IMMULITE 2000 Third Generation TSH, IMMULITE 2000 Free T4

K160647 · Siemens Healthcare Diagnostics, Inc. · JLW · May 31, 2016 · Clinical Chemistry

Device Facts

Record IDK160647
Device NameIMMULITE 2000 Third Generation TSH, IMMULITE 2000 Free T4
ApplicantSiemens Healthcare Diagnostics, Inc.
Product CodeJLW · Clinical Chemistry
Decision DateMay 31, 2016
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1690
Device ClassClass 2
AttributesPediatric

Indications for Use

For in vitro diagnostic use with the IMMULITE® 2000 Systems Analyzers - for the quantitative measurement of thyrotropin (TSH) in serum. Measurements of thyroid stimulating hormone produced by the anterior pituitary are used in the diagnosis of thyroid or pituitary disorders. For in vitro diagnostic use with the IMMULITE® 2000 Systems Analyzers — for the quantitative measurement of nonprotein-bound thyroxine (free T4) in serum and heparinized plasma. Measurements of free thyroxine are used in the diagnosis and treatment of thyroid disease.

Device Story

IMMULITE 2000 systems perform quantitative in vitro diagnostic measurements of TSH and free T4 in serum/plasma samples. Devices utilize automated immunoassay technology to measure hormone levels; results are provided to clinicians to assist in diagnosing and treating thyroid or pituitary disorders. This submission updates the Instructions for Use (Package Inserts) to include established pediatric reference intervals for infants, children, and adolescents. No changes were made to the device design, manufacturing, or analytical performance characteristics. The system remains intended for professional use in clinical laboratory settings.

Clinical Evidence

No clinical diagnostic studies performed. Evidence consists of a reference range study involving 426 pediatric samples for FT4 and 433 pediatric samples for TSH, collected across 8 U.S. sites. Reference intervals were established using non-parametric methods for children/adolescents (n>120) and robust symmetrical methods for infants. 90% confidence intervals provided for infant reference limits.

Technological Characteristics

Solid-phase chemiluminescent immunoassay. Reagents: monoclonal murine anti-T4/anti-TSH antibodies, alkaline phosphatase-conjugated T4/anti-TSH. Solid phase: coated beads. Detection: chemiluminescent substrate. System: IMMULITE 2000 automated analyzer. Software: embedded system control. Sterilization: not applicable (reagents).

Indications for Use

Indicated for quantitative measurement of TSH and free T4 in serum and heparinized plasma (FT4 only) to aid in diagnosis of thyroid or pituitary disorders in pediatric (1 month to 20 years) and adult populations.

Regulatory Classification

Identification

A thyroid stimulating hormone test system is a device intended to measure thyroid stimulating hormone, also known as thyrotrophin and thyrotrophic hormone, in serum and plasma. Measurements of thyroid stimulating hormone produced by the anterior pituitary are used in the diagnosis of thyroid or pituitary disorders.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} 1 # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION MEMORANDUM ASSAY ONLY TEMPLATE A. 510(k) Number: k160647 B. Purpose for Submission: Adding pediatric reference intervals to previously cleared assays (cleared under k083373 and k970227) C. Measurand: Free thyroxine (FT4) and thyroid stimulating hormone (TSH) D. Type of Test: Quantitative, chemiluminescence immunoassay E. Applicant: Siemens Healthcare Diagnostics Inc. F. Proprietary and Established Names: IMMULITE 2000 Third Generation TSH IMMULITE 2000 Free T4 G. Regulatory Information: 1. Regulation section: 21 CFR § 862.1690, Thyroid stimulating hormone test system 21 CFR § 862.1695, Free thyroxine test system 2. Classification: Class II 3. Product code: JLW, CEC {1} 4. Panel: Clinical Chemistry (75) H. Intended Use: 1. Intended use(s): See Indications(s) for use below 2. Indication(s) for use: For in vitro diagnostic use with the IMMULITE 2000 Systems Analyzers for the quantitative measurement of thyrotropin (TSH) in serum. Measurements of thyroid stimulating hormone produced by the anterior pituitary are used in the diagnosis of thyroid or pituitary disorders. For in vitro diagnostic use with the IMMULITE 2000 Systems Analyzers for the quantitative measurement of non-protein-bound thyroxine (free T4) in serum and heparinized plasma. Measurements of free thyroxine are used in the diagnosis of thyroid disease. 3. Special conditions for use statement(s): For prescription use 4. Special instrument requirements: IMMULITE 2000 Systems I. Device Description: IMMULITE 2000 Free T4 set consists of the following reagents: Free T4 Bead Pack – 200 beads coated with monoclonal murine anti-T4 antibody. Free T4 Reagent Wedge – 11.5 mL alkaline phosphatase (bovine calf intestine) conjugated to T4 in buffer. Free T4 Adjustors – Two vials (low and high) of lyophilized free T4 in processed human serum, with preservative. IMMULITE 2000 Third Generation TSH set consists of the following reagents: TSH Bead Pack – 200 beads coated with monoclonal murine anti-TSH antibody. TSH Reagent Wedge – 23 mL alkaline phosphatase (bovine calf intestine) conjugated to TSH in buffer (with preservative). TSH Adjustors – Two vials (low and high) of lyophilized human TSH in a serum/buffer matrix. TSH Sample Diluent – 25 mL of concentrated (ready-to-use) TSH-free serum/buffer matrix. 2 {2} J. Substantial Equivalence Information: 1. Predicate device name(s): IMMULITE 2000 Free T4 IMMULITE 2000 Third Generation TSH 2. Predicate 510(k) number(s): k083373, k970227 3. Comparison with predicate: FT4 assay: | Similarities and Differences | | | | --- | --- | --- | | Item | Candidate device IMMULITE 2000 Free T4 (k160647) | Predicate device IMMULITE 2000 Free T4 (unmodified) (k083373) | | Intended Use | For in vitro diagnostic use with the IMMULITE® 2000 Systems Analyzers — for the quantitative measurement of non-protein-bound thyroxine (free T4) in serum and heparinized plasma. | Same | | Analyte | Free Thyroxine (T4) | Same | | Instruments | IMMULITE 2000 Systems | Same | | Analytical Measuring Range (Assay Range) | 0.30 – 6.0 ng/dL (3.9 – 77.2 pmol/L) | Same | | Infants (1-23 months) Reference Intervals | 0.80 – 1.27 ng/dL (10.3 – 16.3 pmol/L) | None | | Children (2-12 years) Reference Intervals | 0.74 – 1.28 ng/dL (9.5 – 16.5 pmol/L) | None | | Adolescents (13-20 years) Reference Intervals | 0.75 – 1.27 ng/dL (9.7 – 16.3 pmol/L) | None | | Euthyroid Adults (≥21 yrs) Reference Intervals | 0.89 – 1.76 ng/dL (11.5 – 22.7 pmol/L) | Same | {3} TSH assay: | Similarities and Differences | | | | --- | --- | --- | | Item | Candidate device IMMULITE 2000 Third Generation TSH (k160647) | Predicate device IMMULITE 2000 Third Generation TSH (unmodified) (k970227) | | Intended Use | For in vitro diagnostic use with the IMMULITE® 2000 Systems Analyzers — for the quantitative measurement of thyrotropin (TSH) in serum. | Same | | Analyte | Thyroid stimulating hormone (TSH) | Same | | Instruments | IMMULITE 2000 Systems | Same | | Analytical Measuring Range (Assay Range) | 0.004 – 75 μIU/mL (mIU/L) | Same | | Infants(1-23 months) Reference Intervals | 0.83 – 6.5 μIU/mL (mIU/L) | None | | Children(2-12 years) Reference Intervals | 0.58 – 4.1 μIU/mL (mIU/L) | None | | Adolescents (2-12 years) Reference Intervals | 0.39 – 4.0 μIU/mL (mIU/L) | None | | Euthyroid Adults (≥21 yrs) Reference Intervals | 0.4073 – 4.0 μIU/mL (mIU/L) | Same | # K. Standard/Guidance Document Referenced (if applicable): CLSI C28-A3c: Defining, Establishing and Verifying Reference Intervals in the Clinical Laboratory; Approved Guideline - Third Edition # L. Test Principle: IMMULITE 2000 Free T4: IMMULITE 2000 Free T4 is a solid-phase, enzyme-labeled chemiluminescent competitive immunoassay. The solid phase (bead) is coated with monoclonal murine anti-T4 antibody. The liquid phase consists of alkaline phosphatase (bovine calf intestine) conjugated to T4. The patient sample and the reagent are first incubated together with the coated bead. During this time, free T4 in the sample competes with enzyme conjugated T4 in the reagent for a limited number of antibody binding sites on the bead. Unbound patient sample and enzyme conjugate are then removed by centrifugal washes. Finally, chemiluminescent substrate is added to the reaction tube containing the bead and the {4} signal is generated in proportion to the bound enzyme. IMMULITE 2000 Third Generation TSH: IMMULITE 2000 Third Generation TSH is a solid-phase, two-site chemiluminescent immunometric assay. The solid phase (bead) is coated with monoclonal murine anti-TSH antibody. The liquid phase consists of alkaline phosphatase (bovine calf intestine) conjugated to polyclonal goat anti-TSH antibody. The patient sample and the reagent are incubated together with the coated bead for 60 minutes. During this time, TSH in the sample binds to enzyme conjugated anti-TSH antibody in the reagent, and then this complex is captured by the anti-TSH antibody on the bead. Unbound patient sample and enzyme conjugate are then removed by centrifugal washes. Finally, chemiluminescent substrate is added to the reaction tube containing the bead and the signal is generated in proportion to the bound enzyme. ## M. Performance Characteristics (if/when applicable): 1. Analytical performance: a. Precision/Reproducibility: Provided in k083373 and k970227. b. Linearity/assay reportable range: Provided in k083373 and k970227. c. Traceability, Stability, Expected values (controls, calibrators, or methods): Provided in k083373 and k970227. d. Detection limit: Provided in k083373 and k970227. e. Analytical specificity: Provided in k083373 and k970227. f. Assay cut-off: Not applicable 2. Comparison studies: {5} a. Method comparison with predicate device: Provided in k083373 and k970227. b. Matrix comparison: Provided in k083373 and k970227. 3. Clinical studies: a. Clinical Sensitivity: Not applicable b. Clinical specificity: Not applicable c. Other clinical supportive data (when a. and b. are not applicable): Not applicable 4. Clinical cut-off: Not applicable 5. Expected values/Reference range: IMMULITE 2000 Free T4: A pediatric reference range study using population age from 1 month old to 20 years old was conducted. Serum samples were collected at 8 U.S. sites to be representative of the U.S. population diversity. The samples were obtained from apparently healthy individuals who were not taking prescribed or over-the counter medications in the 7 days prior to sample collection, or were not taking contraceptive medication in the 90 days prior to sample collection, or were not pregnant. Subject who are suspicious of premature/delayed puberty, or are positive for the presence of autoantibody against Thyroid Peroxidase (TPO) and Thyroglobulin (TG) are also excluded from the study. All samples were shipped and tested at a single laboratory study site. A total of 426 pediatric patient (81 infants, 197 children, and 148 adolescents) samples were tested in singleton using four (4) reagent lots to establish the IMMULITE 2000 Free T4 reference ranges for the studied pediatric population. Results from the $2.5^{\text{th}}$ to $97.5^{\text{th}}$ percentile were used as the pediatric reference ranges. The reference interval for the infant group was calculated using the robust symmetrical method to accommodate the smaller sample size of this population. A non-parametric approach was used to establish the reference intervals for children and adolescents based on the CLSI C28-A3 recommendation because the sample size was greater than 120. Reference ranges are summarized in the table below. {6} Confidence intervals for the limits of the infant reference range are as follows: | | Lower Limit of reference range | Upper Limit of reference | | --- | --- | --- | | 90% confidence intervals: | 0.76 – 0.84 ng/dL | 1.23 – 1.32 ng/dL | IMMULITE 2000 Third Generation TSH: A pediatric reference range study using population age from 1 month old to 20 years old was conducted. Serum samples were collected at 8 U.S. sites to be representative of the U.S. population diversity. All samples were shipped and tested at a single laboratory study site. A total of 433 pediatric patient (90 infants, 195 children, and 148 adolescents) samples were tested in singleton using three (3) reagent lots to establish the IMMULITE 2000 Third Generation TSH reference ranges for the studied pediatric population. Results from the $2.5^{\text{th}}$ to $97.5^{\text{th}}$ percentile were used as the pediatric reference ranges. The reference interval for the infant group was calculated by performing a log-transformation of the raw data followed by the application of the robust symmetric method to the transformed data. This approach was selected to be most appropriate for the smaller sample size as the data from this population was highly skewed to the right and a transformation was necessary to obtain a normal distribution. The sponsor has language in their package insert explaining the uncertainty in the value for the upper limit of the reference range. A non-parametric approach was used to establish the reference intervals for children and adolescents based on the CLSI C28-A3 recommendation because the sample size was greater than 120. Reference ranges are summarized in the table below. | Age Range | Reference Ranges | | --- | --- | | Infants (01 – 23 months) | 0.83 - 6.5 μIU/mL | | Children (02 – 12 years) | 0.58 - 4.1 μIU/mL | | Adolescents (13 – 20 years) | 0.39 - 4.0 μIU/mL | Confidence intervals for the limits of the infant reference range are as follows: | | Lower Limit of reference range | Upper Limit of reference range | | --- | --- | --- | | 90% confidence intervals: | 0.70 – 0.99 μIU/mL | 5.58 – 7.65 μIU/mL | {7} N. Proposed Labeling: The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10. O. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision. 8
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...