PROTEUS PTA BALLOON CATHETER WITH EMBOLIC CAPTURE FEATURE
Applicant
Angioslide, Ltd.
Product Code
LIT · Cardiovascular
Decision Date
Apr 13, 2012
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.1250
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Angioslide PROTEUS™ Percutaneous Transluminal Angioplasty (PTA) Balloon Catheter with Embolic Capture Feature is indicated for peripheral transluminal angioplasty and for capture and containment of embolic material during angioplasty, for the femoral, iliac, ilio-femoral, popliteal, tibial, peroneal, and profunda arteries. The Angioslide PROTEUS™ PTA Balloon Catheter with Embolic Capture is not intended for use in the renal, cerebral, coronary or carotid vasculature.
Device Story
Over-the-wire dual-lumen PTA balloon catheter; features foldable balloon near distal atraumatic tip. One lumen for balloon inflation; second lumen for guidewire insertion (0.014" or 0.035" depending on size). Radiopaque markers facilitate positioning relative to stenosis. Used by physicians in clinical settings for peripheral artery angioplasty. Embolic capture achieved via proximal vessel occlusion using inflatable balloon, followed by aspiration of embolic material. Device provides inflatable segment of known diameter/length at specified pressure to dilate lesions. Benefits include simultaneous vessel dilation and embolic protection, reducing risk of distal embolization during peripheral interventions.
Clinical Evidence
Bench testing only. Testing included dimensional evaluation, burst strength (RBP ≥ 14atm), balloon compliance, fatigue (≥ 10 cycles), tensile strength, simulated use in tortuous anatomy, and capture efficiency. All tests met pre-defined acceptance criteria with required confidence/reliability levels. No clinical data presented.
Technological Characteristics
Over-the-wire co-axial dual-lumen catheter; foldable balloon; radiopaque markers. Materials: standard medical-grade catheter materials. Energy: pneumatic (inflation). Connectivity: none. Sterilization: Ethylene Oxide (EtO) per ISO 11135-1. Standards: ISO 10555-1, ISO 10555-4, ISO 10993-7.
Indications for Use
Indicated for peripheral transluminal angioplasty and embolic material capture/containment in femoral, iliac, ilio-femoral, popliteal, tibial, peroneal, and profunda arteries. Contraindicated for renal, cerebral, coronary, or carotid vasculature.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
Predicate Devices
Angioslide PROTEUS™ PTA Balloon Catheter with Embolic Capture Feature (K111750)
Boston Scientific Sterling® SL Monorail PTA Balloon, Dilation Catheter (K093720)
{1}------------------------------------------------
# 20805
Angioslide Ltd. - Special 510(k) PROTEUS™ PTA Catheter with Embolic Capture Feature Section 7: 510(k) Summary
APR 1 3 2012
# 510(k) Summary
# PROTEUSTM PTA Catheter with Embolic Capture Feature
#### Introduction:
.
This document contains the 510(k) summary for the modified PROTEUS™ PTA Catheter with Embolic Capture Feature. The content of this summary is based on the requirements of 21 CFR Section 807.92(c).
#### Applicant Name and Address:
Name: Angioslide, Ltd. Address: 7 Halamish St. Business and Industry Park - North Caesarea 30889 Israel Tel: +(972)4-6187200 Fax: +(972)4-6187222 Official Contact: Clay Anselmo CEO Reglera LLC 11925 W. I-70 Frontage Road North Address: Suite 900 Wheat Ridge, CO 80033 Phone: (303) 223-4303 Fax: (303) 832-6700
#### Summary Preparation Date: March 14, 2012
#### Device Name and Classification:
| Trade Name: | PROTEUSTM PTA Balloon Catheter with Embolic Capture Feature |
|----------------------|-------------------------------------------------------------|
| Common Name: | Percutaneous Transluminal Angioplasty Balloon Catheter |
| Classification Name: | Catheter, Percutaneous |
| Classification: | Class II, 21 CFR 870.1250 |
| Product Code: | DQY/LIT |
anselmoc@reglera.com
#### Predicate Devices:
The modified PROTEUS™ Percutaneous Transluminal Angioplasty (PTA) Balloon Catheter with Embolic Capture Feature is claimed to be substantially equivalent to the following legally marketed predicate devices:
{2}------------------------------------------------
- Angioslide PROTEUS™ PTA Balloon Catheter with Embolic Capture Feature . (K111750)
- Boston Scientific Sterling® SL Monorail PTA Balloon,Dilation Catheter (K093720) .
Performance Standards: There are no mandatory performance standards for this device. Device Description: (see Figure 1)
Image /page/1/Figure/4 description: This image shows a technical drawing of a medical device with labeled parts. The drawing includes two detailed views labeled "Detail A" and "Detail B", providing a closer look at specific components. The main view shows the device's overall structure, including a handle, a needle, and a connector. The parts are labeled with numbers 1 through 12 to identify different components of the device.
#### Figure 1
The Angioslide PROTEUS™ Percutaneous Transluminal Angioplasty (PTA) Balloon Catheter with Embolic Capture Feature is an over the wire dual lumen catheter with a foldable balloon (5) located near the distal atraumatic soft tip (9).
One lumen is used for inflation of the balloon and is accessed via the inflation port (2). The other lumen, starting at the guidewire port (10), allows access to the distal tip for guidewire insertion (max. 0.035"). The balloon has two radiopaque markers (8) for positioning the balloon relative to stenosis. The radiopaque markers indicate the dilating section of the balloon and help in balloon placement. The balloon is designed to provide an inflatable segment of known diameter and length at specified pressure.
The Angioslide PROTEUS™ Percutaneous Transluminal Angioplasty (PTA) Balloon Catheter with Embolic Capture Feature is an over the wire dual lumen catheter with a foldable balloon (5) located near the distal atraumatic soft tip (9),
One lumen is used for inflation of the balloon and is accessed via the inflation port (2). The other lumen, starting at the guidewire port (10), allows access to the distal tip for guidewire insertion. The balloon has two radiopaque markers (8) for positioning the balloon relative to stenosis. The radiopaque markers indicate the dilating section of the balloon and help in balloon placement.
{3}------------------------------------------------
The balloon is designed to provide an inflatable segment of known diameter and length at specified pressure.
#### Indications for Use:
The PROTEUS™ Percutaneous Transluminal Angioplasty (PTA) Balloon Catheter with Embolic Capture Feature is indicated for peripheral transluminal angioplasty and for capture and containment of embolic material during angioplasty, for the femoral, iliac, illo-femoral, popliteal, tibial, peroneal, and profunda arteries.
The PROTEUS™ PTA Balloon Catheter with Embolic Capture is not intended for use in the renal, cerebral, coronary or carotid vasculature.
#### Comparison of Modified Indications for Use:
The Indications for Use for the predicate PROTEUS™ Percutaneous Transluminal Angioplasty (PTA) Balloon Catheter with Embolic Capture Feature (K11750) are identical to the modified PROTEUS™ Percutaneous Transluminal Angioplasty (PTA) Balloon Catheter with Embolic Capture Feature, but includes the addition of the 3x100mm balloon size.
#### Comparison of Technological Characteristics:
The PROTEUS™ Percutaneous Transluminal Angioplasty (PTA) Balloon Catheter is an over the wire co-axial dual lumen catheter with a foldable balloon located near the distal atraumatic soft tip. The catheter is compatible with a 0.014" guidewire in the 3x100mm size only, and a 0.035" guidewire for all other sizes, which are currently cleared.
The balloon technological characteristics of the modified PROTEUS™ PTA Balloon Catheter are substantially equivalent to those of the PROTEUS™ PTA Balloon Catheter and the Sterling® SL Monorail PTA Balloon Dilation Catheter. In both devices lession dilation is achieved by means of an inflatable balloon.
The modified PROTEUS™ PTA Balloon Catheter overall length, catheter sheath sizing, balloon diameter, balloon length, balloon nominal pressure, balloon rated burst pressure and end hole diameter are the same or similar to the PROTEUS™ PTA Balloon Catheter and the Sterling® SL Monorail PTA Balloon Dilation Catheter.
The embolic capture technological characteristics of the modified PROTEUS™ PTA Balloon Catheter are identical to those of the PROTEUS™ PTA Balloon Catheter. In both devices the containment and removal of embolic material is achieved by proximal vessel occlusion, by means of an inflatable balloon, and subsequent aspiration of embolic material.
#### Summary of Non-Clinical Testing:
In viro bench testing of the Angioslide PROTEUS™ PTA Balloon Catheter was conducted in accordance with Angioslide's Risk Analysis and all applicable FDA Guidance documents and ISO standards, including:
ISO 10555-1 - Sterile, Single Use Intravascular Catheters- Part 1: General Requirements
ISO 10555-4 - Sterile, Single Use Intravascular Catheters- Part 4: Balloon Dilatation Catheters
{4}------------------------------------------------
FDA Guidance - Non-Clinical Tests and Recommended Labeling for Intravascular Stents and Associated Delivery Systems, January 13, 2005
FDA Guidance - Coronary and Carotid Embolic Protection Devices - Premarket Notification [510(k)] Submissions, February 15, 2008
All bench testing, unless otherwise specified, was conducted using finished devices which were sterilized by the final validated sterilization process. .
#### Design Verification and Validation:
Sample sizes used for the testing were based on required confidence/reliability levels as per FDA Guidance "Non-Clinical Tests and Recommended Labeling for Intravascular Stents and Associated Delivery Systems " and the results of the risk analysis (DFMEA) performed for the PROTEUS™ PTA catheter. The number of units utilized for each test depends on whether the data to be collected was variable data or attribute data, therefore the number of units tested varies from test to test. However, in all cases, the number utilized for testing met the required number of units based on the risk analysis, and the required confidence/reliability levels.
The results of verification and validation testing demonstrated that the modified PROTEUS™ PTA Balloon Catheter with Embolic Capture Feature design met all specifications, and is adequate for its intended use. Additionally, the test results demonstrated substantial equivalence of the modified PROTEUSTM PTA Balloon Catheter with Embolic Capture Feature to its predicate devices.
Verification and validation testing of the modified PROTEUS™ PTA Balloon Catheter with Embolic Capture Feature included catheter and balloon dimensional evaluation, minimum balloon burst strength (RBP), balloon compliance, balloon inflation and deflation time, balloon fatigue, tensile strength, simulated use, and capture efficiency testing.
| Test | Accept/Reject Criteria | Results |
|------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
| Visual inspection - external surface | External surface of catheter effective length,<br>including the distal end, is free from extraneous<br>matter and surface defects (<0.2mm² TAPPI).<br>90% Confidence, 90% Reliability | PASS |
| Visual inspection - distal tip | Distal tip is smooth, rounded, tapered, or<br>similarly finished.<br>90% Confidence, 90% Reliability | PASS |
| Distal Bond Outer Diameter | In Tolerance<br>90% Confidence, 90% Reliability | PASS |
| Distal Balloon Cone to Distal Tip | In Tolerance<br>90% Confidence, 90% Reliability | PASS |
| Distal Tip to Proximal Outer Tube Tip<br>Length | In Tolerance<br>90% Confidence, 90% Reliability | PASS |
| Soft Tip Length | In Tolerance<br>90% Confidence, 90% Reliability | PASS |
| Wrapped Balloon Outer Diameter | In Tolerance<br>90% Confidence, 90% Reliability | PASS |
| Guide Wire Lumen Inner Diameter | In Tolerance<br>90% Confidence, 90% Reliability | PASS |
| Proximal Bond Outer Diameter | In Tolerance<br>90% Confidence, 90% Reliability | PASS |
| Test | Accept/Reject Criteria | Results |
| Marker Band Spacing | In Tolerance<br>90% Confidence, 90% Reliability | PASS |
| Balloon Working Length | In Tolerance<br>90% Confidence, 90% Reliability | PASS |
| Balloon Outer Diameter | In Tolerance<br>90% Confidence, 90% Reliability | PASS |
| Catheter Overall Effective Length | In Tolerance<br>90% Confidence, 90% Reliability | PASS |
| Catheter Overall Length | In Tolerance<br>90% Confidence, 90% Reliability | PASS |
| Minimum Balloon Burst Pressure (RBP) | RBP $\ge$ 14atm<br>95% Confidence, 99.9% Reliability | PASS |
| Balloon Fatigue (Repeated<br>Inflation/Deflations) | Inflation/Deflation Cycles $\ge$ 10 at 14atm (for<br>3x100)<br>No leakage, rupture, and/or herniation<br>Up to max 40 cycles<br>95% Confidence, 90% Reliability | PASS |
| Tensile Strength- Distal Balloon to Inner<br>Tube (Peel) | $\ge$ 5N<br>90% Confidence, 95% Reliability | PASS |
| Tensile Strength- Distal Balloon to Inner<br>Tube (Shear) | $\ge$ 5N<br>90% Confidence, 95% Reliability | PASS |
| Tensile Strength- Proximal Balloon to<br>Outer Tube | $\ge$ 10N<br>90% Confidence, 95% Reliability | PASS |
| Tensile Strength- Cylinder to T-Connector | $\ge$ 15N<br>90% Confidence, 90% Reliability | PASS |
| Tensile Strength- Inflation Tube to T-<br>Connector | $\ge$ 15N<br>90% Confidence, 90% Reliability | PASS |
| Tensile Strength- Outer Tube to T-<br>Connector | $\ge$ 15N<br>90% Confidence, 90% Reliability | PASS |
| Tensile Strength- Inner Shaft to Pulling<br>Rod | $\ge$ 10N<br>90% Confidence, 95% Reliability | PASS |
| Tensile Strength- Pulling Rod to Proximal<br>Luer | $\ge$ 15N<br>90% Confidence, 90% Reliability | PASS |
| Tensile Strength- Pulling Rod to Knob<br>Base | $\ge$ 15N<br>90% Confidence, 90% Reliability | PASS |
| Tensile Strength- Cylinder to O-ring Cap | $\ge$ 15N<br>90% Confidence, 90% Reliability | PASS |
| Liquid Leakage | No leakage<br>90% Confidence, 90% Reliability | PASS |
| Balloon Inflation/Deflation Testing | Inflation Time: $\le$ 14.0s<br>Deflation Time: $\le$ 30.6s<br>90% Confidence, 90% Reliability | PASS |
| Balloon Compliance | Compliance: $\le$ 13%<br>Nominal Pressure: 8atm<br>RBP: 14atm<br>90% Confidence, 90% Reliability | PASS |
| Flow Characteristics- Straight<br>Configuration | Distal flow observed in uninflated and deflated<br>state, occlusion of distal flow in inflated state<br>90% Confidence, 90% Reliability | PASS |
| Flow Characteristics- Extreme Angle<br>Configuration | Distal flow observed in uninflated and deflated<br>state, occlusion of distal flow in inflated state<br>90% Confidence, 90% Reliability | PASS |
| Test | Accept/Reject Criteria | Results |
| Stroke Length | Minimum Stroke Length:<br>≥ 70% of the associated Balloon Working<br>Length<br><br>Maximum Stroke Length:<br>Balloon can be deflated after reaching stroke<br>limit. | PASS |
| Simulated Use in Tortuous Anatomy<br>Model - guide wire compatibility | 90% Confidence / 90% Reliability<br>Catheter can be mounted over a 0.014" guide<br>wire<br>90% Confidence, 90% Reliability | PASS |
| Simulated Use in Tortuous Anatomy<br>Model - Introducer Sheath Compatibility | 90% Confidence, 90% Reliability<br>Completely folded balloon passes through<br>identified Introducer Sheath at the end of<br>procedure.<br>90% Confidence, 90% Reliability | PASS |
| Simulated Use in Tortuous Anatomy<br>Model - Kink Resistance | 90% Confidence, 90% Reliability<br>No permanent deformations (kinks) are present<br>once removed from the tortuous anatomy<br>model.<br>90% Confidence, 90% Reliability | PASS |
| Simulated Use in Tortuous Anatomy<br>Model - Max Advancement Force | N/A - Characterization only | N/A |
| Simulated Use in Tortuous Anatomy<br>Model - Max Collapse Force | N/A - Characterization only | N/A |
| Simulated Use in Tortuous Anatomy<br>Model - Max Anatomy Retraction Force | N/A - Characterization only | N/A |
| Simulated Use in Tortuous Anatomy .<br>Model - Max Removal Retraction Force | N/A - Characterization only | N/A |
| Capture Efficiency Comparison | N/A - Characterization only | N/A |
{5}------------------------------------------------
:
:
.
·
·
:
:
:
:
:
:
{6}------------------------------------------------
## Comparative Capture Efficiency Testing:
To ensure comparability of results, the testing approach was identical to that used to gather baseline data submitted in the original 510(K) for the eXtra™ device family (K090364). The CE testing for the modified PROTEUS™ PTA Balloon Catheter was identical to that of its predicate and is comparable to events that could happen in vivo.
A review of the data in detail revealed no significant anomalies in the raw or analyzed data; and trends in performance based on differences in device size were as expected.
#### Biocompatibility Testing:
Based on Risk Analysis, no additional biocompatibility testing was required for this modification.
#### Sterilization:
The PROTEUS™ PTA Balloon Catheter is packaged and sterilized using substantially equivalent materials, methods, and sterilization parameters used for most commercially available PTA balloon catheter products. The PROTEUS™ PTA Balloon Catheter is sterilized by ethylene oxide (EtO) sterilization providing a Sterility Assurance Level (SAL) of 1x10 °. EtO sterilization validation was completed in accordance with ISO 11135-1:2007 "Medical devices - Validation and routine control of ethylene oxide sterilization". Testing of EtO residuals was performed in
{7}------------------------------------------------
accordance with ISO 10993-7:2008 "Biological Evaluation of Medical Devices - Part 7: Ethylene Oxide Sterilization Residuals." Results of the sterilization were found to meet all acceptance criteria, confirmed a SAL of 1x10 °, and EtO/ECh residuals were found to be within the ISO 10993-7 standard specification. -
#### Packaging:
There have been no changes to the packaging materials or assembly for the modified PROTEUS™ PTA Balloon Catheter from its predicate device, K111750; therefore, no packaging validation was performed.
#### Shelf Life:
The shelf-life of the modified PROTEUS™ PTA Balloon Catheter is identical to the currently validated three (3) year shelf life for its predicate device, PROTEUS™ PTA Balloon Catheter. Therefore, no shelf-life validation or accelerated aging tested were performed.
### Substantial Equivalence Conclusion:
The modified Angioslide PROTEUS™ PTA Balloon Catheter is substantially equivalent with respect to the indications for use, technological characteristics and performance characteristics to the following legally marketed predicate devices:
- PROTEUSTM PTA Balloon Catheter, Angioslide K111750 .
- Sterling® SL Monorail PTA Balloon Catheter, Boston Scientific K093720
{8}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
APR 1 3 2012
Angioslide, LTD. c/o Clay Anselmo Chief Executive Officer Reglera, LLC 11925 West I-70 Frontage Road North, Suite 900 Wheat Ridge, CO 80033
Re: K120805
Trade/Device Name: PROTEUS™ PTA Balloon Catheter with Embolic Capture Feature Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: LIT, DQY Dated: March 14, 2012 Received: March 16, 2012
Dear Mr. Anselmo:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{9}------------------------------------------------
# Page 2 - Mr. Clay Anselmo
Please be advised that FDA's issuance of a substantial equivalence determination does not mean r lease of advised that I Dri 3 150aan.co or device complies with other requirements of the Act that I DA has made a determinations administered by other Federal agencies. You must of ally Federal statutes and regaranents and limited to: registration and listing (21 Comply will an the Her 5 requirements (0); medical device reporting (reporting of medical CI K I at 607), labeling (21 OFR 803); good manufacturing practice requirements as set device in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic (200 form in the quality of ovisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please IT you desire specific advice to: your conters offices/CDRH/CDRHOffices/ucm115809.htm for go to intep.//www.laagonomical Health's (CDRH's) Office of Compliance. Also, please the Center for Devices and Radiological g by reference to premarket notification" (21CFR Part) flore the regulations regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the You may obtain only general miormaneneas and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
M. A. Hillham
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{10}------------------------------------------------
### Indications for Use
510(k) Number (if known):
#### PROTEUSTM Percutaneous Transluminal Angioplasty (PTA) Balloon Catheter with Device Name: Embolic Capture Feature
Indications for Use:
The Angioslide PROTEUS™ Percutaneous Transluminal Angioplasty (PTA) Balloon Catheter with Embolic Capture Feature is indicated for peripheral transluminal angroplasty and for capture and containment of embolic material during angioplasty, for the femoral, iliac, ilio-femoral, popliteal, tibial, peroneal, and profunda arteries.
The Angioslide PROTEUS™ PTA Balloon Catheter with Embolic Capture is not intended for use in the renal, cerebral, coronary or carotid vasculature.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
.A. Hillian
Page 1 of
| (Division Sign-Off) | |
|------------------------------------|--|
| Division of Cardiovascular Devices | |
K120805 510(k) Number
Page 22 of 117
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.