ACE ALBUMIN REAGENT, ACE TOTAL PROTEIN REAGENT, ACE CALCIUM-ARSENAZO REAGENT AND ACE PHOSPHORUS REAGENT

K131488 · Alfa Wassermann Diagnostics Technologies, LLC · CIX · Aug 19, 2013 · Clinical Chemistry

Device Facts

Record IDK131488
Device NameACE ALBUMIN REAGENT, ACE TOTAL PROTEIN REAGENT, ACE CALCIUM-ARSENAZO REAGENT AND ACE PHOSPHORUS REAGENT
ApplicantAlfa Wassermann Diagnostics Technologies, LLC
Product CodeCIX · Clinical Chemistry
Decision DateAug 19, 2013
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1035
Device ClassClass 2

Intended Use

The ACE Albumin Reagent is intended for the quantitative determination of albumin concentration in serum and lithium heparin plasma using the ACE, ACE Alere and ACE Axcel Chemistry Systems. Albumin measurements are used in the diagnosis and treatment of numerous diseased involving primarily the liver or kidneys. This test is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only. The ACE Total Protein Reagent is intended for the quantitative determination of total protein concentration in serum and lithium heparin plasma using the ACE, ACE Alera and ACE Axcel Clinical Chemistry System. Total protein measurements are used in the diagnosis and treatment of a variety of diseases involving the liver, kidney, or bone marrow as well as other metabolic or nutritional disorders. This test is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only. The ACE Calcium-Arsenazo Reagent is intended for the quantitative determination of calcium concentration in serum and lithium heparin plasma using the ACE, ACE Alera and ACE Axcel Clinical Chemistry Systems Calcium measurements are used in the diagnosis and treatment of parathyroid disease, a variety of bone diseases, chronic renal disease and tetany (intermittent muscular contractions or spasms). This test is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only. The ACE Inorganic Phosphorus U.V. Reagent is intended for the quantitative determination of inorganic phosphorus concentration in serum and lithium heparin plasma using the ACE, ACE Alera and ACE Axcel Clinical Chemistry Systems Measurements of inorganic phosphorus are used in the diagnosis and treatment of various disorders, including parathyroid gland and kidney diseases and vitamin D imbalance. This test is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only.

Device Story

Reagents for quantitative photometric/colorimetric analysis of serum and lithium heparin plasma samples; used on ACE, ACE Alera, and ACE Axcel clinical chemistry systems. Reagents react with analytes (albumin, total protein, calcium, phosphorus) to form colored complexes; intensity measured bichromatically. Systems operate in clinical laboratories or physician office laboratories; operated by lab personnel. Output is quantitative concentration value; used by clinicians for diagnosis and treatment monitoring of liver, kidney, bone, and metabolic diseases.

Clinical Evidence

Bench testing only. Precision studies (20-day, n=80; POL studies, n=20) showed total CVs generally <5%. Linearity confirmed across reportable ranges (r² >0.99). Method comparison (n=50-51 per analyte) against predicate showed high correlation (r² >0.99). Matrix comparison (serum vs. lithium heparin plasma) confirmed equivalence. Interference testing performed per CLSI EP07-A2.

Technological Characteristics

Photometric/colorimetric assays. Reagents: Albumin (Bromcresol green), Total Protein (Copper sulfate/tartrate), Calcium (Arsenazo III), Phosphorus (Ammonium molybdate/sulfuric acid). Bichromatic measurement at specific wavelengths (e.g., 629/692 nm, 544/692 nm, 647/692 nm, 340/378 nm). Standalone reagents for use on automated clinical chemistry analyzers. Traceability to NIST SRM 927, 909, 915b, 200, 3139a.

Indications for Use

Indicated for quantitative determination of albumin, total protein, calcium, and inorganic phosphorus in serum and lithium heparin plasma. Used in clinical laboratories or physician office laboratories for diagnosis/treatment of liver, kidney, bone marrow, parathyroid, and metabolic disorders.

Regulatory Classification

Identification

An albumin test system is a device intended to measure the albumin concentration in serum and plasma. Albumin measurements are used in the diagnosis and treatment of numerous diseases involving primarily the liver or kidneys.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K131488 | 510(k) Owner: | Alfa Wassermann Diagnostic Technologies, LLC<br>4 Henderson Drive<br>West Caldwell, NJ 07006 | | | |------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------|--------------| | Contact: | Hkatz@AlfaWassermannUS.com<br>Hyman Katz, Ph.D.<br>Phone: 973-852-0158<br>Fax: 973-852-0237 | | AUG 1 9 2013 | | Date Summary Prepared: | July 12, 2013 | | | | Device: | Trade Name: | ACE Albumin Reagent | | | | Classification: | Class 2 | | | | Common/Classification Name: | Bromcresol Green Dye-Binding, Albumin<br>(21 C.F.R. § 862.1035)<br>Product Code CIX | | | | Trade Name: | ACE Total Protein Reagent | | | | Classification: | Class 2 | | | | Common/Classification Name: | Biuret (Colorimetric), Total Protein<br>(21 C.F.R. § 862.1635)<br>Product Code CEK | | | | Trade Name: | ACE Calcium-Arsenazo Reagent | | | | Classification: | Class 2 | | | | Common/Classification Name: | Azo Dye, Calcium<br>(21 C.F.R. § 862.1445)<br>Product Code CJY | | | | Trade Name: | ACE Inorganic Phosphorus U.V. Reagent | | | | Classification: | Class 1 | | | | Common/Classification Name: | Phosphomolybdate (Colorimetric), Inorganic<br>Phosphorus<br>(21 C.F.R. § 862.1580)<br>Product Code CEO | | | Predicate Devices: | Manufacturer for reagent system predicates:<br>Alfa Wassermann ACE and ACE Axcel Clinical Chemistry Systems and ACE<br>Reagents (K930104, K113253, K113374) | | | {1}------------------------------------------------ | Device<br>Descriptions: | In the ACE Albumin Reagent assay, Bromcresol green binds specifically to albumin to<br>form a green colored complex, which is measured bichromatically at 629 nm/692 nm.<br>The intensity of color produced is directly proportional to the albumin concentration in<br>the sample.<br><br>In the ACE Total Protein Reagent assay, cupric ions react with the peptide bonds of<br>proteins under alkaline conditions to form a violet colored complex, which is measured<br>bichromatically at 544 nm/692 nm. The intensity of color produced is directly<br>proportional to the total protein concentration in the sample.<br><br>In the ACE Calcium-Arsenazo Reagent assay, calcium reacts with Arsenazo III in an<br>acidic solution to form a blue-purple colored complex, which is measured<br>bichromatically at 647 nm/692 nm. The intensity of color produced is directly<br>proportional to the calcium concentration in the sample.<br><br>In the ACE Inorganic Phosphorus U.V. Reagent assay, under acidic conditions,<br>inorganic phosphorus in serum reacts with ammonium molybdate to form an unreduced<br>phosphomolybdate complex, which absorbs strongly at 340 nm. The increase in<br>absorbance, measured bichromatically at 340 nm/378 nm, is directly proportional to the<br>amount of phosphorus in the sample. | |-----------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Intended Use: | Indications for Use:<br><br>ACE Albumin Reagent is intended for the quantitative determination of albumin<br>concentration in serum and lithium heparin plasma using the ACE, ACE Alera, and<br>ACE Axcel Clinical Chemistry Systems. Albumin measurements are used in the<br>diagnosis and treatment of numerous diseases involving primarily the liver or kidneys.<br>This test is intended for use in clinical laboratories or physician office laboratories. For<br><i>in vitro</i> diagnostic use only.<br><br>ACE Total Protein Reagent is intended for the quantitative determination of total<br>protein concentration in serum and lithium heparin plasma using the ACE, ACE Alera,<br>and ACE Axcel Clinical Chemistry Systems. Total protein measurements are used in<br>the diagnosis and treatment of a variety of diseases involving the liver, kidney, or bone<br>marrow as well as other metabolic or nutritional disorders. This test is intended for use<br>in clinical laboratories or physician office laboratories. For <i>in vitro</i> diagnostic use only.<br><br>ACE Calcium-Arsenazo Reagent is intended for the quantitative determination of<br>calcium concentration in serum and lithium heparin plasma using the ACE, ACE Alera,<br>and ACE Axcel Clinical Chemistry Systems. Calcium measurements are used in the<br>diagnosis and treatment of parathyroid disease, a variety of bone diseases, chronic renal<br>disease and tetany (intermittent muscular contractions or spasms). This test is intended<br>for use in clinical laboratories or physician office laboratories. For <i>in vitro</i> diagnostic<br>use only. | | | ACE Inorganic Phosphorus U.V. Reagent is intended for the quantitative determination<br>of inorganic phosphorus concentration in serum and lithium heparin plasma using the<br>ACE, ACE Alera, and ACE Axcel Clinical Chemistry Systems. Measurements of<br>inorganic phosphorus are used in the diagnosis and treatment of various disorders.<br>including parathyroid gland and kidney diseases and vitamin D imbalance. This test is<br>intended for use in clinical laboratories or physician office laboratories. For in vitro<br>diagnostic use only. | | Technological<br>Characteristics: | ACE Albumin Reagent consists of a single reagent bottle. The reagent contains<br>Bromcresol green and acetate buffer. | | | ACE Total Protein Reagent consists of a single reagent bottle. The reagent contains<br>copper sulfate, sodium potassium tartrate, potassium iodide and sodium hydroxide. | | | ACE Calcium-Arsenazo Reagent consists of a single reagent bottle. The Reagent<br>contains Arsenazo III. | | | ACE Inorganic Phosphorus U.V. Reagent consists of a single reagent bottle. The<br>reagent contains ammonium molybdate and sulfuric acid. | . {2}------------------------------------------------ · {3}------------------------------------------------ | | Comparison of similarities and differences with predicate device | | | |----------------------------------------|------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------|-----------------------------------------------------------| | Device Comparison<br>with Predicate | ACE Albumin Reagent | | | | ALB | Candidate Device | Predicate Device<br>K930104<br>(ACE ALB) | | | Intended Use/<br>Indications for Use | The ACE Albumin Reagent is intended<br>for the quantitative determination of<br>albumin concentration. | Same | | | Platforms | ACE, ACE Alera®, and ACE Axcel<br>Clinical Chemistry Systems | ACE Clinical Chemistry System | | | Method | Photometric | Same | | | Calibration<br>Stability | 30 days | Same | | | On-Board Stability | 30 Days | Same | | | Sample Type | Serum and lithium heparin plasma | Serum | | | Sample Volume | 3 µL | Same | | | Reaction Volume | 463 µL | Same | | | Expected Values | 3.5 – 5.2 g/dL | Same | | | Measuring Range | 0.1 – 7.6 g/dL | Same | | | Sample Stability | Specimen stable at 4°C for up to 72 hours<br>and frozen at -20°C for 6 months or<br>indefinitely at -70°C. | Same | | | | Calcium-Arsenazo | Candidate Device | Predicate Device<br>K930104<br>(ACE Calcium-Arsenazo) | | Device<br>Comparison<br>with Predicate | Intended Use/<br>Indications for Use | ACE Calcium-Arsenazo Reagent is<br>intended for the quantitative<br>determination of calcium. | Same | | | Platforms | ACE, ACE Alera®, and ACE Axcel<br>Clinical Chemistry Systems | ACE Clinical Chemistry System | | | Method | Photometric | Same | | | Calibration<br>Stability | 30 Days | Same | | | On-Board Stability | 30 Days | Same | | | Sample Type | Serum and lithium heparin plasma | Serum | | | Sample Volume | 3 µL | Same | | | Reaction Volume | 318 µL | Same | | | Expected Values | 8.5 - 10.2 mg/dL | Same | | | Measuring Range | 0.2 - 16.5 mg/dL | Same | | | Sample Stability | Specimen stable for 7 days at 20-25°C, 3<br>weeks at 4-8°C, and 8 months at -20°C | Same | | | | ACE Inorganic Phosphorus U.V. Reagent | | | | Inorganic<br>Phosphorus | Candidate Device | Predicate Device<br>K930104<br>(ACE Inorganic Phosphorus) | | | Intended Use/<br>Indications for Use | ACE Inorganic Phosphorus U.V.<br>Reagent is intended for the<br>quantitative determination of | Same | | | | inorganic phosphorus. | | | | Platforms | ACE, ACE Alera®, and ACE Axcel<br>Clinical Chemistry Systems | ACE Clinical Chemistry System | | | Method | Photometric | Same | | | Calibration Stability | 30 Days | Same | | | On-Board Stability | 30 Days | Same | | | Sample Type | Serum and lithium heparin plasma | Serum | | | Sample Volume | 3 µL | Same | | | Reaction Volume | 218 µL | Same | | | Expected Values | 2.7 - 4.5 mg/dL | Same | | | Measuring Range | 0.4-21 mg/dL | Same | {4}------------------------------------------------ · {5}------------------------------------------------ ## Performance data for the Alfa Wassermann ACE Reagents run on the Alfa Wassermann ACE, ACE Alera and ACE Axcel Clinical Chemistry Systems | In-House Precision: Serum vs. Plasma – ACE Albumin Reagent | | | | | | | | | | | |------------------------------------------------------------|---------------------|----------------|------------|---------------|----------------|------------|---------------|----------------|------------|--| | | | | | | | | | | | | | | Precision (SD, %CV) | | | | | | | | | | | Albumin<br>g/dL | ACE<br>Mean | Within-<br>Run | Total | Alera<br>Mean | Within-<br>Run | Total | Axcel<br>Mean | Within-<br>Run | Total | | | Serum Low | 4.1 | 0.05, 1.3% | 0.07, 1.6% | 4.1 | 0.04, 0.9% | 0.04, 1.1% | 4.1 | 0.02, 0.5% | 0.04, 1.0% | | | Plasma Low | 3.8 | 0.06, 1.7% | 0.06, 1.7% | 3.7 | 0.03, 0.8% | 0.05, 1.4% | 3.7 | 0.06, 1.6% | 0.06, 1.6% | | | Serum Mid | 5.4 | 0.08, 1.6% | 0.10, 1.8% | 5.3 | 0.05, 1.0% | 0.06, 1.1% | 5.3 | 0.03, 0.6% | 0.03, 0.6% | | | Plasma Mid | 5.0 | 0.05, 1.0% | 0.07, 1.4% | 5.0 | 0.08, 1.7% | 0.08, 1.7% | 4.9 | 0.04, 0.9% | 0.05, 1.1% | | | Serum High | 6.5 | 0.07, 1.1% | 0.11, 1.6% | 6.5 | 0.05, 0.8% | 0.08, 1.3% | 6.4 | 0.06, 1.0% | 0.09, 1.3% | | | Plasma High | 6.2 | 0.09, 1.5% | 0.10, 1.7% | 6.1 | 0.08, 1.3% | 0.10, 1.6% | 6.1 | 0.05, 0.9% | 0.08, 1.3% | | #### In-House Precision: Serum vs. Plasma - ACE Total Protein Reagent | Precision (SD, %CV) | | | | | | | | | | |--------------------------|-------------|----------------|------------|---------------|----------------|------------|---------------|----------------|------------| | Total<br>Protein<br>g/dL | ACE<br>Mean | Within-<br>Run | Total | Alera<br>Mean | Within-<br>Run | Total | Axcel<br>Mean | Within-<br>Run | Total | | Serum Low | 6.7 | 0.06, 1.0% | 0.07, 1.0% | 6.7 | 0.05, 0.7% | 0.05, 0.8% | 6.8 | 0.08, 1.1% | 0.09, 1.3% | | Plasma Low | 7.2 | 0.06, 0.9% | 0.06, 0.9% | 7.1 | 0.08, 1.1% | 0.09, 1.2% | 7.2 | 0.05, 0.8% | 0.07, 0.9% | | Serum Mid | 8.4 | 0.11, 1.3% | 0.11, 1.3% | 8.4 | 0.08, 1.0% | 0.08, 1.0% | 8.4 | 0.07, 0.8% | 0.11, 1.4% | | Plasma Mid | 8.8 | 0.04, 0.5% | 0.06, 0.7% | 8.7 | 0.06, 0.7% | 0.1, 1.2% | 8.8 | 0.07, 0.8% | 0.08, 0.9% | | Serum High | 10.1 | 0.07, 0.7% | 0.08, 0.8% | 10.0 | 0.07, 0.7% | 0.09, 0.9% | 10.1 | 0.07, 0.7% | 0.09, 0.9% | | Plasma High | 10.3 | 0.13, 1.3% | 0.14, 1.4% | 10.2 | 0.11, 1.1% | 0.14, 1.3% | 10.4 | 0.08, 0.8% | 0.10, 1.0% | #### In-House Precision: Serum vs. Plasma - ACE Calcium-Arsenazo Reagent | Precision (SD, %CV) | | | | | | | | | | |-------------------------------|-------------|----------------|------------|---------------|----------------|------------|---------------|----------------|------------| | Calcium-<br>Arsenazo<br>mg/dL | ACE<br>Mean | Within-<br>Run | Total | Alera<br>Mean | Within-<br>Run | Total | Axcel<br>Mean | Within-<br>Run | Total | | Serum Low | 9.3 | 0.12, 1.3% | 0.25, 2.7% | 9.3 | 0.09, 0.9% | 0.22, 2.4% | 9.3 | 0.08, 0.8% | 0.17, 1.8% | | Plasma Low | 8.4 | 0.04, 0.5% | 0.2, 2.4% | 8.3 | 0.10, 1.2% | 0.17, 2.0% | 8.3 | 0.08, 0.9% | 0.11, 1.4% | | Serum Mid | 11.7 | 0.18, 1.6% | 0.2, 1.7% | 11.6 | 0.14, 1.2% | 0.14, 1.2% | 11.6 | 0.1, 0.9% | 0.11, 0.9% | | Plasma Mid | 10.7 | 0.19, 1.7% | 0.20, 1.9% | 10.7 | 0.13, 1.2% | 0.15, 1.4% | 10.7 | 0.12, 1.2% | 0.13, 1.2% | | Serum High | 13.9 | 0.20, 1.4% | 0.2, 1.4% | 13.8 | 0.19, 1.4% | 0.19, 1.4% | 13.8 | 0.09, 0.7% | 0.11, 0.8% | | Plasma High | 13.0 | 0.25, 1.9% | 0.26, 2.0% | 12.9 | 0.13, 1.0% | 0.14, 1.1% | 13.1 | 0.15, 1.2% | 0.18, 1.4% | {6}------------------------------------------------ | In-House Precision: Serum vs. Plasma - ACE Inorganic Phosphorus Reagent | | | | | | | | | | |-------------------------------------------------------------------------|-------------|----------------|------------|---------------|----------------|------------|---------------|----------------|------------| | Precision (SD, %CV) | | | | | | | | | | | Inorganic<br>Phosphorus<br>U.V.<br>mg/dL | ACE<br>Mean | Within-<br>Run | Total | Alera<br>Mean | Within-<br>Run | Total | Axcel<br>Mean | Within-<br>Run | Total | | Serum Low | 3.5 | 0.15, 4.4% | 0.17, 5.0% | 3.4 | 0.11, 3.1% | 0.14, 4.0% | 3.5 | 0.11, 3.1% | 0.14, 4.1% | | Plasma Low | 3.1 | 0.16, 5.1% | 0.18, 5.9% | 3.0 | 0.11, 3.7% | 0.15, 5.0% | 3.1 | 0.15, 5.0% | 0.19, 6.1% | | Serum Mid | 10.2 | 0.04, 0.3% | 0.05, 0.5% | 9.9 | 0.08, 0.8% | 0.08, 0.8% | 10.2 | 0.04, 0.4% | 0.12, 1.2% | | Plasma Mid | 9.8 | 0.09, 0.9% | 0.09, 0.9% | 9.6 | 0.07, 0.8% | 0.08, 0.8% | 9.9 | 0.06, 0.6% | 0.12, 1.2% | | Serum High | 17.0 | 0.26, 1.5% | 0.26, 1.6% | 16.6 | 0.22, 1.3% | 0.22, 1.3% | 17.3 | 0.28, 1.6% | 0.30, 1.7% | | Plasma High | 16.7 | 0.23, 1.4% | 0.24, 1.4% | 16.3 | 0.24, 1.5% | 0.29, 1.8% | 16.9 | 0.30, 1.8% | 0.32, 1.9% | . {7}------------------------------------------------ Performance Data: In-House Precision -Serum vs. Plasma ## Performance data for the Alfa Wassermann ACE Reagents run on the Alfa Wassermann ACE, ACE Alera and ACE Axcel Clinical Chemistry Systems In-House Matrix Comparison: Serum vs. Plasma – ACE Albumin Reagent | System | Range | Results - Serum vs. Plasma | | |-----------------------|--------------|--------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------| | ACE<br>55 pairs | 0.3-6.8 g/dL | Slope:<br>Intercept:<br>Correlation:<br>Std. Error Est:<br>Confidence Interval Slope:<br>Confidence Interval Intercept: | 0.991<br>0.03<br>0.9874<br>0.19<br>0.948 to 1.034<br>-0.15 to 0.20 | | ACE Alera<br>56 pairs | 0.3-6.8 g/dL | Slope:<br>Intercept:<br>Correlation:<br>Std. Error Est:<br>Confidence Interval Slope:<br>Confidence Interval. Intercept: | 1.002<br>-0.01<br>0.9905<br>0.17<br>0.964 to 1.040<br>-0.15 to 0.14 | | ACE Axcel<br>56 pairs | 0.7-6.7 g/dL | Slope:<br>Intercept:<br>Correlation:<br>Std. Error Est:<br>Confidence Interval Slope:<br>Confidence Interval Intercept: | 0.956<br>0.20<br>0.9850<br>0.20<br>0.911 to 1.001<br>0.04 to 0.37 | #### In-House Matrix Comparison: Serum vs. Plasma – ACE Total Protein Reagent | System | Range | Results - Serum vs. Plasma | |-----------|---------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | ACE | 0.5-12.3 g/dL<br>56 pairs | Slope: 1.001<br>Intercept: 0.12<br>Correlation: 0.9798<br>Std. Error Est: 0.40<br>Confidence Interval Slope: 0.946 to 1.056<br>Confidence Interval Intercept: -0.24 to 0.48 | | ACE Alera | 0.5-12.0 g/dL<br>56 pairs | Slope: 0.999<br>Intercept: 0.14<br>Correlation: 0.9840<br>Std. Error Est: 0.35<br>Confidence Interval Slope: 0.950 to 1.047<br>Confidence Interval Intercept: -0.18 to 0.46 | | ACE Axcel | 0.5-13.9 g/dL<br>81 pairs | Slope: 0.994<br>Intercept: 0.34<br>Correlation: 0.9885<br>Std. Error Est: 0.26<br>Confidence Interval Slope: 0.961 to 1.028<br>Confidence Interval Intercept: 0.12 to 0.57 | {8}------------------------------------------------ Performance Data: In-House #### Precision -Serum vs. Plasma #### In-House Matrix Comparison: Serum vs. Plasma – ACE Calcium-Aresnazo Reagent | System | Range | Results - Serum vs. Plasma | | | |-----------------|-----------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | ACE<br>56 pairs | 1.0-13.7 mg/dL | Slope: 1.006<br>Intercept: -0.01<br>Correlation: 0.9824<br>Std. Error Est: 0.39<br>Confidence Interval Slope: 0.955 to 1.058<br>Confidence Interval Intercept: -0.46 to 0.45 | | | | | ACE Alera<br>56 pairs | 1.0-13.7 mg/dL | Slope: 1.008<br>Intercept: -0.06<br>Correlation: 0.9793<br>Std. Error Est: 0.43<br>Confidence Interval Slope: 0.952 to 1.064<br>Confidence Interval Intercept: -0.55 to 0.42 | | | | | ACE Axcel<br>81 pairs | 0.7-15.0 mg/dL | Slope: 0.978<br>Intercept: 0.33<br>Correlation: 0.9911<br>Std. Error Est: 0.23<br>Confidence Interval Slope: 0.949 to 1.007<br>Confidence Interval Intercept: 0.06 to 0.60 | #### In-House Matrix Comparison: Serum vs. Plasma - ACE Inorganic Phosphorus Reagent | System | Range | Results - Serum vs. Plasma | |------------------------|----------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | ACE<br>100 pairs | 1.3-19.3 mg/dL | Slope: 1.042<br>Intercept: -0.26<br>Correlation: 0.9927<br>Std. Error Est: 0.33<br>Confidence Interval Slope: 1.017 to 1.067<br>Confidence Interval Intercept: -0.38 to -0.14 | | ACE Alera<br>102 pairs | 1.3-19.3 mg/dL | Slope: 1.049<br>Intercept: -0.28<br>Correlation: 0.9928<br>Std. Error Est: 0.33<br>Confidence Interval Slope: 1.024 to 1.074<br>Confidence Interval Intercept: -0.40 to -0.16 | | ACE Axcel<br>56 pairs | 0.5-19.8 mg/dL | Slope: 0.999<br>Intercept: 0.04<br>Correlation: 0.9950<br>Std. Error Est: 0.34<br>Confidence Interval Slope: 0.972 to 1.027<br>Confidence Interval Intercept: -0.12 to 0.20 | {9}------------------------------------------------ | POL - Precision for ACE and ACE Alera Clinical Chemistry Systems<br>(Note: Refer to previously cleared submission k113374 for ACE Axcel POL data) | | | | | | | | | |---------------------------------------------------------------------------------------------------------------------------------------------------|----------------|--------|------|----------------------------|--------------|------|----------------------------------|--------------| | Performance Data:<br>In-House<br>Matrix<br>Comparison –<br>Serum vs.<br>Plasma | Albumin<br>Lab | Sample | Mean | ACE Result<br>g/dL SD, %CV | | Mean | ACE Alera Result<br>g/dL SD, %CV | | | | | | | Within-Run | Total | | Within-Run | Total | | | In-House | 1 | 3.5 | 0.05<br>1.4% | 0.07<br>2.0% | 3.5 | 0.02<br>0.6% | 0.04<br>1.1% | | | POL 1 | 1 | 3.5 | 0.04<br>1.3% | 0.04<br>1.3% | 3.5 | 0.05<br>1.4% | 0.06<br>1.7% | | | POL 2 | 1 | 3.5 | 0.06<br>1.7% | 0.07<br>2.0% | 3.6 | 0.05<br>1.4% | 0.05<br>1.5% | | | POL 3 | 1 | 3.5 | 0.08<br>2.3% | 0.08<br>2.4% | 3.5 | 0.05<br>1.6% | 0.05<br>1.6% | | | In-House | 2 | 5.0 | 0.06<br>1.2% | 0.06<br>1.2% | 5.0 | 0.05<br>1.0% | 0.05<br>1.1% | | | POL 1 | 2 | 4.9 | 0.06<br>1.2% | 0.07<br>1.4% | 5.0 | 0.08<br>1.7% | 0.09<br>1.9% | | | POL 2 | 2 | 4.9 | 0.03<br>0.6% | 0.06<br>1.2% | 5.0 | 0.06<br>1.2% | 0.08<br>1.6% | | | POL 3 | 2 | 4.9 | 0.06<br>1.2% | 0.09<br>1.9% | 4.9 | 0.03<br>0.6% | 0.03<br>0.7% | | | In-House | 3 | 6.2 | 0.11<br>1.9% | 0.13<br>2.1% | 6.2 | 0.06<br>1.0% | 0.07<br>1.1% | | | POL 1 | 3 | 6.1 | 0.07<br>1.1% | 0.07<br>1.2% | 6.2 | 0.07<br>1.1% | 0.10<br>1.6% | | | POL 2 | 3 | 6.1 | 0.10<br>1.6% | 0.12<br>1.9% | 6.2 | 0.06<br>1.0% | 0.07<br>1.1% | | | POL 3 | 3 | 6.1 | 0.10<br>1.7% | 0.11<br>1.8% | 6.1 | 0.08<br>1.3% | 0.08<br>1.4% | {10}------------------------------------------------ | | | | | ACE Result | | | ACE Alera Result | | | |-----------------------------------------------------------------------------------|------------------|----------|--------|------------|--------------|-------|------------------|--------------|-------| | Performance<br>Data:<br>In-House<br>Matrix<br>Comparison -<br>Serum vs.<br>Plasma | Total<br>Protein | Lab | Sample | Mean | g/dL SD, %CV | | Mean | g/dL SD, %CV | | | | | | | | Within-Run | Total | | Within-Run | Total | | | | In-House | 1 | 5.3 | 0.05 | 0.06 | 5.3 | 0.08 | 0.10 | | | | | | | 0.9% | 1.2% | | 1.5% | 1.8% | | | | POL 1 | 1 | 5.3 | 0.13 | 0.13 | 5.5 | 0.07 | 0.10 | | | | | | | 2.5% | 2.5% | | 1.4% | 1.8% | | | | POL 2 | 1 | 5.3 | 0.08 | 0.16 | 5.2 | 0.07 | 0.15 | | | | | | | 1.5% | 3.1% | | 1.3% | 2.8% | | | | POL 3 | 1 | 5.6 | 0.10 | 0.12 | 5.6 | 0.07 | 0.12 | | | | | | | 1.7% | 2.1% | | 1.4% | 2.2% | | | | In-House | 2 | 8.3 | 0.10 | 0.12 | 8.3 | 0.10 | 0.11 | | | | | | | 1.2% | 1.4% | | 1.2% | 1.4% | | | | POL 1 | 2 | 8.2 | 0.08 | 0.11 | 8.4 | 0.09 | 0.10 | | | | | | | 1.0% | 1.4% | | 1.1% | 1.2% | | | | POL 2 | 2 | 8.3 | 0.06 | 0.18 | 8.4 | 0.10 | 0.11 | | | | | | | 0.7% | 2.1% | | 1.2% | 1.4% | | | | POL 3 | 2 | 8.6 | 0.04 | 0.10 | 8.2 | 0.09 | 0.14 | | | | | | | 0.5% | 1.1% | | 1.1% | 1.7% | | | | In-House | 3 | 11.2 | 0.14 | 0.17 | 11.3 | 0.14 | 0.15 | | | | | | | 1.3% | 1.5% | | 1.3% | 1.4% | | | | POL 1 | 3 | 11.2 | 0.14 | 0.17 | 11.3 | 0.14 | 0.14 | | | | | | | 1.3% | 1.5% | | 1.2% | 1.2% | | | | POL 2 | 3 | 11.2 | 0.09 | 0.20 | 11.5 | 0.09 | 0.16 | | | | | | | 0.8% | 1.8% | | 0.8% | 1.4% | | | | POL 3 | 3 | 11.4 | 0.22 | 0.23 | 11.1 | 0.26 | 0.31 | | | | | | | 1.9% | 2.0% | | 2.3% | 2.8% | {11}------------------------------------------------ | | | | | ACE Result | | | ACE Alera Result | | |--|----------------------|--------|------|-----------------|-----------------|------|------------------|-----------------| | | Calcium-<br>Arsenazo | | | mg/dL SD, %CV | | | mg/dL SD, %CV | | | | Lab | Sample | Mean | Within-Run | Total | Mean | Within-Run | Total | | | In-House | 1 | 7.0 | 0.12 SD<br>1.7 | 0.17 SD<br>2.4 | 6.9 | 0.08 SD<br>1.2% | 0.15 SD<br>2.1% | | | POL 1 | 1 | 7.0 | 0.14 SD<br>2.0% | 0.15 SD<br>2.1% | 6.9 | 0.07 SD<br>1.0% | 0.19 SD<br>2.7% | | | POL 2 | 1 | 7.0 | 0.16 SD<br>2.3% | 0.17 SD<br>2.4% | 7.0 | 0.19 SD<br>2.7% | 0.19 SD<br>2.7% | | | POL 3 | 1 | 7.0 | 0.16 SD<br>2.3% | 0.17 SD<br>2.4% | 7.0 | 0.14 SD<br>1.9% | 0.14 SD<br>1.9% | | | In-House | 2 | 10.7 | 0.21 SD<br>2.0% | 0.21 SD<br>2.0% | 10.5 | 0.05 SD<br>0.5% | 0.06 SD<br>0.6% | | | POL 1 | 2 | 10.6 | 0.06 SD<br>0.6% | 0.06 SD<br>0.6% | 10.5 | 0.09 SD<br>0.9% | 0.33 SD<br>3.2% | | | POL 2 | 2 | 10.5 | 0.12 SD<br>1.2% | 0.15 SD<br>1.5% | 10.6 | 0.21 SD<br>1.9% | 0.22 SD<br>2.1% | | | POL 3 | 2 | 10.5 | 0.10 SD<br>1.0% | 0.11 SD<br>1.0% | 10.6 | 0.16 SD<br>1.5% | 0.16 SD<br>1.5% | | | In-House | 3 | 13.6 | 0.11 SD<br>0.8% | 0.26 SD<br>1.9% | 13.5 | 0.17 SD<br>1.3% | 0.20 SD<br>1.5% | | | POL 1 | 3 | 13.6 | 0.14 SD<br>1.1% | 0.20 SD<br>1.5% | 13.4 | 0.14 SD<br>1.1% | 0.34 SD<br>2.5% | | | POL 2 | 3 | 13.5 | 0.32 SD<br>2.3% | 0.37 SD<br>2.7% | 13.6 | 0.21 SD<br>1.5% | 0.23 SD<br>1.7% | | | POL 3 | 3 | 13.6 | 0.16 SD<br>1.2% | 0.17 SD<br>1.2% | 13.6 | 0.14 SD<br>1.0% | 0.18 SD<br>1.3% | ・ : {12}------------------------------------------------ | Performance Data:<br>Precision - POL | POL - Precision for ACE and ACE Alera Clinical Chemistry Systems | | | | | | | |--------------------------------------|------------------------------------------------------------------|------|-----------------------------|------------------------|------|-----------------------------|------------------------| | Inorganic<br>Phosphorus<br>U.V. | | | ACE Result | | | ACE Alera Result | | | Lab | Sample | Mean | mg/dL SD, %CV<br>Within-Run | mg/dL SD, %CV<br>Total | Mean | mg/dL SD, %CV<br>Within-Run | mg/dL SD, %CV<br>Total | | In-House | 1 | 2.7 | 0.08 SD<br>3.0% | 0.08 SD<br>3.0% | 2.8 | 0.06 SD<br>2.1% | 0.06 SD<br>2.1% | | POL 1 | 1 | 2.7 | 0.03 SD<br>1.2% | 0.05 SD<br>1.9% | 2.7 | 0.04 SD<br>1.4% | 0.10 SD<br>3.8% | | POL 2 | 1 | 2.6 | 0.06 SD<br>2.3% | 0.09 SD<br>3.6% | 2.5 | 0.02 SD<br>0.9% | 0.11 SD<br>4.4% | | POL 3 | 1 | 2.8 | 0.06 SD<br>2.1% | 0.10 SD<br>3.5% | 2.9 | 0.05 SD<br>1.9% | 0.07 SD<br>2.4% | | In-House | 2 | 7.0 | 0.07 SD<br>1.0% | 0.09 SD<br>1.3% | 7.1 | 0.07 SD<br>0.9% | 0.09 SD<br>1.3% | | POL 1 | 2 | 7.0 | 0.04 SD<br>0.6% | 0.07 SD<br>1.1% | 7.1 | 0.07 SD<br>0.9% | 0.18 SD<br>2.5% | | POL 2 | 2 | 6.7 | 0.08 SD<br>1.2% | 0.14 SD<br>2.1% | 6.7 | 0.07 SD<br>1.1% | 0.22 SD<br>3.2% | | POL 3 | 2 | 7.2 | 0.04 SD<br>0.6% | 0.07 SD<br>1.0% | 7.4 | 0.10 SD<br>1.4% | 0.13 SD<br>1.7% | | In-House | 3 | 11.1 | 0.14 SD<br>1.2% | 0.18 SD<br>1.6% | 11.3 | 0.09 SD<br>0.8% | 0.11 SD<br>0.9% | | POL 1 | 3 | 11.1 | 0.13 SD<br>1.2% | 0.14 SD<br>1.3% | 11.3 | 0.16 SD<br>1.4% | 0.27 SD<br>2.4% | | POL 2 | 3 | 10.9 | 0.12 SD<br>1.1% | 0.21 SD<br>1.9% | 10.6 | 0.15 SD<br>1.4% | 0.21 SD<br>1.9% | | POL 3 | 3 | 11.4 | 0.13 SD<br>1.1% | 0.18 SD<br>1.6% | 11.7 | 0.11 SD<br>0.9% | 0.14 SD<br>1.2% | . . {13}------------------------------------------------ | Performance<br>Data:<br>Precision -<br>POL | POL – Method Comparison for ACE Clinical Chemistry System | | | | | |--------------------------------------------|-----------------------------------------------------------|------------------|------------------------------------------|------------------------------------------|------------------------------------------| | | Reagent | Statistic | ACE In-House (x) vs.<br>ACE<br>POL 1 (y) | ACE In-House (x) vs.<br>ACE<br>POL 2 (y) | ACE In-House (x) vs.<br>ACE<br>POL 3 (y) | | | Albumin | n | 50 | 50 | 50 | | | | Range (g/dL) | 1.0 to 6.4 | 1.0 to 6.4 | 1.0 to 6.4 | | | | Regression | $y = 0.983x + 0.03$ | $y = 0.992x - 0.01$ | $y = 1.006x - 0.03$ | | | | Correlation | 0.9934 | 0.9965 | 0.9971 | | | | Std. Error Est. | 0.10 | 0.08 | 0.07 | | | | CI Slope | 0.950 to 1.016 | 0.968 to 1.016 | 0.984 to 1.028 | | | | CI Intercept | -0.11 to 0.17 | -0.11 to 0.09 | -0.12 to 0.06 | | | Total Protein | n | 51 | 51 | 51 | | | | Range (g/dL) | 0.9 to 13.6 | 0.9 to 13.6 | 0.9 to 13.6 | | | | Regression | $y = 1.008x + 0.02$ | $y = 1.007x + 0.06$ | $y = 1.029x + 0.01$ | | | | Correlation | 0.9957 | 0.9976 | 0.9960 | | | | Std. Error Est. | 0.15 | 0.11 | 0.15 | | | | CI Slope | 0.981 to 1.035 | 0.987 to 1.027 | 1.003 to 1.056 | | | | CI Intercept | -0.17 to 0.22 | -0.08 to 0.20 | -0.18 to 0.20 | | | Calcium-Arsenazo | n | 50 | 50 | 50 | | | | Range (mg/dL) | 1.9 to 13.7 | 1.9 to 13.7 | 1.9 to 13.7 | | | | Regression | $y = 1.004x - 0.07$ | $y = 1.002x - 0.17$ | $y = 0.981x + 0.09$ | | | | Correlation Std. | 0.9915 | 0.9944 | 0.9951 | | | | Error Est. CI | 0.26 | 0.21 | 0.19 | | | | Slope | 0.966 to 1.042 | 0.971 to 1.033 | 0.953 to 1.009 | | | | CI Intercept | -0.42 to 0.27 | -0.45 to 0.11 | -0.17 to 0.34 | | | Inorganic<br>Phosphorus<br>U.V. | n | 50 | 48 | 50 | | | | Range (mg/dL) | 1.0 to 18.4 | 1.0 to 18.4 | 1.0 to 18.4 | | | | Regression | $y = 0.966x + 0.13$ | $y = 1.007x - 0.10$ | $y = 0.975x + 0.11$ | | | | Correlation Std. | 0.9991 | 0.9982 | 0.9987 | | | | Error Est. CI | 0.12 | 0.16 | 0.14 | | | | Slope | 0.954 to 0.978 | 0.989 to 1.025 | 0.960 to 0.989 | | | | CI Intercept | 0.07 to 0.19 | -0.18 to -0.01 | 0.04 to 0.19 | {14}------------------------------------------------ | Performance<br>Data: | POL - Method Comparison for ACE Alera Clinical Chemistry System | | | | | |----------------------|-----------------------------------------------------------------|------------------|------------------------------------------|------------------------------------------|------------------------------------------| | Precision -<br>POL | Reagent | Statistic | ACE In-House (x) vs. ACE Alera POL 1 (y) | ACE In-House (x) vs. ACE Alera POL 2 (y) | ACE In-House (x) vs. ACE Alera POL 3 (y) | | | | n | 50 | 50 | 50 | | | | Range (g/dL) | 1.0 to 6.4 | 1.0 to 6.4 | 1.0 to 6.4 | | | | Regression | y= $1.004x - 0.03$ | y= $1.005x - 0.05$ | y = $0.982x + 0.01$ | | | Albumin | Correlation | 0.9949 | 0.9960 | 0.9967 | | | | Std. Error Est. | 0.09 | 0.08 | 0.07 | | | | CI Slope | 0.975 to 1.034 | 0.979 to 1.031 | 0.959 to 1.005 | | | | Cl Intercept | -0.15 to 0.10 | -0.16 to 0.06 | -0.09 to 0.11 | | | | n | 51 | 51 | 51 | | | | Range (g/dL) | 0.9 to 13.6 | 0.9 to 13.6 | 0.9 to 13.6 | | | | Regression | y= $0.998x + 0.16$ | y = $1.027x - 0.06$ | y = $0.979x + 0.24$ | | | Total Protein | Correlation | 0.9969 | 0.9962 | 0.9964 | | | | Std. Error Est. | 0.13 | 0.14 | 0.14 | | | | CI Slope | 0.976 to 1.020 | 1.002 to 1.053 | 0.955 to 1.003 | | | | CI Intercept | 0.00 to 0.33 | -0.24 to 0.13 | 0.07 to 0.42 | | | | n | 50 | 50 | 50 | | | | Range (mg/dL) | 1.9 to 13.7 | 1.9 to 13.7 | 1.9 to 13.7 | | | | Regression | y = $0.992x - 0.09$ | y = $1.007x - 0.11$ | y = $1.008x - 0.08$ | | | Calcium-<br>Arsenazo | Correlation Std. | 0.9904 | 0.9929 | 0.9929 | | | | Error Est. Cl | 0.27 | 0.23 | 0.23 | | | | Slope | 0.952 to 1.032 | 0.972 to 1.042 | 0.973 to 1.043 | | | | CI Intercept | -0.46 to 0.27 | -0.43 to 0.21 | -0.40 to 0.23 | | | | n | 50 | 50 | 50 | | | | Range (mg/dL) | 1.0 to 18.4 | 1.0 to 18.4 | 1.0 to 18.4 | | | Inorganic<br>Phosphorus<br>U.V. | Regression | y = $1.015x + 0.14$ | y = $0.960x + 0.12$ | y = $0.984x + 0.05$ | | | | Correlation Std. | 0.9992 | 0.9986 | 0.9991 | | | | Error Est. Cl | 0.12 | 0.14 | 0.12 | | | | Slope | 1.003 to 1.027 | 0.945 to 0.974 | 0.972 to 0.996 | | | | Cl Intercept | 0.08 to 0.20 | 0.05 to 0.19 | -0.01 to 0.11 | {15}------------------------------------------------ # Performance Data: Method Comparison -POL on ACE ## Performance data for the Alfa Wassermann ACE Reagents run on the Alfa Wassermann ACE Alera Clinical Chemistry Systems # Detection Limits - ACE Alera Clinical Chemistry System | ACE Alera | ALB<br>(g/dL) | TP<br>(g/dL) | CA<br>(mg/dL) | PHOS<br>(mg/dL) | |-----------|---------------|--------------|---------------|-----------------| | LoB | 0.08 | 0.08 | 0.09 | 0.25 | | LoD | 0.09 | 0.13 | 0.11 | 0.35 | | LoQ | 0.09 | 0.20 | 0.23 | 0.35 | #### Linearity - ACE Alera Clinical Chemistry System | ACE<br>Reagents | Low Level<br>Tested | Upper Level<br>Tested | Linear to: | Linear Regression<br>Equation | |-----------------|---------------------|-----------------------|------------|----------------------------------| | ALB | 0.1 g/dL | 7.6 g/dL | 7.6 g/dL | y = 0.980x + 0.01<br>r2 = 0.9982 | | TP | 0.2 g/dL | 15.1 g/dL | 15.1 g/dL | y=0.991x + 0.04<br>r2 = 0.9979 | | CA | 0.3 g/dL | 16.5 mg/dL | 16.5 mg/dL | y=0.992x +0.27<br>r2 = 0.9990 | | PHOS | 0.2 mg/dL | 21 mg/dL | 21 mg/dL | y=1.001x +0.03<br>r2 = 0.9995 | {16}------------------------------------------------ ### Performance Data: ## ACE Alera #### Interferences - ACE Alera Clinical Chemistry System | Interferents<br>on ACE Alera | No Significant Interference at or below: | | | | |------------------------------|------------------------------------------|------------|------------|------------| | | ALB | TP | CA | PHOS | | Icterus | 60 mg/dL | 56.8 mg/dL | 58.8 mg/dL | 11.5 mg/dL | | Hemolysis | 250 mg/dL | 250 mg/dL | 1000 mg/dL | 250 mg/dL | | Lipemia | 1000 mg/dL | 929 mg/dL | 1000 mg/dL | 306 mg/dL | | Ascorbic Acid | 6 mg/dL | 6 mg/dL | 6 mg/dL | 6 mg/dL | ## Precision - ACE Alera Clinical Chemistry System | on-ACE Alera | | Precision (SD, %CV) | | | |---------------|------------|---------------------|------------|------------| | | | Mean | Within-Run | Total | | ALB<br>g/dL | Serum Low | 2.6 | 0.03, 1.3% | 0.05, 2.0% | | | Serum Mid | 3.4 | 0.07, 1.9% | 0.09, 2.5% | |…
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...