K220410 · Natec Medical , Ltd. · LIT · Jun 29, 2022 · Cardiovascular
Device Facts
Record ID
K220410
Device Name
Ebony HP PTA OTW 0.035 Catheter
Applicant
Natec Medical , Ltd.
Product Code
LIT · Cardiovascular
Decision Date
Jun 29, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1250
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Ebony HP PTA OTW 0.035" Catheter is indicated for; · Percutaneous Transluminal Angioplasty (PTA) in the peripheral vasculature, including iliac, femoral, popliteal, tibial, peroneal, subclavian, and renal arteries and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae. · Post-dilatation of balloon expandable and self-expanding stents in the peripheral vasculature.
Device Story
Over-the-wire (OTW) percutaneous transluminal angioplasty (PTA) catheter; features non-compliant inflatable balloon at distal end; atraumatic tapered tip; dual-lumen shaft; Y-hub proximal connector for inflation and guidewire insertion. Used by physicians in clinical settings for peripheral vascular interventions. Guidewire (max 0.035") facilitates navigation to stenosis. Inflation device connects to lateral port to expand balloon; radio-opaque markers provide visual positioning reference. Hydrophilic coating on distal shaft and balloon reduces friction. Device enables vessel dilation or stent post-dilatation; improves blood flow; restores patency in peripheral arteries and dialysis fistulae.
Clinical Evidence
Bench testing only. No clinical data. Testing included dimensional verification, balloon fatigue (repeat inflations and in-stent), inflation/deflation time, rated burst pressure (RBP), bond strength, deployment/retraction, and particulate evaluation. Biocompatibility testing included cytotoxicity, sensitization, systemic toxicity, hemocompatibility, and thrombogenicity in canine study.
Technological Characteristics
OTW PTA catheter; dual-lumen; non-compliant balloon; hydrophilic coating; radio-opaque markers. Materials biocompatible per ISO 10993. Dimensions: 40, 75, 135 cm lengths; 3.0-12.0 mm diameters. Mechanical operation via manual inflation device. No software or electronic components.
Indications for Use
Indicated for patients requiring percutaneous transluminal angioplasty in peripheral vasculature (iliac, femoral, popliteal, tibial, peroneal, subclavian, renal arteries) or treatment of obstructive lesions in native/synthetic arteriovenous dialysis fistulae; also indicated for post-dilatation of balloon-expandable and self-expanding stents.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
K083919 — AMPHIRION DEEP 0.014 OTW PTA BALLOON CATHETER · Invatec Innovative Technologies, S.R.L. · Mar 13, 2009
K202800 — Pacific Plus PTA Catheter · Medtronic Vascular,Inc(Formerly D.B.A Ev3 Inc.,Covidien, LLC) · Nov 18, 2020
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name written out to the right of the square.
June 29, 2022
Natec Medical Ltd. Roy Devassy Pallippatt Regulatory Affairs & Quality Manager Maeva Centre Building, Silicon Avenue, Ebene Business Park Reduit. 72201 Mauritius
Re: K220410
Trade/Device Name: Ebony HP PTA OTW 0.035" Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous catheter Regulatory Class: Class II Product Code: LIT Dated: May 20, 2022 Received: May 26, 2022
Dear Roy Devassy Pallippatt:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Gregory O'Connell Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K220410
Device Name Ebony HP PTA OTW 0.035" Catheter
Indications for Use (Describe)
The Ebony HP PTA OTW 0.035" Catheter is indicated for;
· Percutaneous Transluminal Angioplasty (PTA) in the peripheral vasculature, including iliac, femoral, popliteal, tibial, peroneal, subclavian, and renal arteries and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae.
· Post-dilatation of balloon expandable and self-expanding stents in the peripheral vasculature.
Type of Use (Select one or both, as applicable)
| <span style="font-family: Arial, sans-serif;"> <span style="font-size: 10.0pt;"> <span style="font-family: Wingdings;">ü</span> Prescription Use (Part 21 CFR 801 Subpart D)</span> </span> | <span style="font-family: Arial, sans-serif;"> <span style="font-size: 10.0pt;"> Over-The-Counter Use (21 CFR 801 Subpart C)</span> </span> |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------|
---
__ Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) SUMMARY
This 510(k) Summary is submitted in accordance with the requirements of 21 CFR Part 807.92.
| Applicant: | Natec Medical Ltd.<br>Maeva Centre Building,<br>Silicon Avenue<br>Ebene Business Park<br>Reduit - 72201, Mauritius |
|-------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Official Correspondent: | Roy Devassy Pallippatt<br>Regulatory Affairs Manager<br>Natec Medical Ltd.<br>Maeva Centre Building,<br>Silicon Avenue<br>Ebene Business Park<br>Reduit - 72201, Mauritius<br>Tel: +230 466 30 54<br>FAX: +230 466 67 70<br>Email: rdpallippatt@natec-medical.com |
- Preparation Date: June 28, 2022
## Device Name:
| Trade Name: | Ebony HP PTA OTW 0.035” Catheter |
|-------------------------|------------------------------------------------|
| Common/Regulatory name: | Percutaneous catheter |
| Classification Name: | Percutaneous transluminal angioplasty catheter |
| Generic name: | PTA Catheter |
| Regulation Number: | 21 CFR 870.1250 |
| Product Code: | LIT |
| Device Class: | Class II |
| Predicate Device: | ■ Charger PTA Balloon Dilatation Catheter - K112697 |
|-------------------|--------------------------------------------------------------------|
| Reference Device: | ■ Ebony PTA 0.035 Peripheral Dilatation Catheter - K143036/K103354 |
## Device Description:
The Ebony HP PTA OTW 0.035" Catheter is a standard Over The Wire (OTW) PTA catheter with a non-compliant inflatable balloon at the distal part with an atraumatic, tapered tip at its distal end to facilitate advancement of the catheter through the stenosis, and a luer lock fitting (Y hub) at the proximal end allowing the connection with an inflation device.
{4}------------------------------------------------
The catheter is a two lumen catheter; one lumen is used for inflation of the balloon and accessed via the lateral port of the Y hub. The second lumen, starting at the straight entry port of the Y hub, allows access to the distal tip of the catheter for guide wire insertion. The maximum recommended guide wire diameter is 0.035". The internal tubing of the balloon has two radio opaque markers to provide visual reference points for balloon positioning relative to the stenosis within the vessel. The working pressure range for the balloon is between the nominal pressure (NP) and the rated burst pressure (RBP).
A hydrophilic coating solution is applied on the distal section of the shaft and on the balloon to improve the pushability of the catheter by reduction of the friction coefficient of the outer body.
The Ebony HP PTA OTW 0.035" catheter is available in following sizes;
| Catheter Usable lengths 40, 75 & 135 cm | | | | | | | | | |
|-----------------------------------------|-------|---------------------|----|----|----|-----|-----|-----|-----|
| Balloon Size | | Balloon Length (mm) | | | | | | | |
| Balloon Diameter (mm) | | 20 | 40 | 60 | 80 | 100 | 120 | 150 | 200 |
| | 3.00 | ✓ | ✓ | ✓ | ✓ | ✓ | ✓ | ✓ | ✓ |
| | 4.00 | ✓ | ✓ | ✓ | ✓ | ✓ | ✓ | ✓ | ✓ |
| | 5.00 | ✓ | ✓ | ✓ | ✓ | ✓ | ✓ | ✓ | ✓ |
| | 6.00 | ✓ | ✓ | ✓ | ✓ | ✓ | ✓ | ✓ | ✓ |
| | 7.00 | ✓ | ✓ | ✓ | ✓ | ✓ | ✓ | ✓ | ✓ |
| | 8.00 | ✓ | ✓ | ✓ | ✓ | ✓ | ✓ | ✓ | ✓ |
| | 9.00 | ✓ | ✓ | ✓ | ✓ | | | | |
| | 10.00 | ✓ | ✓ | ✓ | ✓ | | | | |
| | 12.00 | ✓ | ✓ | ✓ | ✓ | | | | |
Table 1: Balloon Size Matrix
Note: For Catheters with usable length 75 & 135 cm all the sizes are available.
Catheter with usable length 40 cm is available only in balloon length up to 100mm for all available balloon diameter and only for balloon diameter 5.00 & 6.00 it is available in balloon length up to 120mm.
## Device Indication for Use:
The Ebony HP PTA OTW 0.035" Catheter is indicated for:
- Percutaneous Transluminal Angioplasty (PTA) in the peripheral vasculature, including iliac, femoral, popliteal, tibial, peroneal, subclavian, and renal arteries and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae.
- Post-dilatation of balloon expandable and self-expanding stents in the peripheral . vasculature.
{5}------------------------------------------------
## Comparison of Technical Characteristics:
The design, materials and manufacturing of the Ebony HP PTA OTW 0.035" Catheter is same or similar to those used for the predicate devices. The intended use for the Ebony HP PTA OTW 0.035" Catheter is also comparable to the predicate devices.
### Biocompatibility:
All materials used in the Ebony HP PTA OTW 0.035" Catheter are biocompatible based on biocompatibility testing results. The device has been tested according to "ISO 10993, Biological Evaluation of Medical Devices Part-1: Evaluation and Testing" and 21 CFR 58 (GLP regulations).
The following biocompatibility tests were completed on the Ebony HP PTA OTW 0.035" Catheter;
- Acute Systemic Toxicity Study
- 트 Cytotoxicity Study
- Hemocompatibility Study
- Hemolysis Test
- I Partial Thromboplastin Time (PTT)
- Sc5b-9 Complement Activation
- 트 Intracutaneous Reactivity Test
- Material Mediated Pyrogen Test
- I Skin Sensitization Study
- I Thrombogenicity in Canine Study
#### Performance Data:
Substantial equivalence has been demonstrated based on the results of non-clinical testing on the Ebony HP PTA OTW 0.035" Catheter which addressed the following considerations:
- I Dimensional Verification
- Balloon Fatigue test (Repeat Balloon Inflations)
- Catheter Torque Strength evaluation ■
- I Balloon Fatigue in stent (Repeat Balloon Inflations)
- 트 Balloon Inflation and Deflation Time
- 트 Balloon Rated Burst Pressure (RBP) and compliance test
- I Balloon Rated Burst Pressure (RBP in Stent)
- I Catheter Bond (sleeve, hub and tip) Strength
- I Balloon Preparation, Deployment and Retraction
- Coating Inspection I
- PTA Catheter Performance test
- Sizing & counting of particulate matter on catheters (Particulate evaluation) I
{6}------------------------------------------------
## Conclusion:
The subject device, the Ebony HP PTA OTW 0.035" Catheter met all the predetermined acceptance criteria of design verification and validation as specified by applicable standards, FDA guidance documents and test protocols. No new questions of safety or effectiveness were raised during the testing program.
Based on the similarities in the indication for use, device design, materials and the results of the non-clinical testing and analysis, the Ebony HP PTA OTW 0.035" Catheter is considered substantially equivalent to the aforementioned predicate devices.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.