VITROS CHEMISTRY PRODUCTS TIBC KIT, MODEL 886 7541; CALIBRATOR KIT 4, MODEL 120 4668
K061464 · Ortho-Clinical Diagnostics, Inc. · JMO · Jun 15, 2006 · Clinical Chemistry
Device Facts
Record ID
K061464
Device Name
VITROS CHEMISTRY PRODUCTS TIBC KIT, MODEL 886 7541; CALIBRATOR KIT 4, MODEL 120 4668
Applicant
Ortho-Clinical Diagnostics, Inc.
Product Code
JMO · Clinical Chemistry
Decision Date
Jun 15, 2006
Decision
SESE
Submission Type
Special
Regulation
21 CFR 862.1415
Device Class
Class 1
Indications for Use
VITROS Chemistry Products TIBC Kit with VITROS Chemistry Products Fe Slides quantitatively measures total iron-binding capacity (TIBC) in serum. The iron binding capacity is useful in the differential diagnosis of anemia, iron deficiency anemia, thalassemia, sideroblastic anemia, and iron poisoning. VITROS Chemistry Products Calibrator Kit 4 is used to calibrate VITROS Chemistry Systems for the quantitative measurement of ALB, BuBc, Fe, TBIL, TIBC, and TP. For in vitro diagnostic use only.
Device Story
VITROS TIBC assay measures total iron-binding capacity in serum samples. Process involves pretreating sample with excess iron citrate to saturate apotransferrin sites; applying sample to alumina column to adsorb unbound iron; depositing eluate onto VITROS Fe Slide. System performs two reflection density measurements at 600 nm at one and five minutes. Software uses two-point rate math model to calculate TIBC based on reflectance change. Used in clinical laboratory settings by trained personnel. Output provides quantitative TIBC values to assist physicians in diagnosing anemia-related conditions and iron poisoning.
Clinical Evidence
Bench testing only; verification and validation activities performed per design control requirements to confirm performance claims following internal procedural updates.
Technological Characteristics
In vitro diagnostic reagent kit for TIBC analysis. Fundamental scientific technology unchanged. Performance claims updated based on internal calibrator value assignment and reportable range procedures.
Indications for Use
Indicated for in vitro diagnostic quantitative measurement of TIBC in serum to aid in differential diagnosis of anemia, iron deficiency anemia, thalassemia, sideroblastic anemia, and iron poisoning.
Regulatory Classification
Identification
An iron-binding capacity test system is a device intended to measure iron-binding capacity in serum. Iron-binding capacity measurements are used in the diagnosis and treatment of anemia.
K970248 — CANTROL SERUM IRON AND TOTAL IRON BINDING CAPACITY TEST SYSTEM · Canyon Diagnostics, Inc. · Feb 11, 1997
Submission Summary (Full Text)
{0}
SPECIAL 510(k): Device Modification
ODE Review Memorandum (Decision Making Document is Attached)
To: THE FILE
RE: DOCUMENT NUMBER K041464
This 510(k) submission contains information/data on modifications made to the SUBMITTER'S own Class II, Class III or Class I devices requiring 510(k). The following items are present and acceptable (delete/add items as necessary):
1. The name and 510(k) number of the SUBMITTER'S previously cleared device
2. Submitter's statement that the INDICATION/INTENDED USE of the modified device as described in its labeling HAS NOT CHANGED along with the proposed labeling which includes instructions for use, package labeling, and, if available, advertisements or promotional materials (labeling changes are permitted as long as they do not affect the intended use).
3. A description of the device MODIFICATIONS, including user's and/or service manuals in sufficient detail to demonstrate that the FUNDAMENTAL SCIENTIFIC TECHNOLOGY of the modified device has not changed.
This change was for internal updates (i.e., value assignment procedure of TIBC in calibrator and procedures for determining reportable range) and updated performance claims for the TIBC Analysis using the VITROS Chemistry Products TIBC Kit.
4. Comparison Information (similarities and differences) to applicant's legally marketed predicate device including, labeling, intended use, and physical characteristics.
5. A Design Control Activities Summary which includes:
a) Identification of Risk Analysis method(s) used to assess the impact of the modification on the device and its components, and the results of the analysis
b) Based on the Risk Analysis, an identification of the verification and/or validation activities required, including methods or tests used and acceptance criteria to be applied
c) A declaration of conformity with design controls. The declaration of conformity should include:
i) A statement signed by the individual responsible, that, as required by the risk analysis, all verification and validation activities were performed by the designated individual(s) and the results demonstrated that the predetermined acceptance criteria were met, and
ii) A statement signed by the individual responsible, that the manufacturing facility is in conformance with design control procedure requirements as specified in 21 CFR 820.30 and the records are available for review.
6. A Truthful and Accurate Statement, a 510(k) Summary and the Indications for Use Enclosure.
The labeling for this modified subject device has been reviewed to verify that the indication/intended use for the device is unaffected by the modification. In addition, the submitter's description of the particular modifications and the comparative information between the modified and unmodified devices demonstrate that the fundamental scientific technology has not changed. The submitter has provided the design control information as specified in The New 510(k) Paradigm and on this basis, I recommend the device be determined substantially equivalent to the previously cleared (or their preamendment) device.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.