AUTION MAX AX-4060 Urinalysis System

K193514 · Arkray, Inc. · JIL · Apr 13, 2023 · Clinical Chemistry

Device Facts

Record IDK193514
Device NameAUTION MAX AX-4060 Urinalysis System
ApplicantArkray, Inc.
Product CodeJIL · Clinical Chemistry
Decision DateApr 13, 2023
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1340
Device ClassClass 2

Indications for Use

The AUTION MAX AX-4060 urine analyzer, when used with AUTION Sticks 9EB Test Strips is a fully automated urinalysis system intended for the in vitro qualitative or semi-quantitative measurement of the following analytes: glucose, protein, bilirubin, urobilinogen, pH, blood, ketones, nitrites, leukocytes, turbidity, and color. The test results of these parameters can be used in the evaluation of kidney, urinary, liver and other metabolic disorders. This system is intended to be used by trained operators in clinical laboratories. AUTION Sticks 9EB Test strips are test strips for the in vitro qualitative or semi-quantitative measurement of the following analytes: glucose, protein, bilirubin, urobilinogen, pH, blood, ketones, nitrite, and leukocytes with the AUTION MAX AX-4060 urine analyzer. The test results of these parameters can be used in the evaluation of kidney, urinary, liver and other metabolic disorders.

Device Story

The AUTION MAX AX-4060 is a fully automated urine analyzer used in clinical laboratories by trained operators. It processes urine samples using AUTION Sticks 9EB test strips. The system uses spectrophotometry to measure color changes on reagent pads resulting from chemical reactions with urine analytes. Additionally, it employs reflectance refractometry, transmitted light, and scattered light measurements to determine specific gravity, turbidity, and color tone. The device automates the analysis process, providing semi-quantitative or qualitative results that assist healthcare providers in evaluating metabolic, kidney, urinary, and liver disorders. The system includes a sampler, barcode reader, and touch-panel display. It replaces the AX-4030 model, featuring updated LED and nozzle dispensing units to maintain performance equivalence.

Clinical Evidence

Method comparison study conducted at two clinical laboratories using 1,374 samples (1,333 natural, 41 spiked). Results showed high agreement with the predicate device (e.g., 98.1% exact match for glucose, 99.5% for bilirubin, 99.7% for nitrite). Precision (repeatability/reproducibility) was evaluated per CLSI EP5-A3 using three levels of controls over 20 days, demonstrating high exact match percentages. Interference testing identified specific limitations regarding ascorbic acid, hematuria, and high-dose medication effects.

Technological Characteristics

System uses dry chemistry reagent pads on plastic strips. Sensing: spectrophotometry (reflectance) for analytes; reflectance refractometry for specific gravity; transmitted/scattered light for turbidity/color. Energy: electrical. Connectivity: barcode scanner for sample ID. Software: automated processing and color correction algorithm. Sterilization: N/A.

Indications for Use

Indicated for trained operators in clinical laboratories for the in vitro qualitative or semi-quantitative measurement of urine analytes (glucose, protein, bilirubin, urobilinogen, pH, blood, ketones, nitrite, leukocytes, turbidity, and color) to evaluate kidney, urinary, liver, and metabolic disorders. Prescription use only.

Regulatory Classification

Identification

A urinary glucose (nonquantitative) test system is a device intended to measure glucosuria (glucose in urine). Urinary glucose (nonquantitative) measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, hypoglycemia, and hyperglycemia.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD & DRUG ADMINISTRATION # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT ## I Background Information: A 510(k) Number K193514 B Applicant Arkray Inc. C Proprietary and Established Names AUTION MAX AX-4060 Urinalysis System D Regulatory Information | Product Code(s) | Classification | Regulation Section | Panel | | --- | --- | --- | --- | | JIL | Class II | 21 CFR 862.1340 - Urinary Glucose (Nonquantitative) Test System | CH - Clinical Chemistry | | JIO | Class II | 21 CFR 864.6550 - Occult blood test | HE - Hematology | | CDM | Class I | 21 CFR 862.1785 - Urinary urobilinogen (nonquantitative) test system | CH - Clinical Chemistry | | JIR | Class I | 21 CFR 862.1645 - Urinary protein or albumin (nonquantitative) test system | CH - Clinical Chemistry | | JMT | Class I | 21 CFR 862.1510 - Nitrite (nonquantitative) test system | CH - Clinical Chemistry | Food and Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993-0002 www.fda.gov {1} | LJX | Class I | 21 CFR 864.7675 - Leukocyte peroxidase test | HE - Hematology | | --- | --- | --- | --- | | JJB | Class I | 21 CFR 862.1115 - Urinary bilirubin and its conjugates (nonquantitative) test system | CH - Clinical Chemistry | | CEN | Class I | 21 CFR 862.1550 - Urinary pH (nonquantitative) test system | CH - Clinical Chemistry | | JIN | Class I | 21 CFR 862.1435 - Ketones (nonquantitative) test system | CH - Clinical Chemistry | | KQO | Class I | 21 CFR 862.2900 - Automated urinalysis system | CH - Clinical Chemistry | ## II Submission/Device Overview: A Purpose for Submission: New device B Measurand: Measurement of the following in urine samples: glucose, protein, bilirubin, urobilinogen, pH, blood, ketones, nitrite, leukocytes, turbidity, and color. C Type of Test: Qualitative and semi-quantitative urinalysis ## III Intended Use/Indications for Use: A Intended Use(s): See Indications for Use below. K193514 - Page 2 of 20 {2} B Indication(s) for Use: The AUTION MAX AX-4060 Urinalysis System (AUTION MAX) is comprised of the AUTION MAX AX-4060 automated urine analyzer and AUTION Sticks 9EB multi-parameter test strips. The AUTION MAX AX-4060 urine analyzer, when used with AUTION Sticks 9EB Test Strips is a fully automated urinalysis system intended for the in vitro qualitative or semi-quantitative measurement of the following analytes: glucose, protein, bilirubin, urobilinogen, pH, blood, ketones, nitrites, leukocytes, turbidity, and color. The test results of these parameters can be used in the evaluation of kidney, urinary, liver and other metabolic disorders. This system is intended to be used by trained operators in clinical laboratories. AUTION Sticks 9EB Test strips are test strips for the in vitro qualitative or semi-quantitative measurement of the following analytes: glucose, protein, bilirubin, urobilinogen, pH, blood, ketones, nitrite, and leukocytes with the AUTION MAX AX-4060 urine analyzer. The test results of these parameters can be used in the evaluation of kidney, urinary, liver and other metabolic disorders. Special conditions for use statements: Prescription use only. AUTION Sticks 9EB test strips are not to be read visually. C Special Conditions for Use Statement(s): Rx - For Prescription Use Only AUTION Sticks 9EB test strips are not to be read visually. D Special Instrument Requirements: AUTION MAX AX-4060 urine analyzer IV Device/System Characteristics: A Device Description: The AUTION MAX AX-4060 Urinalysis System consists of the AUTION MAX AX-4060 urine analyzer and the AUTION Sticks 9EB test strips. The test system is fully automated and provides a semi-quantitative or qualitative measurement for glucose, protein, bilirubin, urobilinogen, pH, blood, ketones, nitrites and leukocytes. The AUTION sticks 9EB test strips consist of a plastic strip containing 9 dry chemistry reagent pads impregnated with chemical substances for the determination of the above analytes in urine. These substances if present in urine lead to a chemical reaction that results in a color change, which is measured by the AUTION MAX AX-4060 urine analyzer based on spectrophotometry. In addition, three additional parameters, specific gravity, turbidity, and color tone are directly measured based on reflectance refractometry, transmitted, and scattered light measurements. K193514 - Page 3 of 20 {3} B Principle of Operation: AUTION Sticks 9EB test strips Glucose: Based on a double sequential enzyme reaction. Glucose oxidase catalyzes the formation of gluconic acid and hydrogen peroxide from the oxidation of glucose. A second enzyme, peroxidase, catalyzes the oxidative coupling of 4-amino-antipyrine and 1-Naphthol3-6-disulfonic acid by hydrogen peroxide to form a quinone imine purple color. Protein: Based on the protein-error-of-indicators reaction. At a constant pH, the presence of protein causes a change in the color of the indicator to a cyan color. Bilirubin: The test is based on azo-coupling reaction of bilirubin with diazotized 2-methyl5-nitroaniline in acidic medium. Varying bilirubin levels will produce a reddish-brown color. Urobilinogen: The test is based on azo-coupling reaction of urobilinogen with diazotized 3,3'-dimethyloxy-4,4'-biphenyl bis-diazonium tetrafluoroborate in acidic medium. Varying urobilinogen levels will produce a reddish-brown color. pH: Based on a mixed indicator principle that gives a broad range of colors (yellow to cyan) covering the entire urinary pH range. Ketones: Legal reaction. Based on the reaction of sodium nitroprusside with acetoacetic acid to form a purple complex. Nitrites: Griess reaction. Nitrite reacts with sulfanilamide, followed by a diazo-coupling reaction to form a pink colored product. Leukocytes: Granulocytic leukocytes contain esterases that catalyze the hydrolysis of 3-(NToluenesulfonyl-L-alanyloxy) indole (TAI) to form indoxyl, which further reacts with 2-Methoxy-4-(N-morpholino) benzenediazonium (MMB) to form a purple product. Blood: Based on the peroxidase-like activity of hemoglobin, which catalyzes the reaction of Cumene hydroperoxide and 3,3,5,5'-tetramethylbenzidine to form a cyan color. AUTION MAX AX-4060 urine analyzer Color: Urine color tone measurement, red (635 nm), green (525 nm) and blue (470 nm) lights are transmitted to the sample in the cylindrical tube, and the hue and contrast of the sample are obtained from the amount of transmitted light. Urine color is reported as colorless, yellow, orange, brown, red, violet, blue, and green. Light, normal, or dark are also reported for each color (for a total of 22 contrast and hue indications). The urine color measurement is also used in this urinalysis system for a color correction algorithm to compensate for intrinsically colored urine samples when testing for bilirubin, urobilinogen, and ketone analyte pads. The measurements glucose, protein, pH, nitrites, leukocytes, and blood are not compensated for urine color. K193514 - Page 4 of 20 {4} Two wavelengths are used for the reflectance determination of each analyte, except blood which uses a single wavelength. The percent reflectance received by the optical sensor for each analyte on the test strip is then used to calculate a semiquantitative value and a qualitative value. An internal turbidity/color tone measurement unit measures sample turbidity based on the calculation of scattered and transmitted light. Specific gravity is calculated based on the light refraction of the urine sample (reflection-refractometry). ## C Instrument Description Information: 1. Instrument Name: AUTION MAX AX-4060 urine analyzer 2. Specimen Identification: Urine identification can be performed manually or with the analyzer barcode scanner. 3. Specimen Sampling and Handling: Samples tubes are manually placed into a rack, which is loaded into the analyzer. Alternatively, samples can be placed individually into the STAT port. Samples can be labeled with a barcode, which is read with a barcode reader. Test strip picking, sample transport, sample aspiration, sample dispensing onto the test strip, and measurement are automated by the instrument. 4. Calibration: The AUTION MAX AX-4060 urine analyzer requires optical unit calibration. For optical unit calibration, the instrument only needs to be calibrated if the reflectivity for each wavelength falls outside of the defined range in the measurement check and/or trouble with the instrument is suspected. 5. Quality Control: The sponsor states the following in their labeling: "Performance of the AUTION MAX AX-4060 and the AUTION Sticks 9EB should be confirmed regularly utilizing both known negative and known positive urine specimens or control materials." ## V Substantial Equivalence Information: A Predicate Device Name(s): AUTION MAX AX-4030 Urinalysis System B Predicate 510(k) Number(s): K093098 K193514 - Page 5 of 20 {5} C Comparison with Predicate(s): | Device & Predicate Device(s): | K193514 | K093098 | | --- | --- | --- | | Device Trade Name | AUTION MAX AX-4060 Urinalysis System | AUTION MAX AX-4030 Urinalysis System | | General Device Characteristic Similarities | | | | Intended Use/Indications For Use | Intended for the in vitro measurement of the following analytes: glucose, protein, bilirubin, urobilinogen, pH, blood, ketones, nitrite, leukocytes. | same | | Specimen Type | Urine | same | | General Device Characteristic Differences | | | | LED Wavelength | 5 LED wavelengths (450, 525, 562, 635 and 760 nm) | 5 LED wavelengths (430, 500, 565, 635, 760 nm) | VI Standards/Guidance Documents Referenced: Clinical and Laboratory Standards Institute (CLSI) EP05-A3: Evaluation of Precision Performance of Clinical Devices; Approved Guideline—Third Edition CLSI EP12-A2: User Protocol for Evaluation of Qualitative Test Performance; Approved Guideline—Second Edition International Electrotechnical Commission (IEC) 60601-1-2:2014: Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests IEC 61010-1 (Edition 3.1): Safety requirements for electrical equipment for measurement, control, and laboratory use - Part 1: General requirements [Including: Corrigendum 1 (2019)] VII Performance Characteristics (if/when applicable): A Analytical Performance: 1. Precision/Reproducibility: Precision study was conducted in accordance with the general guidelines established in CLSI EP05-A3 and CLSI EP12-A2. K193514 - Page 6 of 20 {6} Three (3) independent AX-4060 Analyzers and Test Strip lot sets were tested according to a 20 day, two (2) runs per day, and two (2) replicates (duplicates). The study was performed by four (4) operators. For each analyte, three sample levels (Negative/Normal, Mid, and High) were tested. Within run (repeatability) and total imprecision (reproducibility) were calculated. The results are displayed below: Precision - Negative | Analyte | Expected Result | Repeatability (N=240) | | Reproducibility (N=240) | | | --- | --- | --- | --- | --- | --- | | | | Exact match | Within ± 1 CB | Exact match | Within ± 1 CB | | Glucose | Normal | 100% | 100% | 100% | 100% | | Protein | Negative | 100% | 100% | 100% | 100% | | Bilirubin | Normal | 100% | 100% | 100% | 100% | | Urobilinogen | Negative | 100% | 100% | 100% | 100% | | pH | 5 | 99% | 100% | 100% | 100% | | Blood | Negative | 100% | 100% | 100% | 100% | | Ketones | Negative | 100% | 100% | 100% | 100% | | Nitrite | Negative | 100% | 100% | 100% | 100% | | Leukocytes | Negative | 100% | 100% | 100% | 100% | | Turbidity | Negative | 100% | 100% | 100% | 100% | | Color | Colorless | 100% | 100% | 100% | 100% | Precision - Positive test | Analyte | Expected Result | Repeatability (N=240) | | Reproducibility (N=240) | | | --- | --- | --- | --- | --- | --- | | | | Exact match | Within ± 1 CB | Exact match | Within ± 1 CB | | Glucose | 1+ | 100% | 100% | 100% | 100% | | Protein | 1+ | 91% | 100% | 95% | 100% | | Bilirubin | 2+ | 100% | 100% | 100% | 100% | | Urobilinogen | 2+ | 100% | 100% | 96% | 100% | | pH | 6.5 | 99% | 100% | 100% | 100% | | Blood | 1+ | 100% | 100% | 100% | 100% | | Ketones | 1+ | 98% | 100% | 99% | 100% | | Nitrite | 1+ | 100% | 100% | 100% | 100% | | Leukocytes | 250 (Leu/μL) | 100% | 100% | 100% | 100% | | Turbidity | 1+ | 100% | 100% | 100% | 100% | | Color | Yellow | 100% | 100% | 100% | 100% | K193514 - Page 7 of 20 {7} Precision – High Positive | Analyte | Expected Result | Repeatability (N=240) | | Reproducibility (N=240) | | | --- | --- | --- | --- | --- | --- | | | | Exact match | Within ± 1 CB | Exact match | Within ± 1 CB | | Glucose | 3+ | 100% | 100% | 100% | 100% | | Protein | 3+ | 91% | 100% | 95% | 100% | | Bilirubin | 3+ | 100% | 100% | 100% | 100% | | Urobilinogen | 4+ | 100% | 100% | 96% | 100% | | pH | 5 | 99% | 100% | 100% | 100% | | Blood | 2+ | 100% | 100% | 100% | 100% | | Ketones | 3+ | 98% | 100% | 99% | 100% | | Nitrite | 2+ | 100% | 100% | 100% | 100% | | Leukocytes | >500 (Leu/μL) | 100% | 100% | 100% | 100% | | Turbidity | 2+ | 100% | 100% | 100% | 100% | | Color | Dark Yellow | 100% | 100% | 100% | 100% | 2. Linearity: The sponsor performed other studies to establish sensitivity/cutoffs. Refer to the Detection Limit and Sensitivity studies in section 6 below, for more information. 3. Analytical Specificity/Interference: Analytical specificity studies were conducted to assess for interfering effects of endogenous substances (including pH and specific gravity) and therapeutic drugs on the AUTION MAX AX-4060 Urinalysis System. Testing was performed on one (1) analyzer using one (1) lot of test strips and conducted with urine samples prepared at two concentrations for each analyte, with six (6) replicates per sample. A high positive sample was prepared by spiking each analyte into pooled negative urine. A negative sample pool was also used. The sample pools were spiked with potential interfering substances. Interference was defined when the color block changed by more than $\pm 1$. The results of testing depicting the highest concentration of the interferent tested with no effect, are provided below: Substances with no interference at the highest levels tested | Interfering Substances | Highest Concentration Tested with No Interference | | --- | --- | | Therapeutic drugs | | | Acetaminophen | 40 mg/dL | | Acetoacetic Acid Lithium | 150 mg/dL | | Ammonium Chloride | 400 mg/dL | | Amoxicillin | 360 mg/dL | | Calcium Chloride | 300 mg/dL | | Citric Acid Monohydrate | 75 mg/dL | | Methyl-dopa | 40 mg/dL | | Oxalic Acid Dihydrate | 70 mg/dL | K193514 - Page 8 of 20 {8} | Interfering Substances | Highest Concentration Tested with No Interference | | --- | --- | | Therapeutic drugs | | | Potassium Chloride | 1000 mg/dL | | Sodium Chloride | 2000 mg/dL | | Sodium Dihydrogen Phosphate | 1000 mg/dL | | Sodium Nitrate | 10 mg/dL | | Sodium Nitrite | 2 mg/dL | | Endogenous substances | | | Albumin | 1000 mg/dL | | Creatinine | 600 mg/dL | | Fructose | 100 mg/dL | | Galactose | 100 mg/dL | | Glucose | 2000 mg/dL | | Glycine | 450 mg/dL | | Lactose | 108 mg/dL | | Low Urine pH | pH 5 | | Riboflavin | 10 mg/dL | | Urea | 3000 mg/dL | For interfering substances whose effects on various analytes were observed in the initial screening, an additional dose-dependency evaluation was performed to establish the concentration limit below which no significant interference is expected. The results are provided in the table (below): Substances interfering with test analytes | Analyte | Substance | Concentration limit with no significant interference | Effect when above the concentration limit | | --- | --- | --- | --- | | Glucose | Ascorbic Acid | 20 mg/dL | At 50 mg/dL decreased from +2 to +1 | | | Urobilinogen | 4 mg/dL | At 8 mg/dL decreased from +2 to +1 | | | MESNA | 100 mg/dL | At 150 mg/dL increased from NORMAL to +/- | | Protein | Sodium Bicarbonate | 200 mg/dL | At 250 mg/dL increased from +3 to +4 | | | Specific Gravity | 1.02 | At 1.025 decreased from +3 to +2 | | | Hemoglobin | 10 mg/dL | At 12 mg/dL increased from - to +/- | | Urobilinogen | Bilirubin | 3.5 mg/dL | At 7 mg/dL increased from NORMAL to +1 | | Blood | Ascorbic Acid | 50 mg/dL | At 100 mg/dL decreased from +2 to +1 | | | MESNA | 50 mg/dL | At 100 mg/dL decreased from +2 to +1 | K193514 - Page 9 of 20 {9} | Analyte | Substance | Concentration limit with no significant interference | Effect when above the concentration limit | | --- | --- | --- | --- | | | Urine pH | pH 8.5 | At pH 9 decreased from +2 to +/- | | | Methyl-dopa | 50 mg/dL | At 67 mg/dL increased from - to +/- | | Ketones | MESNA | 1.5 mg/dL | At 3 mg/dL increased from - to +/- | | Nitrite | Ascorbic Acid | 50 mg/dL | At 100 mg/dL decreased from +2 to +1 | | | MESNA | 150 mg/dL | At 200 mg/dL decreased from +2 to +1 | | | Urobilinogen | 5 mg/dL | At 8 mg/dL decreased from +2 to +1 | | Leukocytes | Bilirubin | 7 mg/dL | At 10.5 mg/dL increased from negative to 25 | | | Hemoglobin | 3 mg/dL | At 6.25 mg/dL decreased from 250 to 75 | | | MESNA | 370 mg/dL | At 660 mg/dL decreased from 250 to 75 | Interference effect of intrinsic urine color: Studies were conducted to verify that the test strip analytes of protein, leukocytes, nitrite, ketones, glucose, urobilinogen and blood were insensitive to interference from red, brown, orange, green and various yellow-intensity colored urine samples. The results demonstrated that red, brown, orange, green and various yellow-intensity urine colors do not significantly impact the results for those analytes tested since all results were within +/- 1 color block. Color Correction: The candidate device includes a feature that corrects for intrinsic urine color for bilirubin, urobilinogen, and ketones. The sponsor also provided information to support this feature. Limitations: The following limitations are included in the labeling: 1. Elevated hematuria (6.25 mg/dL Hb, approximately 2,156 RBC/μL) has been shown to cause falsely lowered Leukocyte readings and should be considered when interpreting Leukocyte results. 2. Samples from patients taking methyl dopa may be falsely elevated for blood, negative to trace at 67 mg/dL and negative to +1 at 225 mg/dL. Patients taking the maximum daily dose of methyl dopa may have approximately 210 mg/dL methyl dopa excreted in the urine. 3. Patients taking SGLT2 or similar inhibitors may have extremely high urinary glucose values. Due to a hook effect on the glucose pad, if a patient's urine glucose level is greater than 20,000 mg/dL, then falsely decreased glucose results may be obtained with the test. K193514 - Page 10 of 20 {10} 4. When acidic or alkaline compounds, such as ascorbic acid and sodium bicarbonate are taken internally for treatment or as supplements, they may affect urine pH. # 4. Assay Reportable Range: Measurement results are represented as qualitative and semi-quantitative values as listed below: Analyte Measurement ranges Qualitative/Semi-Quantitative | Analyte | Reportable range | | | --- | --- | --- | | Glucose | Qualitative: normal, ±, 1+, 2+, 3+, 4+ Semi-quantitative: normal, 30-50, 70-100, 150-200, 300-500, ≥1000 mg/dL | | | Protein | Qualitative: negative, ±, 1+, 2+, 3+, 4+ Semi-quantitative: negative, 10-20, 30-70, 100-200, 300-600, >600 mg/dL | | | Bilirubin | Qualitative: negative, 1+, 2+, 3+, 4+ Semi-quantitative: negative, 0.5-1.0, 2.0-4.0, 6.0-10.0, >10.0 mg/dL | | | Urobilinogen | Qualitative: normal, 1+, 2+, 3+, 4+ Semi-quantitative: normal, 2.0-3.0, 4.0-6.0, 8.0-12.0, >12.0 mg/dL | | | Blood | Qualitative: negative, ±, 1+, 2+, 3 Semi-quantitative: negative, 0.03, 0.06-0.10, 0.20-0.50, ≥1.0 mg/dL | | | Ketones | Qualitative: negative, ±, 1+, 2+, 3+, 4+ Semi-quantitative: Negative, 10-20, 40-60, 80-100, ≥150 mg/dL | | | Nitrite | Qualitative: negative, 1+, 2+ | | | pH | Measured value: 5.0, 5.5, 6.0, 6.5, 7.0, 7.5, 8.0, 8.5, 9.0 | | | Leukocytes | Semi-quantitative: negative, 25, 75, 250, 500 Leu/μL | | # 5. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods): The system is traceable to internal standards. # 6. Detection Limit: Limits of Detection and Minimum Detection Sensitivity were determined using three instruments and three reagent lots. In the table (below), are the quantitative values indicating the lower limit of the block in which $50\%$ or more of the results fall into the specific block: Detection Sensitivity for each Analyte per block | Analyte | Qualitative Block Output | Semi-Quantitative rank | Evaluation concentration (mg/dL) | % Positive | | --- | --- | --- | --- | --- | | Bilirubin | +1 | 0.5-1.0 | 0.5 | 53 | | | +2 | 2.0-4.0 | 1.5 | 93 | | | +3 | 8.0-12.0 | 5 | 93 | | | +4 | Over 12.0 | 12 | 93 | | Blood | +/- | 0.03 | 0.023 | 100 | | | +1 | 0.06-0.10 | 0.045 | 100 | | | +2 | 0.20-0.50 | 0.15 | 97 | K193514 - Page 11 of 20 {11} | Analyte | Qualitative Block Output | Semi-Quantitative rank | Evaluation concentration (mg/dL) | % Positive | | --- | --- | --- | --- | --- | | | +3 | 1.00-over | 0.75 | 73 | | Glucose | +/- | 30-50 | 30 | 100 | | | +1 | 70-100 | 60 | 93 | | | +2 | 150-200 | 125 | 97 | | | +3 | 300-500 | 250 | 100 | | | +4 | 1000-over | 750 | 100 | | Ketones | +/- | Below 9 | 4 | 97 | | | +1 | 10-20 | 7.5 | 100 | | | +2 | 40-60 | 30 | 100 | | | +3 | 80-100 | 70 | 63 | | | +4 | 150-over | 150 | 100 | | Leukocytes | 25 | 25 | 14 cells/μL | 100 | | | 75 | 75 | 48 cells/μL | 70 | | | 250 | 250 | 172 cells/μL | 60 | | | 500 | 500 | 396 cells/μL | 90 | | Nitrites | +1 | | 0.075 | 57 | | | +2 | | 0.3 | 100 | | Protein | +/- | 10-20 | 10 | 67 | | | +1 | 30-70 | 30 | 77 | | | +2 | 100-200 | 85 | 53 | | | +3 | 300-600 | 250 | 73 | | | +4 | Over 600 | 800 | 93 | | pH | 5.5 | 5.5 | 5.3 | 100 | | | 6 | 6 | 5.8 | 100 | | | 6.5 | 6.5 | 6.2 | 100 | | | 7 | 7 | 6.8 | 93 | | | 7.5 | 7.5 | 7.3 | 80 | | | 8 | 8 | 8 | 100 | | | 8.5 | 8.5 | 8.5 | 100 | | | 9 | 9 | 8.8 | 87 | | Urobilinogen | +1 | 2.0-3.0 | 2 | 60 | | | +2 | 4.0-6.0 | 4 | 80 | | | +3 | 8.0-12.0 | 7 | 100 | | | +4 | Over 12.0 | 14 | 100 | 7. Assay Cut-Off: Not applicable. 8. Accuracy (Instrument): Refer to the Method Comparison studies below for more information about the device's accuracy. K193514 - Page 12 of 20 {12} 9. Carry-Over: A study was performed to evaluate carry-over contamination potential caused by the AX-4060 nozzle and pump tubing from one sample to the next. Two sample pools were created at two different analyte concentration levels (low target value [LTV] sample and high target value [HTV] sample) created by spiking the pools with the appropriate analytes and tested to ensure high/low testing rank. Testing found all the measurement results of the LTV sample after HTV measurement agreed with the target block for each measurement item. B Comparison Studies: 1. Method Comparison with Predicate Device: A method comparison study was conducted at two (2) clinical laboratories to evaluate the agreement of results from the AUTION MAX AX-4060 Urinalysis System to the predicate device, the AUTION MAX AX-4030 Urinalysis System. 1044 natural patient samples were tested at two sites, with additional spiked and archived samples used to supplement abnormal analyte levels. Testing was performed in duplicate on the comparator AX-4030. The results below compare the first replicate on the AX-4060 (candidate device) to the first replicate from the AX-4030 (comparator device). Duplicates measured on the comparator were used to identify samples near the cutoff (i.e., where sample replicates were not exact matches but were within one block). Based on duplicate testing on the comparator, some of the discordant samples in the tables below were near the cutoff with duplicate results within one color block. Analyte Concordance Charts are depicted below: Concordance Chart for Bilirubin | Bilirubin | Comparator: AX-4030 | | | | | | | --- | --- | --- | --- | --- | --- | --- | | | | Negative (-) | 1+ | 2+ | 3+ | 4+ | | AX-4060 | Negative (-) | 1181 | 4 | 0 | 0 | 0 | | | 1+ | 4 | 65 | 1 | 0 | 0 | | | 2+ | 0 | 1 | 40 | 0 | 0 | | | 3+ | 0 | 0 | 0 | 24 | 0 | | | 4+ | 0 | 0 | 0 | 0 | 18 | | Total sample # | | 1185 | 70 | 41 | 24 | 18 | | Exact Match % | | 99.7 | 92.9 | 97.6 | 100 | 99.3 | | Within ±1 Color Block % | | 100 | 100 | 100 | 100 | 100 | K193514 - Page 13 of 20 {13} Concordance Chart for Blood | Blood | Comparator: AX-4030 | | | | | | | --- | --- | --- | --- | --- | --- | --- | | | | Negative (-) | ± | 1+ | 2+ | 3+ | | AX-4060 | Negative (-) | 648 | 6 | 0 | 0 | 0 | | | ± | 4 | 126 | 7 | 0 | 0 | | | 1+ | 0 | 18 | 97 | 3 | 0 | | | 2+ | 0 | 0 | 13 | 51 | 2 | | | 3+ | 0 | 0 | 0 | 11 | 58 | | Total sample # | | 652 | 150 | 117 | 65 | 60 | | Exact Match % | | 99.4 | 84.0 | 82.9 | 78.5 | 96.7 | | Within ±1 Color Block % | | 100 | 100 | 100 | 100 | 100 | Concordance Chart for Glucose | Glucose | Comparator: AX-4030 | | | | | | | | --- | --- | --- | --- | --- | --- | --- | --- | | | | Normal | ± | 1+ | 2+ | 3+ | 4+ | | AX-4060 | Normal | 663 | 3 | 0 | 0 | 0 | 0 | | | ± | 2 | 44 | 2 | 0 | 0 | 0 | | | 1+ | 0 | 2 | 40 | 3 | 0 | 0 | | | 2+ | 0 | 0 | 0 | 28 | 6 | 0 | | | 3+ | 0 | 0 | 0 | 2 | 62 | 0 | | | 4+ | 0 | 0 | 0 | 0 | 7 | 180 | | Total | | 665 | 49 | 42 | 33 | 75 | 180 | | Exact Match % | | 99.7 | 89.8 | 95.2 | 84.8 | 82.7 | 100 | | Within ±1 Color Block % | | 100 | 100 | 100 | 100 | 100 | 100 | K193514 - Page 14 of 20 {14} Concordance Chart for Ketones | Ketones | Comparator: AX-4030 | | | | | | | | --- | --- | --- | --- | --- | --- | --- | --- | | | | Negative | ± | 1+ | 2+ | 3+ | 4+ | | AX-4060 | Negative | 1123 | 3 | 0 | 0 | 0 | 0 | | | ± | 2 | 40 | 4 | 0 | 0 | 0 | | | 1+ | 0 | 1 | 45 | 1 | 0 | 0 | | | 2+ | 0 | 0 | 0 | 58 | 3 | 0 | | | 3+ | 0 | 0 | 0 | 2 | 33 | 0 | | | 4+ | 0 | 0 | 0 | 0 | 1 | 20 | | Total # | | 1125 | 45 | 48 | 61 | 37 | 20 | | Exact Match % | | 99.8 | 90.9 | 91.8 | 95.1 | 89.2 | 100.0 | | Within ±1 Color Block % | | 100 | 100 | 100 | 100 | 100 | 100 | Concordance Chart for Leukocytes | Leukocytes | Comparator: AX-4030 | | | | | | | --- | --- | --- | --- | --- | --- | --- | | | | Negative | 25 | 75 | 250 | 500 | | AX-4060 | Negative | 692 | 6 | 0 | 0 | 0 | | | 25 | 5 | 62 | 12 | 0 | 0 | | | 75 | 0 | 5 | 81 | 6 | 0 | | | 250 | 0 | 0 | 6 | 59 | 6 | | | 500 | 0 | 0 | 0 | 4 | 100 | | Total | | 697 | 73 | 99 | 69 | 106 | | Exact Match % | | 99.3 | 84.9 | 81.8 | 85.5 | 94.3 | | Within ±1 Color Block % | | 100 | 100 | 100 | 100 | 100 | Concordance Chart for Nitrite | Nitrite | Comparator: AX-4030 | | | | | --- | --- | --- | --- | --- | | | | Negative | 1+ | 2+ | | AX-4060 | Negative | 926 | 0 | 0 | | | 1+ | 2 | 26 | 1 | | | 2+ | 0 | 0 | 91 | | Total | | 928 | 26 | 92 | | Exact Match % | | 99.8 | 100 | 98.9 | | Within ± 1 Color Block % | | 100 | 100 | 100 | K193514 - Page 15 of 20 {15} Concordance Chart for Protein | Protein | Comparator: AX-4030 | | | | | | | | --- | --- | --- | --- | --- | --- | --- | --- | | | | Neg | ± | 1+ | 2+ | 3+ | 4+ | | AX-4060 | Neg | 410 | 24 | 0 | 0 | 0 | 0 | | | ± | 8 | 213 | 32 | 0 | 0 | 0 | | | 1+ | 0 | 1 | 185 | 10 | 0 | 0 | | | 2+ | 0 | 0 | 1 | 84 | 1 | 0 | | | 3+ | 0 | 0 | 0 | 3 | 51 | 0 | | | 4+ | 0 | 0 | 0 | 0 | 12 | 13 | | Total sample # | | 418 | 238 | 218 | 97 | 64 | 13 | | Exact Match % | | 98.1 | 89.5 | 84.9 | 86.6 | 79.7 | 100 | | Within ± 1 Color Block % | | 100 | 100 | 100 | 100 | 100 | 100 | Concordance Chart for Urobilinogen | Urobilinogen | Comparator: AX-4030 | | | | | | | --- | --- | --- | --- | --- | --- | --- | | | | Normal | 1+ | 2+ | 3+ | 4+ | | AX-4060 | Normal | 1192 | 5 | 0 | 0 | 0 | | | 1+ | 2 | 55 | 2 | 0 | 0 | | | 2+ | 0 | 0 | 44 | 4 | 0 | | | 3+ | 0 | 0 | 0 | 21 | 3 | | | 4+ | 0 | 0 | 0 | 0 | 13 | | Total sample # | | 1194 | 60 | 46 | 25 | 16 | | Exact Match % | | 99.8 | 91.7 | 95.7 | 84.0 | 81.3 | | Within ± 1 Color Block % | | 100 | 100 | 100 | 100 | 100 | K193514 - Page 16 of 20 {16} Concordance Chart for pH | pH | Comparator: AX-4030 | | | | | | | | | | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | | | 5.0 | 5.5 | 6 | 6.5 | 7.0 | 7.5 | 8.0 | 8.5 | 9 | | AX-4060 | 5.0 | 70 | 2 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | | | 5.5 | 15 | 266 | 8 | 0 | 0 | 0 | 0 | 0 | 0 | | | 6.0 | 0 | 9 | 227 | 7 | 0 | 0 | 0 | 0 | 0 | | | 6.5 | 0 | 0 | 8 | 186 | 2 | 0 | 0 | 0 | 0 | | | 7.0 | 0 | 0 | 0 | 11 | 109 | 1 | 0 | 0 | 0 | | | 7.5 | 0 | 0 | 0 | 0 | 6 | 43 | 0 | 0 | 0 | | | 8.0 | 0 | 0 | 0 | 0 | 0 | 14 | 33 | 0 | 0 | | | 8.5 | 0 | 0 | 0 | 0 | 0 | 0 | 6 | 18 | 0 | | | 9 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 11 | 11 | | Total sample # | | 85 | 277 | 243 | 204 | 117 | 58 | 39 | 29 | 11 | | Exact Match % | | 82.4 | 96.0 | 93.4 | 91.2 | 93.2 | 74.1 | 84.6 | 62.1 | 100 | | Within ± 1 Color Block % | | 100 | 100 | 100 | 100 | 100 | 100 | 100 | 100 | 100 | Concordance Chart for Turbidity | Turbidity | Comparator: AX-4030 | | | | | --- | --- | --- | --- | --- | | | | - | 1+ | 2+ | | AX-4060 | - | 827 | 2 | 0 | | | 1+ | 12 | 149 | 1 | | | 2+ | 0 | 0 | 30 | | Total | | 839 | 151 | 31 | | Exact Match % | | 98.6 | 98.7 | 96.8 | | Within ± 1 Block % | | 100% | 100% | 100% | Color Tone: This is a light-transmission-based color detection system which is integrated into the AX-4060 analyzer and is independent of the 9EB test strip. This system reports the color tone results and can detect the following urine colors: Colorless, Yellow, Orange, Brown, Red, Violet, Blue, and Green. As a result of a lack of natural clinical specimens for Brown, Green, Blue, Violet, and Red colors (due to the infrequent occurrence of urine specimens of these colors), these colors were evaluated using contrived samples (dyes added to natural urine samples). K193514 - Page 17 of 20 {17} Concordance Chart for Color Tone | Color Tone | Comparator: AX-4030 | | | | | | | | | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | | | Colorless | Yellow | Orange | Brown | Red | Violet | Blue | Green | | AX-4060 | Colorless | 17 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | | | Yellow | 127 | 822 | 0 | 0 | 0 | 0 | 0 | 7 | | | Orange | 5 | 52 | 20 | 6 | 0 | 0 | 0 | 0 | | | Brown | 0 | 0 | 0 | 108 | 0 | 0 | 0 | 0 | | | Red | 0 | 0 | 0 | 0 | 63 | 0 | 0 | 0 | | | Violet | 0 | 0 | 0 | 0 | 2 | 91 | 0 | 0 | | | Blue | 0 | 0 | 0 | 0 | 0 | 0 | 66 | 0 | | | Green | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 92 | | Total sample # | | 149 | 874 | 20 | 114 | 65 | 91 | 66 | 99 | | Exact Match % | | 11.4 | 94.1 | 100 | 94.7 | 96.9 | 100 | 100 | 92.9 | | Within ± 1 Color Block % | | 100 | 100 | 100 | 100 | 100 | 100 | 100 | 100 | Matrix Comparison: Not applicable. C Clinical Studies: 1. Clinical Sensitivity: Not applicable. 2. Clinical Specificity: Not applicable. 3. Other Clinical Supportive Data (When 1. and 2. Are Not Applicable): Not applicable. D Clinical Cut-Off: Not applicable. E Expected Values/Reference Range: Glucose A small amount of glucose may be detected in normal urine. Generally, the amount of glucose is below the sensitivity level of the method; however, on occasion may produce $\pm$ (trace) results. K193514 - Page 18 of 20 {18} Consistently positive glucose results should be clinically investigated. ## Protein Although very small quantities of protein are normally excreted, the amount is generally below the sensitivity level for detection. Positive results require clinical assessment/additional testing to determine its significance. ## Bilirubin In normal urine, no bilirubin is normally detected. Positive findings should be diagnostically and clinically investigated. ## Urobilinogen Healthy individuals may excrete a small amount of Urobilinogen and may be increased especially after exercise. Concentrations are generally at their peak in the afternoon. ## pH Normal and abnormal urine pH ranges from 5.0 to 8.0 and is influenced by diet. Typical values for first morning specimens, from healthy individuals, are between pH 5.0 - 6.0. ## Blood A blue-green dotted reaction indicates the presence of erythrocytes. Up to 5 erythrocytes / μL may be found in normal urine specimens. Urine from menstruating women may contain blood. A large amount of blood should be clinically investigated. ## Ketones Ketones are not normally detected in urine specimens from healthy individuals. However, urine specimens from individuals who are fasting, pregnant, or who undergo regular strenuous exercise, may exhibit significant amounts of ketones. The presence of ketones, in urine specimens from patients with diabetes, may provide a useful marker for metabolic status. ## Nitrite A nitrite concentration of as low as 0.08 mg/dL may be detected and produce a positive test. A negative result may occur during fasting episodes since nitrates are not appearing in the urine. ## Leukocytes Normal urine specimens should not produce a positive reaction. Small amounts of leukocyte esterase, causing a positive reaction should be repeated, using a fresh urine specimen, from the same patient. Positive results require further testing for pyuria. ## F Other Supportive Instrument Performance Characteristics Data: Not applicable. ## VIII Proposed Labeling: The labeling supports the finding of substantial equivalence for this device. K193514 - Page 19 of 20 {19} IX Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision. K193514 - Page 20 of 20
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...