DESTINATION RENAL GUIDING SHEATH

K081045 · Terumo Medical Corp. · DYB · May 28, 2008 · Cardiovascular

Device Facts

Record IDK081045
Device NameDESTINATION RENAL GUIDING SHEATH
ApplicantTerumo Medical Corp.
Product CodeDYB · Cardiovascular
Decision DateMay 28, 2008
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 870.1340
Device ClassClass 2

Indications for Use

The Destination® Renal Guiding Sheath is intended for the introduction of interventional and diagnostic devices into the human vasculature including but not limited to the Renal Arteries.

Device Story

The Destination® Renal Guiding Sheath functions as both a guiding catheter and an introducer sheath. It is a manual device used by clinicians to facilitate the introduction of diagnostic or interventional instruments into the human vasculature, specifically targeting the renal arteries. The system includes a sheath, dilator, hemostatic valve, and dilator retaining clip. It is operated manually by a physician in a clinical setting. The device provides a pathway for other medical instruments, aiding in minimally invasive vascular procedures. Its design allows for navigation through the vasculature to reach target sites, potentially improving the efficiency and safety of renal artery interventions.

Clinical Evidence

No clinical data. Equivalence demonstrated through bench testing and vessel model testing.

Technological Characteristics

Catheter introducer; 5Fr size; 45-55 cm length; hydrophilic coating on distal 5 cm; distal shape configurations include straight, HS, RDC, LIMA, MP. Materials are biocompatible per ISO 10993. Sterilization via ethylene oxide per EN ISO 11135-1.

Indications for Use

Indicated for the introduction of interventional and diagnostic devices into the human vasculature, including the renal arteries, in patients requiring such procedures.

Regulatory Classification

Identification

A catheter introducer is a sheath used to facilitate placing a catheter through the skin into a vein or artery.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K08/045 100% Lot 4 Terumo Medical Corporation Special 510(k)- Destination® Renal Guiding Sheath Section II. 510(k) Summary MAY 2 8 2008 # SECTION II. 510(k) SUMMARY #### A. DEVICE NAME Proprietary Name: Destination® Renal Guiding Sheath Classification Name: Catheter Introducer Common Name: Guiding Sheath #### B. PREDICATE DEVICE The predicate device is the Renal Guiding Sheath, which is manufactured by Terumo Medical Corporation. The Renal Guiding Sheath is cleared through the premarket notification process (K012854). #### C. INTENDED USE The Destination® Renal Guiding Sheath is intended for the introduction of interventional and diagnostic devices into the human vasculature including but not limited to the Renal Arteries. #### D. DESCRIPTION The Destination® Renal Guiding Sheath is designed to perform as a guiding catheter and an introducer sheath. The Guiding Sheath is packaged with the following components: a Sheath, a Dilator, a Hemostatic Valve, and a Dilator Retaining Clip. {1}------------------------------------------------ ### E. PRINCIPLE OF OPERATION / TECHNOLOGY The Destination® Renal Guiding Sheath is operated manually or by a manual process. ### F. DESIGN / MATERIALS The Destination® Renal Guiding Sheath uses similar materials as the predicate device. Differences in materials between the two devices do not raise any new issues of safety and effectiveness. #### G. SPECIFICATIONS | Sheath Size: | 5Fr. | |------------------------------|-----------------------------| | Sheath Length: | 45-55 cm | | Hydrophilic Coating: | Distal 5 cm | | Distal Shape Configurations: | Straight, HS, RDC, LIMA, MP | #### H. PERFORMANCE The performance of the Destination® Renal Guiding Sheath is substantially equivalent to the performance of the unmodified Renal Guiding Sheath. The equivalence was shown through bench and vessel model testing. {2}------------------------------------------------ Terumo Medical Corporation Special 510(k)- Destination® Renal Guiding Sheath Section II. 510(k) Summary #### I. ADDITIONAL SAFETY INFORMATION Sterilization conditions have been validated in accordance with EN ISO 11135-1 "Sterilization of health care products – Ethylene Oxide – Part 1 : requirements for development, validation and routine control of sterilization process for medical devices." to provide a Sterility Assurance Level of 10-6. Blood contacting materials were tested in accordance with the test recommendations in the FDA General Program Memorandum #G95-1 (5/1/95): Use of International Standard ISO 10993, "Biological Evaluation of Medical Devices – Part I: Evaluation and Testing." The Destination® Renal Guiding Sheath is categorized as "Externally Communicating Device, Circulating Blood, Limited Contact (≤24hrs)". The blood contacting materials were found to be biocompatible. Expiration dating for the Destination® Renal Guiding Sheath will be 30 months. ### J. SUBSTANTIAL EQUIVALENCE The Destination® Renal Guiding Sheath submitted in this 510(k) is substantially equivalent in intended use, design, principle of operation / technology, materials and performance to the Renal Guiding Sheath (K012854), which is manufactured by Terumo Medical Corporation. Differences between the devices do not raise any issues of safety or effectiveness. {3}------------------------------------------------ KOE 1045 page let y Terumo Medical Corporation Special 510(k)- Destination® Renal Guiding Sheath Section II. 510(k) Summary ### K. SUBMITTER INFORMATION ## Name and Address Terumo Medical Corporation 950 Elkton Blvd. Elkton, MD 21921 ## Contact Person Mr. Mark Unterreiner Sr. Regulatory Affairs Specialist Ph: 410-392-7213 Fax: 410-398-6079 Email: mark.unterreiner@terumomedical.com . "A . . . Date Prepared April 11, 2008 . {4}------------------------------------------------ Public Health Service Image /page/4/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird-like figure, with its wings forming a flowing, abstract shape. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES · USA" is arranged in a circular pattern around the bird symbol. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 MAY 2 8 2008 Terumo Medical Corporation c/o Mr. Mark Unterreiner Sr. Regulatory Affairs Specialist 950 Elkton Blvd. Elkton, MD 21921 Re: K081045 Trade/Device Name: Destination Renal Guiding Sheath Regulation Number: 21 CFR 870.1340 Regulation Name: Catheter Introducer Regulatory Class; Class II (two) Product Code: DYB Dated: May 19, 2008 Received: May 20, 2008 Dear Mr. Unterreiner: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {5}------------------------------------------------ ## Page 2 - Mr. Mark Unterreiner Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, uma R.v.luhner A Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health {6}------------------------------------------------ Indications for Use | 510(k) Number (if known): | K081045 | |---------------------------|---------| |---------------------------|---------| Destination® Renal Guiding Sheath Device Name: Indications For Use: The Destination® Renal Guiding Sheath is intended for the introduction of interventional and diagnostic devices into the human vasculature including but not limited to the Renal Arteries. Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Anna R. Kirchner (Division Sign-Off) Division of Cardiovascular Devices 510(k) Number_ Ko81045
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...