EXTRA PTA BALLOON CATHETER WITH EMBOLIC CAPTURE FEATURE; PROTEUS PTA BALLOON CATHETER WITH EMBOLIC CAPTURE FEATURE
Applicant
Angioslide, Ltd.
Product Code
DQY · Cardiovascular
Decision Date
Jul 21, 2011
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.1250
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Angioslide PROTEUS™ Percutaneous Transluminal Angioplasty (PTA) Balloon Catheter with Embolic Capture Feature is indicated for peripheral transluminal angioplasty and for capture and containment of embolic material during angioplasty, for the femoral, iliac, ilio-femoral, popliteal, tibial, peroneal, and profunda arteries. The Angioslide PROTEUS™ PTA Balloon Catheter with Embolic Capture Feature is not intended for use in the renal, cerebral, coronary or carotid vasculature.
Device Story
Over-the-wire coaxial dual-lumen PTA balloon catheter; features foldable balloon with radiopaque markers for positioning; includes embolic capture feature via proximal vessel occlusion and aspiration. Used by physicians during angioplasty procedures. Modification adds Particle Visualization Kit (PVK) accessory for use outside sterile field to visualize captured embolic material. Device facilitates angioplasty while reducing embolic risk to patient.
Clinical Evidence
Bench testing only; simulated use testing performed for the Particle Visualization Kit (PVK) accessory.
Technological Characteristics
Over-the-wire coaxial dual-lumen catheter; foldable balloon; radiopaque markers; compatible with 0.035" guidewire. Embolic capture via proximal balloon occlusion and aspiration. Modification includes non-sterile PVK accessory for particle visualization.
Indications for Use
Indicated for peripheral transluminal angioplasty and embolic material capture/containment in femoral, iliac, ilio-femoral, popliteal, tibial, peroneal, and profunda arteries. Contraindicated for renal, cerebral, coronary, or carotid vasculature.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
Predicate Devices
Angioslide eXtra™ PTA Balloon Catheter with Embolic Capture Feature (K090364)
Related Devices
K133043 — PROTEUS PTA BALLOON CATHETER WITH EMBOLIC CAPTURE FEATURE · Angioslide, Ltd. · Dec 6, 2013
K090364 — EXTRA PTA BALLOON CATHETER WITH EMBOLIC PROTECTION FEATURE · Angioslide, Ltd. · Mar 23, 2010
K120164 — PROTEUS PTA BALLOON CATHETER WITH EMBOLIC CAPTURE FEATURE · Angioslide, Ltd. · Jun 8, 2012
K172494 — PROTEUS PTA Balloon Catheter with Embolic Capture Feature · Angioslide, Ltd. · Dec 15, 2017
Submission Summary (Full Text)
{0}------------------------------------------------
Angioslide Ltd. - Special 510(k) PROTEUS™ PTA Catheter with Embolic Capture Feature Section 7: 510(k) Summary
510(k) Summary
JUL 2 1 2011
# PROTEUSTM PTA Catheter with Embolic Capture Feature
#### Introduction
This document contains the 510(k) summary for the modified PROTEUSTM PTA Catheter with Embolic Capture Feature. The content of this summary is based on the requirements of 21 CFR Section 807.92(c).
#### Applicant Name and Address
| Name: | Angioslide, Ltd. |
|-------------------|--------------------------------------------------------------------------|
| Address: | 32 Maskit St.<br>P.O. Box 12489<br>Herzilya 46733<br>Israel |
| | Tel: +(972) 9-970-0300<br>Fax: +(972) 9-970-0307<br>ilyab@angioslide.com |
| Official Contact: | Ilya Burovoy<br>Vice President, Regulatory Affairs and Quality Assurance |
### Summary Preparation Date: June 16th, 2011
### Device Name and Classification:
| Trade Name: | PROTEUS™ PTA Balloon Catheter with<br>Embolic Capture Feature |
|----------------------|---------------------------------------------------------------|
| Common Name: | Percutaneous Transluminal Angioplasty<br>Balloon Catheter |
| Classification Name: | Catheter, Percutaneous |
| Classification: | Class II, 21 CFR 870.1250 |
| Product Code | DQY/LIT |
### Predicate Device:
The modified PROTEUS™ Percutaneous Transluminal Angioplasty (PTA) Balloon Catheter with Embolic Capture Feature is claimed to be substantially equivalent to the following legally marketed predicate device:
- Angioslide eXtra™ PTA Balloon Catheter with Embolic Capture Feature (K090364) .
### Device Description:
The PROTEUS™ PTA Balloon Catheter with Embolic Capture is an over the wire coaxial dual lumen catheter with a foldable balloon located near the distal atraumatic soft tip.
The PROTEUS™ PTA Balloon Catheter with Embolic Capture Feature consists of a foldable balloon with radiopaque markers which aid in positioning the balloon in the artery during the
{1}------------------------------------------------
Angioslide Ltd. - Special 510(k) PROTEUS™ PTA Catheter with Embolic Capture Feature Section 7: 510(k) Summary
angioplasty procedure, the co-axial shaft comprising the inner shaft and the outer shaft, and the handle comprising the handle enclosure, the T-connector, the inflation port, the pulling rod with the guidewire port, and the pulling rod locks.
Following embolic capture, the user can opt to use the PROTEUS™ Particle Visualization Kit (PVK) to facilitate the visualization of the embolic material captured by the PROTEUS™ PTA Balloon Catheter.
# Indications for Use:
The PROTEUS™ Percutaneous Transluminal Angioplasty (PTA) Balloon Catheter with Embolic Capture Feature is indicated for peripheral transluminal angioplasty and for capture and containment of embolic material during angioplasty, for the femoral, iliac, ilio-femoral, popliteal, tibial, peroneal, and profunda arteries.
The PROTEUS™ PTA Balloon Catheter with Embolic Capture is not intended for use in the renal, cerebral, coronary or carotid vasculature.
# Summary of Technological Characteristics
The PROTEUS™ Percutaneous Transluminal Angioplasty (PTA) Balloon Catheter is an over the wire co-axial dual lumen catheter with a foldable balloon located near the distal atraumatic soft tip. The catheter is compatible with a 0.035" guidewire.
The balloon technological characteristics of the modified version of PROTEUS™ PTA Balloon Catheter are substantially equivalent to those of the unmodified version of the device. The modification that is the subject of this Special 510(k) is the addition of a Particle Visualization Kit (PVK) accessory that is used outside of the sterile field to allow users to visualize embolic particles captured by the embolic capture feature of the catheter.
The modified PROTEUSTM PTA Balloon Catheter overall length, catheter sheath sizing, balloon diameter, balloon length, balloon nominal pressure, balloon rated burst pressure and end hole diameter are identical to the unmodified version of the device.
The embolic capture technological characteristics of the modified PROTEUSTM PTA Balloon Catheter are identical to those of the unmodified version of the device. In both devices the containment and removal of embolic material is achieved by proximal vessel occlusion, by means of an inflatable balloon, and subsequent aspiration of embolic material.
# Summary of Non-Clinical Testing
Verification and validation testing of the PROTEUS™ Particle Visualization Kit included simulated use testing.
# Conclusion
The Angioslide PROTEUS™ PTA Balloon Catheter is substantially equivalent with respect to the indications for use, technological characteristics, and performance characteristics to the following legally marketed predicate device:
- eXtraTM PTA Balloon Catheter, Angioslide K090364 .
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows a black and white logo for the Department of Health & Human Services. The logo features a stylized eagle with three curved lines representing its wings. The text "DEPARTMENT OF HEALTH & HC." is arranged vertically along the left side of the logo.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room-WO66-G609 Silver Spring, MD 20993-0002
Angioslide, LTD. c/o Clay Anselmo President and CEO Reglera, LLC 11925 West I-70 Frontage Road North, Suite 900 Wheatridge, CO 80033
JUL 2 1 2011
Re: K111750
Trade/Device Name: PROTEUS PTA Balloon Catheter with Embolic Capture Feature Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: DQY, LIT Dated: June 20, 2011 Received: June 22, 2011
Dear Mr. Anselmo:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Or us and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. However, we remind you that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 - Mr. Clay Anselmo
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours.
Arlile B. Bow
Bram Zuckerman. M.D. M Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
510(k) Number (if known): K111750
#### Device Name: PROTEUS™ Percutaneous Transluminal Angioplasty (PTA) Balloon Catheter with Embolic Capture Feature
Indications for Use:
The Angioslide PROTEUS™ Percutaneous Transluminal Angioplasty (PTA) Balloon Catheter with Embolic Capture Feature is indicated for peripheral transluminal angioplasty and for capture and containment of embolic material during angioplasty, for the femoral, ilia-femoral, popliteal, tibial, peroneal, and profunda arteries.
The Angioslide PROTEUS™ PTA Balloon Catheter with Embolic Capture Feature is not intended for use in the renal, cerebral, coronary or carotid vasculature.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
#### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Of
Page
Page 1 of 1
(Division Sign-Off)
Division of Cardiovascular Devices
510(k) Number K111750
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.