K073223 · Alfa Wassermann Diagnostic Technologies, Inc. · KHS · Jun 26, 2008 · Clinical Chemistry
Device Facts
Record ID
K073223
Device Name
S-TEST C02
Applicant
Alfa Wassermann Diagnostic Technologies, Inc.
Product Code
KHS · Clinical Chemistry
Decision Date
Jun 26, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1160
Device Class
Class 2
Intended Use
The S-Test Carbon Dioxide Reagent is intended for the quantitative determination of carbon dioxide concentration in serum or heparin plasma using the S40 Clinical Analyzer. Bicarbonate/carbon dioxide measurements are used in the diagnosis and treatment of numerous potentially serious disorders associated with changes in body acid-base balance. This test is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only.
Device Story
Single-use reagent cartridge for S40 Clinical Analyzer; contains PEP, PEPC, MD, and NADH analog. Input: serum or heparin plasma sample (5 μL). Principle: enzymatic endpoint reaction; CO2 reacts with PEP via PEPC to yield oxaloacetic acid; MD reduces cofactor; absorbance measured bichromatically at 405/508 nm. Output: quantitative CO2 concentration (mEq/L). Used in clinical labs or physician's office labs by trained operators. 2-D barcode on cartridge provides lot-specific calibration data. Results assist clinicians in diagnosing/treating acid-base balance disorders.
Clinical Evidence
Bench testing only. Precision studies (21 days, 3 levels) showed within-run CV 1.7-3.0%, total CV 5.7-6.7%. POL site precision showed within-run CV 1.3-6.6%, total CV 2.0-6.6%. Accuracy correlation study (n=94) yielded correlation coefficient 0.984, standard error 1.1. POL site correlation studies showed correlation coefficients 0.937-0.980. Detection limit 5 mEq/L.
Technological Characteristics
Photometric enzymatic endpoint assay. Reagents: PEP, PEPC (microbial), MD (mammalian), NADH analog, buffer (pH 7.2), activators, stabilizers, surfactant. Form factor: single-use cartridge with two reagent cells, cuvette, film seal, 2-D barcode. Connectivity: 2-D barcode for lot-specific calibration data. Standards: NIST SRM351 traceability; CLSI EP5-A2, EP10-A, EP6-A, EP7-A, EP17-A, EP9-A2, C28-A2.
Indications for Use
Indicated for quantitative determination of CO2 in serum or heparin plasma to diagnose and treat acid-base balance disorders. Intended for use in clinical laboratories or physician's office laboratories.
Regulatory Classification
Identification
A bicarbonate/carbon dioxide test system is a device intended to measure bicarbonate/carbon dioxide in plasma, serum, and whole blood. Bicarbonate/carbon dioxide measurements are used in the diagnosis and treatment of numerous potentially serious disorders associated with changes in body acid-base balance.
Predicate Devices
ACE plus ISE/Clinical Chemistry System ACE Carbon Dioxide Reagent (K931786)
{0}------------------------------------------------
## JUN 2 6 2008
K073223
## 510(k) SUMMARY
| 510(k) Owner: | Alfa Wassermann Diagnostic Technology, LLC<br>4 Henderson Drive<br>West Caldwell, NJ 07006 | | |
|---------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------|--|
| Contact: | Dennis Taschek<br>Phone: 973-852-0177<br>Fax: 973-852-0237 | | |
| Date Summary<br>Prepared: | June 19, 2008 | | |
| Device: | Trade Name: | S-Test CO2 Reagent cartridge | |
| | Common/Classification Name: | Carbon dioxide test system<br>(21 C.F.R. § 862.1160)<br>Product Code KHS | |
| | Classification: | Class II | |
| Predicate<br>Devices: | Manufacturers for analyzer/reagent system predicates are:<br>1. ACE plus ISE/Clinical Chemistry System<br>ACE Carbon Dioxide Reagent (K931786)<br>2. Ortho Clinical Diagnostics Fusion Clinical Chemistry Analyzer<br>Carbon Dioxide Reagent (K946090)<br>3. Piccolo® xpress Chemistry Analyzer<br>Carbon Dioxide Reagent (K942782) | | |
| Device<br>Description: | The S-Test Carbon dioxide (CO2) reagent cartridge, used with the S40<br>Clinical Analyzer, is intended for quantitative <i>in vitro</i> diagnostic<br>determination of CO2 in serum or heparin plasma based on a<br>photometric test measuring the formation of an enzymatic cofactor in a<br>coupled enzymatic reaction. | | |
| Intended Use: | The S-Test Carbon Dioxide Reagent is intended for the quantitative<br>determination of carbon dioxide concentration in serum or heparin<br>plasma using the S40 Clinical Analyzer. Bicarbonate/carbon dioxide<br>measurements are used in the diagnosis and treatment of numerous<br>potentially serious disorders associated with changes in body acid-base<br>balance. This test is intended for use in clinical laboratories or<br>physician office laboratories. For <i>in vitro</i> diagnostic use only. | | |
:
{1}------------------------------------------------
| Technological<br>Characteristics: | The S-Test CO2 Reagent is contained in a bi-reagent cartridge.<br>Reagent 1 contains water and a preservative. Reagent 2 contains<br>phosphoenolpyruvate, nicotinamide adenine dinucleotide analog<br>(reduced), phosphoenol pyruvate carboxylase, and malate<br>dehydrogenase. |
|-----------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Performance<br>Data: | Performance data on the S-Test CO2 included precision, accuracy, and<br>sensitivity data.<br><br>Precision: In testing conducted at three CO2 levels for 21 days, the<br>within-run CV ranged from 1.7 to 3.0%, and total CV ranged from 5.7<br>to 6.7%. In precision studies at three separate Physician Office<br>Laboratory (POL) sites and in-house over five or more days, the<br>within-run CV ranged from 1.3 to 6.6% and total CV ranged from 2.0<br>to 6.6%.<br><br>Accuracy: In the correlation study, 94 samples with CO2 values<br>ranging from 6.0 to 41.8 mEq/L were assayed on the S40 Clinical<br>Analyzer using S-Test CO2 and a comparison method. Least-squares<br>regression analysis yielded a correlation coefficient of 0.984, a standard<br>error estimate of 1.1, a confidence interval slope of 0.940 to 1.029, and<br>a confidence interval intercept of -0.6 to 1.5. In patient correlation<br>studies at three separate POL sites using the S40 Clinical Analyzer and<br>a comparative method, least-squares regression analysis yielded<br>correlation coefficients ranging from 0.937 to 0.980, standard error<br>estimates ranging from 1.40 to 2.65, confidence interval slopes ranging<br>from 0.880 to 1.080, and confidence interval intercepts ranging from<br>-1.73 to 2.03.<br><br>Sensitivity: The detection limit was 5 mEq/L. |
| Conclusions: | Based on the foregoing data, the device is safe and effective. These<br>data also indicate substantial equivalence to the predicate devices. |
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with its wings spread, symbolizing protection and care. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular pattern around the eagle, indicating the department's name and national affiliation. The logo is presented in black and white, giving it a formal and official appearance.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
## JUN 2 6 2008
Alfa Wassermann Diagnostic Technologies, Inc. c/o Mr. Daivd Slavin Vice President, Quality and Regulatory Affairs 4 Henderson Drive West Caldwell, NJ 07006
Re: k073223/S1
Trade/Device Name: S Test Carbon Dioxide (CO2) Reagent Cartridge Regulation Number: 21 CFR 862.1160 Regulation Name: Bicarbonate/carbon dioxide test system Regulatory Class: Class II, Product Code: KHS Dated: November 15, 2007 Received: November 15, 2007
Dear Mr. Slavin
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).
{3}------------------------------------------------
Page 2 -
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of in Vitro Diagnostic Device Evaluation and Safety at (240) 276-0490. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address at http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Jean M. Cooper, M.S., D.V.M.
Jean M. Cooper, M.S., D.V.M. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Alfa Wassermann Diagnostic Technology, Inc. 510(k) Submission K073223
S40 Clinical Analyzer S-Test CO2
## Indications for Use Statement
510(k) Number K073223 (if known):
S-Test Carbon Dioxide (CO2) Device Name:
The S-Test Carbon Dioxide Reagent is intended for the quantitative Indications for Use: determination of carbon dioxide concentration in serum or heparin plasma using the S40 Clinical Analyzer. Bicarbonate/carbon dioxide measurements are used in the diagnosis and treatment of numerous potentially serious disorders associated with changes in body acid-base balance. This test is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only.
| Prescription Use | X | And/Or | Over the Counter Use ______ |
|-----------------------------|---|--------|-----------------------------|
| (21 CFR Part 801 Subpart D) | | | (21 CFR Part 801 Subpart C) |
(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OIVD)
Division Sign Off
| June 3, 2008 | Office of In Vitro Diagnostic |
|--------------|-------------------------------|
| | Device Evaluation and Safety |
| | K73223 |
CONFIDENTIAL
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.