The NanoCross™ Elite 0.014" Over-the-Wire PTA Balloon Dilatation Catheter is intended to dilate stenoses in the iliac, femoral, ilio-femoral, popliteal, infra-popliteal, and renal arteries, and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae. This device is also indicated for stent post-dilatation in the peripheral vasculature.
Device Story
Over-the-wire (OTW) coaxial percutaneous transluminal angioplasty (PTA) balloon catheter; compatible with 0.014" guidewires. Features distally mounted semi-compliant inflatable balloon with two radiopaque markers for positioning; atraumatic tapered tip; lubricious coating. Operated by physicians in clinical settings to treat peripheral artery stenoses and arteriovenous dialysis fistulae. Inflation/deflation performed via balloon lumen using contrast/saline mixture. Device tracks over guidewire to lesion site; balloon inflation dilates vessel or stent. Benefits include restoration of blood flow in peripheral vasculature.
Clinical Evidence
Bench testing only. Testing included dimensional verification (crossing profile, balloon OD/length, tip ID), balloon performance (rated burst pressure, compliance, fatigue), mechanical integrity (pull-back force, pushability, torque strength, kink, bond strength), and biocompatibility per ISO 10993-1. Results met all acceptance criteria and were comparable to predicate devices.
Technological Characteristics
OTW coaxial catheter; 0.014" guidewire compatibility. Semi-compliant balloon; radiopaque markers. Lubricious coating. Balloon sizes: 1.5 mm to 6 mm diameter; 20 mm to 210 mm length. Biocompatibility per ISO 10993-1. Sterilization method: standard industry process.
Indications for Use
Indicated for patients requiring dilation of stenoses in iliac, femoral, ilio-femoral, popliteal, infra-popliteal, and renal arteries, treatment of obstructive lesions in native or synthetic arteriovenous dialysis fistulae, and stent post-dilatation in peripheral vasculature.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
{1}------------------------------------------------
K141118 p. 1 of 3
Image /page/0/Picture/1 description: The image shows the logo for "ev3", a company that specializes in endovascular products. The logo features the letters "ev3" in a stylized font, with three circles arranged in a triangular pattern above the letters. To the right of the logo, the text "Your endovascular company." is written in a smaller font.
JUL 1 8 2014
# 510(k) Summary
# NanoCross™ Elite 0.014" Over-the-Wire PTA Balloon Dilatation Catheter
| 510(k) Summary | This 510(k) summary information is submitted in accordance<br>with the requirements of 21 CFR §807.92. | | |
|----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|
| Applicant | ev3 Inc. | | |
| Submitter | ev3 Inc.<br>3033 Campus Drive<br>Plymouth, MN 55441-2651<br>Tel: 763.398.7000<br>Fax: 763.591.3248 | | |
| Contact Person | Laura J. Lind | | |
| Date Prepared | June 18, 2014 | | |
| Device Trade Name | NanoCross™ Elite<br>0.014" Over-the-Wire<br>PTA Balloon Dilatation Catheter | | |
| Device Common Name | PTA Dilatation Catheter | | |
| Classification Name | Catheter, Angioplasty, Peripheral, Transluminal<br>(21 CFR §870.1250, Product Code LIT) | | |
| Classification Panel | Cardiovascular | | |
| Predicate Devices | NanoCross™ Elite 0.014" Over-the-Wire PTA Balloon<br>Dilatation Catheter (K132777), PowerCross™ 0.018" OTW PTA<br>Dilatation Catheter (K093286). | | |
| Intended use | The NanoCross™ Elite 0.014" Over-the-Wire PTA Balloon<br>Dilatation Catheter is intended to dilate stenoses in the iliac,<br>femoral, ilio-femoral, popliteal, infra-popliteal, and renal<br>arteries, and for the treatment of obstructive lesions of native or<br>synthetic arteriovenous dialysis fistulae. This device is also<br>indicated for stent post-dilatation in the peripheral vasculature. | | |
| Device Description | The NanoCross catheter is an over-the-wire (OTW) coaxial<br>lumen percutaneous transluminal angioplasty (PTA) balloon<br>catheter compatible with 0.014"guidewires with a distally<br>mounted semi-compliant inflatable balloon and an atraumatic<br>tapered tip. The distal portion of the catheter has a lubricious<br>coating. The manifold includes a lumen marked "THRU". This<br>is the central lumen of the catheter, which terminates at the distal | | |
ev3 Corporate World Meadquarters / Peripheral Vascular - 3033 Carryus Drive, Plynouth, MN 55411, USA - Ph + 1 763 398 7000
ev3 Neurovascular - 9775 Toledo Way, Irvine, CA 92618 - Ph +1 949 837 3700
ev3 Europe / International Headquarters - 106/108 r.je La Boétie, 75008 Paris, France · Ph +33 156 88 59 10
{2}------------------------------------------------
p. 2 of 3
## NanoCross™ Elite 0.014" Over-the-Wire PTA Balloon Dilatation Catheter Summary
tip. This lumen is used to pass the catheter over a guidewire with a maximum diameter of 0.014". The lumen marked "BALLOON" is the balloon inflation lumen, which is used to inflate and deflate the dilatation balloon with a mixture of contrast medium and saline solution. The balloon has two radiopaque markers for positioning the balloon relative to the stenosis. The radiopaque marker bands indicate the dilating or working section of the balloon.
The NanoCross Elite catheter is available in balloon sizes ranging from 1.5 mm to 6 mm in diameter, and from 20 mm to 210 mm in length; reference labeling for introducer sheath compatibility.
## Performance data
Bench testing was performed to support a determination of substantial equivalence. Results from this testing provide assurance that the proposed device has been designed and tested to assure conformance to the requirements for its intended use.
| • Dimensional Verification:<br>Crossing Profile | • Dimensional Verification:<br>Balloon OD |
|-------------------------------------------------|---------------------------------------------------------|
| • Balloon Rated Burst Pressure | • Balloon Rated Burst Pressure (In Stent) |
| • Balloon Compliance | • Radiopacity |
| • Balloon Pull-back Force | • Presence of Coating |
| • Dimensional Verification:<br>Balloon Length | • Dimensional Verification:<br>Tip/Lesion Entry Profile |
| • Inflation/Deflation Time | • Pushability |
| • Balloon Fatigue | • Dimensional Verification: Tip ID |
| • Catheter Bond Strength | • Wire Movement |
| • Kink | • Re-Insertion Force |
| • Device Tracking | • Catheter Working Length |
| • Insertion Force | • Torque Strength |
| • Particle Generation | • Coating Durability |
Using internal Risk Analysis procedures, the following tests were performed for one or more model sizes:
Using the same Risk Analysis procedures, the following tests were leveraged from predicate devices for one or more model sizes:
Biocompatibility per ISO 10993-1
The NanoCross Elite catheter met all acceptance criteria for the bench testing with results similar to the predicates. Based on the bench test results, comparison to legally marketed predicates, and non-clinical test results, the NanoCross Elite catheter is determined to perform as safely and effectively as the predicates for its intended use.
{3}------------------------------------------------
r
# NanoCross™ Elite 0.014" Over-the-Wire PTA Balloon Dilatation Catheter Summary
| Summary of<br>Substantial<br>Equivalence | The NanoCross Elite catheter has the following similarities to<br>one or more of the predicate devices:<br>Same fundamental scientific technology Identical intended use Same operating principle Identical balloon rated burst pressures Identical balloon nominal pressure Similar balloon diameters Similar balloon lengths Identical catheter lengths A lubricious coating Packaged with the same materials and processes Same sterility assurance level and sterilization method |
|------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | The devices are compatible with 0.014" guidewires. All devices<br>have similar construction and principles of operation. All<br>devices are used by the physician in a similar manner typical of<br>PTA balloon catheters.<br>The NanoCross Elite catheter and the predicates have the same<br>intended use - all devices are intended to treat peripheral arteries.<br>All devices are intended to treat the same target population. The<br>manner in accessing and treating lesions is the same. |
| Conclusion | Based on the intended use, technological characteristics, and<br>results from safety and performance testing, the modified<br>NanoCrossTM Elite 0.014" Over-the-Wire PTA Balloon<br>Dilatation Catheter is considered substantially equivalent to the<br>legally marketed predicate devices NanoCrossTM Elite 0.014"<br>Over-the-Wire PTA Balloon Dilatation Catheter (K132777) and<br>PowerCrossTM .018" OTW PTA Dilatation Catheter (K093286). |
.
.
:
.
·
.
ーPage 3 of 3
.
,
{4}------------------------------------------------
Image /page/3/Picture/1 description: The image shows a circular logo for the U.S. Department of Health & Human Services. The logo features the department's emblem, which consists of a stylized human figure embracing a globe. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the emblem in a circular fashion. The image is black and white.
Food and Drug Administration 0903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 18, 2014
ev3 Inc. Ms. Brenda Johnson Regulatory Affairs Manager 3033 Campus Drive Suite N550 Plymouth, MN 55441
Re: K141118
> Trade/Device Name: NanoCross Elite 0.014" Over-the-Wire PTA Balloon Dilatation Catheter Regulation Number: 21 CFR 871.1250 Regulation Name: Catheter, Angioplasty, Peripheral, Transluminal Regulatory Class: Class II Product Code: LIT Dated: June 18, 2014 Received: June 19, 2014
Dear Ms. Johnson,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{5}------------------------------------------------
Page 2 - Ms. Brenda Johnson
the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours
**Kenneth J. Cavanaugh -S**
for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
Form Approved: OMB No. 0910-0120
Expiration Date: January 31, 2017 See PRA Statement below.
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known) ה/a
Device Name
NanoCross™ Elite 0.014" Over-the-Wire PTA Balloon Dilatation Catheter
#### Indications for Use (Describe)
The NanoCross™ Elite 0.014" Over-the-Wire PTA Balloon Dilatation Catheter is intended to dilate stenoses in the illias, femoral, ilio-femoral, popliteal, and renal arteries, and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae. This device is also indicated for stent post-dilatation in the peripheral vasculature.
| CONSULERS CONNUIL I AN AN AND LINE LINE LE LE SENSIO LEBE | Income of experience of the property of the property of the program and |
|-----------------------------------------------------------|-------------------------------------------------------------------------|
| Type of Use (Select one or both, as applicable) | |
2 Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
# PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY J. 3"
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and mainlain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.