PROTEUS PTA BALLOON CATHETER WITH EMBOLIC CAPTURE FEATURE
Applicant
Angioslide, Ltd.
Product Code
LIT · Cardiovascular
Decision Date
Jun 8, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1250
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Angioslide PROTEUS™ Percutaneous Transluminal Angioplasty (PTA) Balloon Catheter with Embolic Capture Feature is indicated for peripheral transluminal angioplasty and for capture and containment of embolic material during angioplasty, for the femoral, iliac, ilio-femoral, popliteal, tibial, peroneal, and profunda arteries. The Angioslide PROTEUS™ PTA Balloon Catheter with Embolic Capture Feature is not intended for use in the renal, cerebral, coronary or carotid vasculature.
Device Story
Over-the-wire, dual-lumen PTA balloon catheter; features foldable balloon for lesion dilation and embolic capture. Operated by physician in clinical setting. Embolic capture mechanism: proximal vessel occlusion via balloon inflation; inward folding of balloon creates negative pressure in capture cavity to contain/remove debris. Input: guidewire (max 0.035"). Output: mechanical dilation and debris removal. Benefits: reduces embolic risk during angioplasty. Device includes handle, pulling knob, and inflation port. No electronic or software components.
Clinical Evidence
No clinical data. Substantial equivalence supported by extensive in vitro bench testing, including dimensional inspection, burst/fatigue testing, tensile strength, simulated use in tortuous anatomy, and flow characteristics. All tests met pre-defined acceptance criteria.
Technological Characteristics
Over-the-wire, co-axial dual-lumen catheter. Materials: standard medical-grade catheter materials. Energy source: manual hydraulic (inflation). Dimensions: compatible with 0.035" guidewire. Sterilization: Ethylene Oxide (EtO) per ISO 11135. No software or electronic components.
Indications for Use
Indicated for peripheral transluminal angioplasty and embolic material capture/containment in femoral, iliac, ilio-femoral, popliteal, tibial, peroneal, and profunda arteries. Contraindicated for renal, cerebral, coronary, or carotid vasculature.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
Predicate Devices
Angioslide PROTEUS™ PTA Balloon Catheter with Embolic Capture Feature (K111750)
Related Devices
K172494 — PROTEUS PTA Balloon Catheter with Embolic Capture Feature · Angioslide, Ltd. · Dec 15, 2017
K090364 — EXTRA PTA BALLOON CATHETER WITH EMBOLIC PROTECTION FEATURE · Angioslide, Ltd. · Mar 23, 2010
K133043 — PROTEUS PTA BALLOON CATHETER WITH EMBOLIC CAPTURE FEATURE · Angioslide, Ltd. · Dec 6, 2013
K111750 — EXTRA PTA BALLOON CATHETER WITH EMBOLIC CAPTURE FEATURE; PROTEUS PTA BALLOON CATHETER WITH EMBOLIC CAPTURE FEATURE · Angioslide, Ltd. · Jul 21, 2011
Submission Summary (Full Text)
{0}------------------------------------------------
K120164
JUN - 8 2012
# 510(k) Summary
# PROTEUS™ PTA Catheter with Embolic Capture Feature
#### Introduction
This document contains the 510(k) summary for the modified PROTEUS™ PTA Catheter with Embolic Capture Feature. The content of this summary is based on the requirements of 21 CFR Section 807.92(c).
#### Applicant Name and Address:
| Name: | Angioslide, Ltd. |
|-------------------|----------------------------------------------------------------------------------|
| Address: | 7 Halamish St.<br>Business and Industry Park - North<br>Caesarea 30889<br>Israel |
| | Tel: 011-972-4-6187200<br>Fax: 011-972-4-6187222 |
| Official Contact: | Clay Anselmo<br>CEO<br>Reglera LLC |
| Address: | 11925 W. I-70 Frontage Rd. North<br>Suite 900<br>Wheat Ridge, CO 80033 |
| Phone: | 303-223-4303 |
| Fax: | 303-832-6700 |
| E-mail: | anselmoc@reglera.com |
# Summary Preparation Date: 1/18/2012
## Device Name and Classification:
| Trade Name: | PROTEUS™ PTA Balloon Catheter with<br>Embolic Capture Feature |
|----------------------|---------------------------------------------------------------|
| Common Name: | Percutaneous Transluminal Angioplasty<br>Balloon Catheter |
| Classification Name: | Catheter, Percutaneous |
| Classification: | Class II, 21 CFR 870.1250 |
| Product Code | DQY/LIT |
{1}------------------------------------------------
# Predicate Devices:
The modified PROTEUS™ Percutaneous Transluminal Angioplasty (PTA) Balloon Catheter with Embolic Capture Feature is claimed to be substantially equivalent to the following legally marketed predicate device:
- Angioslide PROTEUSTM PTA Balloon Catheter with Embolic Capture Feature . (K111750)
Performance Standards: There are no mandatory performance standards for this device.
Device Description (see Figure 1)
Image /page/1/Figure/6 description: This image shows a diagram of a medical device with labeled parts. The device includes a handle (3), a needle (4, 5), and a connector (2). Detail A shows a close-up of the needle tip (9), while Detail B shows the needle base (6). Other labeled components include 1, 7, 8, 10, 11, and 12.
Figure 1
The Angioslide PROTEUSTM Percutaneous Transluminal Angioplasty (PTA) Balloon Catheter with Embolic Capture Feature is an over the wire dual lumen catheter with a foldable balloon (5) located near the distal atraumatic soft tip (9).
One lumen is used for inflation of the balloon and is accessed via the inflation port (2). The other lumen, starting at the guidewire port (10), allows access to the distal tip for guidewire insertion (max. 0.035"). The balloon has two radiopaque markers (8) for positioning the balloon relative to stenosis. The radiopaque markers indicate the dilating section of the balloon and help in balloon placement. The balloon is designed to provide an inflatable segment of known diameter and length at specified pressure.
{2}------------------------------------------------
The shaft (4) comprises the outer shaft (6) and the inner shaft (7). The distal end of the balloon (A) is connected to the inner shaft and the proximal end of the balloon (B) is connected to the outer shaft. The inner shaft is connected to the proximal hub (10) which is connected to the pulling knob (1) and the outer shaft is connected to the handle grip (3). The pulling knob lock (11) locks the handle grip and the pulling knob together. The distal end of the balloon is folded inwards towards the proximal end of the balloon, by pressing on pulling knob lock ( 1 l ) and pulling the pulling knob away from the handle (1). The embolic capture feature involves a single-use suction mechanism that works through inward folding of the balloon, which creates negative pressure within the capture cavity for debris capture and removal. The reduced pressure in the capture cavity causes some of the particles that are released during the procedure to flow into the cavity for containment and removal.
The balloon size and diameter are printed on the strain relief (12). Refer also to the package label for information about catheter length, balloon nominal and rated burst pressure, balloon size, balloon compliance, guidewire compatibility and sheath compatibility.
#### Indications for Use:
The PROTEUS™ Percutaneous Transluminal Angioplasty (PTA) Balloon Catheter with Embolic Capture Feature is indicated for peripheral transluminal angioplasty and for capture and containment of embolic material during angioplasty, for the femoral, iliac, illo-femoral, poplited, tibial, peroneal, and profunda arteries.
The PROTEUS™ PTA Balloon Catheter with Embolic Capture Feature is not intended for use in the renal, cerebral, coronary or carotid vasculature.
#### Comparison of Technological Characteristics
The PROTEUS™ Percutaneous Transluminal Angioplasty (PTA) Balloon Catheter is an over the wire co-axial dual lumen catheter with a foldable balloon located near the distal atraumatic soft tip. The catheter is compatible with a 0.035" guidewire.
The balloon catheter technological characteristics of the PROTEUS™ PTA Balloon Catheter are identical to those of the PROTEUS™ PTA Balloon Catheter (K111750). In both devices, lesion dilation is achieved by means of an inflatable balloon.
The PROTEUS™ PTA Balloon Catheter overall length, catheter sheath sizing, balloon diameter, balloon length, balloon nominal pressure, balloon rated burst pressure and end hole diameter are the same as those of the PROTEUS™ PTA Balloon Catheter (K111750).
The embolic capture technological characteristics of the PROTEUS™ PTA Balloon Catheter are identical to those of the PROTEUS™ PTA Balloon Catheter (K111750). In both devices, the containment and removal of embolic material is achieved by proximal vessel occlusion, by means of an inflatable balloon, and subsequent capture and removal of embolic material.
### Summary of Non-Clinical Testing
{3}------------------------------------------------
In vitro bench testing of the Angioslide PROTEUS™ PTA Balloon Catheter was conducted in accordance with Angioslide's Risk Analysis and all applicable FDA Guidance documents and ISO standards, including:
ISO 10555-1 - Sterile, Single Use Intravascular Catheters- Part 1: General Requirements
ISO 10555-4 - Sterile, Single Use Intravascular Catheters- Part 4: Balloon Dilatation Catheters
FDA Guidance - Non-Clinical Tests and Recommended Labeling for Intravascular Stents and Associated Delivery Systems , January 13, 2005
FDA Guidance - Coronary and Carotid Embolic Protection Devices - Premarket Notification [510(k)] Submissions, February 15, 2008
All bench testing, unless otherwise specified, was conducted using finished devices which were sterilized by the final validated sterilization process.
#### Design Verification and Validation:
Design verification and validation testing of the PROTEUSTM PTA Balloon Catheter was performed using a "four corners" approach (i.e. 2 x 2 factorial of the largest and smallest balloon diameters and lengths. In addition to these four sizes, a middle catheter size was tested to provide supplemental data to support the robustness of the overall design of the Angioslide PROTEUSTM PTA Balloon Catheter. Sample sizes used for Design Verification and Validation testing were based on required confidence/reliability levels as a result of risk analysis performed for the PROTEUS™ PTA Balloon Catheter, or per recommendations within the FDA Guidance "Non-Clinical Tests and Recommended Labeling for Intravascular Stents and Associated Delivery Systems." Specifically, the number of samples utilized for each test depended on whether the data to be collected was variable data or attribute data in nature.
{4}------------------------------------------------
| Test Description | Accept/Reject Criteria | Results |
|---------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
| Visual Inspection -<br>External Surface | External surface of catheter effective<br>length, including the distal end, is free<br>from extraneous matter and surface<br>defects<br>90% Confidence, 90% Reliability | PASS |
| Visual Inspection -<br>Distal Tip | Distal tip is smooth, rounded, tapered, or<br>similarly finished.<br>90% Confidence, 90% Reliability | PASS |
| Dimensional<br>Inspection - Distal<br>Bond O.D. | In Tolerance<br>90% Confidence, 90% Reliability | PASS |
| Dimensional<br>Inspection - Soft<br>Tip Length | In Tolerance<br>90% Confidence, 90% Reliability | PASS |
| Dimensional<br>Inspection -<br>Guidewire Inner<br>Lumen | In Tolerance<br>90% Confidence, 90% Reliability | PASS |
| Dimensional<br>Inspection -<br>Wrapped Balloon<br>O.D. | The balloon must pass aperture <<br>2.00mm at 37deg C<br>90% Confidence, 90% Reliability | PASS |
| Dimensional<br>Inspection -<br>Catheter Overall<br>Length | In tolerance<br>90% Confidence, 90% Reliability | PASS |
| Dimensional<br>Inspection -<br>Catheter Overall<br>Effective Length | In tolerance<br>90% Confidence, 90% Reliability | PASS |
| Dimensional<br>Inspection - Knob<br>Height | In tolerance<br>90% Confidence, 90% Reliability | PASS |
| Handle Assembly<br>Burst | Burst Pressure ≥ 12 atm<br>90% Confidence, 90% Reliability | PASS |
| Handle Assembly<br>Fatigue | ≥ 10 Inflation/Deflation Cycles to RBP<br>95% Confidence, 90% Reliability | PASS |
| Test Description | Accept/Reject Criteria | Results |
| Handle Assembly<br>Leakage | No Leakage | PASS |
| Balloon Inflation<br>and Deflation Time | 90% Confidence, 90% Reliability<br>No leakage during inflation<br>Inflation Time: ≤ 14.0 sec,<br>Deflation Time: ≤ 30.6 sec | PASS |
| Tensile Strength -<br>Distal Balloon to<br>Inner Shaft (Peel) | 90% Confidence, 90% Reliability<br>Force at Break ≥ 10N | PASS |
| Tensile Strength -<br>Distal Balloon to<br>Inner Shaft (Shear) | 90% Confidence, 95% Reliability<br>Force at Break ≥ 10N | PASS |
| Tensile Strength -<br>Outer Shaft to T<br>Connector: | 90% Confidence, 95% Reliability<br>Force at Break ≥ 10N | PASS |
| Tensile Strength -<br>Inflation Tube to T<br>Connector: | 90% Confidence, 90% Reliability<br>Force at Break ≥ 15N | PASS |
| Tensile Strength –<br>Inflation Tube to<br>Inflation Luer | 90% Confidence, 90% Reliability<br>Force at Break ≥ 15N | PASS |
| Tensile Strength -<br>Cylinder to T<br>Connector | 90% Confidence, 90% Reliability<br>Force at Break ≥ 15N | PASS |
| Tensile Strength -<br>Inner Shaft to<br>Pulling Rod | 90% Confidence, 90% Reliability<br>Force at Break ≥ 10N | PASS |
| Tensile Strength -<br>Pulling Rod to<br>Proximal Luer | 90% Confidence, 95% Reliability<br>Force at Break ≥ 15N | PASS |
| Tensile Strength -<br>Pulling Rod to<br>Knob Base | 90% Confidence, 90% Reliability<br>Force at Break ≥ 15N | PASS |
| Tensile Strength -<br>Distal Cap to Shells | 90% Confidence, 90% Reliability<br>Force at Break ≥ 15N | PASS |
| Tensile Strength -<br>Proximal Cap to<br>Shells | 90% Confidence, 90% Reliability<br>Force at Break ≥ 15N | PASS |
| Test Description | Accept/Reject Criteria | Results |
| Tensile Strength-<br>Knob Base to Knob<br>Cover | Force at Break ≥ 15N<br>90% Confidence, 90% Reliability | PASS |
| Corrosion<br>Resistance | No signs of corrosion after treatment | PASS |
| Stroke Length | Minimum Stroke Length:<br>≥ 70% of the associated Balloon<br>Working Length<br>Maximum Stroke Length:<br>Balloon can be deflated after reaching<br>stroke limit. | PASS |
| Simulated Use in<br>Tortuous Anatomy<br>Model - guidewire<br>compatibility | 90% Confidence / 90% Reliability<br>Catheter can be mounted over a .035”<br>guidewire | PASS |
| Simulated Use in<br>Tortuous Anatomy<br>Model -<br>Advance/Retract/<br>Deploy/Fold<br>Balloon/Withdraw<br>after Procedure | 90% Confidence, 90% Reliability<br>Completely folded balloon passes<br>through identified Introducer Sheath at<br>the end of procedure. | PASS |
| Simulated Use in<br>Tortuous Anatomy<br>Model - Kink<br>Resistance | 90% Confidence, 90% Reliability<br>No permanent deformations (kinks) are<br>present once removed from the tortuous<br>anatomy model. | PASS |
| Flow<br>Characteristics | 90% Confidence, 90% Reliability<br>Distal flow observed in uninflated and<br>deflated state, occlusion of distal flow in<br>inflated state (straight and bend<br>configurations) | PASS |
| Labeling<br>Validation | 90% Confidence, 90% Reliability<br>Instructions for performing type 1 and<br>type 2 recovery methods can be<br>performed successfully per instructions. | PASS |
| Test Description | Accept/Reject Criteria | Results |
| Packaging<br>Validation - Pouch<br>Label Visual<br>Inspection | Label is securely attached to the pouch<br>surface.<br>Across entire label, details/print is<br>legible and free of smudging or creasing | PASS |
| | 90% Confidence, 90% Reliability | |
| Packaging<br>Validation - Pouch<br>Seal Visual<br>Inspection | Pouch Seals at locations A, B, and C are<br>visibly free of wrinkles, channels,<br>visible bubbles, foreign material,<br>transparent areas, and non-uniform seal<br>width.<br>90% Confidence, 90% Reliability | PASS |
| Packaging<br>Validation -<br>Bubble Leak<br>Testing | No leaks are observed from seals or<br>surface of the pouch.<br>90% Confidence, 90% Reliability | PASS |
| Packaging<br>Validation - Seal<br>Strength Testing | Contract Manufacturer and Supplier<br>Seal: Peel Force $\geq$ 5.0N<br>90% Confidence, 90% Reliability | PASS |
PROTEUS™ PTA Balloon Catheter Design Verification and Validation Summary Table
{5}------------------------------------------------
.
。
{6}------------------------------------------------
{7}------------------------------------------------
:
{8}------------------------------------------------
# Biocompatibility Testing:
Based on Risk Analysis no additional biocompatibility testing was required for this modification.
# Sterilization:
The PROTEUS™ PTA Balloon Catheter is packaged and sterilized using substantially equivalent materials, methods, and sterilization parameters used for most commercially available PTA balloon catheter products. The PROTEUSTM PTA Balloon Catheter is sterilized by ethylene oxide (EtO) sterilization providing a Sterility Assurance Level (SAL) of Ix10°. EtO sterilization validation was completed in accordance with ISO 11135 "Medical devices -Validation and routine control of ethylene oxide sterilization". Testing of EtO residuals was performed in accordance with ISO 10993-7 "Biological Evaluation of Medical Devices - Part 7: Ethylene Oxide Sterilization Residuals." Results of the sterilization validation were found to meet all acceptance criteria and confirmed a SAL of 1x10", and EtO/ECh residuals were found to be within the ISO 10993-7 standard specification.
# Packaging:
Packaging validation testing was conducted on the PROTEUS™ PTA Balloon Catheter packaging which included visual inspection (per ASTM F1886), leak testing (per ASTM F2096/ASTM F1929), and pouch seal strength testing (per ASTM F88). Prior to the start of packaging integrity testing, all samples were pre-conditioned by sterilization and subjected to transportation simulation per ASTM D4332-01 and ASTM D4169 respectively. All results of the packaging validation testing were found to meet acceptance criteria.
# Pyrogenicity:
The PROTEUS™ PTA Balloon catheter is non-pyrogenic based successful completion of pyrogenicity testing:
| Description of the method<br>used to make the<br>determination: | The Limulus amebocyte lysate (LAL) test is used for<br>detection and quantification of bacterial endotoxin on each<br>lot prior to release. |
|------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Identification of the testing<br>endpoint reached and<br>rationale for selecting that<br>endpoint: | Must meet spec of 0.5 EU/ml for general blood contacting<br>medical devices |
| Description of the<br>extraction technique used<br>to obtain the test fluid from<br>the test device: | Flow through method, volume of LAL water used:<br>40ml/device |
| Identification of the<br>reference method used: | FDA Guideline on Validation of the Limulus Amebocyte<br>Lysate (LAL) Test as an End-Product Endotoxin Test,<br>Dec 1987,<br>USP <85> Bacterial Endotoxins Test |
{9}------------------------------------------------
# Shelf Life:
The PROTEUS™ PTA Balloon Catheter is currently validated for a three (3) year shelf life. After sterilization and accelerated aging, device function and performance verification/validation testing was conducted on the PROTEUS™ PTA Balloon Catheter and its packaging system that included the following:
- Device
- Visual Inspection o
- Dimensional Inspection
- Balloon Burst Pressure Testing 0
- Balloon Fatigue Testing O
- Balloon Compliance Testing 0
- Balloon Inflation/Deflation Testing o
- Catheter Leakage Testing O
- Bond Tensile Strength Testing o
- Simulated Use Testing 0
- Stroke Length Testing o
- Flow Characteristics Testing o
- o Corrosion Resistance Testing
#### Packaging .
- O Visual Inspection (per ASTM F1886)
- o Leak Testing (per ASTM F2096/ASTM F1929)
- o Pouch Seal Strength Testing (per ASTM F88)
The results of the testing described above were found to meet all acceptance criteria, and demonstrate that the PROTEUS™ PTA Balloon Catheter and its packaging system meets all functional and performance specifications at the product's three year shelf-life. Accelerated aging results are being confirmed through real time aging studies.
# Substantial Equivalence Conclusion:
Non-clinical verification and validation of the PROTEUS™ PTA Balloon Catheter was performed through extensive bench testing, sterilization, packaging and shelf life testing. Results of the testing demonstrated that the PROTEUS™ PTA Balloon Catheter design met all specifications and is adequate for its intended use. Additionally, the test results demonstrated substantial equivalence of the PROTEUS™ PTA Balloon Catheter to its predicate device.
In conclusion, the PROTEUS™ PTA Balloon Catheter with Embolic Capture Feature is substantially equivalent in intended use, technological characteristics, safety, and performance characteristics to the following legally marketed predicate device:
- PROTEUS™ PTA Balloon Catheter with Embolic Capture Feature, Angioslide -K11750
{10}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/10/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular and features an eagle with its wings spread. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are written around the eagle in a circular fashion. The seal is black and white.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
JUN - 8 2012
Angioslide c/o Clay Anselmo CEO Reglera, LLC 11925 West I-70 Frontage Road North Suite 900 Wheat Ridge, CO 80033
Re: K120164
Trade/Device Name: PROTEUS PTA Balloon Catheter with Embolic Capture Feature Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: LIT, DQY Dated: May 11, 2012 Received: May 14, 2012
Dear Mr. Anselmo:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. However, we remind you that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{11}------------------------------------------------
Page 2 - Mr. Clay Anselmo
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801): medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
M.S. Killebrew
Bram Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
{12}------------------------------------------------
### Indications for Use
510(k) Number (if known): 上(ac) 164
Device Name: PROTEUS™ Percutaneous Transluminal Angioplasty (PTA) Balloon Catheter with Embolic Capture Feature
Indications for Use:
The Angioslide PROTEUS™ Percutaneous Transluminal Angioplasty (PTA) Balloon Catheter with Embolic Capture Feature is indicated for peripheral transluminal angioplasty and for capture and containment of embolic material during angioplasty, for the femoral, iliac, ilio-femoral, popliteal, tibial, peroneal, and profunda arteries.
The Angioslide PROTEUS™ PTA Balloon Catheter with Embolic Capture Feature is not intended for use in the renal, cerebral, coronary or carotid vasculature.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Miz Keller
Page 1 of 1
(Division Sign-Off) Division of Cardiovascular Devices
K120164 510(k) Number_
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.