BETTERCARE PLATFORM - ADVANCED WEB CLIENT

K171361 · Better Care S.L. · MWI · Feb 5, 2018 · Cardiovascular

Device Facts

Record IDK171361
Device NameBETTERCARE PLATFORM - ADVANCED WEB CLIENT
ApplicantBetter Care S.L.
Product CodeMWI · Cardiovascular
Decision DateFeb 5, 2018
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.2300
Device ClassClass 2
AttributesSoftware as a Medical Device, Pediatric

Intended Use

The BetterCare Platform - Advanced Web Client is a software application capable of displaying physiological data and other patient information to clinicians in near real-time through a web browser. The information is captured from patient monitors by the Draeger Infinity Gateway interfaced with the BetterCare Platform and sent through the hospital local area network to wired or wireless devices running the Advanced Web Client application. The information must be analyzed and interpreted by an expert clinician only. The Better Care Platform is not indicated for use as a real-time monitor. The BetterCare Platform - Advanced Web Client is intended to display data and waveforms from attached medical devices in a pass-though fashion, the BetterCare Platform - Advanced Web Client is agnostic to the age or condition of patients attached to the medical devices from which data is received. For example, the special requirements for neonatal, pediatric or adult patients are all provided by the medical devices directly attached to the patient. The BetterCare Platform - Advanced Web Client simply reproduces that numeric, waveform and other clinical data provided to it for communications and display. The Better Care Platform - Advanced Web Client is installed on compatible desktop and mobile devices including cellular telephones, and is intended to be used by clinicians to: - remotely view patient physiological information as captured by Draeger Infinity Gateway - remotely view waveforms as captured by Draeger Infinity Gateway - remotely view other patient data as captured by Draeger Infinity Gateway The BetterCare Platform - Advanced Web Client is not patient connected and is not a primary alarm source. It can display the following physiological data captured by Draeger Infinity Gateway: - ECG Waveform - Arterial Pressure Waveform - SpO2 Waveform - CO2 Waveform - Central Venous Pressure Waveform - Pulmonary Arterial Pressure Waveform - Right Auricular Pressure Waveform - Left Auricualar Pressure Waveform - Resp Waveform - Heart Rate - Respiratory Rate - Oxygen Saturation - Central Venous Pressure - Systolic Arterial Pressure Invasive - Mean Arterial Pressure Invasive - Diastolic Arterial Pressure Invasive - Systolic Blood Pressure Non invasive - Mean Bloood Pressure Non invasive - Diastolic Blood Pressure Non invasive - Systolic Pulmonary Arterial Pressure - Mean Pulmonary Arterial Pressure - Diastolic Pulmonary Arterial Pressure - Right Auricular Pressure - Left Auricular Pressure BetterCare Platform Advanced Web Client is intended to be installed on wired and wireless devices. including cellular telephones, and is not intended to be used where such devices are prohibited. BetterCare Platform Advanced Web Client is intended for use by trained clinicians in hospital environments as a diagnostic aid and not as a replacement for direct viewing of any of the monitoring devices interfaced.

Device Story

Software application; displays physiological data/waveforms from Draeger Infinity Gateway; operates via hospital LAN/WLAN; accessed via web browser on desktop/mobile devices. Input: physiological signals (ECG, SpO2, CO2, pressures, heart/respiratory rates) from patient monitors. Output: near real-time visual reproduction of numeric/waveform data. Used by clinicians in hospital settings as diagnostic aid; not patient-connected; not primary alarm source. Clinicians interpret data to support clinical decision-making; does not replace direct monitoring. Benefits: enables remote access to patient status within hospital network.

Clinical Evidence

No clinical data submitted. Bench testing only; verification of waveform and numeric data points performed using Draeger Infinity Gateway interface.

Technological Characteristics

Software-only application; runs on desktop/mobile devices via web browser. Connectivity: Hospital LAN/WLAN. Security: SSL via HTTPS. Standards: ISO 14971 (risk management), ISO 62304 (software lifecycle).

Indications for Use

Indicated for trained clinicians in hospital environments to remotely view near real-time physiological data, waveforms, and patient information captured by Draeger Infinity Gateway. Agnostic to patient age or condition. Not for use as a real-time monitor or primary alarm source. Not for use where wireless devices are prohibited.

Regulatory Classification

Identification

A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue. February 5, 2018 Bettercare S.L. Bernat Sales Chief Executive Officer Xaloc 6 Sabadell, 08203 ES Re: K171361 Trade/Device Name: BetterCare Platform - Advanced Web Client Regulation Number: 21 CFR 870.2300 Regulation Name: Cardiac Monitor (Including Cardiotachometer and Rate Alarm) Regulatory Class: Class II Product Code: MSX Dated: December 18, 2017 Received: December 21, 2017 Dear Bernat Sales: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); {1}------------------------------------------------ and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, M& Millelreme for Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K171361 Device Name BetterCare Platform - Advanced Web Client #### Indications for Use (Describe) The BetterCare Platform - Advanced Web Client is a software application capable of displaying physiological data and other patient information to clinicians in near real-time through a web browser. The information is captured from patient monitors by the Draeger Infinity Gateway interfaced with the BetterCare Platform and sent through the hospital local area network to wired or wireless devices running the Advanced Web Client application. The information must be analyzed and interpreted by an expert clinician only. The Better Care Platform is not indicated for use as a real-time monitor. The BetterCare Platform - Advanced Web Client is intended to display data and waveforms from attached medical devices in a pass-though fashion, the BetterCare Platform - Advanced Web Client is agnostic to the age or condition of patients attached to the medical devices from which data is received. For example, the special requirements for neonatal, pediatric or adult patients are all provided by the medical devices directly attached to the patient. The BetterCare Platform - Advanced Web Client simply reproduces that numeric, waveform and other clinical data provided to it for communications and display. The Better Care Platform - Advanced Web Client is installed on compatible desktop and mobile devices including cellular telephones, and is intended to be used by clinicians to: - remotely view patient physiological information as captured by Draeger Infinity Gateway - remotely view waveforms as captured by Draeger Infinity Gateway - remotely view other patient data as captured by Draeger Infinity Gateway The BetterCare Platform - Advanced Web Client is not patient connected and is not a primary alarm source. It can display the following physiological data captured by Draeger Infinity Gateway: - ECG Waveform - Arterial Pressure Waveform - SpO2 Waveform - CO2 Waveform - Central Venous Pressure Waveform - Pulmonary Arterial Pressure Waveform - Right Auricular Pressure Waveform - Left Auricualar Pressure Waveform - Resp Waveform - Heart Rate - Respiratory Rate - Oxygen Saturation - Central Venous Pressure - Systolic Arterial Pressure Invasive - Mean Arterial Pressure Invasive - Diastolic Arterial Pressure Invasive - Systolic Blood Pressure Non invasive - Mean Bloood Pressure Non invasive - Diastolic Blood Pressure Non invasive - Systolic Pulmonary Arterial Pressure - Mean Pulmonary Arterial Pressure - Diastolic Pulmonary Arterial Pressure {3}------------------------------------------------ - Right Auricular Pressure - Left Auricular Pressure BetterCare Platform Advanced Web Client is intended to be installed on wired and wireless devices. including cellular telephones, and is not intended to be used where such devices are prohibited. BetterCare Platform Advanced Web Client is intended for use by trained clinicians in hospital environments as a diagnostic aid and not as a replacement for direct viewing of any of the monitoring devices interfaced. Type of Use (Select one or both, as applicable) X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the logo for Bcare BETTER CARE, S.L. The logo features a stylized, curved shape in shades of orange and red on the left side. To the right of this shape is the word "care" in a blue sans-serif font. Below the word "care" is the text "BETTER CARE, S.L." in a smaller, sans-serif font, underlined with a thin gray line. ## K171361 - 510(k) Summary DATE OF SUBMISSION: SUBMITTER NAME: SUBMITTER ADDRESS: CONTACT: TELEPHONE: e-mail: COMMON NAME: PRODUCT CODE: 2016-02-26 BETTERCARE S.L. Xaloc, 6 08203 - BARCELONA SPAIN Bernat Sales +34 65 693964 8 fda-submission@bettercare.es DEVICE TRADE NAME: BetterCare Platform - Advanced Web Client System, Network And Communication, Physiological Monitors, Web Client, Web Viewer, MView CLASSIFICATION NAME: Cardiac monitor (870.2300) MSX REGULATION DESCRIPTION: Cardiac monitor (including cardiotachometer and rate alarm) ### PREDICATE DEVICE(S): - K100133 Airstrip RPM (Airstrip Technologies) - Primary Predicate - K061994 Web Viewer, Pocket Viewer and Cellular Viewer (General Electric) Reference Predicate ### DEVICE DESCRIPTION: The BetterCare Platform - Advanced Web Client is a software application capable of displaying physiological data and other patient information to clinicians in near real-time through a web browser. The information is captured by Draeger Infinity Gateway interfaced with the BetterCare Platform and sent through the hospital local area network to wireless devices running the Advanced Web Client application. The information must be analyzed and interpreted by an expert clinician only. ### SUMMARY OF COMPARISON WITH PREDICATE DEVICE: BetterCare Platform - Advanced Web Client is compared with the following previously cleared devices: Airstrip RPM (K122133) - Primary Predicate Web Viewer, Pocket Viewer and Cellular Viewer (K061994) - Reference Predicate Comparison of the proposed devices with the predicate device is summarized in the following table: {5}------------------------------------------------ Image /page/5/Picture/0 description: The image shows the logo for Bcare, which is a company that provides better care. The logo features a stylized letter B in a gradient of red and orange, followed by the word "care" in blue. Below the word "care" is the text "BETTER CARE, S.L." in a smaller font. # K171361 | Item | New Device<br>BetterCare Platform<br>- Advanced Web<br>Client | Primary Predicate<br>K122133<br>Airstrip RPM | Reference Predicate<br>K061994<br>Web viewer, Pocket<br>Viewer and Cellular<br>Viewer | |-----------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Intended use | Display of physiologic<br>data and other patient<br>information to clinicians<br>in near real-time through<br>a web browser as<br>captured by Draeger<br>Infinity Gateway<br>interfaced with the<br>BetterCare platform. It is<br>intended to be used by<br>clinicians to:<br>- remotely view patient<br>physiological information<br>as captured by Draeger<br>Infinity Gateway<br>- remotely view near<br>real-time waveforms as<br>captured by Draeger<br>Infinity Gateway<br>- remotely view other<br>near real-time patient<br>data as captured by<br>Draeger<br>Infinity<br>Gateway. | Display of physiologic and other<br>patient information generated by<br>other medical devices and patient<br>information system.<br><br>AirStrip RPM is intended to be used<br>by clinicians for the following<br>purposes:<br>- by using a cellular telephone or<br>other device on which AirStrip RPM<br>is installed, to review physiologic<br>data of a patient when the clinician<br>is not at the hospital<br>- to view the near real-time<br>waveforms remotely<br>- to remotely review other standard<br>or critical near real-time patient data<br>from the monitored system<br>- to provide a request for remote<br>consultation regarding a patient's<br>waveform or other data. | Display of Information<br>received from other<br>networked devices. It is<br>comprised of a Mobile Care<br>Server and Web Viewer or<br>Pocket Viewer clients | | Target population | Clinicians | Clinicians | Clinicians | | Communication | Secure Sockets Layer<br>(SSL) via HTTPS | Secure Sockets Layer (SSL) via<br>HTTPS | Secure Sockets Layer<br>(SSL) via HTTPS | | Communication<br>Methods | LAN / WLAN Wi-Fi | Cellular Modem, Wi-Fi | Cellular Modem, Wi-Fi | | Data<br>Source<br>Location | Hospital | Hospital | Hospital | | Security<br>Administration | Yes | Yes | Yes | | Where used | Within the hospital<br>where the clinician has<br>wired or wireless access<br>to the hospital local area<br>network. | Anywhere the clinician has remote<br>internet access | Anywhere the clinician has<br>remote internet access. | | Client Platforms /<br>Operating systems | Windows 7 and above;<br>iOS 6 and above,<br>Android 4.0 and above | iOS / iPhone | Windows<br>95/98/ME/NT4.0/2000/XP<br>or MacOS X 10.1 on<br>desktop or portable | | Item | New Device<br>BetterCare Platform<br>- Advanced Web<br>Client | Primary Predicate<br>K122133<br>Airstrip RPM | Reference Predicate<br>K061994<br>Web viewer, Pocket<br>Viewer and Cellular<br>Viewer | | | | | computers,<br>PcoketPC, PDA or cellular<br>phone with Windows<br>Moobile 2003 / 5.0 / Phone<br>Edition | | Ability to view near<br>real-time data | Yes | Yes | Yes | | Physiologic<br>data<br>displayed | Waveform Data<br>- ECG Waveform<br>- Arterial Pressure<br>Waveform<br>- SpO2 Waveform<br>- CO2 Waveform<br>- Central Venous<br>Pressure Waveform<br>- Pulmonary Arterial<br>Pressure Waveform<br>- Right Auricular<br>Pressure Waveform<br>- Left Auricular<br>Pressure Waveform<br>- Resp Waveform<br><br>Near real-time<br>physiological data<br>- Heart Rate<br>- Respiratory Rate<br>- Oxygen Saturation<br>- Central Venous<br>Pressure<br>- Systolic Arterial<br>Pressure Invasive<br>- Mean Arterial<br>Pressure Invasive<br>- Diastolic Arterial<br>Pressure Invasive<br>- Systolic Blood<br>Pressure Non<br>invasive<br>- Mean Blood<br>Pressure Non<br>invasive | Waveform Data<br>- ECG Waveform<br>- Blood Pressure Waveform<br>- O2 Waveform<br>- CO2 Waveform<br><br>Near real-time physiological data<br>- Heart Rate<br>- Respiratory Rate<br>- Oxygen Saturation<br>- Intracranial Pressure<br>- Central Venous Pressure<br>- Pulmonary Capillary Wedge<br>Pressure<br>- Cardiac Index<br>- Cardiac Output<br>- Cerebral Perfusion Pressure<br>- Urine Output<br>- Urine / Stool Mix Output<br>- Systolic Blood Pressure<br>Invasive<br>- Mean Arterial Pressure<br>Invasive<br>- Diastolic Blood Pressure<br>Invasive<br>- Systolic Blood Pressure Cuff<br>- Mean Arterial Pressure Cuff<br>- Diastolic Blood Pressure Cuff<br><br>Other patient information<br>displayed<br>- Vasoactive Infusions<br>- Antiarrythmics<br>- Sedation | Waveform Data<br>- ECG Waveform<br>- Blood Pressure Waveform<br>- O2 Waveform<br>- CO2 Waveform<br><br>Near real-time<br>physiological data<br>- Heart rate<br>- Respiratory Rate<br>- Oxygen Saturation<br>- Intracranial Pressure<br>- Central Venous Pressure<br>- Pulmonary<br>Capillary Wedge Pressure<br>- Cardiac Index<br>- Cardiac Output<br>- Cerebral Perfusion<br>Pressure<br>- Systolic Blood Pressure<br>Cuff<br>- Mean Blood Pressure Cuff<br>- Diastolic Blood Pressure<br>Cuff<br>- Systolic Blood Pressure<br>Invasive<br>- Mean Blood<br>Pressure<br>Invasive<br>- Diastolic Blood Pressure<br>Invasive<br>- Temperature | | Item | New Device<br>BetterCare Platform<br>- Advanced Web Client | Primary Predicate<br>K122133<br>Airstrip RPM | Reference Predicate<br>K061994<br>Web viewer, Pocket<br>Viewer and Cellular<br>Viewer | | | - Diastolic Blood Pressure Non invasive<br>- Systolic Pulmonary Arterial Pressure<br>- Mean Pulmonary Arterial Pressure<br>- Diastolic Pulmonary Arterial Pressure<br>- Right Auricular Pressure<br>- Left Auricular Pressure | Laboratory Data including:<br>o Blood Gas<br>o Chemistry<br>o Hematology<br>o Coagulation<br>- Allergies<br>- Medication | | | Contraindications | BetterCare Platform - Advanced Web Client is intended to be installed on wired and wireless devices, including cellular telephones, and is not intended to be used where such devices are prohibited. | AirStrip RPM software is intended for installation on cellular telephones and other wireless devices and is not intended for use anywhere cellular telephones or wireless devices are prohibited. | N/A | | Principal operation | BetterCare Platform - Advanced Web Client works by retrieving patient data from the hospital monitoring system and providing that data to the user's computer or device over the hospital local area network. | AirStrip RPM works by retrieving patient data from the hospital monitoring system and providing that data to the end user's device via WiFi or cellular modem over the Internet. | Pocket Viewer / Web Viewer works by retrieving data from the hospital monitoring system and providing that data to the end user's device via WiFi or cellular modem over the Internet. | | Client devices | Desktop or portable computers or mobile hand-held wireless devices running a web browser | iPhone | Desktop or portable computers or Windows mobile hand held PDA | {6}------------------------------------------------ Image /page/6/Picture/0 description: The image shows the logo for "Bcare BETTER CARE, S.L.". The logo features a stylized, flowing red line that resembles a cursive letter. To the right of this symbol, the word "care" is written in blue, sans-serif font. Below the word "care", the text "BETTER CARE, S.L." is written in a smaller, sans-serif font, underlined by a thin gray line. # K171361 – 510(k) Summary {7}------------------------------------------------ Image /page/7/Picture/0 description: The image contains the logo for "Bcare BETTER CARE, S.L.". The logo features a stylized, red, beta symbol on the left. To the right of the symbol is the word "care" in a blue, sans-serif font. Below the word "care" is the text "BETTER CARE, S.L." in a smaller, sans-serif font, underlined by a thin gray line. # K171361 – 510(k) Summary Table 5.1 Summary comparison of characteristics and features - proposed and predicate devices {8}------------------------------------------------ ## K171361 – 510(k) Summary ## INTENDED USE / INDICATIONS FOR USE: The BetterCare Platform - Advanced Web Client is a software application capable of displaying physiological data and other patient information to clinicians in near real-time through a web browser. The information is captured by Draeger Infinity Gateway interfaced with the BetterCare Platform and sent through the hospital local area network to wired or wireless running the Advanced Web Client application. The information must be analyzed and interpreted by an expert clinician only. The BetterCare Platform is not indicated for use as a real-time monitor. The BetterCare Platform - Advanced Web Client is installed in compatible desktop and mobile devices, including cellular telephones, and is intended to be used by clinicians to: - remotely view patient physiological information as captured by Draeger Infinity Gateway - remotely view waveforms as captured by Draeger Infinity Gateway - remotely review other patient data as captured by Draeger Infinity Gateway The BetterCare Platform - Advanced Web Client is not patient connected and is not a primary alarm source. It can display the following physiological data captured by Draeger Infinity Gateway: - ECG Waveform - Arterial Pressure Waveform - SpO2 Waveform - CO2 Waveform - Central Venous Pressure Waveform - Pulmonary Arterial Pressure Waveform - Right Auricular Pressure Waveform - Left Auricular Pressure Waveform - Resp Waveform - Heart Rate - Respiratory Rate - Oxygen Saturation - Central Venous Pressure - Systolic Arterial Pressure Invasive - Mean Arterial Pressure Invasive - Diastolic Arterial Pressure Invasive - Systolic Blood Pressure Non invasive - Mean Blood Pressure Non invasive - Diastolic Blood Pressure Non invasive - Systolic Pulmonary Arterial Pressure - Mean Pulmonary Arterial Pressure - Diastolic Pulmonary Arterial Pressure - Right Auricular Pressure - Left AuricularPressure BetterCare Platform - Advanced Web Client is intended to be installed on wired and wireless devices, {9}------------------------------------------------ Image /page/9/Picture/0 description: The image shows the logo for "Bcare BETTER CARE, S.L.". The logo features a stylized, curved line in shades of red and orange on the left side, resembling a check mark. To the right of this mark is the word "care" in a blue, sans-serif font. Below "care" is the text "BETTER CARE, S.L." in a smaller, sans-serif font, underlined with a thin line. 510(k) Premarket Notification BetterCare Platform - Advanced Web Client ## K171361 – 510(k) Summary including cellular telephones, and is not intended to be used where such devices are prohibited. BetterCare Platform - Advanced Web Client is intended for use by trained clinicians in hospital environments as a diagnostic aid and not as a replacement for direct viewing of any of the monitoring devices interfaced. ### SUMMARY DISCUSSION OF NON-CLINICAL DATA: The proposed device has been designed, developed, tested, verified and validated according to documented procedures and specific protocols in line with the following FDA quidance documents: - FDA Guidance on the Content of Premarket Submissions for Software Contained in Medical Devices - FDA Guidance on the Content of Premarket Submissions for Management of Cybersecurity in Medical Devices - FDA Guidance Cybersecurity for Networked Medical Devices Containing Off-the-shelf (OTS) Software - FDA Guidance General Principles of Software Validation - FDA Guidance Off-the-sheld Software Use in Medical Devices Design and development included identification, evaluation and control of potential hazards as per standard ISO 14971. Development, integration, verification testing have been successfully completed following standard ISO 62304. Testing included verification of all available listed waveform and near real-time data points. Known data sets were gathered using Draeger Infinity Gateway interfaced with BetterCare Platform and displayed on Advanced Web Client. All waveforms and near-real time physiological information points were verified and in all cases, Advanced Web Client displayed data points accurately. ### SUMMARY DISCUSSION OF CLINICAL DATA: No clinical data is submitted to support this premarket notification. ### CONCLUSIONS: We believe the intended use, the indications for use and performance of the BetterCare Platform -Advanced Web Client is substantially equivalent to the intended use, indications for use and performance of the primary predicate device. We also believe that the BetterCare Platform - Advanced Web Client does not suppose any new or increased risk compared with the primary predicate device. Based on the information included in this submission, we conclude that BetterCare Platform -Advanced Web Client is substantially equivalent to the stated primary predicate device.
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%