GENCHEM CO2 ACID REAGENT

K040975 · Genchem, Inc. · JFL · Dec 27, 2004 · Clinical Chemistry

Device Facts

Record IDK040975
Device NameGENCHEM CO2 ACID REAGENT
ApplicantGenchem, Inc.
Product CodeJFL · Clinical Chemistry
Decision DateDec 27, 2004
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1160
Device ClassClass 2

Intended Use

The GenChem CO₂ Acid Reagent when used in conjunction with the GenChem ISE Electrolyte Reference, GenChem Electrolyte Buffer, GenChem CO₂ Alkaline Buffer, GenChem Wash Concentrate, and appropriate Calibrators or Calibration Standards is intended for the quantitative determination of Carbon Dioxide in serum and plasma. Carbon Dioxide results are used in the diagnosis and treatment of numerous and potentially serious disorders associated with changes in the body’s acid-base balance.

Device Story

GenChem CO₂ Acid Reagent is a liquid, ready-to-use chemical reagent kit designed for use on the Beckman SYNCHRON CX3 System. It functions via an Ion-Selective Electrode (ISE) principle; the sample is mixed with electrolyte buffer and acid to release dissolved CO₂ gas. This gas diffuses across a silicon rubber membrane into an alkaline buffer, causing a pH change. The rate of pH change is measured and is directly proportional to the original CO₂ concentration in the sample. The device is intended for clinical laboratory use by trained personnel to aid in diagnosing acid-base disorders. Results are used by physicians to guide clinical decision-making regarding patient metabolic status.

Clinical Evidence

Bench testing only. Precision/reproducibility evaluated over 30 days (n=60 per sample); linearity assessed via Least Squares method (r²=1.000). Analytical specificity tested against hemoglobin, bilirubin, and lipemia per NCCLS EP7-A. Method comparison study performed on 80 serum and 80 plasma samples using linear regression against predicate (Serum R²=0.953; Plasma R²=0.987).

Technological Characteristics

Solution containing 0.6 mol/L sulfuric acid and surfactants. Designed for use on Beckman CX3 Synchron Analyzer. Analytical method based on pH rate measurement of CO2. Compatible with sodium heparin, lithium heparin, ammonium heparin, sodium fluoride, and potassium oxalate anticoagulants.

Indications for Use

Indicated for the quantitative determination of Carbon Dioxide in serum and plasma for adult patients to assist in the diagnosis and treatment of disorders associated with acid-base balance changes. Prescription use only.

Regulatory Classification

Identification

A bicarbonate/carbon dioxide test system is a device intended to measure bicarbonate/carbon dioxide in plasma, serum, and whole blood. Bicarbonate/carbon dioxide measurements are used in the diagnosis and treatment of numerous potentially serious disorders associated with changes in body acid-base balance.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # DEC 27 2004 ### 510(k) SUMMARY K040975 #### 1.0 Submitted By: C.C. Allain, Ph.D. Chief Scientific Officer GenChem, Inc. 471 W. Lambert Road, Ste 107 Brea, CA 92821 Telephone: (714) 529-7125 FAX: (714) 529-3339 #### 2.0 Date of Preparation: June 1, 2004 #### 3.0 Regulatory Information: - 3.1 Regulation section: - 3.2 21 CFR § 862.1160, pH Rate Measurement, Carbon Dioxide - 3.3 Clasification : Class II - 3.4 Product Code: JFL - 3.5 Panel: Clinical Chemistry (75) #### 4.0 Device Description: The Device is a solution containing 0.6 mol/L sulfuric acid, nonreactive surfactants and other ingredients necessary for optimum system operation. ### 5.0 Substantial Equivalence Information: - a. Predicate Device Name: Beckman CO2 Reagent for the CX3. - b. Predicate K Number: K014034 - c. Comparison with Predicate: Both Reagents are similar in design, function and chemical principle as well as ingredient composition and concentration. ### Performance Characteristics: All studies were performed on the Beckman CX3 6.0 Synchron Analyzer - 6.1 Precision/Reproducibility: Three control sera were each assayed for total CO2 twice per day in triplicate using flow cell reagents, wash solutions and calibrators on a SYNCHRON CX3® System. Data were collected on ten different days over a thirty day period. {1}------------------------------------------------ Estimates of within run and total imprecision were calculated as described in NCCLS publication EP3-T.10 ### Precision of Total CO2 Recoveries (mmol/L) | Sample | n | mean | SD | %CV | SD | %CV | |---------|----|------|------|-----|------|-----| | Serum 1 | 60 | 12.6 | 0.40 | 3.2 | 0.38 | 3.0 | | Serum 2 | 60 | 22.2 | 0.32 | 1.4 | 0.36 | 1.6 | | Serum 3 | 60 | 30.9 | 0.53 | 1.7 | 0.66 | 2.1 | ### 6.2 Linearity/assay reportable range: Linearity was performed according to NCCLS Guideline EP6-A. Commercially available linearity standards ranging from 0 to 40 mmol/L were analyzed in triplicate on the Beckman CX3® and the results analyzed by the Least Squares method. The results gave a slope of 1.000 with an intercept of 0.15, a standard error of estimate of 0.41 and r2 = 1.000. Samples exceeding these limits should be diluted with normal saline and reanalyzed. Multiply the result by the appropriate dilution factor. | Usable Ranges | | | |---------------|--------------------|----------| | Instrument | Conventional Units | SI Units | | CX3® | 0.0 to 40.0 mmol/L | same | #### 6.3 SENSITIVITY: The sensitivity of this method is 5.0 mmol/L and was documented through the repetitive assay of a diluted serum control. The observed sensitivity limit, calculated as three standard deviations of a 21 replicate within run precision study, is 0.255 mg/dL and is below the claimed limit of 5.0 mg/dL. ### 6.4 Analytical Specificity: Determined according to NCCLS EP7-A. Hemoglobin levels up to 500 mg/dL, Bilirubin levels up to 20 mg/dL, and Lipemia levels up to 1800 mg/dL were tested and did not show any adverse effect on a stock sample with a glucose level of 19 mmol/L. Stock solutions of the substance to be tested were prepared at 20x {2}------------------------------------------------ concentrations and 0.5 ml of this stock was placed in a 10 ml volumetric flask and made up to volume with the base pool. The control stock was prepared similarly but with water as the diluent. Sodium Heparin, Lithium Heparin, Ammonium Heparin, sodium fluoride and potassium oxalate are acceptable anticoagulants. ## Patient Comparison Serum and plasma specimens, collected from adult patients, were assayed for total CO2 on a SYNCHRON CX3® System using GenChem (Y) and Beckman®(X) flow cell reagents, wash solutions and calibrators. Results were compared by least squares linear regression and the following statistics were obtained. | VALUE | SERUM | PLASMA | |----------------|------------|------------| | Intercept | 1.2 | 1.0 | | Slope | 0.949 | 0.979 | | R² Value | 0.953 | 0.987 | | N | 80 | 80 | | Range (mmol/L) | 9.5 – 29.1 | 9.6 – 30.0 | ### Expected Values/ Reference Range: 7.0 The expected values for total CO2 are listed below. Use these ranges only as guides. Each laboratory should establish its own normal ranges. | | Normal Ranges | | |--------------|--------------------|----------| | Specimens | Conventional Units | SI Units | | Serum/Plasma | 22-29 mmol/L | same | {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal features a stylized image of an eagle with three lines representing its wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle. Food and Drug Administration 2098 Gaither Road Rockville MD 20850 DEC 27 2004 C.C. Allain, Ph.D. Chief Scientific Officer GenChem, Inc. 471 W. Lambert Road, Suite 107 Brea. CA 92821 > k040975 Trade/Device Name: GenChem CO2 Acid Reagent Regulation Number: 21 CFR 862.1160 Regulation Name: Bicarbonate/carbon dioxide test system Regulatory Class: Class II Product Code: JFL Dated: October 15, 2004 Received: October 15, 2004 Dear Dr. Allain: Re: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). {4}------------------------------------------------ Page 2 - This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html. Sincerely yours, Cornelia B. Parks Cornelia B. Rooks, MA Acting Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ ### INDICATIONS FOR USE 510(k) Number: K040975 Device Name: GenChem CO2 Acid Reagent Indications For Use: The GenChem CO2 Acid Reagent when used in conjunction with the GenChem ISE Electrolyte Reference, GenChem Electrolyte Buffer, GenChem CO2 Alkaline Buffer, GenChem Wash Concentrate, and appropriate Calibrators or Calibration Standards is intended for the quantitative determination of Carbon Dioxide in serum and plasma on the Beckman CX3®. Carbon Dioxide results are used in the diagnosis and treatment of numerous and potentially serious disorders associated with changes in the body's acidbase balance. Alberto Office of In Vitro Diagno Device Evaluation and S 510(k) K040975 Prescription Use X AND/OR Over-The-Counter Use (Part 21 CFR 801 Subpart D) (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...