The BS-200 Chemistry Analyzer is designed for clinical laboratory use, making direct quantitative measurements of Na+ (sodium), K+ (potassium), Cl- (chloride) in serum, plasma and urine samples and Glucose in serum samples. Additionally, other various chemistry assays may be adaptable to the analyzer depending on the reagent used to induce a photometric reaction. Sodium measurements are used in the diagnosis and treatment diseases involving electrolyte imbalance. Potassium measurements monitor electrolyte balance and in the diagnosis and treatment of diseases conditions characterized by low or high blood potassium levels. Chloride measurements are used in the diagnosis and treatment of electrolyte and metabolic disorders. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.
Device Story
BS-200 Chemistry Analyzer is an automated in vitro diagnostic device for clinical laboratories. It processes serum, plasma, urine, or CSF samples. The device utilizes absorbance photometry for general chemistry assays and an optional Ion-Selective Electrode (ISE) module for electrolyte (Na+, K+, Cl-) measurement. Samples are analyzed using reagent-induced photometric reactions or ISE technology; concentrations are determined via standard curves generated from known calibrators. The system is operated by laboratory personnel. Output consists of quantitative analyte concentrations, which clinicians use to diagnose and monitor electrolyte imbalances and metabolic disorders, including diabetes and hypoglycemia. The device automates sample analysis to provide rapid, accurate diagnostic data for patient management.
Clinical Evidence
Bench testing only. Performance validated via precision studies (20-day, CLSI EP5-A2), linearity studies (CLSI EP6-A), and method comparison studies against predicate devices (Hitachi 917 and EasyElectrolytes). Glucose method comparison (N=60) showed r=0.9992. ISE serum/urine method comparisons (N=40 each) showed high correlation (r=0.9868–0.9992). Interference testing performed for bilirubin, hemoglobin, and intralipid.
Technological Characteristics
Automated discrete chemistry analyzer. Photometric module (tungsten halogen lamp, photodiode detector) and ISE module. Analytical methods: endpoint, fixed-time, kinetic. Calibration: linear and nonlinear. Connectivity: LIS external capability, barcode reader. Power: 100-130V. Software: microprocessor-controlled. Standards: IEC 61010-1, ISO 14971, CLSI EP05-A2, EP06-A, EP07-A2, EP09-A2, EP17-A.
Indications for Use
Indicated for clinical laboratory use to perform direct quantitative measurements of Na+ (sodium), K+ (potassium), Cl- (chloride) in serum, plasma, and urine, and Glucose in serum. Used for diagnosis and treatment of electrolyte imbalance, metabolic disorders, and carbohydrate metabolism disorders (e.g., diabetes mellitus, hypoglycemia, pancreatic islet cell carcinoma).
Regulatory Classification
Identification
A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.
Special Controls
*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
Predicate Devices
Roche, Hitachi 911 analyzer (k953239)
Point Scientific Glucose reagent (k002199)
Medica Easyelectrolyte Analyzer and Rapidlyte Analyzer (k000926)
K100060 — EON 100 CHEMISTRY ANALYZER WITH ISE , MODEL 77105, BICARBONATE REAGENT, MODEL 77475 · Vital Diagnostics (Manufacturing) Ptyltd · Dec 15, 2010
{0}------------------------------------------------
FEB - 1 2008
# 510(K) SUMMARY
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR § 807.92.
The assigned 510(k) number is: _ K072018
#### Submitter:
Shenzhen Mindray Bio-medical Electronics Co., LTD Mindray Building, Keji 12th Road South, Hi-tech Industrial Park, Nanshan, Shenzhen, 518057, P. R. China
Tel: +86 755 2658 2888 Fax: +86 755 2658 2680
- Contact Person: ●
Li Dongling
Shenzhen Mindray Bio-medical Electronics Co., LTD Mindray Building, Keji 12th Road South, Hi-tech Industrial Park, Nanshan, Shenzhen, 518057, P. R. China
#### Date Prepared: ●
Mar. 5, 2007
#### Name of the device:
- . Trade/Proprietary Name: BS-200 Chemistry Analyzer
- . Common Name: Clinical Chemistry Analyzer (with optional ISE Module)
#### Classification Number/Class: ●
75JJE, Class I 75CRF, Class II 75CEM, Class II 75CGE, Class II 75JGS, Class II
{1}------------------------------------------------
#### Legally Marketed Predicate Device:
| K953239 | Boehringer Mannheim/Hitachi 917 Analyzer |
|---------|--------------------------------------------|
| K000926 | EasyElectroLyte/RapidLyte Na/K/Cl Analyzer |
| K002199 | Liquid Glucose (Hexokinase) Reagent Set |
#### Description:
The BS-200 Chemistry Analyzer is an automated clinical chemistry analyzer capable of performing various in vitro photometric assays. The Glucose was cleared under K002199 and is the chosen assay to demonstrate performance for the photometric unit. The BS-200 Chemistry Analyzer has an optional Ion-Selective Electrode (ISE) module which measures the concentration of the electrolytes, sodium, potassium, and chloride, in samples using ion selective electrode technology.
#### Intended Use:
The BS-200 Chemistry Analyzer is an automated chemistry analyzer for in vitro diagnostic use in clinical laboratories. The analyzer is designed for the in vitro quantitative determination of clinical chemistries in serum, plasma, urine or cerebral spinal fluid samples.
#### Indications for Use:
The BS-200 Chemistry Analyzer is designed for clinical laboratory use, making direct quantitative measurements of Na+ (sodium), K+ (potassium), Cl-(chloride) in serum, plasma and urine samples and Glucose in serum samples. Additionally, other various chemistry assays may be adaptable to the analyzer depending on the reagent used to induce a photometric reaction.
Sodium measurements are used in the diagnosis and treatment diseases involving electrolyte imbalance.
Potassium measurements monitor electrolyte balance and in the diagnosis and treatment of diseases conditions characterized by low or high blood potassium levels.
Chloride measurements are used in the diagnosis and treatment of electrolyte and metabolic disorders.
Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.
### Comparison of Technological Characteristics:
Substantial equivalence has been demonstrated between the BS-200 Chemistry Analyzer and Boehringer Mannheim/Hitachi 917 analyzer. Both of them utilize absorbance photometry to
{2}------------------------------------------------
perform and output quantitative results for kinetic and endpoint clinical chemistries. For analytes, the BS-200 Chemistry Analyzer and Boehringer Mannheim/Hitachi 917 analyzer determine the concentration of unknown samples from a standard curve generated with known analyte concentrations. Substantial equivalence has also been demonstrated between the BS-200 Chemistry Analyzer and EasyElectroLyte/RapidLyte Na/K/Cl Analyzer. Both of them are used to analyze for electrolytes. The analyzers are calibrated with known concentration callbrator material. The BS-200 Chemistry Analyzer and EasyElectroLyte/RapidLyte Na/K/Cl Analyzer both utilize Ion-Selective Electrodes technology.
#### Performance Characteristics:
Performance testing of the BS-200 Chemistry Analyzer consisted of running the FDA previously cleared assay and the ISE module on the BS-200 to evaluate precision, linearity, and method comparison, Limits of Detection and Limits of Quantitation, interference.
| Item | Regression<br>Slope | Regression<br>Intercept | Correlation<br>Coefficient(r) | Sample<br>numbers |
|------------------------|---------------------|-------------------------|-------------------------------|-------------------|
| GLU (mg/dL) | 1.015 | -0.0288 | 0.9992 | 60 |
| K (mmol/L) serum(ISE) | 0.9864 | 0.1226 | 0.9980 | 40 |
| Na (mmol/L) serum(ISE) | 0.9903 | 1.3158 | 0.9845 | 40 |
| CL (mmol/L) serum(ISE) | 0.9815 | 3.4021 | 0.9905 | 40 |
| K (mmol/L)Urine(ISE) | 0.9668 | 0.5307 | 0.9991 | 40 |
| Na (mmol/L) Urine(ISE) | 0.9479 | -3.1284 | 0.9977 | 40 |
| CL (mmol/L)Urine(ISE) | 1.001 | -6.9595 | 0.9887 | 40 |
A correlation analysis between the BS-200 Chemistry Analyzer and the predicate devices yielded the following results:
The Within-Run precision test yielded the following results:
Glucose
| Level I | | | Level II | | | Level III | | |
|---------|-------|------|----------|-------|------|-----------|-------|------|
| Mean | SD | CV | Mean | SD | CV | Mean | SD | CV |
| mg/dL | mg/dL | % | mg/dL | mg/dL | % | mg/dL | mg/dL | % |
| 55.9 | 0.51 | 0.9% | 121.5 | 0.73 | 0.6% | 295.63 | 1.33 | 0.5% |
ISE Chemistries
| ltem | Level | | | Level J' | | | |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------|------|------|----------|------|------|--|
| | Mean | ST | CV% | Mean | SC | CV% | |
| K serum(ISE)<br>A MOND AND A CONTRACT AND CONSULT OF COLLECTION OF COLLECTION OF COLLECTION OF COLLECTION OF COLLECTION OF COLLECTION OF COLLECTION OF COLLECTION OF COLLECTION OF COLLECTION | | 0.03 | 0.7% | 6.77 | 0.05 | 0.8% | |
Page 3 of 5
{3}------------------------------------------------
| Na serum(ISE) | 136.13 | 0.61 | 0.4% | 146.74 | 0.72 | 0.5% |
|---------------|--------|------|------|--------|------|------|
| CL serum(ISE) | 99.59 | 0.73 | 0.7% | 115.68 | 0.52 | 0.5% |
| K Urine(ISE) | 37.09 | 0.46 | 1.2% | 67.49 | 0.63 | 0.9% |
| Na Urine(ISE) | 99.23 | 1.13 | 1.1% | 170.84 | 1.36 | 0.8% |
| CL Urine(ISE) | 82.90 | 1.36 | 1.6% | 132.16 | 1.24 | 0.9% |
The Between-run imprecision test yielded the following results:
Glucose
| Level I | | | Level II | | | Level III | | |
|---------|-------|------|----------|-------|------|-----------|-------|------|
| Mean | SD | CV | Mean | SD | CV | Mean | SD | CV |
| mg/dL | mg/dL | % | mg/dL | mg/dL | % | mg/dL | mg/dL | % |
| 55.9 | 1.51 | 2.7% | 121.5 | 2.46 | 2.0% | 295.63 | 5.28 | 1.8% |
ISE Chemistries
| Item | Level I | | | Level II | | |
|---------------------------|---------|------|------|----------|------|------|
| | Mean | SD | CV% | Mean | SD | CV% |
| K (mmol/L)<br>serum(ISE) | 4.11 | 0.07 | 1.6% | 6.77 | 0.06 | 1.0% |
| Na (mmol/L)<br>serum(ISE) | 136.13 | 2.12 | 1.6% | 146.74 | 1.66 | 1.1% |
| CL (mmol/L)<br>serum(ISE) | 99.59 | 1.78 | 1.8% | 115.68 | 1.13 | 1.0% |
| K (mmol/L)<br>Urine(ISE) | 37.09 | 0.48 | 1.3% | 67.49 | 0.70 | 1.0% |
| Na (mmol/L)<br>Urine(ISE) | 99.23 | 1.68 | 1.7% | 170.84 | 2.41 | 1.4% |
| CL (mmol/L)<br>Urine(ISE) | 82.90 | 1.50 | 1.8% | 132.16 | 2.12 | 1.6% |
#### The linearity test yielded the following results:
| | Linear range | |
|------------------------|--------------|-------------|
| Item | Lower limit | Upper limit |
| GLU (mg/dL) | 1 | 686 |
| K (mmol/L) serum(ISE) | 1.1 | 8.6 |
| Na (mmol/L) serum(ISE) | 113 | 194 |
| CL (mmol/L) serum(ISE) | 53 | 154 |
{4}------------------------------------------------
| K (mmol/L)Urine(ISE) | 13 | 184 |
|------------------------|----|-----|
| Na (mmol/L) Urine(ISE) | 27 | 372 |
| CL (mmol/L)Urine(ISE) | 42 | 422 |
The Limit of Detection test yielded the following results:
| Item | LoB | LoD | LoQ |
|-------------------------|------|------|------|
| GLU (mg/dl) | 0.04 | 0.22 | 0.97 |
| K (mmol/L) serum (ISE) | 0.1 | 0.16 | 0.42 |
| Na (mmol/L) serum(ISE) | 1.15 | 2.67 | 5.3 |
| CL (mmol/L) serum (ISE) | 2.95 | 5.0 | 12.0 |
| K (mmol/L)Urine(ISE) | 0.26 | 0.58 | 1.34 |
| Na (mmol/L) Urine(ISE) | 5.85 | 8.27 | 22.2 |
| CL (mmol/L)Urine(ISE) | 4.5 | 6.29 | 17.1 |
The Interference test yielded the following results:
| | Interference materials | | |
|------------------------|------------------------|-----------|--------------|
| Item | Hemoglobin | Bilirubin | Intralipids® |
| GLU (mg/dl) | 250mg/dl | 18 mg/dl | 300mg/dl |
| K (mmol/L) serum(ISE) | >500mg/dl | >20mg/dl | >1000 mg/dl |
| Na (mmol/L) serum(ISE) | >500mg/dl | >20mg/dl | >1000 mg/dl |
| CL (mmol/L) serum(ISE) | >500mg/dl | >20mg/dl | >1000 mg/dl |
## Conclusion:
The data demonstrates that the BS-200 Chemistry Analyzer is substantially equivalent to Boehringer Mannheim/Hitachi 917 Analyzer and EasyElectroLyte/RapidLyte Na/K/C1 Analyzer.
{5}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with its wings spread, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" is arranged in a circular pattern around the eagle. The logo is black and white.
Public Health Service
FEB - 1 2008
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
Shenzhen Mindray Bio-Medical Electronics Co., Ltd. c/o Susan D. Goldstein-Falk mdi Consultants, Inc. 55 Northern Blvd., Suite 200 Great Neck, NY 11021 US
Re: k072018 Trade/Device Name: BS-200 Chemistry Analyzer Regulation Number: 21 CFR 862.1345 Regulation Name: Glucose Test System Regulatory Class: Class II Product Code: CFR, JGS, CEM, CGZ, JJE Dated: January 25, 2008 Received: January 28, 2008
Dear Ms. Goldstein Falk:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).
{6}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240) 276-0490. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address at http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Jean M. Cooper, M.S., D.V.M.
Jean M. Cooper, M.S., D.V.M. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
### Indication for Use
510(k) Number (if known): K072018 Device Name: Nat, K+, Cl , glucose tests on for BS-200 Analyze Indication For Use:
The BS-200 Chemistry Analyzer is designed for clinical laboratory use, making direct quantitative measurements of Na+ (sodium), K+ (potassium), Cl-(chloride) in serum, plasma and urine samples and Glucose in serum samples. Additionally, other various chemistry assays may be adaptable to the analyzer depending on the reagent used to induce a photometric reaction.
Sodium measurements are used in the diagnosis and treatment diseases involving electrolyte imbalance.
Potassium measurements monitor electrolyte balance and in the diagnosis and treatment of diseases conditions characterized by low or high blood potassium levels.
Chloride measurements are used in the diagnosis and treatment of electrolyte and metabolic disorders.
Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of panereatic islet cell carcinoma.
Prescription Use x (21 CFR Part 801 Subpart D) And/Or
Over the Counter Use (21 CFR Part 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OIVD)
Signature
Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety
s10(k) K072018
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.