AXSYM FREE TRIIODOTHYRONINE (FREE T3)
Device Facts
| Record ID | K963827 |
|---|---|
| Device Name | AXSYM FREE TRIIODOTHYRONINE (FREE T3) |
| Applicant | Abbott Laboratories |
| Product Code | CDP · Clinical Chemistry |
| Decision Date | Oct 30, 1996 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 862.1710 |
| Device Class | Class 2 |
Intended Use
AxSYM Free T₃ is a microparticle enzyme immunoassay for the quantitative determination of free T₃ in human serum or plasma (lithium heparin, sodium heparin, or dipotassium EDTA).
Device Story
Microparticle enzyme immunoassay (MEIA) for quantitative measurement of free T₃ in human serum or plasma; performed on Abbott AxSYM System; provides clinicians with thyroid status assessment; utilizes Abbott calibrators and controls for accuracy/precision verification.
Clinical Evidence
Bench testing only. Correlation study comparing AxSYM Free T₃ to Ciba Corning ACS Free T₃ using 745 specimens; results: r=0.948, slope=1.19, SEE=0.775, intercept=-0.96 pg/mL.
Technological Characteristics
Microparticle enzyme immunoassay (MEIA) technology; automated system; compatible with human serum, lithium heparin plasma, sodium heparin plasma, and dipotassium EDTA plasma.
Indications for Use
Indicated for quantitative determination of free T₃ in human serum or plasma (lithium heparin, sodium heparin, or dipotassium EDTA) to aid in the assessment of thyroid status.
Regulatory Classification
Identification
A total triiodothyronine test system is a device intended to measure the hormone triiodothyronine in serum and plasma. Measurements obtained by this device are used in the diagnosis and treatment of thyroid diseases such as hyperthyroidism.
Predicate Devices
- Ciba Corning Automated Chemiluminescence System (ACS) Free T₃ Immunoassay
Related Devices
- K963833 — IMX FREE TRIIODOTHYRONINE (FREE T3) · Abbott Laboratories · Oct 30, 1996
- K983439 — ABBOTT ARCHITECT FREE T3 · Abbott Laboratories · Nov 20, 1998
- K983434 — ABBOTT ARCHITECT TOTAL T3 · Abbott Laboratories · Nov 20, 1998
- K983417 — ABBOTT ARCHITECT FREE T4 · Abbott Laboratories · Nov 20, 1998
- K962355 — FREE TRIIODOTHYRONINE(FT3) ASSAY FOR THE TECHNICON IMMUNO 1 SYSTEM (IN-VITRO DIAGNOSTIC SYSTEM) · Bayer Corp. · Jul 12, 1996