EXTRA PTA BALLOON CATHETER WITH EMBOLIC PROTECTION FEATURE
Applicant
Angioslide, Ltd.
Product Code
LIT · Cardiovascular
Decision Date
Mar 23, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1250
Device Class
Class 2
Attributes
Therapeutic, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K090364 · Mar 23, 2010
EXTRA PTA BALLOON CATHETER WITH EMBOLIC PROTECTION FEATURE
Angioslide, Ltd.
Historical clinical literature
Historical literature describing PTA performance was used to derive performance goals for the evaluation of the MC-LEADER and MC-LEADER Supplemental clinical studies.
The Angioslide eXtraTM Percutaneous Transluminal Angioplasty (PTA) Balloon Catheter with Embolic Capture Feature is indicated for peripheral translyminal angioplasty and containment of embolic material during angioplasty, for the femoral, iliac, ilio-femoral, popliteal, tibial, peroneal, and profunda arteries. The Angioslide eXtra™ PTA Balloon Catheter with Embolic Capture is not intended for use in the renal, cerebral, coronary or carotid vasculature.
Device Story
Single-use, over-the-wire, co-axial dual-lumen PTA balloon catheter; features foldable balloon near distal tip. Operates via inflation port for balloon expansion and guidewire port for 0.035" guidewire access. Mechanical mechanism: pulling rod retracts to fold balloon inward, creating a cavity for embolic debris collection/containment during angioplasty. Used by physicians in clinical settings for peripheral artery procedures. Provides clinicians with combined angioplasty and embolic protection; reduces risk of distal embolization during vessel dilation. Radiopaque markers facilitate precise positioning relative to stenosis.
Clinical Evidence
Pooled data from two prospective, multi-center, non-randomized, single-arm studies (MC-LEADER and MC-LEADER Supplemental). Evaluated safety and efficacy for lower extremity PTA and embolic capture. Endpoints included acute device/procedural success, adverse event rates, 30-day clinical success (ABI/RB improvement), and 30-day target vessel revascularization. Results met performance goals derived from historical literature.
Technological Characteristics
Over-the-wire co-axial dual-lumen catheter; foldable balloon; 0.035" guidewire compatible. Materials: radiopaque markers for positioning. Mechanical actuation: pulling rod mechanism for inward balloon folding. Sterilization: double-sterilized. Standards: ISO 10555-1, ISO 10555-4, ISO 10993-1, ASTM F756-00.
Indications for Use
Indicated for peripheral transluminal angioplasty and embolic material capture/containment in femoral, iliac, ilio-femoral, popliteal, tibial, peroneal, and profunda arteries. Contraindicated for renal, cerebral, coronary, or carotid vasculature.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
{1}------------------------------------------------
510(k) Summary: Angioslide eXtra™ PTA Balloon Catheter with Embolic Capture Feature (K090364)
# 510(k) Summary: Angioslide eXtra™ PTA Balloon Catheter with Embolic Capture Feature
This summary of substantial equivalence information is being submitted in accordance with the requirements set forth in 21 CFR 807.92.
Submitter: Angioslide Ltd.
#### Establishment Registration Number: Not Yet Assigned
MAR 2 8 2010
#### Contact Information:
Ilya Burovoy
Vice President of Regulatory Affairs and Quality Assurance
Angioslide Ltd.
32 Maskit St., P.O. Box 12489
Herzliya 46733
Israel
Tel: 972-09-9556500
Fax: 972-09-9556700
Email: ilyab@angioslide.com
Date Summary Prepared: March 1, 2010
Name of Device: Angioslide eXtra™ PTA Balloon Catheter with Embolic Capture Feature
Classification Name: Percutaneous Catheter
Device Classification:
Classification: Il
Classification Panel: Cardiovascular
Regulation Number: 21 CFR 870.1250
Product Code: LIT, DQY
#### Predicate Devices:
- 1. AgilTrac . 035 Peripheral Dilatation Catheter (K023320)
- 2. ATB Advance PTA Balloon Catheter (K052036)
- Spider FX Embolic Protection Device (K063785) 3.
{2}------------------------------------------------
# 510(k) Summary: Angioslide eXtra™ PTA Balloon Catheter with Embolic Capture Feature (K090364)
#### Device Description:
The Angioslide eXtra™ Percutaneous Transluminal Angioplasty (PTA) Balloon Catheter is a single-use disposable over-the-wire co-axial dual lumen catheter with a foldable balloon near the distal tip. The balloon catheter consists of the balloon located near the distal atraumatic soft tip, the telescopic shaft and the handle.
One lumen is used for inflation of the balloon and is accessed via the inflation port. The other lumen, starting at the quidewire port. allows access to the distal tip for guidewire insertion (max. 0.035"). The balloon has two radiopague markers for positioning the balloon relative to stenosis. The radiopaque markers indicate the dilating section of the balloon and help in balloon placement. The balloon is designed to provide an inflatable segment of known diameter and length at specified pressure.
The shaft comprises the outer shaft and the inner shaft. The distal end of the balloon is connected to the inner shaft and the proximal end of the balloon is connected to the outer shaft. The inner shaft is connected to the pulling rod and the outer shaft is connected to the handle body. The pulling rod lock locks the handle body and the pulling rod together. The distal end of the balloon is folded inwards towards the proximal end of the balloon, by unlocking the pulling rod lock counter-clockwise and retracting the pulling rod. The inward-folding of the balloon forms a cavity and allows for collection of embolic material.
### Indications for Use:
The Angioslide eXtra™ Percutaneous Transluminal Angioplasty (PTA) Balloon Catheter with Embolic Capture Feature is indicated for peripheral transluminal angioplasty and for capture and containment of embolic material during angioplasty, for the femoral, iliac, ilio-femoral, popliteal, tibial, peroneal, and profunda arteries.
The Angioslide eXtra™ PTA Balloon Catheter is not intended for use in the renal, cerebral, coronary or carotid vasculature.
# Technological Characteristics:
The Angioslide eXtra™ Percutaneous Transluminal Angioplasty (PTA) Balloon Catheter is an over the wire co-axial dual lumen catheter with a foldable balloon located near the distal atraumatic soft tip. The catheter is compatible with 0.035" guidewire.
The technological characteristics of the eXtra"™ PTA Balloon Catheter are substantially equivalent to those of the AgilTrac .035 Peripheral Dilatation Catheter (K023320), the ATB Advance PTA Dilatation Catheter (K052036), and the Spider FX Embolic Protection Device (K063785).
# Summary of Bench Testing:
In vitro bench testing of the Angioslide eXtra™ PTA Balloon Catheter was conducted in accordance with Angioslide's Risk Analysis and all applicable FDA Guidance documents and ISO standards, including:
ISO 10555-1 - Sterile, Single Use Intravascular Catheters- Part 1: General Requirements
ISO 10555-4 - Sterile, Single Use Intravascular Catheters- Part 4: Balloon Dilatation Catheters
FDA Guidance - Non-Clinical Tests and Recommended Labeling for Intravascular Stents and Associated Delivery Systems , January 13, 2005 (http://www.fda.gov/cdrh/ode/guidance/1545.html)
FDA Guidance - Coronary and Carotid Embolic Protection Devices - Premarket Notification [510(k)] Submissions, February 15, 2008 (http://www.fda.gov/cdrh/ode/guidance/1658.html)
{3}------------------------------------------------
## 510(k) Summary: Angioslide eXtra™ PTA Balloon Catheter with Embolic Capture Feature (K090364)
The following bench tests were conducted. All bench testing, unless otherwise specified, was conducted using finished devices which were double-sterilized by the final validated sterilization process.
- Catheter Dimensional Verification .
- . Balloon Burst
- Balloon Compliance .
- Catheter Inflation / Deflation Time .
- . Balloon Fatigue
- . Corrosion
- . Leakage
- Bond Tensile Strength (all bonds) .
- Distal Bond Peel Strength ●
- Marker Radiopacity .
- Simulated Use including; Guidewire Compatibility, Deployment and Retraction, Kink . Resistance, Advancement Force, Retraction Force, Tortuous Anatomy Compatibility
- . Introducer Sheath Compatibility
- . Catheter Flow Characteristics
- . Comparative Capture Efficiency
### Biocompatibility Testing:
The Angioslide eXtra™ Percutaneous Transluminal Angioplasty catheter is an Externally Communicating Device, which contacts circulating blood for the Limited Contact Duration (< 24 hours).
Biocompatibility testing of the Angioslide eXtra™ balloon catheter was performed on finished and sterilized devices in accordance with the ISO-10993-1, "Biological Evaluation of Medical Devices Part-1: Evaluation and Testing," as specified in the FDA Blue Book Memorandum #G95-1.
All testing documented below was conducted in accordance with the provisions of the FDA GLP regulations 21 CFR Part 58. All the tests passed.
- . Cytotoxicity per EN ISO10993-5
- . Sensitization - Murine LLNA per EN ISO10993-10
- Irritation Test -ISO Intracutaneous per EN ISO10993-10 .
- Acute Systemic Toxicity ISO Systemic Toxicity study per EN ISO10993-11 .
- . USP Pyrogen study ~ material mediated per EN ISO10993-11
- . Hemocompatibility - ASTM Hemolysis per ASTM F756-00 and EN ISO10993-4
- Hemocompatibility C3a Complement Activation Assay per EN ISO10993-4 .
- Hemocompatibility -SC5b-9 Complement Activation Assay per EN ISO10993-4 .
- In Vivo Thromboresistance Study in the Dog Femoral Artery .
#### Summary of Animal Testing:
A chronic GLP study was conducted to evaluate the safety and performance of the Angioslide eXtra™ PTA Balloon Catheter compared to an appropriate control I. The objective of the study was to demonstrate safety and performance of the eXtra™ PTA Balloon Catheter when used for peripheral percutaneous transluminal angioplasty (PTA) and post-stent dilatation in an in vivo porcine model. Safety of the test and control articles was assessed by the study pathologist via gross and microscopic evaluation post-procedure, while functional performance of the test and control articles was assessed by
{4}------------------------------------------------
### 510/k) Summary: Angioslide eXtra™ PTA Balloon Catheter with Embolic Capture Feature (K090364)
investigators during PTA and post-stent dilation. The results of this study demonstrated that the safety and performance of the Angioslide eXtra™ PTA Balloon Catheter was substantially equivalent to that of the control article when used in an in vivo model.
#### Summary of Clinical Testing:
Two clinical studies were conducted to demonstrate safety and efficacy of the Angioslide eXtra™ Percutaneous Transluminal Angioplasty (PTA) Balloon Catheter with an Embolic Protection/Capture Feature. The Multi-Center Studies for Lower Extremity Angioplasty with DEbris Removal (MC-LEADER and MC-LEADER Supplemental) were prospective, multi-center (three and two centers respectively, outside of the US), non-randomized, single arm studies. The objective of the studies was to demonstrate safety and efficacy of the Angioslide eXtra™ Percutaneous Transluminal Angioplasty (PTA) Balloon Catheter with an Embolic Protection/Capture Feature when used for lower extremity percutaneous transluminal angioplasty, and embolic debris capture. The studies were conducted under similar protocols for the purpose of pooling the data to support the evaluation of a statistically powered safety hypothesis. The endpoints for each individual study included; acute device success, acute procedural success (<50% residual stenosis), adverse event rates, clinical success at 30 days (ABI and RB improvement), and target vessel revascularization at 12 months (MC-LEADER only). The endpoints evaluated for the pooled clinical data included; target vessel revascularization at 30 days, serious adverse event rate, acute device success, acute procedural success, and acute distal embolization rate. The results of the MC-LEADER and MC-LEADER Supplemental Studies met performance goals derived from historical literature describing PTA performance.
### Substantial Equivalence:
Non-clinical and clinical evaluation demonstrated substantial equivalence of the eXtra PTA Balloon Catheter with its predicates.
{5}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular arrangement of text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA". Inside the circle is a stylized image of an eagle or bird with its wings spread, facing to the right. The bird is drawn with flowing lines, giving it a sense of movement and freedom.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-O66-0609 Silver Spring, MD 20993-0002
MAR 2 3 2010
AngioSlide, LTD c/o Dr. Elisa Harvey Cardiomed Device Consultants 18905 Celebrity Lane Sandy Spring. MD 20860
Re: K090364
eXtra PTA Balloon Catheter with Embolic Capture Feature Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: LIT, DQY Dated: January 29, 2010 Received: January 29, 2010
Dear Dr. Harvey:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. However, we remind you that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{6}------------------------------------------------
Page 2 - Dr. Elisa Harvey
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
Dmma R. V. Aumer
Bram Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
# Indications for Use
510(k) Number (if known): K090364
# Device Name: eXtra™ Percutaneous Transluminal Angioplasty (PTA) Balloon Catheter with Embolic Capture Feature
Indications for Use:
The Angioslide eXtraTM Percutaneous Transluminal Angioplasty (PTA) Balloon Catheter with Embolic Capture Feature is indicated for peripheral translyminal angioplasty and containment of embolic material during angioplasty, for the femoral, iliac, ilio-femoral, popliteal, tibial, peroneal, and profunda arteries.
The Angioslide eXtra™ PTA Balloon Catheter with Embolic Capture is not intended for use in the renal, cerebral, coronary or carotid vasculature.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
una P. Vanes
Page 1 of
(Division Sign-Off) Division of Cardiovascular Devices
510(k) Number K090364
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.