K141811 · Mortara Instrument, Inc. · MHX · Nov 19, 2014 · Cardiovascular
Device Facts
Record ID
K141811
Device Name
MORTARA MONITORING WAVEFORM VIEWER
Applicant
Mortara Instrument, Inc.
Product Code
MHX · Cardiovascular
Decision Date
Nov 19, 2014
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 870.1025
Device Class
Class 2
Attributes
Software as a Medical Device
Intended Use
The Monitoring Waveform Viewer is a software component intended to be utilized within another software application. The software component can continuously receive and display data generated by multiparameter patient monitoring systems in near real time, display reported alarm states, display static waveforms for review of existing data, and perform measurements with the support of a caliper tool. The Monitoring Waveform Viewer software component is not intended to be used as an alarm device. The data and analysis provided by the Monitoring Waveform Viewer is reviewed, confirmed, and used by trained medical personnel in the diagnosis of patients with various conditions.
Device Story
Monitoring Waveform Viewer is a software component integrated into host medical applications; receives waveform and parameter data (ECG, ART, PA, ICP, IABP, RR, EtCO2, SpO2) from multi-parameter patient monitoring systems. Displays data in near real-time; provides visual indicators for alarm states (High/Medium/Low priority). Features include magnification tool for signal inspection and manual caliper tool for horizontal/time and vertical/signal amplitude measurements. Used by trained medical personnel in clinical settings to review patient data, confirm findings, and support diagnostic decision-making. Does not function as an alarm system; no alarm silencing or notification capabilities. Output used for printing or inclusion in patient records.
Clinical Evidence
No clinical data. Bench testing only. Software verification and validation were performed to confirm compliance with predetermined functional and performance requirements.
Technological Characteristics
Software-only component; no physical materials, sterilization, or electrical/EMC requirements. Operates as a drawing surface with parameter and waveform display areas. Supports manual caliper and magnification tools. Connectivity via integration into host software applications. Algorithm type: rule-based data visualization and measurement processing.
Indications for Use
Indicated for use by trained medical personnel as a software component within a host application to display waveforms, parameters, and alarm states from multi-parameter patient monitoring systems. Supports review of static waveforms, signal magnification, and manual caliper measurements for amplitude and interval analysis. Not for use as a primary or secondary alarm device; does not provide alarm acknowledgement, silencing, or audible/visual alarm notifications.
Regulatory Classification
Identification
The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.
K101449 — INTELLIVUE PATIENT MONITOR · Philips Medizinsysteme Boeblingen GmbH, Cardiac AN · Jun 18, 2010
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is an abstract image of three faces in profile, stacked on top of each other, representing the department's focus on people and their well-being.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 19, 2014
Mortara Instrument, Inc. Ms. Amy Yang Sr. Regulatory Affairs Engineer 7865 N. 86th Street Milwaukee, WI 53224
Re: K141811
Trade/Device Name: Mortara Monitoring Waveform Viewer Regulation Number: 21 CFR 870.1025 Regulation Name: Monitor, physiological, patient (with arrhythmia detection or alarms) Regulatory Class: Class II Product Code: MHX Dated: October 2, 2014 Received: October 6, 2014
Dear Ms. Yang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{1}------------------------------------------------
Page 2 – Ms. Amy Yang
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Melissa A. Torres -S
For Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K141811
Device Name Monitoring Waveform Viewer
Indications for Use (Describe)
The Monitoring Waveform Viewer is indicated for use as follows:
The Monitoring Waveform Viewer is a software component utilized within another software application.
The Monitoring Waveform Viewer performs the following:
· Display of data generated by multi-parameter patient monitoring systems continuously producing waveforms and parameters.
· Display of reported alarm states produced by such monitoring systems.
· Display of static waveforms for review of existing data, with the possibility to activate a magnification tool. The magnification tool will enlarge the signal located under the magnification icon.
· Perform measurements with the support of a caliper tool (providing horizontal/time measurements and vertical/signal amplitude measurements). The caliper tool can be activated when the Monitoring Waver displays still waveforms.
· Provide selected waveform episodes to the host application for printing or inclusion in the patient record.
The Monitoring Waveform Viewer:
- · Is not a primary or secondary alarm device.
· Does not provide a user interface for acknowledging or silencing active alarms.
· Does not generate audible or visual (flashing) notifications of active alarms.
The Monitoring Waveform Viewer does not provide any support for implementing an application that is part of an alarm system.
Type of Use (Select one or both, as applicable)
| Description Use (Part 91 CFR 901 Subpart D) | Over-The-Counter Use (91 CFR 901 Subpart C) |
|---------------------------------------------------|---------------------------------------------|
| <span style="text-decoration: overline;">☒</span> | ☐ |
> Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Abbreviated 510(k) Notification
## 510(k): Mortara Monitoring Waveform Viewer Summary K141811
# Submitter:
Date: November 18, 2014
| Amy Yang, Sr. Regulatory Affairs Engineer | |
|-------------------------------------------|----------------------------------|
| Mortara Instrument, Inc. | |
| 7865 N. 86th Street | |
| Milwaukee, WI 53224 | |
| FAX: | (414) 354-4760 |
| Phone: | (414) 354-1600 |
| Contact: | Amy Yang (see above) |
| Trade Name: | Mortara Monitoring Waveform View |
| Trade Name: | Mortara Monitoring Waveform Viewer |
|-----------------------------|----------------------------------------------------------------------------|
| Common Name: | Monitoring Waveform Viewer |
| Classification Name: | Monitor, Physiological, Patient (with Arrhythmia Detection or Alarms) |
| Classification Description: | Arrhythmia detector and alarm (including ST-segment measurement and alarm) |
| Classification Regulation: | 21 CFR §870.1025 |
| Product Code: | MHX |
# Legally marketed devices to which S.E. is claimed:
| Mortara Monitoring Waveform<br>Viewer | Predicate 510(k) Number | Predicate Manufacturer / Model |
|---------------------------------------|-------------------------|--------------------------------------------------------|
| Surveyor Central Station | K131929 | Mortara Instrument, Inc. / Surveyor<br>Central Station |
# Description:
The Monitoring Waveform Viewer is a software component that includes a system for reviewing snapshots of monitoring data and offers a measurement module with user controlled calipers and magnifier glass.
Medical device data or information system applications may utilize the Monitoring Waveform Viewer software component that includes a stand-alone executable module that functions as a host application. The Monitoring Waveform Viewer is organized as a drawing surface with three areas:
- a parameters area located at the bottom of the window, .
- a parameter waveform area located at the right and
- . a main waveform area.
The parameters area is dedicated to report the values of parameters not associated with displayed waveforms. The parameter waveform area is dedicated to report the values of parameters associated to the displayed waveforms.
The application supports several numerical parameters such as: ECG (electrocardiogram), ART (arterial pressure), PA (pulmonary artery invasive blood pressure), ICP (intracranial pressure), IABP (intra-aotic balloon pump pressure), RR (respiration rate), EtCO2 (end tidal carbon dioxide), SpO2 (oxygen saturation), etc.
The Monitoring Waveform Viewer receives data parameters and waveforms supplied from multi-parameter patient monitoring systems. The Monitoring Waver is able to display alarm states by highlighting the corresponding parameter box. The Monitoring Waveform Viewer operating in near real time mode supports alarm priorities of High Priority, Medium Priority, and Low Priority.
{4}------------------------------------------------
# Abbreviated 510(k) Notification
The Monitoring Waveform Viewer allows the user to right-click over any ECG lead or the ECG parameter box to activate a context menu for selecting the ECG gain and displayed leads. The user is able to select a list of active (with data) ECG leads received from the host application. It also allows the user to select gains of 5, 10, 20 and 40mm/mV. The Monitoring Waveform Viewer provides a caliper tool for performance amplitude and interval measurements on the displayed traces. The caliper can be placed anywhere on the waveform display area.
Further description and details concerning use of the Monitoring Waveform Viewer can be found in:
- Tab 2 Device Description, Monitoring Waveform Viewer Block Diagram .
- . Tab 12 Proposed Labeling documents, Monitoring Waveform Viewer product brochure, Monitoring Waveform Viewer User Manual.
## Technology Comparison:
The Mortara Monitoring Waver utilizes the same or similar technology for each parameter as utilized by the predicate devices.
## Intended Use:
The Monitoring Waveform Viewer is a software component intended to be utilized within another software application. The software component can continuously receive and display data generated by multiparameter patient monitoring systems in near real time, display reported alarm states, display static waveforms for review of existing data, and perform measurements with the support of a caliper tool.
The Monitoring Waveform Viewer software component is not intended to be used as an alarm device. The data and analysis provided by the Monitoring Waveform Viewer is reviewed, confirmed, and used by trained medical personnel in the diagnosis of patients with various conditions.
## Indications for Use:
The Monitoring Waveform Viewer is indicated for use as follows: The Monitoring Waveform Viewer is a software component utilized within another software application.
The Monitoring Waveform Viewer performs the following:
- Display of data generated by multi-parameter patient monitoring systems continuously producing waveforms and parameters.
- . Display of reported alarm states produced by such monitoring systems.
- . Display of static waveforms for review of existing data, with the possibility to activate a magnification tool. The magnification tool will enlarge the signal located under the magnification icon.
- . Perform measurements with the support of a caliper tool (providing horizontal/time measurements and vertical/signal amplitude measurements). The caliper tool can be activated when the Monitoring Waveform Viewer displays still waveforms.
- . Provide selected waveform episodes to the host application for printing or inclusion in the patient record.
The Monitoring Waveform Viewer:
- Is not a primary or secondary alarm device. .
- . Does not provide a user interface for acknowledging or silencing active alarms.
- . Does not generate audible or visual (flashing) notifications of active alarms.
The Monitoring Waveform Viewer does not provide any support for implementing an application that is part of an alarm system.
{5}------------------------------------------------
# Abbreviated 510(k) Notification
## Performance Testing:
Sterilization Validation - The Mortara Monitoring Waveform Viewer is not sterilized or sterilizable, and therefore this section does not apply to the Waveform Viewer itself.
Shelf Life Testing – The Mortara Monitoring Waveform Viewer is not sterilizable, and therefore this section does not apply to the Waveform Viewer itself.
Biocompatibility Testing - The Moritoring Waveform Viewer is software only, and therefore this section does not apply to the Waveform Viewer itself.
Software Testing - Software for the Mortara Monitoring Waveform Viewer was designed and developed according to a robust software development process, and was rigorously verified and validated. Test results indicated that the Mortara Monitoring Waver complies with its predetermined specification.
Electrical Safety - The Mortara Monitoring Waver is software only, and therefore this section does not apply to the Waveform Viewer itself.
Electromagnetic Compatibility Testing - The Mortara Monitoring Waveform Viewer is software only, and therefore this section does not apply to the Waveform Viewer itself.
Performance Testing – Bench – The Mortara Monitoring Waveform Viewer was tested in accordance with internal requirements and procedures, and test results indicated that the device complies with the predetermined requirements. This testing includes performance and functional.
Performance Testing – Animal – Animal performance testing was not performed and is not necessary to demonstrate safety and effectiveness of the Mortara Monitoring Waveform Viewer.
Performance Testing – Clinical performance testing was not performed and is not necessary to demonstrate safety and effectiveness of the Mortara Monitoring Waveform Viewer.
Conclusion - The results of these activities demonstrate that the Mortara Monitoring Waver is as safe, as effective, and performs as well as or better than the predicate device.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.