively and continuously monitor ECG and heart rate (using accessory ECG leads).
When connected to the N-3200 Display/Printer, the purpose and function of the combined device is expanded to:
- display and print out associated ECG and Plethysmographic waveforms, SpO2, pulse rate and heart rate.
When further connected to the N-3100 Blood Pressure Monitor, the intended use of the interconnected N-3000E/N-3100/N-3200 is also to:
- noninvasively and automatically measure systolic, diastolic and mean arterial blood pressure and to derive pulse rate from these measurements (using accessory blood pressure cuffs); and display and print out systolic, diastolic and mean arterial blood pressure (on the N-3200 display/printer)
When the N-3100 is connected to the N-3200 Display/Printer, the purpose and function of the combined device (compared to K945947) is expanded to:
- display and print out associated pulse rate and systolic, diastolic and mean arterial blood pressure.
The N-3000E Patient Monitor is intended for use in hospital, hospital-type and hospital transport environments as a "standalone" product and when connected to the N-3200 display/printer and/or the N-3100 blood pressure monitor. It is also intended for use to monitor neonatal, pediatric, or adult patients. The N-3000E, N-3200 and N-3100 are for prescription use only.
The N-3000E, the N-3000E connected to the N-3200, the N-3000E connected to the N-3100 and the N-3000E connected to both the N-3200 and N-3100 have the following similar intended uses to the predicate devices, in terms of function/purpose, environment of use and patient population:
Datascope PASSPORT - pulse oximetry, ECG, NIBP, display/printing functions; hospital and hospital-type environments; adult/pediatric patients
Protocol Systems PROPAQ 104 - pulse oximetry, ECG, NIBP, display/printing functions; hospital, hospital-type, environments; adult/pediatric patients
NELLCOR SYMPHONY™ N-3000 Pulse Oximeter - pulse oximetry function; hospital, hospital-type; adult/pediatric/neonate patients
{1}
NELLCOR SYMPHONY™ N-3100 Blood Pressure Monitor, standalone and in combination with the NELLCOR SYMPHONY™ N-3000 Pulse Oximeter; pulse oximetry, NIBP functions; hospital, hospital-type environments; adult/pediatric/neonate patients.
The N-3000E, N-3200 and N-3100 differ from the Datascope PASSPORT and Protocol Systems PROPAQ 104 in having fewer measurement/monitoring parameters available. They also differ in that they include neonates in their intended patient population but do not include home or mobile environments.
The N-3000E/N-3200/N-3100 devices have the same patient population as the predicate N-3000 pulse oximeter and predicate N-3100 blood pressure monitor, including adult, pediatric and neonate patients. They differ only in that they feature ECG as an additional measurement parameter and include a display/printer function, but do not include home or mobile environments.
(g) Technological Characteristics
Intrinsic electrocardiac signals are collected by conventional patient surface electrodes and conducted via a 3-lead ECG lead set and cable to the N-3000E's ECG processing module. These signals then undergo amplification and noise filtering. A software algorithm operates on the processed ECG signal to identify sequential QRS complexes and, hence, determine heart rate for numerical presentation on the N-3000E's front panel display. The processed ECG signal may also be supplied to the N-3200 Display/Printer for graphical presentation and printout. The ECG module is designed to comply with the AAMI/ANSI EC-13 standard, with the exception of clause 3.2.6.1. Test results confirming compliance with this standard have been included in the 510(k) submission.
The technological characteristics of the pulse oximetry function of the N-3000E are identical to those in the predicate N-3000 pulse oximeter. The same algorithm is used in both products. Testing has been conducted to confirm that modification of the N-3000 pulse oximeter to add an ECG function and access to the N-3200 display/printer have not affected the safety or effectiveness of the oximetry function of the N-3000E.
Safety and effectiveness of the N-3000E, alone and in combination with the N-3200 display/printer and/or the N-3100 blood pressure monitor, have been confirmed by complying with the requirements of the Reviewer Guidance for Premarket Notification Submissions, November 1993, through design, testing and labeling.
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.