SafeCross Transseptal Puncture and Introducer (TSP/I) System

K203459 · East End Medical I, LLC · DYB · Jul 6, 2021 · Cardiovascular

Device Facts

Record IDK203459
Device NameSafeCross Transseptal Puncture and Introducer (TSP/I) System
ApplicantEast End Medical I, LLC
Product CodeDYB · Cardiovascular
Decision DateJul 6, 2021
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1340
Device ClassClass 2

Indications for Use

The SafeCross Transseptal Puncture Device and Introducer (TSP/I) System is used to introduce various cardiovascular catheters to the heart, including the left side of the heart. The system enables left heart access through a puncture of the atrial septum during a transseptal catheterization procedure. In addition, the device can be used for monitoring intracardiac pressures, sampling blood, and infusing solutions.

Device Story

The SafeCross TSP/I System is a three-component kit comprising a steerable introducer sheath, an RF puncture member, and an access dilator. Used in clinical settings by physicians during transseptal catheterization, the device facilitates access to the left atrium. The steerable sheath features a distal compliant balloon for septal protection and positioning at the fossa ovalis. The RF puncture member connects to a third-party electrosurgical generator (50W, CUT mode) to create an atrial septal puncture via RF energy. The device is radiopaque for fluoroscopic visualization. By providing a stabilized, controlled puncture, the system enables the introduction of cardiovascular catheters, pressure monitoring, and fluid management, potentially improving procedural precision and patient outcomes during complex cardiac interventions.

Clinical Evidence

Bench testing only. No clinical data presented. Testing included dimensional verification, structural integrity (leakage, joint strength, torque), deliverability, radiopacity, balloon performance, electrical compatibility with electrosurgical generators, arc integrity, coring tests, and biocompatibility (ISO 10993). Usability validation was performed to confirm device handling and performance specifications.

Technological Characteristics

System components: Steerable introducer sheath (biocompatible materials, 8.5F ID, 75cm length, bidirectional >90° curve, inflatable positioning balloon), RF puncture member (103cm, 8.5F, Bovie D-shape 4mm connector), and access dilator. Energy source: External third-party electrosurgical generator (50W, CUT mode). Connectivity: Mechanical/Electrical interface to generators. Sterilization: Validated to SAL 10^-6. Radiopaque markers for fluoroscopy.

Indications for Use

Indicated for patients requiring transseptal catheterization for left heart access, intracardiac pressure monitoring, blood sampling, or fluid infusion.

Regulatory Classification

Identification

A catheter introducer is a sheath used to facilitate placing a catheter through the skin into a vein or artery.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. July 6, 2021 East End Medical I LLC % Diane Horwitz Owner Mandell Horwitz Consultants LLC 5 Lake Como Ct. Greenville, South Carolina 29609 Re: K203459 Trade/Device Name: SafeCross Transseptal Puncture and Introducer (TSP/I) System Regulation Number: 21 CFR 870.1340 Regulation Name: Catheter introducer Regulatory Class: Class II Product Code: DYB, DXF Dated: June 7, 2021 Received: June 8, 2021 Dear Diane Horwitz: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. Rachel Neubrander Assistant Director DHT2B: Division of Circulatory Support, Structural and Vascular Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration # Indications for Use 510(k) Number (if known) K203459 Device Name SafeCross Transseptal Puncture and Introducer (TSP/I) System Indications for Use (Describe) The SafeCross Transseptal Puncture Device and Introducer (TSP/I) System is used to introduce various cardiovascular catheters to the heart, including the left side of the heart. The system enables left heart access through a puncture of the atrial septum during a transseptal catheterization procedure. In addition, the device can be used for monitoring intracardiac pressures, sampling blood, and infusing solutions. | Type of Use (Select one or both, as applicable) | |-------------------------------------------------------------------------------------------------------------------------------------------------------| | <div style="display:flex; align-items:center;"><span><span style="font-size:16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D)</span></div> | | <div style="display:flex; align-items:center;"><span><span style="font-size:16px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C)</span></div> | ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the logo for EastEnd Medical. The word "EastEnd" is in dark gray, and the word "Medical" is in light gray. To the left of the text is a circular graphic in shades of orange. # 510(k) SUMMARY #### GENERAL INFORMATION 1. K203459 #### 1.1 Submitter and 510(k) Owner East End Medical I LLC 1157 Sawgrass Corporate Parkway Sunrise, FL 33323 #### 1.2 Official Correspondent Diane Horwitz, Ph.D. 5 Lake Como Ct. Greenville, SC 29609 #### 1.3 Date of Preparation July 1, 2021 ## 2. NAME OF THE DEVICE #### 2.1.1 Trade/Proprietary Name SafeCross Transseptal Puncture and Introducer (TSP/I) System #### 2.1.2 Common/Usual Name Introducer Catheter Septostomy Catheter #### Classification Information 2.1.3 | Classification Name: | Introducer Catheter | |----------------------------|---------------------------| | Classification Regulation: | 21 CFR §870.1340 | | Class: | 2 | | Product Code: | DYB, Introducer, Catheter | | Panel: | Cardiovascular | | Classification Name: | Septostomy Catheter | | Classification Regulation: | 21 CFR §870.5175 | | Class: | 2 | | Product Code: | DXF, Catheter, Septostomy | | Panel: | Cardiovascular | #### PREDICATE DEVICES 3. Primary Predicate Device: Agilis NXT Steerable Introducer, K081645, St. Jude Medical Secondary Predicate Device: NRG Transseptal Needle, K073326, Baylis Medical Company, Inc. Reference Device: AcQGuide Flex with AcQCross QX, AcQGuide Mini with AcQCross QX, K193509, Acutus Medical {4}------------------------------------------------ Image /page/4/Picture/0 description: The image contains the logo for EastEndMedical. The logo features a light red circular graphic to the left of the text. The text "EastEnd" is in a dark gray color, while "Medical" is in a lighter gray color. #### 4. DESCRIPTION OF THE DEVICE The SafeCross Transseptal Puncture and Introducer (TSP/I) System is used to access the heart left atrium through a puncture of the atrial septum. The System includes three (3) components: the Steerable Introducer Sheath, the RF Puncture Member, and an Access Dilator. The Steerable Introducer Sheath has a compliant balloon at the tip for protection of the septum, and the RF Puncture member is compatible with five (5) of the most common electrosurgical generators. #### న. INTENDED USE The SafeCross Transseptal Puncture Device and Introducer (TSP/I) System is used to introduce various cardiovascular catheters to the heart, including the left side of the heart. The system enables left heart access through a puncture of the atrial septum during a transseptal catheterization procedure. In addition, the device can be used for monitoring intracardiac pressures, sampling blood, and infusing solutions. #### 6. INTENDED USE COMPARED TO THE PREDICATES The TSP/I System has an intended use statement that is comprised of a combination of the intended uses from both Predicates. The statement is similar with the exception of several word changes specific to the subject devices also share the same target patient population, the same users and conditions of use (Table 1). | | Subject Device<br>SafeCross<br>TSP/I System<br>East End Medical I Inc. | Primary Predicate<br>Agilis NXT Steerable<br>Introducer<br>St. Jude Medical<br>K081645 | Secondary Predicate<br>NRG Transseptal Needle<br>Baylis Medical Company,<br>Inc.<br>K073326 | |---------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Intended Use<br>Statement | The SafeCross Transseptal<br>Puncture Device and<br>Introducer (TSP/I) System is<br>used to introduce various<br>cardiovascular catheters to<br>the heart, including the left<br>side of the heart. The system<br>enables left heart access<br>through a puncture of the<br>atrial septum during a<br>transseptal catheterization<br>procedure. In addition, the<br>device can be used for<br>monitoring intracardiac<br>pressures, sampling blood,<br>and infusing solutions. | The Agilis ™ NXT Steerable<br>Introducer is indicated for<br>introducing various<br>cardiovascular catheters into the<br>heart, including the left side of<br>the heart through the interatrial<br>septum. | Creation of an atrial septal<br>defect in the heart. Secondary<br>applications include transseptal<br>heart access, monitoring<br>intracardiac pressures,<br>sampling blood, and infusing<br>solutions. | ### Table 1. Intended Use / Indications for Use Comparison #### 7. TECHNOLOGICAL CHARACTERISTICS COMPARED TO THE PREDICATES A comparison of the technological features between the TSP/I System and the Predicates is shown in Table 2 below for the Steerable Introducer Catheter and in Table 3 for the RF Puncture Member. {5}------------------------------------------------ Image /page/5/Picture/0 description: The image contains the logo for EastEndMedical. The words "EastEnd" are in a dark gray color, while the word "Medical" is in a lighter gray color. To the left of the text is a circular graphic with a gradient of orange and red colors. | | Subject Device - Steerable<br>Introducer Sheath<br>SafeCross TSP/I System<br>East End Medical I Inc. | Primary Predicate<br>Agilis NXT Steerable<br>Introducer<br>St. Jude Medical<br>K081645 (L Curve) | Same or Different | |--------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Deflectable Distal<br>Segment Length | S: 5.5 mm<br>M: 6.6 mm<br>L: 9.5 mm | K081645<br>L: 8.4 mm | Same | | Materials | Biocompatible materials | Biocompatible materials | Same | | Radiopacity | Shaft is radiopaque<br>Positioning balloon is inflated<br>with 1 cc of 20% contrast - 80%<br>saline through the Positioning<br>Balloon Inflation Port | "The sheath is filled with<br>radiopaque material for<br>visualization under fluoroscopy" | Similar. Visibility under<br>fluoroscopy was confirmed<br>with bench and usability<br>testing | | Proximal end | Hemostasis valve and 2 side<br>ports with 3-way stopcock for<br>Positioning Balloon Inflation<br>Port and Flushing Line Port<br>(injection /aspiration of fluids) | Hemostasis valve and a side port<br>with 3-way stopcock for<br>injection or aspiration of fluids | Same | | Distal end/tip | Inflatable positioning balloon at<br>the distal tip to facilitate<br>positioning at the fossa ovalis<br>and stabilize the introducer<br>system for precise use of the RF<br>Puncture Member | No balloon | Different. Use of an<br>atraumatic compliant balloon<br>is common in cardiovascular<br>procedures.<br>Bench testing and use<br>validation study confirm<br>requirements and<br>specifications were met.<br>This change does not raise<br>new issues of safety or<br>effectiveness | | OD | 0.160" | 0.161" | Similar | | ID | 8.5F compatible | 8.5F compatible | Same | | Sheath effective length | 75 cm | 71 cm | Similar | | Compatible guidewire | 0.035" | 0.032" | Similar | | Distal curve(s) | > 90° bidirectional | > 90° bidirectional | Same | | Access Dilator | 103 cm TSP/I AD provided with<br>kit; compatible with 0.035"<br>guidewire | 94 cm AD provided with kit;<br>compatible with 0.032"<br>guidewire | Same | # Table 2. TSP/I Steerable Introducer Sheath and Access Dilator Technology Comparison to Predicates {6}------------------------------------------------ Image /page/6/Picture/0 description: The image contains the logo for EastEndMedical. The words "EastEnd" are in a dark gray color, while the word "Medical" is in a lighter gray color. To the left of the text is a circular graphic with a gradient of orange and red colors. | | Subject Device - RF<br>Puncture Member<br>SafeCross TSP/I System<br>East End Medical I Inc. | Secondary Predicate<br>NRG TRANSSEPTAL NEEDLE<br>Baylis Medical Company, Inc.<br>K073326 | Same or Different | |------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Working length | 103 cm | 98 cm | Similar | | Outer shaft diameter | 0.110" (8.5F) | 0.053" and requires 8.5F dilator for use | Same | | Visibility under fluoroscopy | Visible due to metal material | Visible due to metal material | Same | | Compatible guidewire | 0.035" | 0.032" | Similar | | Compatible Generator | Third-party FDA-cleared electrosurgical generators:<br>• Bovie OR/PRO 300<br>• CONMED System 5000<br>• ERBE VIO 300D<br>• Ethicon / Megadyne Mega Power<br>• Medtronic/ Covidien/ Valleylab Force FX/FX-C | The companion generator is cleared by the Baylis Medical Company, Inc., Models RFP-100 and RFP-100A | Different. The Predicate device is used with one generator from the same manufacturer, while the Subject device may be used with several qualified generators.<br>All generators are set to 50 W Power; Subject device details the generator settings in IFU and in training.<br>A Reference Device was selected, AcQGuide Flex (Acutus Medical, Inc., K193509) as it is similarly cleared for use with an electrosurgical generator from a Third Party manufacturer (which is the same as the Subject device compatible generator). | | Mode | Monopolar mode, CUT modality, 50 W | Up to 50W (RFP-100A Generator) | Each compatible generator has been tested with the Subject device and been characterized to have equivalent electrical performance with the Predicate device. | | Connector | Bovie D-shape 4mm connector | Proprietary connector | Similar | # Table 3. TSP/I RF Puncture Member Technology Comparison to Predicate {7}------------------------------------------------ Image /page/7/Picture/0 description: The image contains the logo for EastEndMedical. The word "EastEnd" is in dark gray, while the word "Medical" is in light gray. To the left of the text is a circular shape that is light orange. #### 7.1 Similarities and Differences in Technology Comparison The SafeCross TSP/I System is equivalent to the combination of the Agilis NXT Steerable Introducer (K081645) and the NRG Transseptal Needle (K073326) in terms of components and operational use. The technology of the Subject Steerable Introducer Sheath is identical to the Predicate Introducer Sheaths with the exception that the Subject device includes a Positioning Balloon on its distal end. Both the SafeCross Steerable Introducer Sheath and the Predicate are co-packaged with an access dilator. The technology of the SafeCross RF Puncture Member is almost identical to the Predicate Puncture Member. Both devices use RF energy technology to cross the septal wall, using an electrosurgical generator. Both devices are radiopaque for visualization during the procedure using fluoroscopy. Both devices utilize similar output powers from similar electrosurgical generators. The SafeCross RF Puncture Member has the advantage of being compatible with the most common electrosurgical generators. Electrical safety and compatibility have been demonstrated. #### 8. PERFORMANCE TESTING A series of bench tests was performed to demonstrate that the TSP/I System meets its performance specifications using final finished, sterilized and preconditioned product. Comprehensive verification and validation activities were successfully completed, raising no new issues of safety or effectiveness. All testing passed the acceptance criteria. Performance testing was conducted against known standards and product specifications and evaluated the following: Performance and Physical Requirements of the TSP/I System - Validation testing to demonstrate compatibility with accessory devices and radiopacity ● - . Dimensional verification - Deliverability and retraction ● - . Structural integrity (leakage, joint strength, flexibility, torque strength) - Particulate testing ● - Usability validation testing Performance and Physical Requirements of the TSP/I Steerable Introducer Sheath - Balloon dimensional and performance characteristics, deployment and retraction ● - . Flow and pressure testing ### Performance and Physical Requirements of the TSP/I RF Puncture Member - Electrical verification and compatibility according to IEC 60601-1, IEC 60601-1-2, IEC ● 60601-2-2 - Arc integrity verification ● - Coring test ● {8}------------------------------------------------ Image /page/8/Picture/0 description: The image shows the logo for EastEnd Medical. The word "EastEnd" is in dark gray, and the word "Medical" is in light gray. To the left of the text is a circular graphic in a gradient of orange. # Biological Safety Testing The TSP/I System was subjected to a series of biocompatibility tests in accordance with FDA guidance, using International Standard ISO 10993-1. - . Cytotoxicity - . Sensitization - Intracutaneous Reactivity ● - . Acute Systemic Toxicity - Hemocompatibility (Hemolysis, Complement Activation, Partial Thromboplastin Time, . Thromboresistance) - Material Mediated Pyrogenicity ● # Sterilization Sterilization and sterilization were performed to ensure a SAL of 10°, according to international sterilization standards. ## Packaging Validation and Shelf Life Visual Inspection, Bubble Leak and Seal Strength testing was used to evaluate integrity of the packaging configuration. Testing was conducted after sterilization, environmental conditioning including aging, and simulated shipping and distribution. The results of the performance testing conclude the SafeCross TSP/I System is safe and is substantially equivalent to the predicate devices. #### 9. CONCLUSIONS The information presented in this 510(k) submission demonstrates that the East End Medical I LLC SafeCross TSP/I System is substantially equivalent to the predicate devices.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...