SIGMA DIAGNOSTICS CO2 ACID REAGENT
Device Facts
| Record ID | K963541 |
|---|---|
| Device Name | SIGMA DIAGNOSTICS CO2 ACID REAGENT |
| Applicant | Sigma Diagnostics, Inc. |
| Product Code | JFL · Clinical Chemistry |
| Decision Date | Oct 24, 1996 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 862.1160 |
| Device Class | Class 2 |
Indications for Use
The Sigma Diagnostics CO₂ Acid Reagent, Procedure Number C7558, is intended for the quantitative determination of carbon dioxide in serum or plasma on the SYNCHRON CX®3 System.
Device Story
Reagent kit for quantitative measurement of CO₂ in serum or plasma; used on SYNCHRON CX®3 System. Principle involves chemical reaction to measure dissolved CO₂ and bicarbonate anion. Used in clinical laboratory settings by technicians. Output is CO₂ concentration (mmol/L) used by physicians to assess acid-base balance and electrolyte status. Benefits include diagnostic support for metabolic/respiratory disorders.
Clinical Evidence
Bench testing only. Comparison study of Sigma Diagnostics reagent vs. predicate using serum samples yielded correlation coefficient of 0.992 and regression equation y = 0.95x + 1.54. Precision testing (within-run and total) showed %CV < 6.5%. Linearity established from 5.0 to 40.0 mmol/L.
Technological Characteristics
Chemical reagent for CO₂ determination. Operates via enzymatic/chemical reaction principle on SYNCHRON CX®3 automated analyzer. Linear range: 5.0 to 40.0 mmol/L.
Indications for Use
Indicated for the determination of plasma CO₂ content as part of an electrolyte profile to assess metabolic or respiratory acidosis or alkalosis in patients with conditions such as diabetes mellitus, glomerulonephritis, pyloric obstruction, or diarrhea.
Regulatory Classification
Identification
A bicarbonate/carbon dioxide test system is a device intended to measure bicarbonate/carbon dioxide in plasma, serum, and whole blood. Bicarbonate/carbon dioxide measurements are used in the diagnosis and treatment of numerous potentially serious disorders associated with changes in body acid-base balance.
Predicate Devices
- Beckman CO₂ Acid Reagent Kit, Part No. 443330
Related Devices
- K963538 — SIGMA DIAGNOSTICS CO2ALKALINE BUFFER · Sigma Diagnostics, Inc. · Oct 11, 1996
- K073223 — S-TEST C02 · Alfa Wassermann Diagnostic Technologies, Inc. · Jun 26, 2008
- K071340 — CARBON DIOXIDE (CO2), SCAL, NORTROL AND ABTROL · Thermofisher Scientific OY · Oct 5, 2007
- K070251 — CARBON DIOXIDE LIQUID STABLE REAGENT · Pointe Scientific, Inc., · Apr 30, 2007