Pacific Plus PTA Catheter

K202800 · Medtronic Vascular,Inc(Formerly D.B.A Ev3 Inc.,Covidien, LLC) · LIT · Nov 18, 2020 · Cardiovascular

Device Facts

Record IDK202800
Device NamePacific Plus PTA Catheter
ApplicantMedtronic Vascular,Inc(Formerly D.B.A Ev3 Inc.,Covidien, LLC)
Product CodeLIT · Cardiovascular
Decision DateNov 18, 2020
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1250
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Pacific Plus PTA Catheter is intended to dilate stenoses in the illiofemoral, popliteal, infrapopliteal, and renal arteries, and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae. This device is also indicated for stent postdilatation in the peripheral arteries.

Device Story

Over-the-wire (OTW) peripheral balloon catheter; used for percutaneous transluminal angioplasty (PTA). Coaxial dual-lumen design; one lumen for 0.018-inch guidewire; one lumen for balloon inflation using contrast/saline mixture. Physician-operated in clinical setting. Balloon inflated to dilate stenosed vessel segments or post-dilate stents. Distal section features hydrophilic coating. Benefits include vessel patency restoration and improved blood flow in peripheral arteries and dialysis fistulae.

Clinical Evidence

Bench testing only. Performance testing included dimensional verification, balloon preparation/deployment/retraction, rated burst pressure, fatigue, compliance, inflation/deflation time, bond strength, tip pull, flexibility, kink test, torque strength, coating integrity, and particulate evaluation. Biocompatibility testing conducted per ISO 10993-1 for externally communicating, limited-exposure (<24h) blood-contacting devices, including cytotoxicity, pyrogenicity, systemic toxicity, sensitization, irritation, and haemocompatibility (hemolysis, complement activation, PTT, thromboresistance).

Technological Characteristics

Coaxial dual-lumen OTW catheter; balloon diameters 4.0-7.0 mm; lengths 20-300 mm; catheter lengths 90, 150, 200 cm. Features lubricious hydrophilic coating. Biocompatibility per ISO 10993-1. Sterilization method not specified but confirmed same as predicate.

Indications for Use

Indicated for patients requiring dilation of stenoses in iliac, femoral, iliofemoral, popliteal, infrapopliteal, and renal arteries, and treatment of obstructive lesions in native or synthetic arteriovenous dialysis fistulae. Also indicated for stent postdilatation in peripheral arteries.

Regulatory Classification

Identification

A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. November 18, 2020 Medtronic Vascular, Inc (formerly d.b.a ev3 Inc., Covidien Ilc) Rupali Gupta Principal Regulatory Specialist 3033 Campus Drive Maple Grove, Minnesota 55441 Re: K202800 Trade/Device Name: Pacific Plus PTA Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: LIT Dated: September 21, 2020 Received: September 23, 2020 #### Dear Rupali Gupta: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. Gregory O'Connell Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K202800 Device Name Pacific Plus PTA Catheter #### Indications for Use (Describe) The Pacific Plus PTA Catheter is intended to dilate stenoses in the illiofemoral, popliteal, infrapopliteal, and renal arteries, and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae. This device is also indicated for stent postdilatation in the peripheral arteries. | Type of Use (Select one or both, as applicable) | |------------------------------------------------------------------------------------------------------------------| | <span> <span style="text-decoration: overline;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> | | <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the word "Medtronic" in a bold, blue font. The letters are all capitalized except for the "d", "t", "r", "o", "n", and "i". The word is likely a logo for the company Medtronic, which is a medical technology company. Pacific Plus PTA Catheter 510(k) K202800 - Summary # 510(k) Summary ## Pacific™ Plus PTA Catheter This 510(k) Summary is being submitted in accordance with the requirements of 21 C.F.R § 807.92. # 1. Submitter Information | Applicant | Medtronic, Inc<br>710 Medtronic Parkway<br>Minneapolis, MN 55432<br>Tel: 763-514-4000 | |----------------|---------------------------------------------------------------------------------------| | Contact Person | Rupali Gupta<br>Principal Regulatory Affairs Specialist<br>Tel: 763-398-7000 | | Date Prepared | September 21, 2020 | ### 2. Subject Device | Device Trade Name | Pacific Plus PTA Catheter | |------------------------|-------------------------------------------------| | Device Common Name | Catheter, Angioplasty, Peripheral, Transluminal | | Classification Name | Percutaneous Catheter | | Product Code | LIT | | Classification | Class II | | Classification Panel & | Cardiovascular | | Regulation Number | 21 CFR§870.1250 | ## 3. Predicate Device | Device Trade Name | Pacific Plus PTA Catheter (Primary Predicate) | |-----------------------|-----------------------------------------------| | 510(k) Number | K123358 | | 510(k) Clearance Date | January 29, 2013 | ## 4. Reference Devices | Device Trade Name | PowerCross .018" OTW PTA Dilatation Catheter | |-----------------------|------------------------------------------------| | 510(k) Number | K093286 | | 510(k) Clearance Date | November 13, 2009 | | | | | Device Trade Name | Pacific Xtreme PTA Balloon Dilatation Catheter | | 510(k) Number | K103464 | {4}------------------------------------------------ #### acific Plus PTA Catheter 510(k) K202800 - Summary | 510(k) Clearance Date | December 22, 2010 | |-----------------------|------------------------------------------------| | | | | Device Trade Name | Ultraverse 018 PTA Balloon Dilatation Catheter | | 510(k) Number | K192318 | | 510(k) Clearance Date | October 3, 2019 | #### 5. Device Description The Pacific Plus PTA catheter is an Over the Wire (OTW) peripheral balloon catheter, specifically designed for percutaneous transluminal angioplasty (PTA) in stenosed vessel segments. The OTW catheter is used to guide the balloon to the stenosed vessel segment. The balloon is then inflated to dilate the vessel. The catheter is a coaxial dual lumen device. The lumen marked "WIRE" is the central lumen of the catheter, which terminates at the distal tip. This lumen is used to pass the catheter over a guidewire with a maximum outer diameter of 0.018 in (0.46 mm). The lumen marked "BALLOON" is the balloon inflation lumen, which is used to inflate and deflate the dilatation balloon with a mixture of contrast medium and saline solution. The Pacific Plus PTA catheter is available in balloon diameters 4.0 mm - 7.0 mm, and balloon lengths from 20 mm to 300 mm, with 90 cm, 150 cm, and 200 cm catheter lengths. The distal section of the catheter includes a lubricious hydrophilic coating. #### 6. Indications for Use The Pacific Plus PTA catheter is intended to dilate stenoses in the iliac, femoral, iliofemoral, popliteal, infrapopliteal, and renal arteries, and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae. The device is also indicated for stent postdilatation in the peripheral arteries. #### 7. Comparison of Technological Characteristics with Predicate Device The Pacific Plus PTA Catheter uses similar design and materials as the predicate device. Modifications have been made to the Pacific Plus PTA Catheter design and materials to accommodate expanded product size offerings. The overall design and fundamental scientific technology (operating principle or mechanism of action) of the subject device are the same as the predicate device. All devices are used by the physician in a similar manner typical of a PTA balloon catheter. The proposed and predicate devices share the following technological characteristics: - Same intended use - Similar indications for use ● - Same fundamental scientific technology {5}------------------------------------------------ ## Pacific Plus PTA Catheter 510(k) K202800 - Summary - Same principles of operation - Similar balloon diameters - Similar balloon lengths ● - Similar rated burst pressure ● - Same lubricious coating - Similar packaging materials ● - Same sterility assurance level and method of sterilization . #### 8. Performance Data To demonstrate substantial equivalence of the Pacific Plus PTA Catheter to the predicate and reference devices, bench testing and biocompatibility evaluation was performed. Results from these testing provide assurance that the proposed device has been designed and tested to assure conformance to the requirements for its intended use. #### Performance Bench Testing Using internal risk analysis procedures, the following performance tests were performed: - Dimensional Verification ● - Balloon Preparation, Deployment and Retraction - Balloon Rated Burst Pressure ● - Balloon Fatigue - Balloon Compliance - Balloon Inflation/Deflation Time - Catheter Bond Strength and tip pull test - Catheter Flexibility and Kink Test ● - Catheter Torque Strength - Catheter coating Integrity - Particulate Evaluation . The results from these tests demonstrate that the technological characteristics and performance criteria of the Pacific Plus PTA Catheter devices are comparable to the predicate and reference devices and that the Pacific Plus PTA Catheter performs in a manner equivalent to the predicate and reference devices currently on the market with the same intended use. #### Biocompatibility Testing The biocompatibility evaluation for the Pacific Plus PTA Catheter was conducted in accordance with the principles of the ISO 10993-1:2018, Biological evaluation of medical devices -- Part 1: Evaluation and testing within a risk management process and FDA guidance: Class II Special Controls Guidance Document for Certain Percutaneous Transluminal Coronary Angioplasty (PTCA) Catheters. Sept 2010 and CDRH's 2016 Biocompatibility Guidance, Use of International Standard ISO 10993-1, Biological {6}------------------------------------------------ ## Pacific Plus PTA Catheter 510(k) K202800 - Summary evaluation of medical devices-Part 1: Evaluation and testing within a risk management process as recognized by FDA. The biocompatibility testing included the following tests: - Cytotoxicity . - Pyrogenicity ● - Systemic Toxicity ● - Sensitization ● - Intracutaneous Reactivity (irritation) ● - Haemocompatibility -Hemolysis ● - Haemocompatibility -Complement Activation - Haemocompatibility -Partial Thromboplastin Time (PTT) ● - Haemocompatibility Thromboresistance . The Pacific Plus PTA Catheter is considered biocompatible for its intended use under ISO 10993-1 category: externally communicating device, circulating blood contact with limited (<24 hour) exposure. #### 9. Conclusions Based on the same intended use, similar indications for use, similar technological characteristics, and safety and performance data, Medtronic Vascular, Inc considers the Pacific Plus PTA Balloon Catheter substantially equivalent to the legally marketed predicate device (Pacific Plus, K123358) and reference devices (PowerCross, K093286, Pacific Xtreme, K103464 and Ultraverse, K192318).
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...