K030600 · International Remote Imaging Systems · KQO · Jul 11, 2003 · Clinical Chemistry
Device Facts
Record ID
K030600
Device Name
AUTION JET AJ-4270 URINE ANALYZER
Applicant
International Remote Imaging Systems
Product Code
KQO · Clinical Chemistry
Decision Date
Jul 11, 2003
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.2900
Device Class
Class 1
Indications for Use
The AUTION™ JET AJ-4270 Urine Analyzer (AUTIONTM JET) is a semiautomated urine analyzer and is intended for use with the AUTION™ Sticks 10EA multi-parameter test strips for the in vitro measurement of the following analytes: glucose, protein, bilirubin, urobilinogen, pH. specific gravity, blood, ketones, nitrite, leukocytes, and color. The AUTION™ JET Analyzer and AUTION™ Sticks 10EA test strips are intended for mutually exclusive use as a system.
Device Story
AUTION™ JET AJ-4270 is a semiautomated urine analyzer. User manually dips test strip into urine sample and places it on analyzer tray. Device automatically times reaction and moves strip to internal Optical Block. Reflectance spectroscopy measures analyte concentrations and urine color. Output provided to healthcare provider for clinical decision-making regarding patient health status. System designed for use with AUTION™ Sticks 10EA test strips. Optional hand-held bar code reader available for sample identification. Device intended for clinical laboratory use.
Clinical Evidence
Bench testing only. Correlation study compared AUTION™ JET results against a commercially available automated urine analyzer, demonstrating excellent agreement. Additional studies evaluated precision and linearity, confirming acceptable performance for the intended use.
Technological Characteristics
Semiautomated urine analyzer using reflectance spectroscopy for analyte and color measurement. Operates with AUTION™ Sticks 10EA multi-parameter test strips. Includes tray-based strip transport mechanism. Optional external bar code reader. No specific material standards or software architecture details provided.
Indications for Use
Indicated for in vitro diagnostic measurement of urine analytes (glucose, protein, bilirubin, urobilinogen, pH, specific gravity, blood, ketones, nitrite, leukocytes, and color) using AUTION™ Sticks 10EA test strips in a clinical setting.
Regulatory Classification
Identification
An automated urinalysis system is a device intended to measure certain of the physical properties and chemical constituents of urine by procedures that duplicate manual urinalysis systems. This device is used in conjunction with certain materials to measure a variety of urinary analytes.
{1}------------------------------------------------
# JUL 1 1 2003
# 510(k) Summary for AUTION™ JET AJ-4270 Urine Analyzer
#### 1. SPONSOR
International Remote Imaging Systems, Inc. 9162 Eton Avenue Chatsworth, CA 91311
| Contact Person: | William M. Dougherty |
|-----------------|-----------------------------------|
| | Manager |
| | Quality Assurance and Evaluations |
| Telephone: | 818-709-1244, ext. 140 |
June 12, 2003 Date Prepared:
### 2. DEVICE NAME
| Proprietary Name: | AUTION TM JET AJ-4270 Urine Analyzer |
|----------------------|---------------------------------------|
| Common/Usual Name: | Urine Analyzer |
| Classification Name: | Automated Urinalysis System |
#### 3. PREDICATE DEVICES
- Chemstrip 101 Urine Analyzer (K983510) .
- AUTION MAX™ AX-4280 Urine Analyzer (K013783) .
- Clinitek Atlas Automated Urine Chemistry Analyzer (K946183) .
#### 4. DEVICE DESCRIPTION
The AUTION™ JET AJ-4270 Urine Analyzer (AUTION™ JET) consists of an analyzer and accessories. ARKRAY also offers a hand-held Bar Code Reader for use with the AUTION™ JET. The AUTION™ JET Analyzer is designed for use only with the AUTIONTM Sticks 10EA multi-parameter test strips.
{2}------------------------------------------------
# ഗ് INTENDED USE
The AUTION™ JET AJ-4270 Urine Analyzer (AUTIONTM JET) is a semiautomated urine analyzer and is intended for use with the AUTION™ Sticks 10EA multi-parameter test strips for the in vitro measurement of the following analytes: glucose, protein, bilirubin, urobilinogen, pH. specific gravity, blood, ketones, nitrite, leukocytes, and color. The AUTION™ JET Analyzer and AUTION™ Sticks 10EA test strips are intended for mutually exclusive use as a system.
## 6. TECHNOLOGICAL CHARACTERISTICS AND SUBSTANTIAL EQUIVALENCE
The indicated use of the proposed and predicate urine analyzers is identical. All of these urine analyzers provide qualitative and semi-quantitative measurement of urine analytes using multi-parameter test strips for the measurement of the urine analytes.
The overall design of the proposed AUTION™ JET Analyzer is identical to the predicate Chemstrip 101 Urine Analyzer. Both of these analyzers are semiautomated, requiring that the user dip a test strip in the urine and place the test strip on a tray in the analyzer. The analyzers then time the reaction on the strip and automatically move the test strip to the Optical Block for analysis. The predicate AUTION MAX™ and Clinitek Atlas Analyzers are fully automated and contain mechanical systems for urine specimen handling and sampling in addition to those for test strip handling and analysis.
The proposed AUTION™ JET and the predicate Chemstrip 101 Analyzers use reflectance spectroscopy for the measurement of all urine analyte species. The proposed AUTIONTM JET also uses reflectance spectroscopy for urine color determination. Urine color determination is not performed by the Chemstrip 101.
The predicate AUTION MAX™ and Clinitek Atlas Analyzers use reflectance spectroscopy for the measurement of all urine analyte species and urine color determination, with the exception of urine SG. The AUTION MAX™ and Clinitek Atlas Analyzers use the refractive index method for SG determination.
The AUTION™ JET is indicated for use with the AUTION™ Sticks 10EA multiparameter test strips. The proposed AUTIONTM Sticks 10EA are identical to the AUTION™ Sticks 9EB that are indicated for use with the predicate AUTION MAX™, with the exception that the AUTION™ Sticks 10EA test strips contain a reagent pad for SG determination. The AUTION™ Sticks 9EB do not contain a reagent pad for the measurement of SG. The AUTION MAX™ and Clinitek Atlas determine urine SG by the refractive index method.
{3}------------------------------------------------
The mechanisms of action for the chemical reactions used for the determination of urinary analytes, including glucose and occult blood, are similar for the proposed AUTION™ Sticks 10EA test strips and the predicate Clinitek Atlas reagent strips and Chemstrip 10UA test strips. The chemical constituents of the individual reagent formulations are similar for the proposed and predicate reagent pads, with minor differences in the choice of enzyme substrate and chromogens used for production of the colored endproduct.
### 7. PERFORMANCE TESTING
Several studies were conducted to evaluate the performance characteristics of the AUTION™ JET Urine Analyzer. A correlation study demonstrated excellent agreement when results from the AUTION™ JET were compared to those from a commercially available automated urine analyzer. Additional studies for precision and linearity demonstrated acceptable performance of the AUTION™ JET System for its intended use.
{4}------------------------------------------------
Image /page/3/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of three human profiles facing right, stacked on top of each other. The profiles are connected by a flowing line that resembles hair or fabric. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the image.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
JUL 1 1 2003
International Remote Imaging System c/o Ms. Cynthia J.M. Nolte, Ph.D., RAC Staff Consultant Medical Device Consultants, Inc. 49 Plain Street North Attleboro, MA 02760
Re: k030600 Trade/Device Name: AUTION™ JET AJ-4270 Urine Analyzer Regulation Number: 21 CFR 862.2900 Regulation Name: Automated urinalysis system Regulatory Class: Class I Product Code: KOO: JIL; JIO Dated: June 12, 2003 Received: June 13, 2003
Dear Dr. Nolte:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).
{5}------------------------------------------------
# Page 2 -
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Steven Putman
Steven I. Gutman, M.D., M.B.A. Director Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
510(k) Number (if known): K030600
Device Name: AUTION™ JET AJ-4270 Urine Analyzer
Indications for Use:
The AUTION™ JET AJ-4270 Urine Analyzer (AUTION™ JET) is a semi-automated urine analyzer and is intended for use with the AUTION™ Sticks 10EA multi-parameter test strips for the in vitro measurement of the following analytes: glucose, protein, bilirubin, urobilinogen, pH, specific gravity, blood, ketones, nitrite, leukocytes, and color. The AUTION™ JET Analyzer and AUTION™ Sticks 10EA test strips are intended for mutually exclusive use as a system.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NECESSARY)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Division Sign-Off
תערו Office of In Vitro Diagnostic Device
**Evaluation and Safety**
510(k) K030600
Prescription Use
(Per 21 CFR 801.109)
OR
Over-The-Counter Use
(Optional Format 1-2-96)
June 12, 2003
Appendix A • Page 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.