XL-200 CLINICAL CHEMISTRY ANALYZER, JAS GLUCOSE REAGENT, ISE REAGENT PACK

K130915 · Jas Diagnostics, Inc. · JJE · May 15, 2014 · Clinical Chemistry

Device Facts

Record IDK130915
Device NameXL-200 CLINICAL CHEMISTRY ANALYZER, JAS GLUCOSE REAGENT, ISE REAGENT PACK
ApplicantJas Diagnostics, Inc.
Product CodeJJE · Clinical Chemistry
Decision DateMay 15, 2014
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.2160
Device ClassClass 1

Intended Use

The XL-200 Clinical Chemistry Analyzer is an automated, random access. computer controlled, bench top clinical chemistry analyzer for clinical chemistry tests. The instrument provides in vitro quantitative measurements for glucose, sodium, potassium, and chloride in serum. This device is intended for clinical laboratory use. The JAS Glucose Reagent in intended for the in vitro quantitative measurement of glucose in serum on the XL-200 Clinical Chemistrv Analyzer. This device is intended for clinical laboratory use. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders, including diabetes mellitus, neonatal hypoglycemia and of pancreatic islet cell carcinoma. The ISE Reagent Pack is intended for the in vitro quantitative measurement of sodium, potassium, and chloride concentrations in serum on the XL-200 Clinical Chemistry Analyzer. This device is intended for clinical laboratory use. Sodium measurements are used in the diagnosis and treatment of aldosteronism, diabetes insipidus and other diseases involving electrolyte imbalance. Potassium measurements are used to monitor electrolyte balance in the diagnosis and treatment of disease conditions characterized by low or high blood potassium levels. Chloride measurements are used in the diagnosis and treatment of electrolyte and metabolic disorders such as cystic fibrosis and diabetic acidosis.

Device Story

The XL-200 is an automated, random-access, benchtop clinical chemistry analyzer. It processes serum samples using colorimetric, turbidimetric, and ion-selective electrode (ISE) methods. The system includes a sampling arm, reagent arm, reaction station, sample/reagent plates, photometer, and wash station. Operators load samples and reagents; the system performs automated analysis, including re-dilution of abnormal samples. The photometer uses a halogen lamp and silicone photodiode array to measure absorbance at multiple wavelengths (340-700nm). ISE modules measure sodium, potassium, and chloride. Output is quantitative concentration data, which clinicians use to diagnose and monitor metabolic and electrolyte disorders. The device supports LIS connectivity and barcode scanning for sample/reagent tracking. It benefits patients by providing rapid, automated diagnostic testing for conditions like diabetes and cystic fibrosis.

Clinical Evidence

Bench testing only. Precision evaluated per CLSI EP5-A2 (N=80 total precision). Linearity evaluated per CLSI EP06-A. Interference testing per CLSI EP07-A2. Method comparison performed against predicate (N=88-103 samples per analyte) showing high correlation (R² 0.9589-0.9970). Limits of quantitation/detection established per CLSI EP17-A.

Technological Characteristics

Benchtop analyzer; photometric (halogen lamp, silicone photodiode array) and ISE sensing. Wavelengths: 340, 405, 505, 546, 578, 600, 660, 700 nm. 45 quartz semi-disposable cuvettes. Reagent refrigeration included. Connectivity: LIS external. Power: 110V/220V. Software-controlled automated sampling, mixing, and wash cycles.

Indications for Use

Indicated for in vitro quantitative measurement of glucose, sodium, potassium, and chloride in serum for clinical laboratory use. Used in diagnosis/treatment of carbohydrate metabolism disorders (diabetes mellitus, neonatal/idiopathic hypoglycemia, pancreatic islet cell carcinoma) and electrolyte/metabolic disorders (aldosteronism, diabetes insipidus, cystic fibrosis, diabetic acidosis).

Regulatory Classification

Identification

A discrete photometric chemistry analyzer for clinical use is a device intended to duplicate manual analytical procedures by performing automatically various steps such as pipetting, preparing filtrates, heating, and measuring color intensity. This device is intended for use in conjunction with certain materials to measure a variety of analytes. Different models of the device incorporate various instrumentation such as micro analysis apparatus, double beam, single, or dual channel photometers, and bichromatic 2-wavelength photometers. Some models of the device may include reagent-containing components that may also serve as reaction units.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # 1 5 2014 ### 510(k) Summary ### 1.) Safety and effectiveness as required by 21 CFR 807.92 STATEMENT This summary of the 510(k) safety and effectiveness information is being submitted in accordance with the requirement 21 CFR 807,92. ### 2.) Submitter's / Owner of the 510(k) and Contact Information Name: JAS Diagnostics Address: 14100 N.W. 57th Court Miami Lakes, FL 33014 Phone: 305 418-2320 Fax: 305 418-2321 ### Official Correspondent David Johnston E-mail: djohnston@jasdiagnostics.com Date of Preparation April 14, 2014 510(k) Application Number K130915 3.) 510k Number, Device, Proprietary Name, Common Name, Purpose for Submission, Regulatory Classification, Panel, Product Code And CFR Number | Regulation Number: | Device Description | |-----------------------|----------------------------------------------------------| | 21CFR Sec. - 862.2160 | Discrete Photometric Chemistry Analyzer for Clinical Use | | 21CFR Sec. - 862.1345 | Glucose Test System | | 21CFR Sec. - 862.1665 | Sodium Test System | | 21CFR Sec. - 862.1600 | Potassium Test System | | 21CFR Sec. - 862.1170 | Chloride Test System | {1}------------------------------------------------ MOT A : YAM #### Classifications: Class II and Class I (Analyzer) ### Product Codes - JJE Analyzer, Chemistry (Photometric, Discrete), for Clinical Use - CFR Hexokinase, Glucose JGS - Electrode, Ion Specific, Sodium CEM - Electrode, Ion Specific, Potassium - CGZ Electrode, Chloride #### Panel Chemistry (75) ### 4.) Proprietary Name and Common Name #### Trade/Proprietary Name XL-200 Clinical Chemistry Analyzer, JAS Glucose Reagent, ISE Reagent Pack #### Common Name/Usual Name XL-200 Clinical Chemistry Analyzer, JAS Glucose Reagent, ISE Reagent Pack ### 5.) Indications for Use The XL-200 Clinical Chemistry Analyzer is an automated, random access. computer controlled, bench top clinical chemistry analyzer for clinical chemistry tests. The instrument provides in vitro quantitative measurements for glucose, sodium, potassium, and chloride in serum. This device is intended for clinical laboratory use. The JAS Glucose Reagent in intended for the in vitro quantitative measurement of glucose in serum on the XL-200 Clinical Chemistrv Analyzer. This device is intended for clinical laboratory use. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders, including diabetes mellitus, neonatal hypoglycemia and of pancreatic islet cell carcinoma. The ISE Reagent Pack is intended for the in vitro quantitative measurement of sodium, potassium, and chloride concentrations in serum on the XL-200 Clinical Chemistry Analyzer. This device is intended for clinical laboratory use. Sodium measurements are used in the diagnosis and treatment of aldosteronism, diabetes insipidus and other diseases involving electrolyte imbalance. {2}------------------------------------------------ Potassium measurements are used to monitor electrolyte balance in the diagnosis and treatment of disease conditions characterized by low or high blood potassium levels. Chloride measurements are used in the diagnosis and treatment of electrolyte and metabolic disorders such as cystic fibrosis and diabetic acidosis. ### 6.) Device Description: JAS XL-200 Clinical Chemistry Analyzer The XL-200 Clinical Chemistry Analyzer is an automated bench top, random access, open analyzer for clinical chemistry and immunoturbidimetric analysis on serum, urine, and other body fluids. The analyzer mainly uses colorimetric, turbidimetric, and ion selective electrode methods for analysis of samples. The instrument includes of the following main parts; - . Sampling Arm (for sample addition to cuvettes) - Reagent Arm (for reagent addition to cuvettes) . - Reaction Station (cuvettes) . - Sample Plate Station (for loading samples) . - Reagent Plate Station (for on board reagents) . - . Photometer (for reaction analysis reading) - Wash Station (for cleaning of reaction cuvettes) . - Electronic Boards (for controlling the open functions) . ### 7.) Substantial Equivalence Information: - 1.) Predicate 510(k) number: K981743 - 2.) Olympus AU400 Clinical Chemistry Analyzer: - 3.) Comparison to predicate: | | Device Name | Device:<br>XL-200<br>Clinical Chemistry<br>Analyzer | Predicate Device:<br>OLYMPUS AU400<br>Clinical Chemistry<br>Analyzer | |---|-------------|-----------------------------------------------------|----------------------------------------------------------------------| | 1 | 510(k) | Pending | K981743 | {3}------------------------------------------------ | 2 | System Principles | | | |----|-----------------------------|-----------------------------------------------------------------|--------------------------------------------------------------------------------------------------------| | | QC/Calibration | Automatic and Manual | Automatic and Manual | | | LIS external connectivity | YES | YES | | | Barcode | Samples | Samples | | | | Reagents | Reagents | | | Re-dilution on abnormals | YES | YES | | 3 | Throughput | 200 test/hour | 400 test/hour | | | | 400 with ISE | 800 with ISE | | 4 | Principle of analysis | | | | | Mode of detection | Photometric / ISE | Photometric / ISE | | | Analytical Methods | Endpoint, Kinetic, Ion<br>Selective Electrodes | Endpoint, Kinetic, Ion<br>Specific Electrodes | | | Calibration | Linear and non linear | Linear and non linear | | 5 | Optical<br>measurement unit | | | | | Measurement modes | Absorbance | Absorbance | | | Optical modes | Bichromatic,<br>turbidimetric | Bichromatic, turbidimetric | | | Wavelengths | 340nm, 405nm,<br>505nm, 546nm,<br>578nm, 600nm,<br>660nm, 700nm | 340nm, 380nm, 410nm,<br>450 nm, 480nm, 520nm,<br>540nm, 570nm, 600nm,<br>660nm , 700nm, 750nm<br>800nm | | | Linear absorbance range | 0.0 to 2.5 A | 0 to 2.5 A | | | Light source | halogen lamp | halogen lamp | | | Detector | Silicone photodiode<br>array | Silicon photodiodes | | | | | | | 6 | Reaction unit | | | | | Reaction cuvettes | 45 quartz semi-disposable | 88 quartz semi-disposable | | | Reaction volume | 180 to 570 uL | 180 to 610 uL | | | Path length | Calculated at 10mm | 10mm conversion | | | Temperature | Room, 30°C, 37°C | Room, 30°C, 37°C | | 7 | Sample and<br>reagent unit | | | | | Reagent positions | 25 for Rgt1, 50mL<br>bottles<br>25 for Rgt2, 20mL<br>bottles | 38 for Rgt1, 60mL bottles<br>38 for Rgt2, 15 & 30mL<br>bottles | | | Sample positions | 39 | 48 | | | Pipetter system | plunger, stepper motor<br>driven | Micro-syringe, motor<br>driven | | | Mixing | Immersion mixing by<br>rotating mixers | Rotating mix bar | | | Reagent refrigeration | YES | YES | | | Sample dispensing | 2 - 70 microliters | 2 - 50 microliters | | 8 | Power | AC220V. 50 Hz or<br>110v, 60Hz | AC210V, 60Hz (U.S.) | | 9 | Environmental | 15 to 30°C,<br>Humidity 40 to 80% | 15 to 30°C,<br>Humidity 40 to 80% | | | conditions | Humidity 40 to 80% | Humidity 40 to 80% | | | | | | | 10 | ISE Principle | Ion selective, direct<br>measurement | Ion selective, direct<br>measurement | | 11 | ISE sample type | Serum | Serum, Urine | | 12 | Available tests | Na, K, Cl | Na, K, Cl | | 13 | ISE calibration | Two-point and single point | Two-point and single point | | 14 | Sample size | 70 uL | 70 uL | : : ・ : {4}------------------------------------------------ • ・ . . . . . . . · . . : ર {5}------------------------------------------------ ### 8.) Device Description: JAS Glucose Reagent The JAS Glucose Reagent is intended for the quantitative measurement of glucose in serum on the XL-200 Clinical Chemistry Analyzer. The Reagent is a single vial liquid that is placed for use on the XL-200 Clinical Chemistry Analyzer reagent carousel. The reagent uses the enzymatic (Hexokinase/G-6-P) UV (340nm) method. This device is for clinical laboratory use. ### 9.) Substantial Equivalence Information: Predicate 510(k) number: K981743 - Olympus AU400 Clinical Chemistry Analyzer, Glucose Reagent ### 10.) Comparison to predicate: | Manufacturer: | JAS Diagnostics | Beckman Coulter Olympus | |----------------|------------------------------------------------------------------|-----------------------------------------------------------------------------------| | Device Name | JAS Glucose Reagent for<br>XL-200 Clinical<br>Chemistry Analyzer | Olympus Glucose Reagent for<br>Olympus AU 400 & 600<br>series Chemistry Analyzers | | | JAS Diagnostics | Beckman Coulter (Olympus) | | FDA 510k # | K130915 (CFR)<br>(Pending) | K981743 (CFR) | | Classification | Moderate<br>(Pending) | Moderate | {6}------------------------------------------------ | Intended Use | quantitative measurement<br>of glucose in serum<br>on the XL-200 Clinical<br>Chemistry analyzer<br>For clinical laboratory use | quantitative measurement<br>of glucose in serum, plasma<br>urine, on Beckman Coulter<br>clinical chemistry analyzers<br>For clinical laboratory use | |-----------------------|--------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------| | Methodology | enzymatic (Hexokinase /<br>G-6-P), UV (340nm) | enzymatic (Hexokinase /<br>G-6-P), UV (340nm) | | Reagent Description | Single vial liquid reagent | Two vial liquid reagent | | Normal Range<br>Adult | 74 - 106 mg/dL | 70 - 105 mg/dL | ### 11.) Performance Characteristics - Glucose . #### Linearity Protocol: 9 Serial Dilutions of a 500 mg/dL Glucose Standard were prepared and run on the XL-200, along with the 500 mg/dL stock Standard. Results were then plotted as a regression graph against theoretical recovery. Results: Acceptable Slope 1.001 and Intercept of -0.1mg/dL, and data linear within allowable nonlinearity of 3mg/dL or 5% resulted; validating the product's claimed linearity range of 9 to 500mg/dL. ### Calibrator Traceability: Protocol: Verichem Laboratories Matrix Plus Chemistry Reference Kit (9500) five level reference material was run on the XL-200 Clinical Chemistry Analyzer, using JAS Glucose Reagent. This material is "Verified and lot certified using available standard reference materials for the National Institute of Standards and Technology (NIST)." #### Results: Program's Evaluation of glucose Results states accuracy test passed and results linear; with a slope of 0.972 and intercept of 2.1mg/dL, with error of 1.4%. #### Interferences Protocol: Serial dilutions were prepared using a high interferent sample material, with the same sample material (glucose level) without the interferent spiked. Acceptable interference level is the is the highest interferent sample level were the glucose values remains within 10% of the sample value unspiked with the intereferent. Results: The following highest interferent levels for the sample, with < 10% interferent were obtained. 7 {7}------------------------------------------------ | Hemoglobin | up to 400 mg/dL | |---------------|------------------| | Bilirubin | up to 21.4 mg/dL | | Lipemia | up to 412 mg/dL | | Ascorbic Acid | up to 10.0mg/dL | ### Correlation Protocol: Multiple serum samples, at various glucose levels, were run on multiple days through the test's analytical range on the XL-200 clinical chemistry analyzer, using JAS Glucose reagent and also on the Olympus 400, using Olympus Glucose reagent. Results were compared using a correlation plot. Results: The following acceptable statistics resulted, i.e. R > 0.95, Slope between 0.90 and 1.10, and intercept close to 0 mg/dL level. JAS Glucose reagent on the XL-200 Clinical Chemistry Analyzer versus Olympus glucose reagent of Olympus 400: | Number of sample pairs: | 103 | |-------------------------|-----------------| | Range tested: x method | 11 to 497 mg/dL | | y method | 9 to 501 mg/dL | | Corr. Coef. (R) | 0.9970 | | Slope | 1.002 | | Intercept | 1.8 mg/dL | ### Precision/Reproducibility: Protocol: Precision was performed using "EP Evaluator's NCCLS EP5-T2 Precision document (NCCLS.EP5)" program. Three glucose sample levels were run 20 times in duplicate, using 20 runs over a 20 day period on the XL-200 Clinical Chemistry Analyzer, for within run and for total imprecision calculation/determination. Results: Results were entered into Rhoads EP Evaluator Software for determining precision S.D. and CV levels. All results below meet requirement of CV (%) < 5% for both within and total precision. | Glucose | | | | |--------------|---------|---------|---------| | Within Run | Level 1 | Level 2 | Level 3 | | Mean (mg/dL) | 49.8 | 96.2 | 303.4 | | S.D. (mg/dL) | 0.9 | 1.1 | 3.6 | | C.V. (%) | 1.7 | 1.2 | 1.2 | | Total | | | | | S.D. (mg/dL) | 1.3 | 1.8 | 3.8 | | C.V. (%) | 2.6 | 1.8 | 1.3 | {8}------------------------------------------------ ### 12.) Device Description: ISE Reagent Pack (for use on the XL-200 Clinical Chemistry Analyzer) | | | XL-200<br>Clinical Chemistry<br>Analyzer | Beckman Coulter Olympus<br>AU400 | |---|-----------------|------------------------------------------|--------------------------------------| | 1 | ISE Principle | Ion selective, direct<br>measurement | Ion selective, direct<br>measurement | | 2 | ISE sample type | Serum | Serum, Urine | | 3 | Available tests | Na, K, Cl | Na, K, Cl | | 4 | ISE calibration | Two-point and single<br>point | Two-point and single point | | 5 | Sample size | 70 uL | 70 uL | ### Comparison to Beckman Coulter Olympus 400: Device Description: The JAS ISE Module consists of ion selective electrodes for sodium, potassium, and chloride, a reference electrode and accessory reagents. XL-200 Clinical Chemistry Analyzer ISE Solutions: *ISE Reagent Pack: REF: 5423-0030: Calibrant A, 520mL: (Na 140.0, K 4.00, Cl 125.0 mmol/L) Calibrant B, 190mL: (Na 70.0, K 8.00, Cl 41.0 mmol/L) ### *REF 5421 Cleaning Solution ### Beckman Coulter Olympus ISE Solutions: ISE Buffer, PN: AUH1011 ISE Reference, PN: AUH1013 ISE Mid Standard, PN: AUH1012 Internal Reference Solution, PN: AUH1017 ISE Low Standard, PN: 1014 ISE High Standard, PN: 1015 Na Selectivity Check, PN: AUH1018 {9}------------------------------------------------ XL-200 Clinical Chemistry Analyzer; Electrodes: *REF 5201: Na Electrode *REF 5202: K Electrode *REF 5207: Cl Electrode *REF 5204: Reference Electrode ### 14.) Substantial Equivalence Information for ISE Reagent Pack: 1.) Predicate 510k number: K981743 - 2.) Manufacturer: Beckman Coulter (Olympus) - 3.) Comparison to predicate: See below comparison. | Device Name | JAS<br>ISE Reagent Pack for<br>XL-200 Clinical Chemistry<br>Analyzer | OLYMPUS<br>ISE Reagents for<br>Olympus AU 400 & 600<br>series chemistry analyzers | |---------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Manufacturer | JAS Diagnostics | Beckman Coulter (Olympus) | | FDA 510k # | K130915<br>(Pending) | K981743 | | | JAS | OLYMPUS | | Classification | Moderate<br>(Pending) | Moderate | | Intended Use | quantitative measurement<br>of sodium, potassium and<br>chloride concentrations<br>in serum on the XL-200<br>clinical chemistry<br>analyzer.<br>For clinical laboratory use | quantitative measurement<br>of sodium, potassium and<br>chloride in serum and urine<br>on Beckman Coulter<br>clinical chemistry analyzers.<br>For clinical laboratory use | | Methodology | Ion Selective Electrodes<br>(ISEs) | Ion Selective Electrodes<br>(ISEs) | | Reagent Description | *ISE Reagent Pack:<br>Calibrant A, 520mL:<br>(Na 140.0, K 4.00, Cl 125.0 mmol/L)<br><br>Calibrant B, 190mL:<br>(Na 70.0, K 8.00, Cl 41.0 mmol/L)<br><br>*REF 5421 Cleaning Solution | ISE Buffer, PN: AUH1011<br>ISE Reference, PN: AUH1013<br>ISE Mid Standard,<br>(PN:AUH1012)<br>Internal Reference Solution,<br>(PN: AUH1017)<br>ISE Low Standard, PN: 1014<br>ISE High Standard, PN: 1015<br>Na Selectivity Check, PN: AUH1018 | {10}------------------------------------------------ ### ELECTRODES XL-200 Clinical Chemistry Analyzer *REF 5201: Na Electrode *REF 5202: K Electrode *REF 5207: Cl Electrode *REF 5204: Reference Electrode Beckman Coulter Olympus 400 Electrodes MU9194: Na Electrode MU9195: K Electrode MU9196: Cl Electrode JAS Expected values/Reference ranges: Sodium: 136 to 145mmol/L Potassium: 3.5 to 5.1mmol/L Chloride: 98 to 107 mmol/L Olympus Expected values/ Reference ranges: 136 to 145mEq/L 3.5 to 5.1mEq/L 98 to 107 mEq/L ### 15.) Performance Characteristics - ISE #### Linearity (serum) Protocol: Serial Dilutions of a Stock Standard were prepared for Sodium, Potassium and Chloride and run on the XL-200 Clinical Chemistry Analyzer. Results were then plotted as a regression graph against theoretical recovery. Results: Slopes and Intercepts resulted for Sodium, Potassium and Chloride, linear within allowable nonlinearity of 3mg/dL or 5%; validating the products' below claimed linearity ranges. mEq/L mEq/L | Sodium | 100 to 200 mEq/L | |-----------|-------------------| | Potassium | 1.0 to 10.0 mEq/L | | Chloride | 50 to 150 mEq/L | ### Calibrator Traceability: Protocol: Verichem Laboratories Matrix Plus Chemistry Reference Kit (9500) five level reference material was run on the XL-200 Clinical Chemistry Analyzer, for XL-200 Sodium, Potassium, and Chloride. This material is "Verified and lot certified using available standard reference materials for the National Institute of Standards and Technology (NIST); for Calibration Verification" #### Results: Program's Evaluation of Results states Sodium, Potassium, and Chloride accuracy tests passed and results linear; with the following statistics: {11}------------------------------------------------ | Test | Slope | Intercept | Error | |-----------|-------|-----------|-------| | Sodium | 1.007 | 4.4mmol/L | 0.1% | | Potassium | 0.902 | 0.1mmol/L | 1.3% | | Chloride | 1.012 | 2.0mmol/L | 1.3% | ### Precision: Protocol: Precision was performed using "EP Evaluator's NCCLS EP5-T2 Precision document (NCCLS.EP5)" program. Three sodium, potassium and chloride sample levels were run 20 times in duplicate, using 20 runs over a 20 day period on the XL-200 Clinical Chemistry Analyzer, for within run and for total imprecision calculation/determination. Results: Results were entered into Rhoads EP Evaluator Software for determining precision S.D. and CV levels. All results below meet requirement of CV (%) < 5% for both within and total precision. | Sodium | | | | |---------------|---------|---------|---------| | Within Run: | Level 1 | Level 2 | Level 3 | | Mean (mmol/L) | 118.2 | 131.5 | 166.6 | | S.D. (mmol/L) | 0.340 | 0.40 | 0.82 | | C.V. (%) | 0.3 | 0.3 | 0.5 | | Total: | | | | | S.D. (mmol/L) | 0.808 | 1.08 | 1.36 | | C.V. (%) | 0.7 | 0.8 | 0.8 | | Potassium | | | | | Within Run: | Level 1 | Level 2 | Level 3 | | Mean (mmol/L) | 1.84 | 3.43 | 6.13 | | S.D. (mmol/L) | 0.035 | 0.011 | 0.038 | | C.V. (%) | 1.9 | 0.3 | 0.6 | | Total: | | | | | S.D. (mmol/L) | 0.038 | 0.026 | 0.066 | | C.V. (%) | 2.1 | 0.8 | 1.1 | | Chloride | | | | | Within Run: | Level 1 | Level 2 | Level 3 | | Mean (mmol/L) | 88.3 | 108.5 | 120.0 | | S.D. (mmol/L) | 0.45 | 0.68 | 0.301 | | C.V. (%) | 0.5 | 0.6 | 0.3 | | Total: | | | | | S.D. (mmol/L) | 0.73 | 0.80 | 0.559 | {12}------------------------------------------------ | C.V. (%) | 0.8 | 0.7 | 0.5 | |----------|-----|-----|-----| |----------|-----|-----|-----| ### Interferences Protocol: Serial dilutions were prepared using a high interferent sample material, with the same sample material (sodium, potassium and chloride level respectively) without the interferent spiked. Acceptable interference level is the is the highest interferent sample level were the glucose values remains within 10% of the sample value unspiked with the intereferent. Results: The following highest interferent levels for the sample, with < 10% interferent were obtained. | Reagent | Hemolysis | Lipemia | Bilirubin | |-----------|--------------------------------------|------------|------------| | Sodium | Hemolyzed samples should not be used | 1084 mg/dL | 22.5 mg/dL | | Potassium | Hemolyzed samples should not be used | 1084 mg/dL | 22.5 mg/dL | | Chloride | Hemolyzed samples should not be used | 1084 mg/dL | 22.5 mg/dL | Method Comparison Studies: (versus Olympus reagent on Olympus 400 analyzer) Protocol: Multiple serum samples, at various analyte levels, were run on multiple days through the test's analytical range on the XL-200 clinical chemistry analyzer using JAS reagent packs (y method) and also on the Beckman Coulter Olympus 400, using Olympus ISE solution as a reference instrument (y method). Results were compared using a correlation plot. Results: The following acceptable statistics resulted, i.e. R > 0.95, Slope between 0.90 and 1.10, and intercept close to 0 mg/dL level ### Sodium | Number of sample pairs: | 101 | |-------------------------|-------------------| | Range tested: x method | 103 to 193 mmol/L | | Range tested: y method | 101 to 190 mmol/L | | Correlation Coefficient | 0.9917 | | Slope: | 0.984 | | Intercept: | 2.256 mmol/L | ### Potassium | Number of sample pairs: | 90 | |-------------------------|--------------------| | Range tested: x method | 1.2 to 9.90 mmol/L | | Range tested: y method | 1.1 to 10.0 mmol/L | | Correlation Coefficient | 0.9954 | | Slope: | 0.999 | | Intercept: | 0.05 mmol/L | {13}------------------------------------------------ ### Chloride 88 Number of sample pairs: 51 to 140 mmol/L Range tested: x method Range tested: y method 51 to 145 mmol/L Correlation Coefficient 0.9856 Slope: 1.070 - 4.3 mmol/L Intercept: > The submitted information in this pre-market notification is complete and supports a substantial equivalence decision. I . {14}------------------------------------------------ #### DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/14/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is an abstract image of an eagle with three lines representing its wings. Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G60 Silver Spring, MI) 20993-00002 May 15, 2014 JAS DIAGNOSTICS. INC. DAVID JOHNSTON VP. RESEARCH AND DEVELOPMENT 14100 N.W. 57TH COURT MIAMI LAKES FL 33014 Re: K130915 Trade/Device Name: JAS Glucose Reagent ISE Reagent Pack XL-200 Clinical Chemistry Analyzer Regulation Number: 21 CFR 862.1345 Regulation Name: Glucose test system Regulatory Class: II Product Code: CFR, JGS. CEM. CGZ. JJE Dated: April 30, 2014 Received: May 2. 2014 Dear Mr. David Johnston: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976. the enactment date of the Medical Device Amendments. or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH docs not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA 's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Parts 801 and 809): medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the {15}------------------------------------------------ Page 2-Mr. David Johnston electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-five number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.lda.gov/MedicalDevices/Safety/ReportalProblem/defiult.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resources/or You/Industry/default.htm. Sincerely yours, ## Courtney H. Lias -S Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {16}------------------------------------------------ #### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration ### Indications for Use 510(k) Number (if known) k130915 #### Device Name XL-200 Clinical Chemistry Analyzer, JAS Glucose Reagent, ISE Reagent Pack #### Indications for Use (Describe) The XL-200 Clinical Chemistry Analyzer is an automated random access, computer controlled, bench top, clinical analyzer for clinical chemistry tests. The instrument provides in vitro quantitative measurements for glucose, sodium, potassium and chloride in serum. This device is intended for clinical laboratory use. The JAS Glucose Reagent is intended for the in vitro quantitative measurement of glucose in serum on the XL-200 clinical chemistry analyzer. This device is intended for clinical laboratory use. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia and idiopathic hypoglycemia and of pancreatic islet cell Carcinoma. The ISE Reagent Pack is intended for the in vitro quantitative measurement of sodium, and chloride concentrations in serum on the XL-200 clinical chemistry analyzer. This device is intended for clinical laboratory use. Sodium measurements are used in the diagnosis and treatment of aldosteronism, diabetes insipidus and other diseases involving electrolyte imbalance. Potassium measurements are used to monitor electrolyte balance in the diagnosis and treatment of disease conditions characterized by low or high blood potassium levels. Chloride measurements are used in the diagnosis and treatment of electrolyte and metabolic disorders such as cystic fibrosis and diabetic acidosis. Type of Use (Select one or both, as applicable) 2 Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) . ‹ • : . ### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED. FOR FDA USE ONLY : Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) . . {17}------------------------------------------------ This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW." The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...