K172893 · Pressure Products Medical Device Manufacturing, LLC · DRC · Dec 21, 2017 · Cardiovascular
Device Facts
Record ID
K172893
Device Name
SafeSept Needle Free Transseptal Guidewire
Applicant
Pressure Products Medical Device Manufacturing, LLC
Product Code
DRC · Cardiovascular
Decision Date
Dec 21, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1390
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The SafeSept® is indicated for use in procedures where access to the left atrium via the transseptal technique is desired. The SafeSept® Transseptal Guidewire is intended for single use only.
Device Story
SafeSept Needle Free Transseptal Guidewire is a 0.031" diameter, 180cm long nitinol guidewire used in hospital transseptal procedures. Device is inserted through a transseptal dilator and sheath via the femoral vein. Sharp distal tip perforates the fossa ovalis when supported by the dilator/sheath; becomes atraumatic when unsupported. Radiopaque coil and proximal marker bands allow physician fluoroscopic visualization and positioning relative to the needle. Device replaces traditional transseptal needles to gain left atrial access. Benefits include atraumatic navigation and reduced risk during septal puncture. Used by physicians in clinical settings.
Clinical Evidence
Bench testing only. Testing included visual/dimensional inspection, particulate testing (USP <788>), surface finish, corrosion resistance, tensile strength, fracture/flex, curve integrity, torque, packaging, and radiopacity. Biocompatibility testing performed per ISO 10993 (cytotoxicity, sensitization, irritation, systemic toxicity, pyrogenicity, genotoxicity, hemocompatibility). Sterilization validated per ANSI/AAMI/ISO 11135-1.
Technological Characteristics
Nitinol super-elastic wire; 0.031" diameter, 180cm length. Features radiopaque coil and printed markers. Sterilized via 100% ethylene oxide (ANSI/AAMI/ISO 11135-1). Biocompatibility per ISO 10993. Mechanical device; no software or energy source.
Indications for Use
Indicated for patients requiring transseptal access to the left atrium via the right side of the heart. Intended for single use.
Regulatory Classification
Identification
A trocar is a sharp-pointed instrument used with a cannula for piercing a vessel or chamber to facilitate insertion of the cannula.
Special Controls
*Classification.* Class II (special controls). Except for trocars that are reprocessed for multiple use, the device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
K241042 — SafeSharp Transseptal Needle (CLICK071, CLICK171, CLICK098, CLICK198) · Pressure Products Medical Device Manufacturing, LLC · May 17, 2024
K081986 — SAFESEPT TRANSSEPTAL TROCAR GUIDEWIRE · Oscor, Inc. · Nov 14, 2008
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 21, 2017
Pressure Products Medical Device Manufacturing LLC Andrew Armour Managing Director 1 School Street Morton, Pennsylvania 19070
Re: K172893
Trade/Device Name: SafeSept Needle Free Transseptal Guidewire Regulation Number: 21 CFR 870.1390 Regulation Name: Trocar Regulatory Class: Class II Product Code: DRC Dated: September 21, 2017 Received: September 22, 2017
Dear Mr. Armour:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good
{1}------------------------------------------------
manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Nicole G. Ibrahim -S
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K172893
Device Name
SafeSept Needle Free Transseptal Guidewire
Indications for Use (Describe)
The SafeSept® is indicated for use in procedures where access to the left atrium via the transseptal technique is desired. The SafeSept® Transseptal Guidewire is intended for single use only.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### 510(k) Summary
#### Submitter
Pressure Products Medical Device Manufacturing LLC. 1 School Street Morton, PA 19070 Phone: 610-285-9858 Fax: 610-285-9859
Contact Person: Andrew Armour Prepared: October 23, 2017
#### Identification of the Device
Proprietary-Trade Name: SafeSept® Needle Free Transseptal Guidewire Device Class: Class II Classification Name: Trocar (CFR 870.1390) Common/Usual Name: Needle Free Transseptal Guidewire Product Code: DRC
#### Equivalent Legally Marketed Devices
Oscor, SafeSept® Needle Free Transseptal Trocar Guidewire, K130843
#### Description of the Devices
The SafeSept® Needle Free Transseptal Guidewire (SafeSept NF) is 0.031" in diameter and 180cm long and consists of a flexible memory wire material called nitinol. The components of the SafeSept NF include the 0.031" transseptal guidewire, dispenser and straightener, and sterile packaging and labeling. There is one model number for the SafeSept NF, SSNF. The SafeSept NF is sterilized by 100% ethylene oxide cvcle and is for single-use only. The SafeSept is used in a healthcare facility/hospital.
The SafeSept NF is used in transseptal procedures to gain access to the left atrium through the right side of the heart. The device is used by inserting the device through a transseptal dilator and sheath located in the femoral vein. The SafeSept NF is then advanced to the fossa ovalis with the support of the dilator and sheath. The SafeSept NF's sharp distal tip is able to perforate through the fossa ovalis, a thin wall between the right and left atrium, when it is supported by a transseptal dilator and sheath. When the wire is no longer supported, it is atraumatic and operates as a typical 0.031" diameter guidewire. A radiopaque coil located along the shaft allows for fluoroscopic visualization of the wire within the left atrium and proximal marker bands help determine the location of the quidewire tip in relation to the needle. The guidewire duration in the body is less than 24 hours.
#### Indications for Use
The SafeSept is indicated for use in procedures where access to the left atrium via the transseptal technique is desired. The SafeSept®Transseptal Guidewire is intended for single use only.
The Indications for Use statement for the SafeSept Transseptal Guidewire device is identical to the predicate device. Both devices have the same intended use and are used in conjunction with a transseptal dilator and sheath to create the primary puncture in the interatrial septum. The devices quide the dilator and introducer through the septum from the right side of the heart to the left side.
#### Comparison of Technological Characteristics with the Predicate Device
The technological characteristics of this device are similar to the predicate device. The main difference between the subject device and the predicate device is the guidewire. The predicate SafeSept NF is 150cm in length and the proposed device is 180cm. The subject and predicate device are based on the follow technological elements:
- Nitinol wire that is super-elastic so that it can have a 'J' curve with a sharp tip .
{4}------------------------------------------------
- Radiopaque coil used by the physician to locate SafeSept and to guide supporting devices like . the transseptal dilator and sheath across the fossa
- Device inserted through the transseptal dilator so that it can be used in the transseptal procedure
- Creates the primary puncture of the fossa with the sharp tip
- Taper is used to dilate the fossa further so that the dilator can then be inserted through the fossa . rather than using a transseptal needle
- Printed markers are along the body for the physician to use as a guide to understand how far the ● SafeSept is in relation to the supporting devices
#### Performance Testing
The following performance tests were performed in support of the substantial equivalence to the predicate:
- Visual and dimensional inspection
- Particulate testing (USP <788>, Light Obscuration Method) ●
- Surface finish assessment ●
- Corrosion resistance ●
- Guidewire tensile strength ●
- Fracture and Flex testing
- . Curve Integrity testing
- Torque Testing
- Packaging testing ●
- . Radiopacity assessment
In addition, the sterilization conditions have been validated in accordance with ANSI/AAMI/ISO 11135- 1. Sterilization of health care products - Ethylene oxide - Part 1: Requirements for development, validation and routine control of sterilization process for medical device is sterilized to a SAL of 10-6. The results of the testing demonstrated the subject device performed equivalently to the predicate device.
#### Biocompatibility Data
The biocompatibility evaluation for the SafeSept NF was conducted in accordance with FDA 510(k) Memorandum- #G95-1 "Use of International Standard ISO-10993, 'Biological Evaluation of Medical Devices Part 1: Evaluation and Testing,"" June 16, 2016, and International Standard ISO 10993-"Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by the FDA. The following tests were performed:
- ISO 10993-5 Cytotoxicity ●
- . ISO 10993-10 Sensitization
- ISO 10993-10 Irritation/Intracutaneous Reactivity ●
- ISO 10993-11 Acute Systemic Toxicity
- ISO 10993-11 Pyrogenicity ●
- ISO 10993-3 Genotoxicity
- ISO 10993-4 Hemocompatibility ASTM Hemolysis Complete ●
- ISO 10993-4 Hemocompatibility In-vivo Dog Thromboresistance .
- ISO 10993-4 Hemocompatibility Complement Activation Complete with C3a & SC5b-9 ●
The SafeSept NF is considered an external communicating device with circulating blood contact and limited exposure (less than 24 hours). The SafeSept NF met the requirements set forth in ISO-10993.
#### Conclusion
When compared to the predicate device, the SafeSept Needle Free Transseptal Guidewire is substantially equivalent in design, technological characteristics, materials and performance testing.
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.