AQT90 FLEX ANALYZER, AQT90 FLEX MYOGLOBIN TEST KIT, AQT90 FLEX LQC MULTI-CHECK, LVLS 1-3, AQT90 FLEX MYO CAL CARTRIDGE..

K112161 · Radiometer Medical Aps · KHO · Jan 20, 2012 · Clinical Chemistry

Device Facts

Record IDK112161
Device NameAQT90 FLEX ANALYZER, AQT90 FLEX MYOGLOBIN TEST KIT, AQT90 FLEX LQC MULTI-CHECK, LVLS 1-3, AQT90 FLEX MYO CAL CARTRIDGE..
ApplicantRadiometer Medical Aps
Product CodeKHO · Clinical Chemistry
Decision DateJan 20, 2012
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.2560
Device ClassClass 1

Intended Use

AQT90 FLEX analyzer is for in vitro diagnostic use. The instrument is an immunoassay instrument based on the quantitative determination of time-resolved fluorescence to estimate the concentrations of clinically relevant markers on whole-blood and plasma specimens to which a relevant anticoagulant has been added. It is intended for use in point-of-care and laboratory settings. AQT90 FLEX Myo Test is an in vitro diagnostic assay for the quantitative determination of myoglobin in EDTA or lithium-heparin whole blood or plasma specimens on the AQT90 FLEX analyzer in point of care and laboratory settings. It is indicated for use as an aid in the rapid diagnosis of heart disease, e.g. acute myocardial infarction. AQT90 FLEX Myo CAL cartridge is for in vitro diagnostic use. For calibration adjustment of the Myo Test, as indicated on the cartridge, on the AQT90 FLEX analyzer. AQT90 FLEX LQC Multi-CHECK, Levels 1-3, is for in vitro diagnostic use. For use with the AQT90 FLEX analyzer as a liquid quality control serum (LQC) to monitor the precision of laboratory testing procedures for the analytes listed on the specification insert.

Device Story

AQT90 FLEX is a cartridge-based, fully automated immunoassay analyzer using time-resolved fluorescence. Input: whole blood or plasma samples added to test cups containing capture and tracer antibodies (Europium-chelate labeled). Process: sample/buffer incubation (7-15 min at 37°C), washing to remove unbound material, drying, and excitation light exposure. Output: emitted light measured via single photon counting (up to 3,300 cycles); total count compared to calibration curve for quantitative analyte concentration. Used in point-of-care and laboratory settings by clinicians/technicians. Provides rapid myoglobin results to assist in diagnosing acute myocardial infarction. Benefits include rapid turnaround (under 20 minutes) and closed-system operation.

Clinical Evidence

Bench testing only. Comparison study performed against ARCHITECT STAT Myoglobin using 157 whole blood samples (range 24-856 ng/mL) and 165 plasma samples (range 22-882 ng/mL). Passing-Bablok regression yielded y = 1.07x + 15 (r²=0.99) for whole blood and y = 1.02x + 13 (r²=0.99) for plasma. Within-run CV ≤ 2.5% (plasma) and ≤ 3.7% (whole blood); total CV ≤ 5.2% (plasma) and ≤ 3.7% (whole blood).

Technological Characteristics

Cartridge-based immunoassay; time-resolved fluorescence detection using Europium (Eu) chelate label. Single photon counting measurement. Closed, fully automated system. Analyzes EDTA or lithium-heparin whole blood/plasma. Dimensions/form factor: cartridge-based. Connectivity: barcode-based sample identification. Calibration: automated via calibration cartridges. Storage: 2-8°C.

Indications for Use

Indicated for quantitative determination of myoglobin in EDTA or lithium-heparin whole blood or plasma specimens to aid in the rapid diagnosis of heart disease, such as acute myocardial infarction, in point-of-care and laboratory settings.

Regulatory Classification

Identification

A fluorometer for clinical use is a device intended to measure by fluorescence certain analytes. Fluorescence is the property of certain substances of radiating, when illuminated, a light of a different wavelength. This device is used in conjunction with certain materials to measure a variety of analytes.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ JAN 2 0 2012 \$\CH/2i6\$ 510(k) Premarket Notification Submission # RADIOMETER R AQT90 FLEX, Reagent Pack, Myo Test Kit, Myo CAL Cartridge and LQC Multi-CHECK ## 510(k) Summary AQT90 FLEX AQT90 FLEX Myoqlobin Test Kit AQT90 FLEX Myoglobin CAL Cartridge AQT90 FLEX LQC Multi-CHECK Manufacturer Radiometer Medical ApS Aakandevej 21, DK-2700 Bronshoj, Denmark Contact information Radiometer Medical ApS Aakandevej 21 DK-2700 Bronshoj Denmark Att. Ms Jana S. Hellmann Vice President Global RA/QA Phone +45 3827 3827 Fax: +45 3827 2727 E-mail: jana.hellmann@radiometer.dk Signature Tan J. Hine Jaña S. Hellmann Date of preparation January 16, 2011 {1}------------------------------------------------ Image /page/1/Picture/1 description: The image contains the word "RADIOMETER" in bold, block letters. To the right of the word is a circled letter R, which is a registered trademark symbol. The letters are all capitalized and evenly spaced. AOT90 FLEX, Reagent Pack, Myo Test Kit, Myo CAL Cartridge and LQC Multi-CHECK ## Table of contents | 1. | Device names and classifications | | |----|-----------------------------------------------------------------|--| | 2. | Predicate device | | | 3. | Device description / test principle | | | 4. | Device intended use | | | 5. | Medical device to which substantial equivalence is claimed | | | 6. | Technological characteristics in comparison to predicate device | | | 7. | Conclusion | | #### 1. Device names and classifications - · Proprietary name: AQT90 FLEX Class I Classification name: fluorometer for clinical use (21 CFR 862.2560) Product code: KHO - . Proprietary name: AQT90 FLEX Myo Test Kit Class II Classification name: myoglobin, antiserum, control use (21 CFR. 866.5680) Product code: DDR - · Proprietary name: AQT90 FLEX Myo CAL Cartridge Class II Classification name: calibrator, secondary (21 CFR. 862.1150) Product code: JIT - · Proprietary name: AQT90 LQC FLEX Multi-CHECK, Levels 1, 2 and 3 Class I Classification name: quality control material (assayed and unassayed) (21 CFR 862.1660) Product code: JJY ## 2. Predicate device - · ARCHITECT STAT Myoglobin, 510(k) number K042924 - . ARCHITECT STAT Myoglobin Calibrators, 510(k) number K042924 - · Liquicheck Cardiac Markers Plus Control LT, 510(k) number K050537 ## 3. Device description / test principle The AOT90 FLEX is a cartridge-based immunoassay, based on time-resolved fluorescence using a europium (Eu) chelate as the fluorescent label. The test receptacles for the assay are 300 µL test cups, which contain the antibodies used for capture of the analyte, and the Eu chelate labeled antibodies used to trace the captured analyte. The sample is added to the test cup together with assay buffer. The cup is then incubated to allow formation of the immuno-complex, and subsequently washed to remove unbound antibodies and sample material. Finally, the cup is exposed to excitation light, and after a delay the emitted light generated by the fluorescent label Page 2 of 9 {2}------------------------------------------------ AQT90 FLEX, Reagent Pack, Myo Test Kit, Myo CAL Cartridge and LQC Multi-CHECK is measured by single photon counting; this measurement cycle is repeated up to 3,300 times. The total count is then compared to an assay calibration curve to obtain a quantitative measurement of the analyte's concentration in the sample. This technology uses dried reagents deposited in the test cups and in the calibration adjustment cups - no liguids other than the sample itself together with the assay buffer are required. Total assay time is less than 20 minutes. In summary, the procedure is as follows: - 1. Metering of an exact amount of sample and assay buffer and dispensing into a test cup - 2. Incubating for 7-15 minutes at 37 °C - 3. Washing of the test cup to remove unbound tracer antibodies and sample material - 4. Drying the test cup - 5. Measuring #### 4. Device intended use AQT90 FLEX is for in vitro diagnostic use. The AQT90 FLEX analyzer is an immunoassay instrument based on the quantitative determination of time-resolved fluorescence to estimate the concentrations of clinically relevant markers on whole-blood and plasma specimens to which a relevant anticoaqulant has been added. It is intended for use in point-of-care and laboratory settings. AQT90 FLEX Myo Test Kit includes 10 AQT90 FLEX Myo Test cartridges and one AQT90 FLEX Myo CAL cartridge. AQT90 FLEX Myo Test is an in vitro diagnostic assay for the guantitative determination of myoqlobin in EDTA or lithium-heparin whole blood or plasma specimens on the AOT90 FLEX analyzer in point of care and laboratory settings. It is indicated for use as an aid in the rapid diagnosis of heart disease, e.g. acute myocardial infarction. AQT90 FLEX Myo CAL cartridge is for in vitro diagnostic use. For calibration adjustment of the Myo Test, as indicated on the cartridge, on the AQT90 FLEX analyzer. AQT90 FLEX LQC Multi-CHECK, Levels 1, 2 and 3, is for in vitro diagnostic use. For use with the AOT90 FLEX analyzer as a liquid quality control serum (LQC) to monitor the precision of laboratory testing procedures for the analytes listed on the specification insert. #### 5. Medical device to which substantial equivalence is claimed The AOT90 FLEX Myo Test Kit and Myo CAL cartridges are substantially equivalent in features and characteristics to the ARCHITECT STAT Myoglobin test and ARCHITECT STAT Myoglobin Calibrators, 510(k) number K042924. The AOT90 LQC FLEX Multi-CHECK, Levels 1, 2 and 3, is substantially equivalent in features and characteristics to the Liquicheck Cardiac Markers Plus Control LT, 510(k) number K050537. {3}------------------------------------------------ AQT90 FLEX, Reagent Pack, Myo Test Kit, Myo CAL Cartridge and LQC Multi-CHECK ## 6. Technological characteristics in comparison to predicate device AQT90 FLEX analyzer ARCHITECT i System Item Intended use The AQT90 FLEX analyzer is The ARCHITECT i System is for for in vitro diagnostic use. It is in vitro diagnostic use. It is an an immunoassay instrument an immunoassay instrument based on the quantitative based on the quantitative determination of time-resolved determination of fluorescence to estimate chemiluminescense to concentrations of analytes in measure and quantify analyte whole-blood and plasma concentration in plasma and specimens. It is intended for serum. use in point-of-care and laboratory settings. Test format A completely closed, fully The ARCHITECT i System is an automated system, using open, fully-automated, cartridge-based immunoassay system allowing immunoassays in disposable random and continuous cups. Offers possibility of access, and priority running 15 different processing. parameters at a time and has a total capacity of 200 tests without replenishment. Specimen Samples are identified from Samples are identified from identification the barcode on blood the barcode on blood collection test tube. The code is read collection test tube. The code is read automatically by the automatically by the analyzer. analyzer. Specimen sampling Blood is drawn into normal Blood is drawn into normal and handling vacuum test. The blood vacuum test tubes, which are collection tube is placed centrifuged to obatin plasma. directly in the analyzer. A The blood collection tube is volume of sample is placed directly in the analyzer. automatically drawn from the A volume of sample is tube through its rubber seal. automatically drawn from the The still closed tube is tube through its rubber seal. discarded. The used The still closed tube is immunoassay cup is discarded discarded. into a closed waste bin in the Comparison of features for AQT90 FLEX analyzer and predicate device {4}------------------------------------------------ | Item | AQT90 FLEX analyzer | ARCHITECT i System | |-----------------|---------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------| | | analyzer. | | | Calibration | The analyzer can be<br>automatically calibrated by<br>means of calibration cartridges<br>at the time of test cartridge lot<br>change. | Calibration must be carried out<br>each time a new reagent lot<br>number is used. Calibration is<br>run automatically. | | Quality control | AQT FLEX LQC Multi-CHECK,<br>Levels 1, 2 and 3 is a liquid<br>quality control. Its barcode<br>includes upper and lower limits<br>for the control range. | Architect STAT Controls,<br>Levels 1, 2 and 3. | AQT90 FLEX, Reagent Pack, Myo Test Kit, Myo CAL Cartridge and LQC Multi-CHECK Comparison of features for AQT90 FLEX Myo Test and predicate device | Item | AQT90 FLEX Myo Test | ARCHITECT STAT Myoglobin | |-----------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Intended use | The AQT90 FLEX Myo Test is<br>an in vitro diagnostic assay for<br>the quantitative determination<br>of myoglobin in EDTA or<br>lithium-heparin whole blood or<br>plasma specimens on the<br>AQT90 FLEX analyzer in point<br>of care and laboratory settings.<br>It is indicated for use as an aid<br>in the rapid diagnosis of heart<br>disease, e.g. acute myocardial<br>infarction. | ARCHITECT STAT Myoglobin is<br>a chemiluminescent<br>microparticle immunoassay<br>(CMIA) for the quantitative<br>determination of myoglobin in<br>human serum and plasma on<br>the ARCHITECT i System with<br>STAT protocol capability.<br>Myoglobin values are used to<br>assist in the diagnosis of<br>myocardial infarction (MI). | | Test format | Cartridge with 16 test cups,<br>each coated with anti-<br>myoglobin capture antibody<br>and containing a separating<br>layer as well as Eu-chelate<br>anti-myoglobin tracer. Sample<br>and assay buffer are added to<br>the cup. After an incubation<br>period the cup is washed with<br>assay buffer and dried. When<br>exposed to an excitation light<br>the bound europium emits a<br>fluorescence, which is<br>measured in cycles of single<br>photon counting. The total | In the 1st step, sample and<br>anti-myoglobin coated<br>paramagnetic microparticles<br>are mixed and incubated.<br>Myoglobin present in the<br>sample binds to the anti-<br>myoglobin coated<br>microparticles. After washing,<br>anti-myoglobin acridinium-<br>labeled conjugate is added in<br>the 2nd step. Following another<br>incubation and wash, pre-<br>trigger and trigger solutions<br>are added to the reaction<br>mixture. The resulting | | Item | AQT90 FLEX Myo Test | ARCHITECT STAT Myoglobin | | | count is then compared to an<br>assay calibration curve to<br>obtain a quantitative<br>measurement of the analyte's<br>concentration in the sample. | chemiluminescent reaction,<br>which is measured as relative<br>lights units (RLUs). A direct<br>relationship exists between the<br>amount of myoglobin in the<br>sample and the RLUs detected. | | Traceability | Scripps M0725 | N/A | | Antibodies | Mouse monoclonals for capture<br>and tracer antibody | Mouse monoclonals for capture<br>and tracer antibody | | Sample type | Human whole blood and<br>plasma | Human serum and plasma | | Anticoagulants | EDTA, Li-heparin | Li-heparin, Na-heparin, EDTA | | Controls | Recommended | Recommended | | In-use stability | 16 days on-board | 30 days on-board | | Storage temperature | 2-8 °C | 2-8 °C | | Reportable /<br>calibration range | Reportable range 20-<br>900 ng/mL (µg/L) | Calibration range 0.0-<br>1,200.0 ng/mL | | Analytical sensitivity | Limit of quantitation 1 ng/mL<br>(µg/L) | ≤ 1.0 ng/mL at the 95% level<br>of confidence | | Reference range | 97.5th percentile for females is<br>75 ng/mL (µg/L), for males<br>142 ng/mL (µg/L) | 99th percentile for females is<br>106.0 ng/mL, for males<br>154.9 ng/mL, total<br>140.1 ng/mL | | Imprecision | Across the reportable range,<br>CV(%)within-run is ≤ 2.5% for<br>plasma and ≤ 3.7% for whole<br>blood;<br>CV(%)total is ≤ 5.2% for<br>plasma and ≤ 3.7% for whole<br>blood | ≤ 10% for myoglobin<br>concentrations ≥ 40 ng/mL.<br>CV(%)total for controls is<br>3.5-4.6% at 56.7-62.3 ng/mL,<br>3.4-4.6% at 324.9-<br>354.9 ng/mL, and 3.2-5.4% at<br>785.2-848.1 ng/mL | | Interference | No significant interferences | No significant interferences | | Comparison with<br>predicate | Comparison with Abbott<br>ARCHITECT STAT Myoglobin<br>assay, with 157 whole blood<br>samples in the range 24-856<br>µg/L (ng/mL), and 165 plasma<br>samples in the range 22-882 | Comparison with Abbott<br>AxSYM myoglobin assay, 234<br>samples in the range 7.6 -<br>1,188.7 ng/mL<br>y = 0.99 x AXSYM + 2.88, r =<br>0.99 | | Item | AQT90 FLEX Myo Test | ARCHITECT STAT Myoglobin | | | AQT90 FLEX Myo assay:<br>$y = 1.07 x ARCHITECT + 15,$<br>$r^2 = 0.99$ with whole blood; $y$<br>$= 1.02 x ARCHITECT + 13, r^2$<br>$= 0.99$ with plasma | | {5}------------------------------------------------ Image /page/5/Picture/0 description: The image shows the word "RADIOMETER" in all caps, with a registered trademark symbol to the right of it. The text is in a bold, sans-serif font. The registered trademark symbol is a capital R inside of a circle, which is inside of a square. AQT90 FLEX, Reagent Pack, Myo Test Kit, Myo CAL Cartridge and LQC Multi-CHECK {6}------------------------------------------------ v - start - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - ## 510(k) Premarket Notification Submission AQT90 FLEX, Reagent Pack, Myo Test Kit, Myo CAL Cartridge and LQC Multi-CHECK ## Comparison of features for AQT90 FLEX Myo CAL and predicate device | Item | AQT90 FLEX Myoglobin CAL | ARCHITECT STAT<br>Myoglobin Calibrators | |------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Intended use | The AQT90 FLEX Myo CAL<br>Cartridge is for in vitro<br>diagnostic use. For calibration<br>adjustment of the Myo Test,<br>as indicated on the cartridge,<br>on the AQT90 FLEX analyzer. | The ARCHITECT STAT<br>Myoglobin Calibrators are for<br>calibration of the ARCHITECT i<br>System with STAT protocol<br>capability when used for the<br>quantitative determination of<br>myoqlobin in human serum or<br>plasma. | | Constituents | Each CAL Cartridge contains<br>eight analyte-specific<br>background cups and eight<br>cups with added antigen | Purified cardiac myoglobin in<br>Tris buffer with stabilizers | | Calibration adjustment<br>interval | Once per lot of AQT90 FLEX<br>Myo Test cartridges and as<br>often as required by relevant<br>regulations | Upon new assay reagent lot<br>number | | In-use stability | 24 hours on-board | 30 days at 2-8 °C | | Storage temperature | 2-8 °C | -10 °C | {7}------------------------------------------------ AQT90 FLEX, Reagent Pack, Myo Test Kit, Myo CAL Cartridge and LQC Multi-CHECK Comparison of features for AQT90 FLEX Multi-CHECK and predicate device, the Liquicheck Cardiac Markers Plus Control LT | Item | AQT90 LQC FLEX Multi-<br>CHECK | Liquicheck Cardiac Markers<br>Plus Control LT | |---------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Intended use | The AQT90 FLEX LQC Multi-<br>CHECK, Levels 1-3 is for <i>in vitro</i> diagnostic use. For use<br>with the AQT90 FLEX analyzer<br>as a liquid quality control,<br>serum (LQC) to monitor the<br>precision of laboratory testing<br>procedures for the analytes<br>listed on the specification<br>insert. | Liquicheck Cardiac Markers<br>Plus Control LT is intended for<br>use as quality control serum to<br>monitor the precision of<br>laboratory testing procedures<br>listing in the package insert. | | Analytes contained | Myoglobin | B-type Natriuretic Peptide<br>(BNP), Creatine Kinase (Total),<br>C-Reactive Protein (CRP),<br>Homocysteine, Digitoxin, N-<br>terminal pro-B-type Natriuretic<br>Peptide (NT-proBNP), CKMB,<br>Myoglobin, Troponin I,<br>Troponin T. | | Matrix | Human serum | Human serum | | Storage temperature | $ {\leq} $ -18 $ ^{\circ} $ C | -20 $ ^{\circ} $ C to -70 $ ^{\circ} $ C until<br>expiration date | | In-use stability | 4 days if stored unused at<br>2-8 $ ^{\circ} $ C; 2 hours if stored<br>unused at room temperature | 20 days at 2-8 $ ^{\circ} $ C | ## Summary of clinical performance data The AQT90 FLEX Myo assay (y) was compared to the predicate device, the ARCHITECT STAT Myoglobin (x) using Li-heparin whole blood samples in the range 24-856 µg/L (ng/mL) and plasma samples in the range of 22-882 ng/mL (ug/L) with the AQT90 FLEX Myo assay. The relationship between the two methods was determined by Passing-Bablok regression. The regression lines and correlation coefficients were found to be y = 1.07 x ARCHITECT + 15, r2 = 0.99 with whole blood, and y = 1.02 x ARCHITECT + 13, r2 = 0.99 with plasma. {8}------------------------------------------------ AQT90 FLEX, Reagent Pack, Myo Test Kit, Myo CAL Cartridge and LQC Multi-CHECK . ## 7. Conclusion The products listed in the table are substantially equivalent based on their indications for use and performance characteristics. | New Device | Predicate Device | |----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------| | AQT90 FLEX Myo Test Kit. Class II.<br>Classification name: myoglobin, antigen,<br>antiserum, control use,<br>(21 CFR. 866.5680), product code DDR | ARCHITECT STAT Myoglobin, 510(k) number<br>K042924 | | AQT90 FLEX Myo CAL Cartridge. Class II.<br>Classification name: calibrator, secondary,<br>(21 CFR. 862.1150), product code JIT | ARCHITECT STAT Myoglobin Calibrators, 510(k)<br>number K042924 | | AQT90 FLEX LQC Multi-CHECK, Levels 1, 2<br>and 3. Class I. Classification name: quality<br>control material (assayed and unassayed)<br>(21 CFR 862.1660), product code JJY | Liquicheck Cardiac Markers Plus Control LT, 510(k)<br>number K050537 | · ・ {9}------------------------------------------------ Image /page/9/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features the department's name encircling a symbol. The symbol consists of a stylized caduceus, a traditional symbol of medicine, with three intertwined snakes and a staff. 10903 New Hampshire Avenue Silver Spring, MD 20993 Radiometer Medical ApS c/o Jana Hellmann Vice President Global RA/QA Aakandevej 21 Bronshoj, Denmark DK-2700 JAN 2 0 2012 k112161 Re: Trade Name: Radiometer Medical ApS AQT90 Flex Myo Test, AQT90 FLEX Myo CAL cartridge, AQT90 FLEX LQC Multi-CHECK, Levels 1-3 and AQT90 FLEX analyzer Regulation Number: 21 CFR §866.5680 Regulation Name: Myoglobin immunological test system. Regulatory Class: Class II Product Code: DDR, JIT, JJY, KHO Dated: January 16, 2012 Received: January 20, 2012 Dear Ms. Hellmann: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 {10}------------------------------------------------ ### Page 2 If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (301) 796-5760. For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/Medical Devices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance ... You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-5680 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm Sincerely yours, Counney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health Enclosure {11}------------------------------------------------ Indications for Use 510(k) number (if known): K112161 Device name: AQT90 FLEX analyzer AQT90 FLEX Myo Test Kit AQT90 FLEX LQC Multi-CHECK, Levels 1-3 AQT90 FLEX Myo CAL Cartridge Intended use: AQT90 FLEX analyzer is for in vitro diagnostic use. The instrument is an immunoassay instrument based on the quantitative determination of time-resolved fluorescence to estimate the concentrations of clinically relevant markers on whole-blood and plasma specimens to which a relevant anticoagulant has been added. It is intended for use in point-of-care and laboratory settings. AQT90 FLEX Myo Test is an in vitro diagnostic assay for the quantitative determination of myoglobin in EDTA or lithium-heparin whole blood or plasma specimens on the AQT90 FLEX analyzer in point of care and laboratory settings. It is indicated for use as an aid in the rapid diagnosis of heart disease, e.g. acute myocardial infarction. AOT90 FLEX Myo CAL cartridge is for in vitro diagnostic use. For calibration adjustment of the Myo Test, as indicated on the cartridge, on the AQT90 FLEX analyzer. AQT90 FLEX LQC Multi-CHECK, Levels 1-3, is for in vitro diagnostic use. For use with the AQT90 FLEX analyzer as a liquid quality control serum (LQC) to monitor the precision of laboratory testing procedures for the analytes listed on the specification insert. Prescription Use × (21 CFR Part 801 Subpart D) And/Or Over the Counter Use _ (21 CFR Part 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OIVD) Russe Shear Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety 510(k):
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...