MODIFICATION TO ONE TOUCH ULTRA BLOOD GLUCOSE MONITORING SYSTEM
K062195 · Lifescan, Inc. · NBW · Aug 16, 2006 · Clinical Chemistry
Device Facts
Record ID
K062195
Device Name
MODIFICATION TO ONE TOUCH ULTRA BLOOD GLUCOSE MONITORING SYSTEM
Applicant
Lifescan, Inc.
Product Code
NBW · Clinical Chemistry
Decision Date
Aug 16, 2006
Decision
SESE
Submission Type
Special
Regulation
21 CFR 862.1345
Device Class
Class 2
Indications for Use
The OneTouch Ultra Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose in fresh capillary whole blood. The OneTouch Ultra System is intended for use outside the body (in vitro diagnostic use) by people with diabetes at home and/or by healthcare professionals in a clinical setting as an aid to monitor the effectiveness of diabetes control. The OneTouch Ultra Blood Glucose Monitoring System is specifically indicated for use on the finger, forearm or palm.
Device Story
System measures glucose in fresh capillary whole blood; used by patients at home or clinicians in clinical settings. Device consists of meter, test strips, control solution, and blood sampler with lancets. User applies blood sample to test strip; meter performs electrochemical analysis to quantify glucose concentration. Output displayed on meter screen; aids in monitoring diabetes control. Modifications to predicate include meter hardware and firmware updates; electrical safety testing updated to ISO 15197:2003(E).
Clinical Evidence
Bench testing only. Verification and validation activities performed per design control procedures to ensure P1 parameter stability following physical impact (dropping/jarring).
Technological Characteristics
Electrochemical glucose measurement. System includes meter, test strips, control solution, and blood sampler. Modifications include meter hardware and firmware updates. Electrical safety testing performed per ISO 15197:2003(E).
Indications for Use
Indicated for use by patients with diabetes for the quantitative measurement of glucose in capillary whole blood.
Regulatory Classification
Identification
A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.
Special Controls
*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
Predicate Devices
OneTouch® Ultra® Blood Glucose Monitoring System (K062195)
Related Devices
K072543 — ONETOUCH SELECT BLOOD GLUCOSE MONITORING SYSTEM · Lifescan, Inc. · Oct 5, 2007
K053529 — ONETOUCH ULTRA 2 BLOOD GLUCOSE MONITORING SYSTEM · Lifescan, Inc. · Jan 17, 2006
K092902 — U-RIGHT TD-4238 AND TD-4246 BLOOD GLUCOSE MONITORING SYSTEM FORA G31 BLOOD GLUCOSE MONITORING SYSTEM, TD-4256 BLOOD GLUC · Taidoc Technology Corporation · Mar 2, 2010
K093712 — FORA G71A BLOOD GLUCOSE MONITORING SYSTEM, TD-4274 BLOOD GLUCOSE MONITORING SYSTEM · Taidoc Technology Corporation · Feb 19, 2010
Submission Summary (Full Text)
{0}
SPECIAL 510(k): Device Modification
ODE Review Memorandum (Decision Making Document is Attached)
To: THE FILE
RE: DOCUMENT NUMBER K062195
This 510(k) submission contains information/data on modifications made to the SUBMITTER'S own Class II, Class III or Class I devices requiring 510(k). The following items are present and acceptable (delete/add items as necessary):
1. The name and 510(k) number of the SUBMITTER'S previously cleared device. (For a preamendments device, a statement to this effect has been provided.) k043197
2. Submitter's statement that the INDICATION/INTENDED USE of the modified device as described in its labeling HAS NOT CHANGED along with the proposed labeling which includes instructions for use, package labeling, and, if available, advertisements or promotional materials (labeling changes are permitted as long as they do not affect the intended use).
3. A description of the device MODIFICATION(S), including clearly labeled diagrams, engineering drawings, photographs, user's and/or service manuals in sufficient detail to demonstrate that the FUNDAMENTAL SCIENTIFIC TECHNOLOGY of the modified device has not changed.
This change was for a modification to the OneTouch Ultra Blood Glucose Meter to correct "P1 Corruption" related to the Units of Measure (UOM). The meter could change the units of measure due to corruption of the P1 parameters in the meter register when the meter was dropped or jarred while in use. The affected meter settings included: hours, day, year, month, date format, calibration code, next address pointer, unit of measure and/or blood glucose averages.
4. Comparison Information (similarities and differences) to applicant's legally marketed predicate device including, labeling, intended use, physical characteristics, accuracy testing, and precision and linearity studies.
5. A Design Control Activities Summary which includes:
a) Identification of Risk Analysis method(s) used to assess the impact of the modification on the device and its components, and the results of the analysis
b) Based on the Risk Analysis, an identification of the verification and/or validation activities required, including methods or tests used and acceptance criteria to be applied
c) A declaration of conformity with design controls. The declaration of conformity should include:
i) A statement signed by the individual responsible, that, as required by the risk analysis, all verification and validation activities were performed by the designated individual(s) and the results demonstrated that the predetermined acceptance criteria were met, and
ii) A statement signed by the individual responsible, that the manufacturing facility is in conformance with design control procedure requirements as specified in 21 CFR 820.30 and the records are available for review.
6. A Truthful and Accurate Statement, a 510(k) Summary or Statement and the Indications for Use Enclosure (and Class III Summary for Class III devices).
The labeling for this modified subject device has been reviewed to verify that the indication/intended use for the device is unaffected by the modification. In addition, the submitter's description of the particular modification(s) and the comparative information between the modified and unmodified devices demonstrate that the fundamental scientific technology has not changed. The submitter has provided the design control information as specified in The New 510(k) Paradigm and on this basis, I recommend the device be determined substantially equivalent to the previously cleared (or their preamendment) device.
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.