SIGMA DIAGNOSTICS CO2ALKALINE BUFFER
Device Facts
| Record ID | K963538 |
|---|---|
| Device Name | SIGMA DIAGNOSTICS CO2ALKALINE BUFFER |
| Applicant | Sigma Diagnostics, Inc. |
| Product Code | JFL · Clinical Chemistry |
| Decision Date | Oct 11, 1996 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 862.1160 |
| Device Class | Class 2 |
Indications for Use
The Sigma Diagnostics CO₂ Alkaline Buffer, Procedure Number C7683, is used to measure carbon dioxide concentrations in serum or plasma on the SYNCHRON CX®3 System.
Device Story
Reagent buffer for in vitro diagnostic measurement of CO₂ in serum/plasma; used on SYNCHRON CX®3 System. Principle: enzymatic/chemical reaction to quantify bicarbonate/dissolved CO₂. Used in clinical laboratory settings by trained technicians. Output: quantitative CO₂ concentration (mmol/L). Clinicians use results to assess acid-base balance and electrolyte status, aiding diagnosis of metabolic/respiratory disorders.
Clinical Evidence
Bench testing only. Comparison study with predicate using serum samples yielded correlation coefficient of 0.992 and regression equation y = 0.95x + 1.54. Precision testing (within-run and total) showed %CV < 6.5%. Linearity established from 5.0 to 40.0 mmol/L.
Technological Characteristics
Chemical reagent buffer for automated clinical chemistry analyzer. Operates via colorimetric/enzymatic reaction principle on SYNCHRON CX®3 System. Linear range: 5.0-40.0 mmol/L.
Indications for Use
Indicated for the determination of plasma CO₂ content as part of an electrolyte profile to assess metabolic or respiratory acidosis or alkalosis in patients with conditions such as diabetes mellitus, glomerulonephritis, pyloric obstruction, or diarrhea.
Regulatory Classification
Identification
A bicarbonate/carbon dioxide test system is a device intended to measure bicarbonate/carbon dioxide in plasma, serum, and whole blood. Bicarbonate/carbon dioxide measurements are used in the diagnosis and treatment of numerous potentially serious disorders associated with changes in body acid-base balance.
Predicate Devices
- Beckman CO₂ Alkaline Buffer Kit (Part No. 443320)
Related Devices
- K963541 — SIGMA DIAGNOSTICS CO2 ACID REAGENT · Sigma Diagnostics, Inc. · Oct 24, 1996
- K073223 — S-TEST C02 · Alfa Wassermann Diagnostic Technologies, Inc. · Jun 26, 2008
- K040975 — GENCHEM CO2 ACID REAGENT · Genchem, Inc. · Dec 27, 2004
- K070251 — CARBON DIOXIDE LIQUID STABLE REAGENT · Pointe Scientific, Inc., · Apr 30, 2007
- K963540 — SIGMA DIAGNOSTICS ELECTROLYTE BUFFER · Sigma Diagnostics, Inc. · Sep 27, 1996