The ADVIA Chemistry® Creatine Kinase assay is for in vitro diagnostic use in the quantitative determination of creatine kinase activity in human plasma (lithium heparin) or serum on ADVIA Chemistry systems. The assay can be used to aid in the diagnosis and treatment of myocardial infarction and muscle diseases, such as Duchenne progressive muscular dystrophy. ADVIA Chemistry® Enzyme 3 Calibrator is intended for in vitro diagnostic use in the calibration of the ADVIA Chemistry Creatine Kinase (CK_L) assay on the ADVIA Chemistry systems.
Device Story
ADVIA Chemistry Creatine Kinase (CK_L) assay is a ready-to-use liquid reagent for automated ADVIA Chemistry systems. Input: human serum or lithium heparin plasma. Principle: CK reacts with creatine phosphate and ADP to form ATP; coupled to hexokinase-G6PD reaction to generate NADPH; NADPH concentration measured via absorbance increase at 340/596 nm. Output: quantitative CK activity (U/L). Used in clinical laboratories by technicians/pathologists. Results aid clinicians in diagnosing myocardial infarction and muscle disorders. Includes ADVIA Chemistry Enzyme 3 Calibrator (liquid frozen human serum albumin-based, containing human heart CK-MM).
Clinical Evidence
Bench testing only. Precision evaluated per CLSI EP05-A2 (N=80 per lot, 20 days). Linearity (1-1372 U/L) and LoQ (15 U/L) established per CLSI EP06-A and EP17-A2. Method comparison (N=116) against predicate yielded Deming regression Y=1.01x-1.8 (R=1.00). Comparison against IFCC reference method (N=100) yielded Y=1.05x-6.9 (R=1.00). Interference testing (CLSI EP07) showed no significant interference (≤10% bias) for hemoglobin, bilirubin, lipemia, ascorbic acid, sulfasalazine, or sulfapyridine.
Technological Characteristics
Liquid reagent assay; human serum albumin-based calibrator (5%) with human heart CK-MM. Sensing principle: kinetic spectrophotometry (absorbance at 340/596 nm). Analytical range: 15–1300 U/L. Connectivity: automated analyzer integration. Storage: -25 to -15°C for calibrator.
Indications for Use
Indicated for quantitative determination of creatine kinase activity in human serum or lithium heparin plasma to aid in diagnosis and treatment of myocardial infarction and muscle diseases (e.g., Duchenne progressive muscular dystrophy). For in vitro diagnostic and prescription use only.
Regulatory Classification
Identification
A creatine phosphokinase/creatine kinase or isoenzymes test system is a device intended to measure the activity of the enzyme creatine phosphokinase or its isoenzymes (a group of enzymes with similar biological activity) in plasma and serum. Measurements of creatine phosphokinase and its isoenzymes are used in the diagnosis and treatment of myocardial infarction and muscle diseases such as progressive, Duchenne-type muscular dystrophy.
K980900 — N-ASSAY CPK-L · Crestat Diagnostics, Inc. · Mar 26, 1998
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes entwined around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
SIEMENS HEALTHCARE DIAGNOSTICS, INC. KATHLEEN DRAY-LYONS REGULATORY AND CLINICAL AFFAIRS SPECIALIST 500 GBC DRIVE NEWARK DE 19714
September 7, 2016
Re: K160724
Trade/Device Name: ADVIA Chemistry® Creatine Kinase (CK L) Assay. ADVIA Chemistry® Enzyme 3 Calibrator Regulation Number: 21 CFR 862.1215 Regulation Name: Creatine phosphokinase/creatine kinase or isoenzymes test system Regulatory Class: II Product Code: CGS, JIT Dated: August 2, 2016 Received: August 3, 2016
Dear Kathleen Dray-Lyons:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Katherine Serrano -S
For : Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K160724
Device Name ADVIA Chemistry® Creatine Kinase (CK L) Assay ADVIA Chemistry® Enzyme 3 Calibrator
### Indications for Use (Describe)
ADVIA Chemistry® Creatine Kinase (CK L) Assay:
The ADVIA Chemistry® Creatine Kinase (CK_L) assay is for in vitro diagnostic use in the quantitative determination of creatine kinase activity in human plasma (lithium heparin) or serum on ADVIA Chemistry systems. The assay can be used to aid in the diagnosis and treatment of myocardial infarction and muscle diseases, such as Duchenne progressive muscular dystrophy.
ADVIA Chemistry® Enzyme 3 Calibrator:
ADVIA Chemistry® Enzyme 3 Calibrator is intended for in vitro diagnostic use in the calibration of the ADVIA Chemistry Creatine Kinase (CK L) assay on the ADVIA Chemistry systems.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995,
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### 510(k) Summary
This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.
The assigned 510(k) number is: K160724
- 1. Manufacturer's Name, Address, Telephone, and Contact Person, Date of Preparation
| Applicant: | |
|------------|--|
|------------|--|
Contact Information:
Siemens Healthcare Diagnostics Inc. 500 GBC Drive Newark, DE 19714
Siemens Healthcare Diagnostics Inc. P.O. Box 6101 Newark, DE 19714 Attn: Kathleen Dray-Lyons Tel: 781-826-4551 Email: kathleen.a.dray-lyons@siemens.com
### Date of Preparation:
September 6, 2016
#### 2. Device Names:
- ADVIA Chemistry® Creatine Kinase (CK_L) Assay
ADVIA Chemistry® Enzyme 3 Calibrator
-
### Classification:
- 21 CFR §862.1215; Creatine phosphokinase/creatine kinase or O isoenzymes test system, Class II
- 21 CFR §862.1150; Calibrator Secondary, Class II O
### Product Code:
- o CGS
- o JIT
### Panel:
- o Clinical Chemistry
- Clinical Chemistry o
#### 3. Identification of the Predicate Devices:
ADVIA Chemistry Creatine Kinase (CKNAC) Assay - K991576 Dimension Vista ENZ 6 CAL - K083579
{4}------------------------------------------------
#### 4. Device Description:
### ADVIA Chemistry Creatine Kinase Assay:
ADVIA Chemistry Creatine Kinase (CK L) assay is a ready-to-use liquid reagent packaged for use on the automated ADVIA Chemistry systems. Creatine Kinase reacts with creatine phosphate and ADP to form adenosine triphosphate (ATP), which is coupled to the hexokinase-G6PD reaction, generating NADPH. The concentration of NADPH is measured by the increase in absorbance at 340/596 nm.
### ADVIA Chemistry ENZ 3 CAL:
ENZ 3 CAL is a liquid frozen human serum albumin based product containing creatine kinase MM from human heart. Enzyme 3 Calibrator kit consists of six vials of the same calibrator which is ready for use (no preparation is required).
#### 5. Device Intended Use:
### ADVIA Chemistry Creatine® Kinase (CK_L) Assay:
The ADVIA Chemistry® Creatine Kinase (CK L) assay is for in vitro diagnostic use in the quantitative determination of creatine kinase activity in human plasma (lithium heparin) or serum on ADVIA Chemistry systems. The assay can be used to aid in the diagnosis and treatment of myocardial infarction and muscle diseases, such as Duchenne progressive muscular dystrophy.
### ADVIA Chemistry ENZ 3 CAL:
ADVIA Chemistry® Enzyme 3 Calibrator is intended for in vitro diagnostic use in the calibration of the ADVIA Chemistry Creatine Kinase (CK L) assay on the ADVIA Chemistry systems.
#### 6. Medical device to which equivalence is claimed:
### ADVIA Chemistry Creatine® Kinase (CK L) Assay:
The modified ADVIA Chemistry CK L assay is substantially equivalent in intended use and technology to the predicate ADVIA Chemistry Creatine Kinase (CKNAC) assay cleared under K991576. Both assays utilize creatine kinase, which reacts with creatine phosphate and ADP to form ATP that is coupled to the hexokinase-G6PD reaction, generating NADPH. The concentration of NADPH is measured by the increase in absorbance. A comparison of the similarities and differences between the currently marketed ADVIA Chemistry CKNAC (predicate) versus the proposed modified ADVIA Chemistry CK L assay is provided in the table below.
{5}------------------------------------------------
### Similarities and Differences ADVIA Chemistry CK Assay (K991576) versus Modified ADVIA Chemistry CK_L Assay
| Feature | Predicate<br>ADVIA Chemistry CKNAC<br>K991576 | Proposed<br>ADVIA Chemistry CK_L<br>Modified Device |
|---------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use<br>Statement | For <i>in vitro</i> diagnostic use in the<br>quantitative determination of<br>creatine kinase activity in human<br>serum and plasma on the ADVIA<br>Chemistry systems. Such<br>measurements are used mainly<br>in the diagnosis and treatment of<br>myocardial infarction and muscle<br>diseases such as Duchenne<br>progressive muscular dystrophy. | The ADVIA Chemistry® Creatine<br>Kinase (CK_L) assay is for <i>in<br/>vitro</i> diagnostic use in the<br>quantitative determination of<br>creatine kinase activity in human<br>plasma (lithium heparin) or<br>serum on ADVIA Chemistry<br>systems. The assay can be<br>used to aid in the diagnosis and<br>treatment of myocardial<br>infarction and muscle diseases,<br>such as Duchenne progressive<br>muscular dystrophy. |
| Principle | Creatine Kinase reacts with<br>creatine phosphate and ADP to<br>form ATP which is coupled to the<br>hexokinase-G6PD reaction,<br>generating NADPH. The<br>concentration of NADPH is<br>measured by the increase in<br>absorbance at 340/410 nm. | Creatine Kinase reacts with<br>creatine phosphate and ADP to<br>form ATP which is coupled to the<br>hexokinase-G6PD reaction,<br>generating NADPH. The<br>concentration of NADPH is<br>measured by the increase in<br>absorbance at 340/596 nm. |
| Kit Components | Reagent 1 - lyophilized<br>Reagent 1 mix - lyophilized | Reagent 1 - liquid<br>Reagent 2 - liquid |
| Calibrator | None - fixed calibrator value | ADVIA Enzyme 3 Calibrator |
| Sample Type | Serum and lithium heparin<br>plasma | Same |
| Analytical Range | 0 to 1300 U/L | 15 to 1300 U/L |
| Extended Range | 1300 to 7800 U/L | Same |
| Analyzers | ADVIA Chemistry 1650, 1200,<br>1800, 2400 and XPT Systems | ADVIA Chemistry 1800 System |
### ADVIA Chemistry ENZ 3 CAL:
The ENZ 3 CAL is substantially equivalent in intended use to the predicate Dimension Vista ENZ 6 CAL (K083579). A comparison of the similarities and differences between the currently Dimension Vista ENZ 6 CAL versus the proposed ADVIA Chemistry ENZ 3 CAL is provided in the table below.
{6}------------------------------------------------
### Similarities and Differences Dimension Vista ENZ 6 CAL (K083579) versus ADVIA Chemistry ENZ 3 CAL
| Feature | Predicate<br>Dimension Vista® ENZ 6 CAL | Proposed<br>ADVIA Chemistry® ENZ 3 CAL |
|-------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | The ENZ 6 CAL is an in vitro<br>diagnostic product for the calibration<br>of Creatine Kinase (CKI) and<br>Creatine Kinase MB (MBI) methods<br>on the Dimension Vista® System | ADVIA Chemistry® Enzyme 3<br>Calibrator is intended for in vitro<br>diagnostic use in the calibration of<br>ADVIA Chemistry Creatine Kinase<br>(CK_L) assay on the ADVIA<br>Chemistry® systems. |
| Traceable to: | IFCC Reference Method | Same |
| Composition | ENZ 6 CAL is a liquid frozen human<br>serum albumin based product<br>containing creatinine kinase BB from<br>porcine brain and creatine kinase<br>MM from human heart. | ADVIA Chemistry ENZ 3 CAL is a<br>liquid frozen human serum albumin<br>(5%) and preservatives with lot-<br>specific concentrations of creatine<br>kinase MM from human heart. |
| Packaging Content | 3 vials: Calibrator A, 2mL per vial | 6 vials of calibrator, 2mL per vial |
| Storage | -20 °C or below | -25 to -15°C |
#### 7. Performance Characteristics
### Method Comparison:
A split sample method comparison between the ADVIA Chemistry CK_L assay versus the predicate ADVIA Chemistry CKNAC (K991576) assay was performed with 116 native de-identified human serum samples across the assay range (15 to 1300 U/L) on the ADVIA Chemistry 1800 system. Analysis of the results using Deming (Orthogonal) regression yielded the following:
| Method | n | Predicate<br>Method<br>Sample<br>Range<br>U/L | Proposed<br>Method<br>Sample<br>Range<br>U/L | Slope<br>(95% CI) | Intercept<br>U/L<br>(95% CI) | Correlation<br>Coefficient<br>(R) |
|----------------------------------|-----|-----------------------------------------------|----------------------------------------------|---------------------------|------------------------------|-----------------------------------|
| ADVIA CKNAC<br>versus ADVIA CK_L | 116 | 23 to 1253 | 22 to 1280 | 1.01<br>(1.00 to<br>1.02) | -1.8<br>(-3.3 to -<br>0.3) | 1.00 |
### Precision:
Testing was performed over twenty (20) days, two (2) runs per day, a single test from two (2) independent cups were analyzed for each test material using the ADVIA
{7}------------------------------------------------
Chemistry CK L assay on the ADVIA Chemistry 1800 System. Analysis of variance (ANOVA) was used to evaluate the data consistent with the recommendations of CLSI EP05-A2. Typical precision is summarized below.
| | | Repeatability | | Within-Lab | |
|---------------|----------|---------------|-----|------------|-----|
| Product | Mean U/L | SD | %CV | SD | %CV |
| BIORAD QC 1 | 75 | 1.3 | 1.7 | 1.9 | 2.6 |
| BIORAD QC 2 | 232 | 1.5 | 0.6 | 3.1 | 1.4 |
| BIORAD QC 3 | 639 | 2.6 | 0.4 | 8.1 | 1.3 |
| PLASMA POOL 1 | 1202 | 4.7 | 0.4 | 6.3 | 0.5 |
| SERUM POOL 1 | 85 | 1.7 | 2.0 | 2.9 | 3.4 |
| SERUM POOL 2 | 194 | 3.0 | 1.5 | 3.2 | 1.7 |
| SERUM POOL 3 | 938 | 5.2 | 0.6 | 5.6 | 0.6 |
# Repeatability and Within-Lab Results ADVIA CK_L on the ADVIA 1800
## Linearity
Linearity across the assay range was confirmed according to CLSI EP06-A, Evaluation of the Linearity of Quantitative Measurement, by testing a sample with a high concentration of creatine kinase on the ADVIA Chemistry CK_L assay and diluting it with low serum sample. Data were analyzed using weighted linear regression analysis [x-axis: theoretical concentration versus y-axis: measured concentration]. A 200 - and 36 -order polynomial regressions of the mean observed analyte values vs. expected concentrations were generated. These data support the measuring range of 15 to 1300 U/L. A summary of the linearity data is presented below.
| Range of<br>Samples | | Linear<br>Regression<br>Equation | Regression<br>Constant | R | P-Value | | |
|---------------------|------|----------------------------------|------------------------|------|--------------|----------|----------|
| Low | High | | | | Quadratic X2 | Cubic X2 | Cubic X3 |
| 1 | 1372 | Y=1.00X+0.5 | 0.5 | 1.00 | 0.2346 | 0.1962 | 0.2610 |
| | Expected | Observed Results (U/L) | | | | | | Linear | % Bias from the<br>Predicted | |
|------------------|------------------|------------------------|----------|----------|----------------|-----|------|--------------------|------------------------------|---------------|
| Sample<br>Number | Results<br>(U/L) | Result 1 | Result 2 | Result 3 | Mean<br>Values | SD | CV | Predicted<br>(U/L) | % Bias | Bias<br>(U/L) |
| 1 | 1 | 0 | 1 | 1 | 1 | 0.6 | 86.6 | 1 | N/A | 0 |
| 2 | 172 | 177 | 175 | 180 | 177 | 2.5 | 1.4 | 172 | 3 | 6 |
| 3 | 344 | 346 | 347 | 343 | 345 | 2.1 | 0.6 | 343 | 1 | 2 |
| 4 | 516 | 513 | 516 | 518 | 516 | 2.5 | 0.5 | 514 | 0 | 1 |
| 5 | 687 | 686 | 689 | 682 | 686 | 3.5 | 0.5 | 685 | 0 | 1 |
| 6 | 859 | 861 | 865 | 867 | 864 | 3.1 | 0.4 | 856 | 1 | 8 |
| 7 | 1031 | 1025 | 1023 | 1014 | 1021 | 5.9 | 0.6 | 1027 | -1 | -6 |
{8}------------------------------------------------
| Sample<br>Number | Expected<br>Results<br>(U/L) | Observed Results (U/L) | | | Mean<br>Values | SD | CV | Linear<br>Predicted<br>(U/L) | % Bias from the<br>Predicted | |
|------------------|------------------------------|------------------------|----------|----------|----------------|-----|-----|------------------------------|------------------------------|---------------|
| | | Result 1 | Result 2 | Result 3 | | | | | % Bias | Bias<br>(U/L) |
| 8 | 1202 | 1195 | 1196 | 1197 | 1196 | 1.0 | 0.1 | 1197 | 0 | -1 |
| 9 | 1374 | 1371 | 1370 | 1375 | 1372 | 2.6 | 0.2 | 1369 | 0 | 3 |
### Detection Capability
The LoB and LoD values are determined with proportions of false positives (α) less than 5% and false negatives (β) less than 5%, based on 450 determinations with 225 blank and 225 low-level sample replicates.
The LoB is the highest measurement result that is likely to be observed on a blank sample. The LoB for the ADVIA Chemistry CK_L assay is 3 U/L.
The LoD is the smallest amount that this assay can reliably detect to determine presence or absence of an analyte. The LoD for the ADVIA Chemistry CK L assay is 6 U/L.
The LoQ is the lowest amount of creatine kinase that can be determined quantitatively within a defined total error. The LoQ for the ADVIA Chemistry CK_L assay is 15 U/L.
### Interference Testing
Interference testing was performed according to CLSI/NCCLS EP07-A2: Approved Guideline Interference Testing in Clinical Chemistry to determine the effect of various endogenous and exogenous substances on the ADVIA Chemistry CK_L assay. For all interferents, the percent bias was determined by testing a control sample without the interferent and comparing it to the value obtained from a test sample to which the potential interferent had been added. Interferents were tested at two levels of creatine kinase, 95 ± 14 U/L and 265 ± 40 U/L. For each spiked sample, the % recovery was determined. A summary of the non-interfering substance data is presented below.
| Interferent | Interferent Test<br>Concentration | CK Test Level<br>(U/L) | % Difference |
|---------------------------|-----------------------------------|------------------------|--------------|
| Hemolysate | 125 (mg/dL) | 100 | 7% |
| | 125 (mg/dL) | 284 | 1% |
| Bilirubin<br>Conjugated | 60 (mg/dL) | 93 | 1% |
| | 60 (mg/dL) | 256 | 2% |
| Bilirubin<br>Unconjugated | 60 (mg/dL) | 94 | 2% |
| | 60 (mg/dL) | 264 | 2% |
| Lipemia<br>(Intralipid) | 1000 (mg/dL) | 94 | 5% |
| | 1000 (mg/dL) | 276 | 0% |
| Ascorbid Acid | 6 (mg/dL) | 91 | 4% |
### Interference Testing CK L on the ADVIA 1800
{9}------------------------------------------------
| Interferent | Interferent Test<br>Concentration | CK Test Level<br>(U/L) | % Difference |
|---------------|-----------------------------------|------------------------|--------------|
| | 6 (mg/dL) | 271 | 1% |
| Sulfasalazine | 300 (mg/L) | 94 | 0% |
| | 300 (mg/L) | 264 | 2% |
| Sulfapyridine | 300 (mg/L) | 93 | 1% |
| | 300 (mg/L) | 263 | 2% |
### Method Comparison versus IFCC Reference Method
A split sample method comparison between the ADVIA Chemistry CK_L assay on the ADVIA Chemistry 1800 system versus the IFCC certified reference method on the Konelab 30i analyzer (Thermo Fisher Scientific) was performed with 100 native de-identified human serum samples across the assay range (15 to 1300 U/L). Analysis of the results yielded the following:
| Method | n | Predicate<br>Method<br>Sample<br>Range<br>U/L | Proposed<br>Method<br>Sample<br>Range<br>U/L | Slope<br>(95% CI) | Intercept<br>U/L<br>(95% CI) | Correlation<br>Coefficient<br>(R) |
|--------------------------------------------|-----|-----------------------------------------------|----------------------------------------------|---------------------------|------------------------------|-----------------------------------|
| ADVIA CK_L versus<br>IFCC Reference Method | 100 | 21 to 1178 | 16 to 1245 | 1.05<br>(1.03 to<br>1.07) | -6.9<br>(-8.8 to -<br>4.9) | 1.00 |
### Matrix Comparison
To confirm the equivalency of sample types, the performance of the ADVIA Chemistry CK L assay was compared for serum and lithium heparin plasma samples. Testing was performed using one lot of reagents and a single replicate from a matched set of serum and plasma samples in accordance with CLSI EP09-A3.
| Specimen<br>Type | n | Serum<br>Sample<br>Range<br>U/L | Plasma<br>(Lithium<br>heparin)<br>Sample<br>Range<br>U/L | Slope | Intercept<br>U/L | Correlation<br>Coefficient<br>(R) |
|------------------------|----|---------------------------------|----------------------------------------------------------|-------|------------------|-----------------------------------|
| Serum versus<br>Plasma | 55 | 37–1282 | 39 to 1284 | 1.01 | -0.6 | 1.00 |
#### 8. Conclusion:
The proposed modified ADVIA Chemistry CK_L assay is substantially equivalent to the current legally marketed device based on intended use, principle and the performance characteristics above.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.