Real Time Nucleic Acid Amplification System

Page Type
Product Code
Definition
The system is a clinical multiplex instrument intended to measure and sort multiple signals generated my multiple probes, intercalating dyes, or other ligands in an assay from a clinical sample. Signals may be generated by fluorescence or other phenomena and may be measured using filters on a photodiode or other detector. It may integrate sample and/or reagent handling, amplification, dedicated instrument control, data acquisition software, raw data storage mechanisms and other essential hardware components along with the signal reader unit. The system is used with specific assays to comprise an assay test system.
Physical State
Should not include microarray or electrophoresis detection methods or instruments.
Technical Method
A real-time thermocycler is intended to identify and/or quantify the presence of specific sequences of double stranded DNA, amplified from a biological source and labeled with fluorescently labeled probes or through the use of intercalating dyes and detect using a high-power light-emitting diode (LED). Fluorescence emission is detected through the use of filters on a photodiode. The emission filters are optimized for use with specific fluorescent dyes.
Target Area
N/A
Regulation Medical Specialty
Chemistry
Review Panel
Chemistry
Submission Type
510(K) Exempt
Device Classification
Class 2
Regulation Number
862.2570
GMP Exempt?
No
Summary Malfunction Reporting
Eligible
Implanted Device
No
Life-Sustain/Support Device
No
Third Party Review
Not Third Party Eligible

CFR § 862.2570 Instrumentation for clinical multiplex test systems

§ 862.2570 Instrumentation for clinical multiplex test systems.

(a) Identification. Instrumentation for clinical multiplex test systems is a device intended to measure and sort multiple signals generated by an assay from a clinical sample. This instrumentation is used with a specific assay to measure multiple similar analytes that establish a single indicator to aid in diagnosis. Such instrumentation may be compatible with more than one specific assay. The device includes a signal reader unit, and may also integrate reagent handling, hybridization, washing, dedicated instrument control, and other hardware components, as well as raw data storage mechanisms, data acquisition software, and software to process detected signals.

(b) Classification. Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9. The special control is FDA's guidance document entitled “Class II Special Controls Guidance Document: Instrumentation for Clinical Multiplex Test Systems.” See § 862.1(d) for the availability of this guidance document.

[70 FR 11868, Mar. 10, 2005, as amended at 84 FR 71797, Dec. 30, 2019]

Real Time Nucleic Acid Amplification System

Page Type
Product Code
Definition
The system is a clinical multiplex instrument intended to measure and sort multiple signals generated my multiple probes, intercalating dyes, or other ligands in an assay from a clinical sample. Signals may be generated by fluorescence or other phenomena and may be measured using filters on a photodiode or other detector. It may integrate sample and/or reagent handling, amplification, dedicated instrument control, data acquisition software, raw data storage mechanisms and other essential hardware components along with the signal reader unit. The system is used with specific assays to comprise an assay test system.
Physical State
Should not include microarray or electrophoresis detection methods or instruments.
Technical Method
A real-time thermocycler is intended to identify and/or quantify the presence of specific sequences of double stranded DNA, amplified from a biological source and labeled with fluorescently labeled probes or through the use of intercalating dyes and detect using a high-power light-emitting diode (LED). Fluorescence emission is detected through the use of filters on a photodiode. The emission filters are optimized for use with specific fluorescent dyes.
Target Area
N/A
Regulation Medical Specialty
Chemistry
Review Panel
Chemistry
Submission Type
510(K) Exempt
Device Classification
Class 2
Regulation Number
862.2570
GMP Exempt?
No
Summary Malfunction Reporting
Eligible
Implanted Device
No
Life-Sustain/Support Device
No
Third Party Review
Not Third Party Eligible

CFR § 862.2570 Instrumentation for clinical multiplex test systems

§ 862.2570 Instrumentation for clinical multiplex test systems.

(a) Identification. Instrumentation for clinical multiplex test systems is a device intended to measure and sort multiple signals generated by an assay from a clinical sample. This instrumentation is used with a specific assay to measure multiple similar analytes that establish a single indicator to aid in diagnosis. Such instrumentation may be compatible with more than one specific assay. The device includes a signal reader unit, and may also integrate reagent handling, hybridization, washing, dedicated instrument control, and other hardware components, as well as raw data storage mechanisms, data acquisition software, and software to process detected signals.

(b) Classification. Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9. The special control is FDA's guidance document entitled “Class II Special Controls Guidance Document: Instrumentation for Clinical Multiplex Test Systems.” See § 862.1(d) for the availability of this guidance document.

[70 FR 11868, Mar. 10, 2005, as amended at 84 FR 71797, Dec. 30, 2019]